Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands

Detalhes bibliográficos
Autor(a) principal: Heron, Michiel
Data de Publicação: 2023
Outros Autores: Schreurs, Marco W.J., Haagen, Inez-Anne, China, Bernard, Faria, Ana Paula, Vanhanen, Anna-Riitta, Thelen, Marc, Weykamp, Cas W.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/10400.18/8852
Resumo: Objectives: This paper evaluates 16 year results of the Allergy EQA program shared by EQA organisers in Belgium, Finland, Portugal, and The Netherlands. Methods: The performance of Thermo Fisher and Siemens user groups (in terms of concordance between both groups, between laboratory CV, prevalence of clinically significant errors) and suitability of samples (stability and validity of dilution of patient samples) are evaluated using data of 192 samples in the EQA programs from 2007 to 2022. Measurands covered are total IgE, screens and mixes, specific IgE extracts and allergen components. Results: There is perfect (53 %), acceptable (40 %) and poor (6 %) concordance between both method groups. In case of poor concordance the best fit with clinical data is seen for Thermo Fisher (56 %) and Siemens (26 %) respectively. The between laboratory CV evolves from 7.8 to 6.6 % (Thermo Fisher) and 7.3 to 7.7 % (Siemens). The prevalence of blunders by individual laboratories is stable for Siemens (0.4 %) and drops from 0.4 to 0.2 % for Thermo Fisher users. For IgE, the between year CV of the mean of both user groups is 1 %, and a fifteen-fold dilution of a patient sample has an impact of 2 and 4 % on the recovery of Thermo Fisher and Siemens user groups. Conclusions: The analytical performance of Thermo Fisher is slightly better than that of Siemens users but the clinical impact of this difference is limited. Stability of the sample and the low impact of dilution on the recovery of measurands demonstrates the suitability for purpose of the EQA program.
id RCAP_405b221e908ce056717f156c31de74e5
oai_identifier_str oai:repositorio.insa.pt:10400.18/8852
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The NetherlandsAllergyEQA ProgramAEQBetween Laboratory CVSample QualityExternal Quality AssessmentThermo FisherSiemensPrograma Nacional de Avaliação Externa da QualidadePNAEQAlergiasAvaliação Externa da QualidadeObjectives: This paper evaluates 16 year results of the Allergy EQA program shared by EQA organisers in Belgium, Finland, Portugal, and The Netherlands. Methods: The performance of Thermo Fisher and Siemens user groups (in terms of concordance between both groups, between laboratory CV, prevalence of clinically significant errors) and suitability of samples (stability and validity of dilution of patient samples) are evaluated using data of 192 samples in the EQA programs from 2007 to 2022. Measurands covered are total IgE, screens and mixes, specific IgE extracts and allergen components. Results: There is perfect (53 %), acceptable (40 %) and poor (6 %) concordance between both method groups. In case of poor concordance the best fit with clinical data is seen for Thermo Fisher (56 %) and Siemens (26 %) respectively. The between laboratory CV evolves from 7.8 to 6.6 % (Thermo Fisher) and 7.3 to 7.7 % (Siemens). The prevalence of blunders by individual laboratories is stable for Siemens (0.4 %) and drops from 0.4 to 0.2 % for Thermo Fisher users. For IgE, the between year CV of the mean of both user groups is 1 %, and a fifteen-fold dilution of a patient sample has an impact of 2 and 4 % on the recovery of Thermo Fisher and Siemens user groups. Conclusions: The analytical performance of Thermo Fisher is slightly better than that of Siemens users but the clinical impact of this difference is limited. Stability of the sample and the low impact of dilution on the recovery of measurands demonstrates the suitability for purpose of the EQA program.De GruyterRepositório Científico do Instituto Nacional de SaúdeHeron, MichielSchreurs, Marco W.J.Haagen, Inez-AnneChina, BernardFaria, Ana PaulaVanhanen, Anna-RiittaThelen, MarcWeykamp, Cas W.2023-11-282024-11-29T00:00:00Z2023-11-28T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.18/8852engClin Chem Lab Med. 2023 Nov 28. doi: 10.1515/cclm-2023-0862. Online ahead of print.1434-662110.1515/cclm-2023-0862info:eu-repo/semantics/embargoedAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2024-01-06T01:33:17Zoai:repositorio.insa.pt:10400.18/8852Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-20T01:30:29.554273Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
title Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
spellingShingle Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
Heron, Michiel
Allergy
EQA Program
AEQ
Between Laboratory CV
Sample Quality
External Quality Assessment
Thermo Fisher
Siemens
Programa Nacional de Avaliação Externa da Qualidade
PNAEQ
Alergias
Avaliação Externa da Qualidade
title_short Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
title_full Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
title_fullStr Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
title_full_unstemmed Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
title_sort Allergy: Evaluation of 16 years (2007–2022) results of the shared external quality assessment program in Belgium, Finland, Portugal and The Netherlands
author Heron, Michiel
author_facet Heron, Michiel
Schreurs, Marco W.J.
