In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants

Detalhes bibliográficos
Autor(a) principal: Carvalho, Elsa Rodrigues
Data de Publicação: 2014
Outros Autores: Figueiral, Maria Helena, Fonseca, Patrícia, Vaz, Mário Augusto, Branco, Fernando Morais
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista odonto ciência (Online)
Texto Completo: https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/11950
Resumo: Objective: Implant-supported overdentures are a valid treatment option for edentulous patients. The use of two mandibular implants is considered the minimum standard of care. This study evaluated the retentive force of two attachments and their respective connection systems in a mandibular implant-supported overdenture. Methods: Two acrylic blocks were prepared: two parallel implants set 22 mm apart were embedded in one of the blocks, while in the other were positioned the female systems for each attachment system tested (Dalbo Ball BTI® and Locator® with female components in both clear and pink nylon). A specific machine was designed to measure the retentive force of the attachments over the cycles applied. The speed of insertion/disinsertion was 0.5 Hz, applied constantly. Results: The clear nylon Locator® attachments registered the highest initial retention value (85.7 N), followed by the Dalbo-Ball® system (62.1 N) and pink nylon Locator® (49.6 N). After 5400 cycles, the mean force exerted was highest in the clear nylon Locator® attachments (42.3 N) followed by Dalbo-Ball® (34.3 N) and pink nylon Locator® (24.6 N). Conclusion: The initial retention force decreases over time for the attachment systems tested.
id PUC_RS-22_24c9e6b76f13f5b20595675dc732a5af
oai_identifier_str oai:ojs.revistaseletronicas.pucrs.br:article/11950
network_acronym_str PUC_RS-22
network_name_str Revista odonto ciência (Online)
repository_id_str
spelling In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implantsEstudo in vitro do efeito de inserção e desinserção na retenção de dois sistemas de attachments numa sobredentadura sobre dois implantesDental prosthesisimplant-supported overdenturesattachment systemPrótese dentáriasobredentadura sobre implantessistema de attachmentObjective: Implant-supported overdentures are a valid treatment option for edentulous patients. The use of two mandibular implants is considered the minimum standard of care. This study evaluated the retentive force of two attachments and their respective connection systems in a mandibular implant-supported overdenture. Methods: Two acrylic blocks were prepared: two parallel implants set 22 mm apart were embedded in one of the blocks, while in the other were positioned the female systems for each attachment system tested (Dalbo Ball BTI® and Locator® with female components in both clear and pink nylon). A specific machine was designed to measure the retentive force of the attachments over the cycles applied. The speed of insertion/disinsertion was 0.5 Hz, applied constantly. Results: The clear nylon Locator® attachments registered the highest initial retention value (85.7 N), followed by the Dalbo-Ball® system (62.1 N) and pink nylon Locator® (49.6 N). After 5400 cycles, the mean force exerted was highest in the clear nylon Locator® attachments (42.3 N) followed by Dalbo-Ball® (34.3 N) and pink nylon Locator® (24.6 N). Conclusion: The initial retention force decreases over time for the attachment systems tested.Objetivo: As sobredentaduras sobre implantes são uma opção válida de tratamento para os desdentados totais. A utilização de dois implantes mandibulares é considerada o minimum standart of care. Este estudo avaliou a força de retenção de dois sistemas de attachments e seus respectivos sistemas de conexão numa sobredentadura mandibular retida por dois implantes. Métodos: Foram confecionados dois blocos de acrílico: num dos blocos incorporaram-se dois implantes paralelos distanciados de 22 mm; no outro foram instalados os respectivos sistemas fêmea para cada sistema de attachment testado (Bola-Dalbo BTI® e Locator®, com fêmeas em nylon transparentes e cor-de-rosa). Desenvolveu-se uma máquina específica para avaliar a retenção dos attachments ao longo dos ciclos aplicados. A velocidade de inserção/desinserção foi de 0,5 Hz, aplicada de forma constante. Resultados: Os Locator® com nylon transparente foram os que apresentaram maior retenção inicial (85,7 N), seguidos do sistema de Bola-Dalbo® (62,1 N) e dos Locator® com nylon cor-de-rosa (49,6 N). Ao fim de 5400 ciclos, a força média exercida foi mais elevada para os Locator® com nylon transparente (42,3 N) seguido do sistema de Bola-Dalbo® (34,3 N) e, por último, dos Locator® com nylon cor-de-rosa (24,6 N). Conclusão: A força de retenção inicial diminui ao longo do tempo para os sistemas de attachments testados.EDIPUCRS - Editora Universitária da PUCRS2014-07-28info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPeer-reviewed