Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study

Detalhes bibliográficos
Autor(a) principal: Saatchi,Masoud
Data de Publicação: 2016
Outros Autores: Aminozarbian,Mohammad Ghasem, Noormohammadi,Hamid, Baghaei,Badri
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Dental Journal
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402016000300336
Resumo: Abstract The aim of this study was to evaluate in vivo the accuracy of the Raypex 5 and Root ZX electronic foramen locators (EFLs) in the presence of blood in the root canal space. Forty single-canal teeth scheduled for extraction were selected. Access cavity was prepared and coronal enlargement was carried out. Approximately two drops of blood were collected by finger prick and injected into the root canal space. The electronic working length (EWL) of each tooth by each device was established twice before (NB group) and after (WB group) injecting blood into the root canal. The tooth was extracted and the actual working length (AWL) was determined. Data were analyzed using McNemar's test. The accuracy rates of Raypex 5 and Root ZX within 0.5 mm in the NB group were 88.9% and 91.5%, with 83.3% and 86.2% in the WB group, respectively. There were no significant differences between the accuracy of each EFL in the two groups (p>0.05). Considering the NB and WB groups, there were no statistically significant differences in the accuracy of the EFLs (p>0.05). The presence of blood in the root canal space did not influence the accuracy of the EFLs.
id FUNORP-1_e44b26c974c136832a0e7a3779aea2e2
oai_identifier_str oai:scielo:S0103-64402016000300336
network_acronym_str FUNORP-1
network_name_str Brazilian Dental Journal
repository_id_str
spelling Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Studybloodelectronic apex locatorsendodonticsworking length.Abstract The aim of this study was to evaluate in vivo the accuracy of the Raypex 5 and Root ZX electronic foramen locators (EFLs) in the presence of blood in the root canal space. Forty single-canal teeth scheduled for extraction were selected. Access cavity was prepared and coronal enlargement was carried out. Approximately two drops of blood were collected by finger prick and injected into the root canal space. The electronic working length (EWL) of each tooth by each device was established twice before (NB group) and after (WB group) injecting blood into the root canal. The tooth was extracted and the actual working length (AWL) was determined. Data were analyzed using McNemar's test. The accuracy rates of Raypex 5 and Root ZX within 0.5 mm in the NB group were 88.9% and 91.5%, with 83.3% and 86.2% in the WB group, respectively. There were no significant differences between the accuracy of each EFL in the two groups (p>0.05). Considering the NB and WB groups, there were no statistically significant differences in the accuracy of the EFLs (p>0.05). The presence of blood in the root canal space did not influence the accuracy of the EFLs.Fundação Odontológica de Ribeirão Preto2016-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402016000300336Brazilian Dental Journal v.27 n.3 2016reponame:Brazilian Dental Journalinstname:Fundação Odontológica de Ribeirão Preto (FUNORP)instacron:FUNORP10.1590/0103-6440201600678info:eu-repo/semantics/openAccessSaatchi,MasoudAminozarbian,Mohammad GhasemNoormohammadi,HamidBaghaei,Badrieng2016-05-18T00:00:00Zoai:scielo:S0103-64402016000300336Revistahttps://www.scielo.br/j/bdj/https://old.scielo.br/oai/scielo-oai.phpbdj@forp.usp.br||sergio@fosjc.unesp.br1806-47600103-6440opendoar:2016-05-18T00:00Brazilian Dental Journal - Fundação Odontológica de Ribeirão Preto (FUNORP)false
dc.title.none.fl_str_mv Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
title Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
spellingShingle Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
Saatchi,Masoud
blood
electronic apex locators
endodontics
working length.
title_short Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
title_full Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
title_fullStr Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
title_full_unstemmed Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
title_sort Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study
author Saatchi,Masoud
author_facet Saatchi,Masoud
Aminozarbian,Mohammad Ghasem
Noormohammadi,Hamid
Baghaei,Badri
author_role author
author2 Aminozarbian,Mohammad Ghasem
Noormohammadi,Hamid
Baghaei,Badri
author2_role author
author
author
dc.contributor.author.fl_str_mv Saatchi,Masoud
Aminozarbian,Mohammad Ghasem
Noormohammadi,Hamid
Baghaei,Badri
dc.subject.por.fl_str_mv blood
electronic apex locators
endodontics
working length.
topic blood
electronic apex locators
endodontics
working length.
description Abstract The aim of this study was to evaluate in vivo the accuracy of the Raypex 5 and Root ZX electronic foramen locators (EFLs) in the presence of blood in the root canal space. Forty single-canal teeth scheduled for extraction were selected. Access cavity was prepared and coronal enlargement was carried out. Approximately two drops of blood were collected by finger prick and injected into the root canal space. The electronic working length (EWL) of each tooth by each device was established twice before (NB group) and after (WB group) injecting blood into the root canal. The tooth was extracted and the actual working length (AWL) was determined. Data were analyzed using McNemar's test. The accuracy rates of Raypex 5 and Root ZX within 0.5 mm in the NB group were 88.9% and 91.5%, with 83.3% and 86.2% in the WB group, respectively. There were no significant differences between the accuracy of each EFL in the two groups (p>0.05). Considering the NB and WB groups, there were no statistically significant differences in the accuracy of the EFLs (p>0.05). The presence of blood in the root canal space did not influence the accuracy of the EFLs.
publishDate 2016
dc.date.none.fl_str_mv 2016-06-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402016000300336
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402016000300336
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/0103-6440201600678
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Fundação Odontológica de Ribeirão Preto
publisher.none.fl_str_mv Fundação Odontológica de Ribeirão Preto
dc.source.none.fl_str_mv Brazilian Dental Journal v.27 n.3 2016
reponame:Brazilian Dental Journal
instname:Fundação Odontológica de Ribeirão Preto (FUNORP)
instacron:FUNORP
instname_str Fundação Odontológica de Ribeirão Preto (FUNORP)