Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes

Detalhes bibliográficos
Autor(a) principal: Bisognin, Edecir Décio Cargnin
Data de Publicação: 2009
Tipo de documento: Dissertação
Idioma: por
Título da fonte: Biblioteca Digital de Teses e Dissertações da UNIGRANRIO
Texto Completo: http://localhost:8080/tede/handle/tede/60
Resumo: The best fit between the implant and the prosthetic component appears to be important for periimplant homeostasis. Considering that the oral cavity is a naturally contaminated environment, it is important that the rehabilitation with implants does not promote the emergence of areas for microorganisms deposition that might cause periimplant diseases, thus compromising the long term osseointegration. The objectives of this study were: evaluate the misfit and bacterial infiltration at the implant-abutment interface of five implant systems with external hexagon, manufactured and marketed in Brazil, and to analyze the vertical and horizontal misfit according to the classification proposed by Kano, Binon e Curtis (2007). Eight samples of each implant system: Conexão®, Dentoflex®, Neodent®, SIN® and Titanium Fix® were tested. Initially, the sets were inoculated with 0.3μl of a suspension containing Escherichia coli and placed in culture medium for analysis of bacterial infiltration. The samples analyses were performed with 1, 2, 5, 7, and 14 days for assessment of contamination through the turbidity of the culture medium. After the microbiological analysis all, samples of each system were prepared for misfits evaluation through scanning electron microscopy (SEM), with 1000x and 3000x of magnification. The results showed that the bacterial infiltration was observed in 25% of samples of the Dentoflex® system. Misfit was observed in all systems. The misfit average values of each system were: Neodent® - 0.51μm ± 0.39; Dentoflex® - 1.44μm ± 0.73; Titanium Fix® - 1.88μm ± 1.28; SIN® - 2, 46μm ± 3.38; Conexão® - 2.68μm ± 3.02. According to the classification of Kano, Binon e Curtis (2007), it was observed a predominance types II and IV of misfit, respectively, horizontal and horizontal and vertical simultaneously. It was concluded that all groups showed misfit at least in one of the points evaluated along the interface, and this increase is not the only determining factor for the occurrence of bacterial infiltration. It was also observed that the extent of the area of coupling interface should also take into account, which showed variations among the implant systems evaluated.
id UGRI_dcb46b8f4235e1990297b352af24c27b
oai_identifier_str oai:localhost:tede/60
network_acronym_str UGRI
network_name_str Biblioteca Digital de Teses e Dissertações da UNIGRANRIO
repository_id_str
spelling Vidigal Júnior, Guaracilei MacielSoares, Glória de AlmeidaAndrade Filho, Hugo deMendes, Luiz MotaMoreira, Edson Jorge LimaBisognin, Edecir Décio Cargnin2016-09-26T22:52:02Z2009-09-24Bisognin, Edecir Décio Cargnin. Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes. 2009. 82 f. Dissertação( Programa de Pós-Graduação em Odontologia) - Universidade do Grande Rio, Duque de Caxias.http://localhost:8080/tede/handle/tede/60The best fit between the implant and the prosthetic component appears to be important for periimplant homeostasis. Considering that the oral cavity is a naturally contaminated environment, it is important that the rehabilitation with implants does not promote the emergence of areas for microorganisms deposition that might cause periimplant diseases, thus compromising the long term osseointegration. The objectives of this study were: evaluate the misfit and bacterial infiltration at the implant-abutment interface of five implant systems with external hexagon, manufactured and marketed in Brazil, and to analyze the vertical and horizontal misfit according to the classification proposed by Kano, Binon e