Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability

Detalhes bibliográficos
Autor(a) principal: Valente,Mariana Lima da Costa
Data de Publicação: 2019
Outros Autores: Castro,Denise Tornavoi de, Shimano,Antônio Carlos, Reis,Andréa Cândido dos
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Dental Journal
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402019000100047
Resumo: Abstract The purpose of thisin vitrostudy was to evaluate the influence of a new proposal of implant design and surgical protocol on primary stability in different bone densities. Four groups were tested (n=9): G1 - tapered, cone morse, Ø 4.3 mm x 10 mm in length (Alvim CM); G2 - experimental tapered; G3 - cylindrical, cone morse, Ø 4.0 mm x 11 mm in length (Titamax CM) and G4 - experimental cylindrical. The experimental implants were obtained from a design change in the respective commercial models. The insertion was performed in polyurethane (PU) blocks 0.24 g/cm3(20 pcf) and 0.64 g/cm3(40 pcf), according to different surgical protocols. The primary stability was measured by means of insertion torque (IT) and pullout test. Data were analyzed by ANOVA, Tukey’s test (α=0.05) and Pearson’s correlation. For IT and pullout, conventional and experimental implants showed no difference between them when inserted in the 20 pcf PU (p>0.05). In the 40 pcf PU, the modified implants exhibited greater IT (p<0.05) and lower pullout (p<0.05) compared to the respective conventional models. The implant design tested associated with the surgical protocol, positively influenced primary stability in higher density bones.
id FUNORP-1_5cb02757522b2cc446eec2d820b97a9b
oai_identifier_str oai:scielo:S0103-64402019000100047
network_acronym_str FUNORP-1
network_name_str Brazilian Dental Journal
repository_id_str
spelling Influence of an Alternative Implant Design and Surgical Protocol on Primary Stabilitydental implantsdesignbone densitiessurgical protocoltorqueAbstract The purpose of thisin vitrostudy was to evaluate the influence of a new proposal of implant design and surgical protocol on primary stability in different bone densities. Four groups were tested (n=9): G1 - tapered, cone morse, Ø 4.3 mm x 10 mm in length (Alvim CM); G2 - experimental tapered; G3 - cylindrical, cone morse, Ø 4.0 mm x 11 mm in length (Titamax CM) and G4 - experimental cylindrical. The experimental implants were obtained from a design change in the respective commercial models. The insertion was performed in polyurethane (PU) blocks 0.24 g/cm3(20 pcf) and 0.64 g/cm3(40 pcf), according to different surgical protocols. The primary stability was measured by means of insertion torque (IT) and pullout test. Data were analyzed by ANOVA, Tukey’s test (α=0.05) and Pearson’s correlation. For IT and pullout, conventional and experimental implants showed no difference between them when inserted in the 20 pcf PU (p>0.05). In the 40 pcf PU, the modified implants exhibited greater IT (p<0.05) and lower pullout (p<0.05) compared to the respective conventional models. The implant design tested associated with the surgical protocol, positively influenced primary stability in higher density bones.Fundação Odontológica de Ribeirão Preto2019-02-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402019000100047Brazilian Dental Journal v.30 n.1 2019reponame:Brazilian Dental Journalinstname:Fundação Odontológica de Ribeirão Preto (FUNORP)instacron:FUNORP10.1590/0103-6440201902324info:eu-repo/semantics/openAccessValente,Mariana Lima da CostaCastro,Denise Tornavoi deShimano,Antônio CarlosReis,Andréa Cândido doseng2020-01-27T00:00:00Zoai:scielo:S0103-64402019000100047Revistahttps://www.scielo.br/j/bdj/https://old.scielo.br/oai/scielo-oai.phpbdj@forp.usp.br||sergio@fosjc.unesp.br1806-47600103-6440opendoar:2020-01-27T00:00Brazilian Dental Journal - Fundação Odontológica de Ribeirão Preto (FUNORP)false
dc.title.none.fl_str_mv Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
title Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
spellingShingle Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
Valente,Mariana Lima da Costa
dental implants
design
bone densities
surgical protocol
torque
title_short Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
title_full Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
title_fullStr Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
title_full_unstemmed Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
title_sort Influence of an Alternative Implant Design and Surgical Protocol on Primary Stability
author Valente,Mariana Lima da Costa
author_facet Valente,Mariana Lima da Costa
Castro,Denise Tornavoi de
Shimano,Antônio Carlos
Reis,Andréa Cândido dos
author_role author
author2 Castro,Denise Tornavoi de
Shimano,Antônio Carlos
Reis,Andréa Cândido dos
author2_role author
author
author
dc.contributor.author.fl_str_mv Valente,Mariana Lima da Costa
Castro,Denise Tornavoi de
Shimano,Antônio Carlos
Reis,Andréa Cândido dos
dc.subject.por.fl_str_mv dental implants
design
bone densities
surgical protocol
torque
topic dental implants
design
bone densities
surgical protocol
torque
description Abstract The purpose of thisin vitrostudy was to evaluate the influence of a new proposal of implant design and surgical protocol on primary stability in different bone densities. Four groups were tested (n=9): G1 - tapered, cone morse, Ø 4.3 mm x 10 mm in length (Alvim CM); G2 - experimental tapered; G3 - cylindrical, cone morse, Ø 4.0 mm x 11 mm in length (Titamax CM) and G4 - experimental cylindrical. The experimental implants were obtained from a design change in the respective commercial models. The insertion was performed in polyurethane (PU) blocks 0.24 g/cm3(20 pcf) and 0.64 g/cm3(40 pcf), according to different surgical protocols. The primary stability was measured by means of insertion torque (IT) and pullout test. Data were analyzed by ANOVA, Tukey’s test (α=0.05) and Pearson’s correlation. For IT and pullout, conventional and experimental implants showed no difference between them when inserted in the 20 pcf PU (p>0.05). In the 40 pcf PU, the modified implants exhibited greater IT (p<0.05) and lower pullout (p<0.05) compared to the respective conventional models. The implant design tested associated with the surgical protocol, positively influenced primary stability in higher density bones.
publishDate 2019
dc.date.none.fl_str_mv 2019-02-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402019000100047
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402019000100047
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/0103-6440201902324
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Fundação Odontológica de Ribeirão Preto
publisher.none.fl_str_mv Fundação Odontológica de Ribeirão Preto
dc.source.none.fl_str_mv Brazilian Dental Journal v.30 n.1 2019
reponame:Brazilian Dental Journal
instname:Fundação Odontológica de Ribeirão Preto (FUNORP)
instacron:FUNORP
instname_str Fundação Odontológica de Ribeirão Preto (FUNORP)
instacron_str FUNORP
institution FUNORP
reponame_str Brazilian Dental Journal
collection Brazilian Dental Journal
repository.name.fl_str_mv Brazilian Dental Journal - Fundação Odontológica de Ribeirão Preto (FUNORP)
repository.mail.fl_str_mv bdj@forp.usp.br||sergio@fosjc.unesp.br
_version_ 1754204095340609536