Haagen, Inez-Anne
China, Bernard
Faria, Ana Paula
Vanhanen, Anna-Riitta
Thelen, Marc
Weykamp, Cas W.
author_role author
author2 Schreurs, Marco W.J.
Haagen, Inez-Anne
China, Bernard
Faria, Ana Paula
Vanhanen, Anna-Riitta
Thelen, Marc
Weykamp, Cas W.
author2_role author
author
author
author
author
author
author
dc.contributor.none.fl_str_mv Repositório Científico do Instituto Nacional de Saúde
dc.contributor.author.fl_str_mv Heron, Michiel
Schreurs, Marco W.J.
Haagen, Inez-Anne
China, Bernard
Faria, Ana Paula
Vanhanen, Anna-Riitta
Thelen, Marc
Weykamp, Cas W.
dc.subject.por.fl_str_mv Allergy
EQA Program
AEQ
Between Laboratory CV
Sample Quality
External Quality Assessment
Thermo Fisher
Siemens
Programa Nacional de Avaliação Externa da Qualidade
PNAEQ
Alergias
Avaliação Externa da Qualidade
topic Allergy
EQA Program
AEQ
Between Laboratory CV
Sample Quality
External Quality Assessment
Thermo Fisher
Siemens
Programa Nacional de Avaliação Externa da Qualidade
PNAEQ
Alergias
Avaliação Externa da Qualidade
description Objectives: This paper evaluates 16 year results of the Allergy EQA program shared by EQA organisers in Belgium, Finland, Portugal, and The Netherlands. Methods: The performance of Thermo Fisher and Siemens user groups (in terms of concordance between both groups, between laboratory CV, prevalence of clinically significant errors) and suitability of samples (stability and validity of dilution of patient samples) are evaluated using data of 192 samples in the EQA programs from 2007 to 2022. Measurands covered are total IgE, screens and mixes, specific IgE extracts and allergen components. Results: There is perfect (53 %), acceptable (40 %) and poor (6 %) concordance between both method groups. In case of poor concordance the best fit with clinical data is seen for Thermo Fisher (56 %) and Siemens (26 %) respectively. The between laboratory CV evolves from 7.8 to 6.6 % (Thermo Fisher) and 7.3 to 7.7 % (Siemens). The prevalence of blunders by individual laboratories is stable for Siemens (0.4 %) and drops from 0.4 to 0.2 % for Thermo Fisher users. For IgE, the between year CV of the mean of both user groups is 1 %, and a fifteen-fold dilution of a patient sample has an impact of 2 and 4 % on the recovery of Thermo Fisher and Siemens user groups. Conclusions: The analytical performance of Thermo Fisher is slightly better than that of Siemens users but the clinical impact of this difference is limited. Stability of the sample and the low impact of dilution on the recovery of measurands demonstrates the suitability for purpose of the EQA program.
publishDate 2023
dc.date.none.fl_str_mv 2023-11-28
2023-11-28T00:00:00Z
2024-11-29T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.18/8852
url http://hdl.handle.net/10400.18/8852
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Clin Chem Lab Med. 2023 Nov 28. doi: 10.1515/cclm-2023-0862. Online ahead of print.
1434-6621
10.1515/cclm-2023-0862
dc.rights.driver.fl_str_mv info:eu-repo/semantics/embargoedAccess
eu_rights_str_mv embargoedAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv De Gruyter
publisher.none.fl_str_mv De Gruyter
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799136787066519552