Curtis (2007). Eight samples of each implant system: Conexão®, Dentoflex®, Neodent®, SIN® and Titanium Fix® were tested. Initially, the sets were inoculated with 0.3μl of a suspension containing Escherichia coli and placed in culture medium for analysis of bacterial infiltration. The samples analyses were performed with 1, 2, 5, 7, and 14 days for assessment of contamination through the turbidity of the culture medium. After the microbiological analysis all, samples of each system were prepared for misfits evaluation through scanning electron microscopy (SEM), with 1000x and 3000x of magnification. The results showed that the bacterial infiltration was observed in 25% of samples of the Dentoflex® system. Misfit was observed in all systems. The misfit average values of each system were: Neodent® - 0.51μm ± 0.39; Dentoflex® - 1.44μm ± 0.73; Titanium Fix® - 1.88μm ± 1.28; SIN® - 2, 46μm ± 3.38; Conexão® - 2.68μm ± 3.02. According to the classification of Kano, Binon e Curtis (2007), it was observed a predominance types II and IV of misfit, respectively, horizontal and horizontal and vertical simultaneously. It was concluded that all groups showed misfit at least in one of the points evaluated along the interface, and this increase is not the only determining factor for the occurrence of bacterial infiltration. It was also observed that the extent of the area of coupling interface should also take into account, which showed variations among the implant systems evaluated.A melhor adaptação entre o implante e seu componente protético parece ser importante para a homeostase perimplantar. Considerando-se que a cavidade oral é um ambiente naturalmente contaminado, é de importância que ao reabilitarem-se os implantes não se favoreça o surgimento de áreas para a deposição destes microorganismos que, poderiam causar doenças perimplantares comprometendo a osseointegração em longo prazo. Os objetivos deste estudo foram: avaliar a desadaptação e a infiltração bacteriana na interface implante-pilar protético de cinco sistemas de implante com hexágono externo, fabricados e comercializados no Brasil; e analisar a desadaptação vertical e horizontal de acordo com a classificação proposta por Kano, Binon e Curtis (2007). Foram testados oito implantes de cada uma das marcas Conexão®, Dentoflex®, Neodent®, SIN® e Titanium Fix®. Inicialmente, os conjuntos foram inoculados com 0,3μl de uma suspensão contendo Escherichia coli e colocados em meio de cultura para análise da infiltração bacteriana. A leitura das amostras foi realizada com 1, 2, 5, 7 e 14 dias para avaliação da contaminação pela turvação do meio de cultura. Após a análise microbiológica, as amostras de cada sistema foram preparadas para avaliação da adaptação através de microscopia eletrônica de varredura, com aumentos de 1000 e 3000 vezes. Os resultados revelaram que a infiltração bacteriana foi observada em 25% das amostras do sistema Dentoflex®. A desadaptação foi observada em todos os sistemas. Os valores médios de desadaptação de cada sistema foram de: Neodent® - 0,51μm ± 0,39; Dentoflex® - 1,44μm ± 0,73; Titanium Fix® - 1,88μm ± 1,28; SIN® - 2,46μm ± 3,38; Conexão® - 2,68μm ± 3,02. De acordo com a classificação de Kano, Binon e Curtis (2007), foi observada uma predominância dos tipos II e IV, de desadaptação, respectivamente, horizontal e horizontal e vertical simultaneamente. Concluiu-se que todos os grupos apresentaram desadaptação, em pelo menos um dos pontos avaliados ao longo da interface, e o aumento desta não é o único fator determinante para a ocorrência do infiltrado bacteriano, devendo-se levar em consideração também a extensão da área de acoplamento da interface, que apresentou variações entre os sistemas avaliados.Submitted by Patricia Vieira Silva (patricia.silva@unigranrio.com.br) on 2016-09-26T22:52:02Z No. of bitstreams: 1 Dissertação_Edecir Décio.pdf: 5659314 