Articleapplication/pdfhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/1195010.15448/1980-6523.2014.1.11950Revista Odonto Ciência; Vol. 29 No. 1 (2014); 01-05Revista Odonto Ciência; v. 29 n. 1 (2014); 01-051980-65230102-946010.15448/1980-6523.2014.1reponame:Revista odonto ciência (Online)instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)instacron:PUC_RSenghttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/11950/11775Carvalho, Elsa RodriguesFigueiral, Maria HelenaFonseca, PatríciaVaz, Mário AugustoBranco, Fernando Moraisinfo:eu-repo/semantics/openAccess2015-08-04T18:11:30Zoai:ojs.revistaseletronicas.pucrs.br:article/11950Revistahttps://revistaseletronicas.pucrs.br/ojs/index.php/foPRIhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/oai||odontociencia@pucrs.br1980-65230102-9460opendoar:2015-08-04T18:11:30Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)false
dc.title.none.fl_str_mv In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
Estudo in vitro do efeito de inserção e desinserção na retenção de dois sistemas de attachments numa sobredentadura sobre dois implantes
title In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
spellingShingle In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
Carvalho, Elsa Rodrigues
Dental prosthesis
implant-supported overdentures
attachment system
Prótese dentária
sobredentadura sobre implantes
sistema de attachment
title_short In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
title_full In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
title_fullStr In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
title_full_unstemmed In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
title_sort In vitro study of the insertion and disinsertion effect on retention of two attachment systems of an overdenture on two implants
author Carvalho, Elsa Rodrigues
author_facet Carvalho, Elsa Rodrigues
Figueiral, Maria Helena
Fonseca, Patrícia
Vaz, Mário Augusto
Branco, Fernando Morais
author_role author
author2 Figueiral, Maria Helena
Fonseca, Patrícia
Vaz, Mário Augusto
Branco, Fernando Morais
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Carvalho, Elsa Rodrigues
Figueiral, Maria Helena
Fonseca, Patrícia
Vaz, Mário Augusto
Branco, Fernando Morais
dc.subject.por.fl_str_mv Dental prosthesis
implant-supported overdentures
attachment system
Prótese dentária
sobredentadura sobre implantes
sistema de attachment
topic Dental prosthesis
implant-supported overdentures
attachment system
Prótese dentária
sobredentadura sobre implantes
sistema de attachment
description Objective: Implant-supported overdentures are a valid treatment option for edentulous patients. The use of two mandibular implants is considered the minimum standard of care. This study evaluated the retentive force of two attachments and their respective connection systems in a mandibular implant-supported overdenture. Methods: Two acrylic blocks were prepared: two parallel implants set 22 mm apart were embedded in one of the blocks, while in the other were positioned the female systems for each attachment system tested (Dalbo Ball BTI® and Locator® with female components in both clear and pink nylon). A specific machine was designed to measure the retentive force of the attachments over the cycles applied. The speed of insertion/disinsertion was 0.5 Hz, applied constantly. Results: The clear nylon Locator® attachments registered the highest initial retention value (85.7 N), followed by the Dalbo-Ball® system (62.1 N) and pink nylon Locator® (49.6 N). After 5400 cycles, the mean force exerted was highest in the clear nylon Locator® attachments (42.3 N) followed by Dalbo-Ball® (34.3 N) and pink nylon Locator® (24.6 N). Conclusion: The initial retention force decreases over time for the attachment systems tested.
publishDate 2014
dc.date.none.fl_str_mv 2014-07-28
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Peer-reviewed Article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/11950
10.15448/1980-6523.2014.1.11950
url https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/11950
identifier_str_mv 10.15448/1980-6523.2014.1.11950
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/11950/11775
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv EDIPUCRS - Editora Universitária da PUCRS
publisher.none.fl_str_mv EDIPUCRS - Editora Universitária da PUCRS
dc.source.none.fl_str_mv Revista Odonto Ciência; Vol. 29 No. 1 (2014); 01-05
Revista Odonto Ciência; v. 29 n. 1 (2014); 01-05
1980-6523
0102-9460
10.15448/1980-6523.2014.1
reponame:Revista odonto ciência (Online)
instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
instacron:PUC_RS
instname_str Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
instacron_str PUC_RS
institution PUC_RS
reponame_str Revista odonto ciência (Online)
collection Revista odonto ciência (Online)
repository.name.fl_str_mv Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
repository.mail.fl_str_mv ||odontociencia@pucrs.br
_version_ 1754820876038045696