A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis

Detalhes bibliográficos
Autor(a) principal: Gehrke, Sergio Alexandre
Data de Publicação: 2021
Outros Autores: Dedavid, Berenice Anina, Fernandes, Gustavo Vicentis de Oliveira
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/10400.14/32186
Resumo: The objective of this study was to evaluate a new design of multifunctional abutment for Morse taper implant connections, relative to the retentive stability after the application of cyclic loads in cemented and screwed crowns. Multifunctional abutments with two different angulations in the seating portion of the crown were tested, forming 2 groups (n = 30 samples per group): Group Abut11, where Smart abutments with an angle of 11.42° were used; Group Abut5, where Ideale abutments with an angle of 5° were used. Fifteen samples from each group received cemented crowns (CC) and another fifteen screwed crowns (SC). All crown samples were subjected to the mechanical cycling test at 360,000 cycles at a frequency of 4 Hz and 150 N of the load. The samples with CC were subjected to the tensile test to remove the crowns, while in the samples with SC, the detorque value of the fastening crown screws was measured. The mean tensile strength value of CC in the Abut11 group was 131.9 ± 13.5 N and, in the Abut5 group was 230.9 ± 11.3 N; while the detorque mean value in samples with SC 5.8 ± 1.8 N for the Abut11 group and, 7.6 ± 1.1 N for Abut5 group. Significant statistical differences were found between the two groups in both situations tested (p < 0.05). The multifunctional abutments, presenting a lesser angulation of the crown-seating portion, showed higher values of retention of the CC and a lesser screw loosening of torque of the fixing screws in the SC after the application of cyclic loads when compared to the abutments with more angulation in the crown-seating portion.
id RCAP_1021e03c29db86fef9fdab730aeac2a9
oai_identifier_str oai:repositorio.ucp.pt:10400.14/32186
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysisAbutment designDental implantsMechanical cyclingMorse taper connectionScrew detorqueTensile strengthThe objective of this study was to evaluate a new design of multifunctional abutment for Morse taper implant connections, relative to the retentive stability after the application of cyclic loads in cemented and screwed crowns. Multifunctional abutments with two different angulations in the seating portion of the crown were tested, forming 2 groups (n = 30 samples per group): Group Abut11, where Smart abutments with an angle of 11.42° were used; Group Abut5, where Ideale abutments with an angle of 5° were used. Fifteen samples from each group received cemented crowns (CC) and another fifteen screwed crowns (SC). All crown samples were subjected to the mechanical cycling test at 360,000 cycles at a frequency of 4 Hz and 150 N of the load. The samples with CC were subjected to the tensile test to remove the crowns, while in the samples with SC, the detorque value of the fastening crown screws was measured. The mean tensile strength value of CC in the Abut11 group was 131.9 ± 13.5 N and, in the Abut5 group was 230.9 ± 11.3 N; while the detorque mean value in samples with SC 5.8 ± 1.8 N for the Abut11 group and, 7.6 ± 1.1 N for Abut5 group. Significant statistical differences were found between the two groups in both situations tested (p < 0.05). The multifunctional abutments, presenting a lesser angulation of the crown-seating portion, showed higher values of retention of the CC and a lesser screw loosening of torque of the fixing screws in the SC after the application of cyclic loads when compared to the abutments with more angulation in the crown-seating portion.Veritati - Repositório Institucional da Universidade Católica PortuguesaGehrke, Sergio AlexandreDedavid, Berenice AninaFernandes, Gustavo Vicentis de Oliveira2021-03-10T17:09:15Z2021-042021-04-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.14/32186eng1751-616110.1016/j.jmbbm.2021.1043478509978414633513461000624346200006info:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2023-12-19T01:37:02Zoai:repositorio.ucp.pt:10400.14/32186Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-19T18:25:58.798144Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
title A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
spellingShingle A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
Gehrke, Sergio Alexandre
Abutment design
Dental implants
Mechanical cycling
Morse taper connection
Screw detorque
Tensile strength
title_short A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
title_full A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
title_fullStr A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
title_full_unstemmed A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
title_sort A new design of a multifunctional abutment to morse taper implant connection: experimental mechanical analysis
author Gehrke, Sergio Alexandre
author_facet Gehrke, Sergio Alexandre
Dedavid, Berenice Anina
Fernandes, Gustavo Vicentis de Oliveira
author_role author
author2 Dedavid, Berenice Anina
Fernandes, Gustavo Vicentis de Oliveira
author2_role author
author
dc.contributor.none.fl_str_mv Veritati - Repositório Institucional da Universidade Católica Portuguesa
dc.contributor.author.fl_str_mv Gehrke, Sergio Alexandre
Dedavid, Berenice Anina
Fernandes, Gustavo Vicentis de Oliveira
dc.subject.por.fl_str_mv Abutment design
Dental implants
Mechanical cycling
Morse taper connection
Screw detorque
Tensile strength
topic Abutment design
Dental implants
Mechanical cycling
Morse taper connection
Screw detorque
Tensile strength
description The objective of this study was to evaluate a new design of multifunctional abutment for Morse taper implant connections, relative to the retentive stability after the application of cyclic loads in cemented and screwed crowns. Multifunctional abutments with two different angulations in the seating portion of the crown were tested, forming 2 groups (n = 30 samples per group): Group Abut11, where Smart abutments with an angle of 11.42° were used; Group Abut5, where Ideale abutments with an angle of 5° were used. Fifteen samples from each group received cemented crowns (CC) and another fifteen screwed crowns (SC). All crown samples were subjected to the mechanical cycling test at 360,000 cycles at a frequency of 4 Hz and 150 N of the load. The samples with CC were subjected to the tensile test to remove the crowns, while in the samples with SC, the detorque value of the fastening crown screws was measured. The mean tensile strength value of CC in the Abut11 group was 131.9 ± 13.5 N and, in the Abut5 group was 230.9 ± 11.3 N; while the detorque mean value in samples with SC 5.8 ± 1.8 N for the Abut11 group and, 7.6 ± 1.1 N for Abut5 group. Significant statistical differences were found between the two groups in both situations tested (p < 0.05). The multifunctional abutments, presenting a lesser angulation of the crown-seating portion, showed higher values of retention of the CC and a lesser screw loosening of torque of the fixing screws in the SC after the application of cyclic loads when compared to the abutments with more angulation in the crown-seating portion.
publishDate 2021
dc.date.none.fl_str_mv 2021-03-10T17:09:15Z
2021-04
2021-04-01T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.14/32186
url http://hdl.handle.net/10400.14/32186
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 1751-6161
10.1016/j.jmbbm.2021.104347
85099784146
33513461
000624346200006
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799131975991164928