bytes, checksum: b6fe4d7fc9719a0353728f8149806bc6 (MD5)Made available in DSpace on 2016-09-26T22:52:02Z (GMT). No. of bitstreams: 1 Dissertação_Edecir Décio.pdf: 5659314 bytes, checksum: b6fe4d7fc9719a0353728f8149806bc6 (MD5) Previous issue date: 2009-09-24application/pdfporUniversidade do Grande RioPrograma de Pós-Graduação em OdontologiaUNIGRANRIOBrasilUnigranrio::OdontologiaOdontologiaImplante dentárioOsseointegraçãoInfiltração dentária-MicrobiologiaImplante dentário endoósseoPerda óssea alveolarBiomecânicaEstudos de avaliaçãoCirurgia Buco-Maxilo-FacialAvaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantesinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/masterThesisinfo:eu-repo/semantics/openAccessreponame:Biblioteca Digital de Teses e Dissertações da UNIGRANRIOinstname:Universidade do Grande Rio (UNIGRANRIO)instacron:UNIGRANRIOLICENSElicense.txtlicense.txttext/plain; charset=utf-81982http://localhost:8080/tede/bitstream/tede/60/1/license.txt4a50535e8405f611398e6e2d408dbd1bMD51ORIGINALDissertação_Edecir Décio.pdfDissertação_Edecir Décio.pdfapplication/pdf5659314http://localhost:8080/tede/bitstream/tede/60/2/Disserta%C3%A7%C3%A3o_Edecir+D%C3%A9cio.pdfb6fe4d7fc9719a0353728f8149806bc6MD52tede/602016-09-26 19:52:02.7oai:localhost: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Biblioteca Digital de Teses e Dissertaçõeshttp://tede.unigranrio.edu.br/PUBhttp://tede.unigranrio.edu.br/oai/requestrepositorio@instituicao.br||repositorio@instituicao.bropendoar:2016-09-26T22:52:02Biblioteca Digital de Teses e Dissertações da UNIGRANRIO - Universidade do Grande Rio (UNIGRANRIO)false
dc.title.por.fl_str_mv Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
title Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
spellingShingle Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
Bisognin, Edecir Décio Cargnin
Odontologia
Implante dentário
Osseointegração
Infiltração dentária-Microbiologia
Implante dentário endoósseo
Perda óssea alveolar
Biomecânica
Estudos de avaliação
Cirurgia Buco-Maxilo-Facial
title_short Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
title_full Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
title_fullStr Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
title_full_unstemmed Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
title_sort Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes
author Bisognin, Edecir Décio Cargnin
author_facet Bisognin, Edecir Décio Cargnin
author_role author
dc.contributor.advisor1.fl_str_mv Vidigal Júnior, Guaracilei Maciel
dc.contributor.advisor2.fl_str_mv Soares, Glória de Almeida
dc.contributor.referee1.fl_str_mv Andrade Filho, Hugo de
dc.contributor.referee2.fl_str_mv Mendes, Luiz Mota
dc.contributor.referee3.fl_str_mv Moreira, Edson Jorge Lima
dc.contributor.author.fl_str_mv Bisognin, Edecir Décio Cargnin
contributor_str_mv Vidigal Júnior, Guaracilei Maciel
Soares, Glória de Almeida
Andrade Filho, Hugo de
Mendes, Luiz Mota
Moreira, Edson Jorge Lima
dc.subject.por.fl_str_mv Odontologia
Implante dentário
Osseointegração
Infiltração dentária-Microbiologia
Implante dentário endoósseo
Perda óssea alveolar
Biomecânica
Estudos de avaliação
topic Odontologia
Implante dentário
Osseointegração
Infiltração dentária-Microbiologia
Implante dentário endoósseo
Perda óssea alveolar
Biomecânica
Estudos de avaliação
Cirurgia Buco-Maxilo-Facial
dc.subject.cnpq.fl_str_mv Cirurgia Buco-Maxilo-Facial
