Errors associated with IOLmaster biometry as a function of internal ocular dimensions

Detalhes bibliográficos
Autor(a) principal: Ribeiro, Miguel Faria
Data de Publicação: 2014
Outros Autores: Ferreira, Daniela Lopes, Gil, Norberto Lopez, Jorge, Jorge, González-Méijome, José Manuel
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/1822/37098
Resumo: To evaluate the error in the estimation of axial length (AL) with the IOLMaster partial coherence interferometry (PCI) biometer and obtain a correction factor that varies as a function of AL and crystalline lens thickness (LT). Methods: Optical simulations were produced for theoretical eyes using Zemax-EE software. Thirty-three combinations including eleven different AL (from 20 mm to 30 mm in 1 mm steps) and three different LT (3.6 mm, 4.2 mm and 4.8 mm) were used. Errors were obtained comparing the AL measured for a constant equivalent refractive index of 1.3549 and for the actual combinations of indices and intra-ocular dimensions of LT and AL in each model eye. Results: In the range from 20 mm to 30 mm AL and 3.6---4.8 mm LT, the instrument measurements yielded an error between −0.043 mm and +0.089 mm. Regression analyses for the three LT condition were combined in order to derive a correction factor as a function of the instrument measured AL for each combination of AL and LT in the theoretical eye. Conclusions: The assumption of a single ‘‘average’’ refractive index in the estimation of AL by the IOLMaster PCI biometer only induces very small errors in a wide range of combinations of ocular dimensions. Even so, the accurate estimation of those errors may help to improve accuracy of intra-ocular lens calculations through exact ray tracing, particularly in longer eyes and eyes with thicker or thinner crystalline lenses.
id RCAP_b21a2c8304d2721be5b5a180961d3da2
oai_identifier_str oai:repositorium.sdum.uminho.pt:1822/37098
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str
spelling Errors associated with IOLmaster biometry as a function of internal ocular dimensionsIntra-ocular lens calculationIOL calculationAxial length measurementTo evaluate the error in the estimation of axial length (AL) with the IOLMaster partial coherence interferometry (PCI) biometer and obtain a correction factor that varies as a function of AL and crystalline lens thickness (LT). Methods: Optical simulations were produced for theoretical eyes using Zemax-EE software. Thirty-three combinations including eleven different AL (from 20 mm to 30 mm in 1 mm steps) and three different LT (3.6 mm, 4.2 mm and 4.8 mm) were used. Errors were obtained comparing the AL measured for a constant equivalent refractive index of 1.3549 and for the actual combinations of indices and intra-ocular dimensions of LT and AL in each model eye. Results: In the range from 20 mm to 30 mm AL and 3.6---4.8 mm LT, the instrument measurements yielded an error between −0.043 mm and +0.089 mm. Regression analyses for the three LT condition were combined in order to derive a correction factor as a function of the instrument measured AL for each combination of AL and LT in the theoretical eye. Conclusions: The assumption of a single ‘‘average’’ refractive index in the estimation of AL by the IOLMaster PCI biometer only induces very small errors in a wide range of combinations of ocular dimensions. Even so, the accurate estimation of those errors may help to improve accuracy of intra-ocular lens calculations through exact ray tracing, particularly in longer eyes and eyes with thicker or thinner crystalline lenses.(undefined)Elsevier DoymaUniversidade do MinhoRibeiro, Miguel FariaFerreira, Daniela LopesGil, Norberto LopezJorge, JorgeGonzález-Méijome, José Manuel20142014-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/1822/37098engFaria-Ribeiro, Miguel; Lopes-Ferreira, Daniela; López-Gil, Norberto; Jorge, Jorge; González-Méijome, José Manuel. Errors Associated with IOLMaster Biometry as a Function of Internal Ocular Dimensions, Journal of Optometry, 7, 2, 75-78, 2014.1888-429610.1016/j.optom.2013.12.00524766863info:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2023-07-21T12:15:44ZPortal AgregadorONG
dc.title.none.fl_str_mv Errors associated with IOLmaster biometry as a function of internal ocular dimensions
title Errors associated with IOLmaster biometry as a function of internal ocular dimensions
spellingShingle Errors associated with IOLmaster biometry as a function of internal ocular dimensions
Ribeiro, Miguel Faria
Intra-ocular lens calculation
IOL calculation
Axial length measurement
title_short Errors associated with IOLmaster biometry as a function of internal ocular dimensions
title_full Errors associated with IOLmaster biometry as a function of internal ocular dimensions
title_fullStr Errors associated with IOLmaster biometry as a function of internal ocular dimensions
title_full_unstemmed Errors associated with IOLmaster biometry as a function of internal ocular dimensions
title_sort Errors associated with IOLmaster biometry as a function of internal ocular dimensions
author Ribeiro, Miguel Faria
author_facet Ribeiro, Miguel Faria
Ferreira, Daniela Lopes
Gil, Norberto Lopez
Jorge, Jorge
González-Méijome, José Manuel
author_role author
author2 Ferreira, Daniela Lopes
Gil, Norberto Lopez
Jorge, Jorge
González-Méijome, José Manuel
author2_role author
author
author
author
dc.contributor.none.fl_str_mv Universidade do Minho
dc.contributor.author.fl_str_mv Ribeiro, Miguel Faria
Ferreira, Daniela Lopes
Gil, Norberto Lopez
Jorge, Jorge
González-Méijome, José Manuel
dc.subject.por.fl_str_mv Intra-ocular lens calculation
IOL calculation
Axial length measurement
topic Intra-ocular lens calculation
IOL calculation
Axial length measurement
description To evaluate the error in the estimation of axial length (AL) with the IOLMaster partial coherence interferometry (PCI) biometer and obtain a correction factor that varies as a function of AL and crystalline lens thickness (LT). Methods: Optical simulations were produced for theoretical eyes using Zemax-EE software. Thirty-three combinations including eleven different AL (from 20 mm to 30 mm in 1 mm steps) and three different LT (3.6 mm, 4.2 mm and 4.8 mm) were used. Errors were obtained comparing the AL measured for a constant equivalent refractive index of 1.3549 and for the actual combinations of indices and intra-ocular dimensions of LT and AL in each model eye. Results: In the range from 20 mm to 30 mm AL and 3.6---4.8 mm LT, the instrument measurements yielded an error between −0.043 mm and +0.089 mm. Regression analyses for the three LT condition were combined in order to derive a correction factor as a function of the instrument measured AL for each combination of AL and LT in the theoretical eye. Conclusions: The assumption of a single ‘‘average’’ refractive index in the estimation of AL by the IOLMaster PCI biometer only induces very small errors in a wide range of combinations of ocular dimensions. Even so, the accurate estimation of those errors may help to improve accuracy of intra-ocular lens calculations through exact ray tracing, particularly in longer eyes and eyes with thicker or thinner crystalline lenses.
publishDate 2014
dc.date.none.fl_str_mv 2014
2014-01-01T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/1822/37098
url http://hdl.handle.net/1822/37098
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Faria-Ribeiro, Miguel; Lopes-Ferreira, Daniela; López-Gil, Norberto; Jorge, Jorge; González-Méijome, José Manuel. Errors Associated with IOLMaster Biometry as a Function of Internal Ocular Dimensions, Journal of Optometry, 7, 2, 75-78, 2014.
1888-4296
10.1016/j.optom.2013.12.005
24766863
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Elsevier Doyma
publisher.none.fl_str_mv Elsevier Doyma
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv
repository.mail.fl_str_mv
_version_ 1777303717025939456