Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry

Detalhes bibliográficos
Autor(a) principal: Fernandes, Sara
Data de Publicação: 2021
Outros Autores: Barreiros, Luísa, Sá, Paula, Miró, Manuel, Segundo, Marcela A.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/10400.22/18503
Resumo: An automatic micro-solid-phase extraction (μSPE) method using on-line renewable sorbent beads followed by liquid chromatography–tandem mass spectrometry (LC–MS/MS) was established for the determination of tranexamic acid (TXA) in urine. The μSPE method was based on the bead injection (BI) concept combined with the mesofluidic lab-on-valve (LOV) platform. All steps of the μSPE–BI–LOV were implemented by computer programming, rendering enhanced precision on time and flow events. Several parameters, including the type of sorbent, volume and composition of the conditioning solution, washing solution, and eluent composition, were evaluated to improve the extraction efficiency. The best results were obtained with a hydrophilic–lipophilic balanced mixed-mode sorbent, decorated with sulfonic acid groups (Oasis MCX), and 99% acetonitrile–water (50:50, v/v)–1% ammonium hydroxide as eluent. Chromatographic separation was performed using a BEH amide column coupled to MS/MS detection in positive ionization mode. Good linearity was achieved (R2 > 0.998) for TXA concentrations in urine ranging from 300 to 3000 ng mL−1, with LOD and LOQ of 30 and 65 ng mL−1, respectively. Dilution integrity was observed for dilution factors up to 20,000 times, providing the extension of the upper limit of quantification to 12 mg mL−1. The method was validated according to international guidelines and successfully applied to urine samples collected during scoliosis surgery of pediatric patients treated with TXA.
id RCAP_fe849ebe37a4fce4454e12ac6c56547a
oai_identifier_str oai:recipp.ipp.pt:10400.22/18503
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometryBead injectionLab-on-valveAutomationMass spectrometryTranexamic acidUrineAn automatic micro-solid-phase extraction (μSPE) method using on-line renewable sorbent beads followed by liquid chromatography–tandem mass spectrometry (LC–MS/MS) was established for the determination of tranexamic acid (TXA) in urine. The μSPE method was based on the bead injection (BI) concept combined with the mesofluidic lab-on-valve (LOV) platform. All steps of the μSPE–BI–LOV were implemented by computer programming, rendering enhanced precision on time and flow events. Several parameters, including the type of sorbent, volume and composition of the conditioning solution, washing solution, and eluent composition, were evaluated to improve the extraction efficiency. The best results were obtained with a hydrophilic–lipophilic balanced mixed-mode sorbent, decorated with sulfonic acid groups (Oasis MCX), and 99% acetonitrile–water (50:50, v/v)–1% ammonium hydroxide as eluent. Chromatographic separation was performed using a BEH amide column coupled to MS/MS detection in positive ionization mode. Good linearity was achieved (R2 > 0.998) for TXA concentrations in urine ranging from 300 to 3000 ng mL−1, with LOD and LOQ of 30 and 65 ng mL−1, respectively. Dilution integrity was observed for dilution factors up to 20,000 times, providing the extension of the upper limit of quantification to 12 mg mL−1. The method was validated according to international guidelines and successfully applied to urine samples collected during scoliosis surgery of pediatric patients treated with TXA.SpringerRepositório Científico do Instituto Politécnico do PortoFernandes, SaraBarreiros, LuísaSá, PaulaMiró, ManuelSegundo, Marcela A.2021-09-23T11:08:28Z20222022-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.22/18503engFernandes, S.R., Barreiros, L., Sá, P. et al. Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry. Anal Bioanal Chem 414, 649–659 (2022). https://doi.org/10.1007/s00216-021-03606-y1618-265010.1007/s00216-021-03606-ymetadata only accessinfo:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2023-03-13T13:10:18Zoai:recipp.ipp.pt:10400.22/18503Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-19T17:38:06.401549Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
title Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
spellingShingle Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
Fernandes, Sara
Bead injection
Lab-on-valve
Automation
Mass spectrometry
Tranexamic acid
Urine
title_short Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
title_full Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
title_fullStr Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
title_full_unstemmed Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
title_sort Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry
author Fernandes, Sara
author_facet Fernandes, Sara
Barreiros, Luísa
Sá, Paula
Miró, Manuel
Segundo, Marcela A.
author_role author
author2 Barreiros, Luísa
Sá, Paula
Miró, Manuel
Segundo, Marcela A.
author2_role author
author
author
author
dc.contributor.none.fl_str_mv Repositório Científico do Instituto Politécnico do Porto
dc.contributor.author.fl_str_mv Fernandes, Sara
Barreiros, Luísa
Sá, Paula
Miró, Manuel
Segundo, Marcela A.
dc.subject.por.fl_str_mv Bead injection
Lab-on-valve
Automation
Mass spectrometry
Tranexamic acid
Urine
topic Bead injection
Lab-on-valve
Automation
Mass spectrometry
Tranexamic acid
Urine
description An automatic micro-solid-phase extraction (μSPE) method using on-line renewable sorbent beads followed by liquid chromatography–tandem mass spectrometry (LC–MS/MS) was established for the determination of tranexamic acid (TXA) in urine. The μSPE method was based on the bead injection (BI) concept combined with the mesofluidic lab-on-valve (LOV) platform. All steps of the μSPE–BI–LOV were implemented by computer programming, rendering enhanced precision on time and flow events. Several parameters, including the type of sorbent, volume and composition of the conditioning solution, washing solution, and eluent composition, were evaluated to improve the extraction efficiency. The best results were obtained with a hydrophilic–lipophilic balanced mixed-mode sorbent, decorated with sulfonic acid groups (Oasis MCX), and 99% acetonitrile–water (50:50, v/v)–1% ammonium hydroxide as eluent. Chromatographic separation was performed using a BEH amide column coupled to MS/MS detection in positive ionization mode. Good linearity was achieved (R2 > 0.998) for TXA concentrations in urine ranging from 300 to 3000 ng mL−1, with LOD and LOQ of 30 and 65 ng mL−1, respectively. Dilution integrity was observed for dilution factors up to 20,000 times, providing the extension of the upper limit of quantification to 12 mg mL−1. The method was validated according to international guidelines and successfully applied to urine samples collected during scoliosis surgery of pediatric patients treated with TXA.
publishDate 2021
dc.date.none.fl_str_mv 2021-09-23T11:08:28Z
2022
2022-01-01T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.22/18503
url http://hdl.handle.net/10400.22/18503
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Fernandes, S.R., Barreiros, L., Sá, P. et al. Automatic and renewable micro-solid-phase extraction based on bead injection lab-on-valve system for determination of tranexamic acid in urine by UHPLC coupled with tandem mass spectrometry. Anal Bioanal Chem 414, 649–659 (2022). https://doi.org/10.1007/s00216-021-03606-y
1618-2650
10.1007/s00216-021-03606-y
dc.rights.driver.fl_str_mv metadata only access
info:eu-repo/semantics/openAccess
rights_invalid_str_mv metadata only access
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Springer
publisher.none.fl_str_mv Springer
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799131470849114112