description The best fit between the implant and the prosthetic component appears to be important for periimplant homeostasis. Considering that the oral cavity is a naturally contaminated environment, it is important that the rehabilitation with implants does not promote the emergence of areas for microorganisms deposition that might cause periimplant diseases, thus compromising the long term osseointegration. The objectives of this study were: evaluate the misfit and bacterial infiltration at the implant-abutment interface of five implant systems with external hexagon, manufactured and marketed in Brazil, and to analyze the vertical and horizontal misfit according to the classification proposed by Kano, Binon e Curtis (2007). Eight samples of each implant system: Conexão®, Dentoflex®, Neodent®, SIN® and Titanium Fix® were tested. Initially, the sets were inoculated with 0.3μl of a suspension containing Escherichia coli and placed in culture medium for analysis of bacterial infiltration. The samples analyses were performed with 1, 2, 5, 7, and 14 days for assessment of contamination through the turbidity of the culture medium. After the microbiological analysis all, samples of each system were prepared for misfits evaluation through scanning electron microscopy (SEM), with 1000x and 3000x of magnification. The results showed that the bacterial infiltration was observed in 25% of samples of the Dentoflex® system. Misfit was observed in all systems. The misfit average values of each system were: Neodent® - 0.51μm ± 0.39; Dentoflex® - 1.44μm ± 0.73; Titanium Fix® - 1.88μm ± 1.28; SIN® - 2, 46μm ± 3.38; Conexão® - 2.68μm ± 3.02. According to the classification of Kano, Binon e Curtis (2007), it was observed a predominance types II and IV of misfit, respectively, horizontal and horizontal and vertical simultaneously. It was concluded that all groups showed misfit at least in one of the points evaluated along the interface, and this increase is not the only determining factor for the occurrence of bacterial infiltration. It was also observed that the extent of the area of coupling interface should also take into account, which showed variations among the implant systems evaluated.
publishDate 2009
dc.date.issued.fl_str_mv 2009-09-24
dc.date.accessioned.fl_str_mv 2016-09-26T22:52:02Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/masterThesis
format masterThesis
status_str publishedVersion
dc.identifier.citation.fl_str_mv Bisognin, Edecir Décio Cargnin. Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes. 2009. 82 f. Dissertação( Programa de Pós-Graduação em Odontologia) - Universidade do Grande Rio, Duque de Caxias.
dc.identifier.uri.fl_str_mv http://localhost:8080/tede/handle/tede/60
identifier_str_mv Bisognin, Edecir Décio Cargnin. Avaliação da adaptação e capacidade de infiltração microbiana na interface implante-pilar protético em diferentes sistemas de implantes. 2009. 82 f. Dissertação( Programa de Pós-Graduação em Odontologia) - Universidade do Grande Rio, Duque de Caxias.
url http://localhost:8080/tede/handle/tede/60
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade do Grande Rio
dc.publisher.program.fl_str_mv Programa de Pós-Graduação em Odontologia
dc.publisher.initials.fl_str_mv UNIGRANRIO
dc.publisher.country.fl_str_mv Brasil
dc.publisher.department.fl_str_mv Unigranrio::Odontologia
publisher.none.fl_str_mv Universidade do Grande Rio
dc.source.none.fl_str_mv reponame:Biblioteca Digital de Teses e Dissertações da UNIGRANRIO
instname:Universidade do Grande Rio (UNIGRANRIO)
instacron:UNIGRANRIO
instname_str Universidade do Grande Rio (UNIGRANRIO)
instacron_str UNIGRANRIO
institution UNIGRANRIO
reponame_str Biblioteca Digital de Teses e Dissertações da UNIGRANRIO
collection Biblioteca Digital de Teses e Dissertações da UNIGRANRIO
bitstream.url.fl_str_mv http://localhost:8080/tede/bitstream/tede/60/1/license.txt
http://localhost:8080/tede/bitstream/tede/60/2/Disserta%C3%A7%C3%A3o_Edecir+D%C3%A9cio.pdf
bitstream.checksum.fl_str_mv 4a50535e8405f611398e6e2d408dbd1b
b6fe4d7fc9719a0353728f8149806bc6
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Biblioteca Digital de Teses e Dissertações da UNIGRANRIO - Universidade do Grande Rio (UNIGRANRIO)
repository.mail.fl_str_mv repositorio@instituicao.br||repositorio@instituicao.br
_version_ 1811013916659023872