In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass

Detalhes bibliográficos
Autor(a) principal: Marupudi,Neelima
Data de Publicação: 2016
Outros Autores: Wang,Shigang, Canêo,Luiz Fernando, Jatene,Fabio Biscegli, Kunselman,Allen R., Undar,Akif
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Journal of Cardiovascular Surgery (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382016000500343
Resumo: Abstract Objective: Usually only FDA-approved oxygenators are subject of studies by the international scientific community. The objective of this study is to evaluate two types of neonatal membrane oxygenators in terms of transmembrane pressure gradient, hemodynamic energy transmission and gaseous microemboli capture in simulated cardiopulmonary bypass systems. Methods: We investigated the Braile Infant 1500 (Braile Biomédica, São José do Rio Preto, Brazil), an oxygenator commonly used in Brazilian operating rooms, and compared it to the Dideco Kids D100 (Sorin Group, Arvada, CO, USA), that is an FDA-approved and widely used model in the USA. Cardiopulmonary bypass circuits were primed with lactated Ringer's solution and packed red blood cells (Hematocrit 40%). Trials were conducted at flow rates of 500 ml/min and 700 ml/min at 35ºC and 25ºC. Real-time pressure and flow data were recorded using a custom-based data acquisition system. For gaseous microemboli testing, 5cc of air were manually injected into the venous line. Gaseous microemboli were recorded using the Emboli Detection and Classification Quantifier. Results: Braile Infant 1500 had a lower pressure drop (P<0.01) and a higher total hemodynamic energy delivered to the pseudopatient (P<0.01). However, there was a higher raw number of gaseous microemboli seen prior to oxygenator at lower temperatures with the Braile oxygenator compared to the Kids D100 (P<0.01). Conclusion: Braile Infant 1500 oxygenator had a better hemodynamic performance compared to the Dideco Kids D100 oxygenator. Braile had more gaseous microemboli detected at the pre-oxygenator site under hypothermia, but delivered a smaller percentage of air emboli to the pseudopatient than the Dideco oxygenator.
id SBCCV-1_868c7870fc426090fcf9f5f934f7727d
oai_identifier_str oai:scielo:S0102-76382016000500343
network_acronym_str SBCCV-1
network_name_str Brazilian Journal of Cardiovascular Surgery (Online)
repository_id_str
spelling In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary BypassCardiopulmonary BypassPediatricsOxygenators, MembraneAbstract Objective: Usually only FDA-approved oxygenators are subject of studies by the international scientific community. The objective of this study is to evaluate two types of neonatal membrane oxygenators in terms of transmembrane pressure gradient, hemodynamic energy transmission and gaseous microemboli capture in simulated cardiopulmonary bypass systems. Methods: We investigated the Braile Infant 1500 (Braile Biomédica, São José do Rio Preto, Brazil), an oxygenator commonly used in Brazilian operating rooms, and compared it to the Dideco Kids D100 (Sorin Group, Arvada, CO, USA), that is an FDA-approved and widely used model in the USA. Cardiopulmonary bypass circuits were primed with lactated Ringer's solution and packed red blood cells (Hematocrit 40%). Trials were conducted at flow rates of 500 ml/min and 700 ml/min at 35ºC and 25ºC. Real-time pressure and flow data were recorded using a custom-based data acquisition system. For gaseous microemboli testing, 5cc of air were manually injected into the venous line. Gaseous microemboli were recorded using the Emboli Detection and Classification Quantifier. Results: Braile Infant 1500 had a lower pressure drop (P<0.01) and a higher total hemodynamic energy delivered to the pseudopatient (P<0.01). However, there was a higher raw number of gaseous microemboli seen prior to oxygenator at lower temperatures with the Braile oxygenator compared to the Kids D100 (P<0.01). Conclusion: Braile Infant 1500 oxygenator had a better hemodynamic performance compared to the Dideco Kids D100 oxygenator. Braile had more gaseous microemboli detected at the pre-oxygenator site under hypothermia, but delivered a smaller percentage of air emboli to the pseudopatient than the Dideco oxygenator.Sociedade Brasileira de Cirurgia Cardiovascular2016-10-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382016000500343Brazilian Journal of Cardiovascular Surgery v.31 n.5 2016reponame:Brazilian Journal of Cardiovascular Surgery (Online)instname:Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)instacron:SBCCV10.5935/1678-9741.20160075info:eu-repo/semantics/openAccessMarupudi,NeelimaWang,ShigangCanêo,Luiz FernandoJatene,Fabio BiscegliKunselman,Allen R.Undar,Akifeng2016-12-07T00:00:00Zoai:scielo:S0102-76382016000500343Revistahttp://www.rbccv.org.br/https://old.scielo.br/oai/scielo-oai.php||rosangela.monteiro@incor.usp.br|| domingo@braile.com.br|| brandau@braile.com.br1678-97410102-7638opendoar:2016-12-07T00:00Brazilian Journal of Cardiovascular Surgery (Online) - Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)false
dc.title.none.fl_str_mv In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
spellingShingle In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
Marupudi,Neelima
Cardiopulmonary Bypass
Pediatrics
Oxygenators, Membrane
title_short In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_full In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_fullStr In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_full_unstemmed In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_sort In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
author Marupudi,Neelima
author_facet Marupudi,Neelima
Wang,Shigang
Canêo,Luiz Fernando
Jatene,Fabio Biscegli
Kunselman,Allen R.
Undar,Akif
author_role author
author2 Wang,Shigang
Canêo,Luiz Fernando
Jatene,Fabio Biscegli
Kunselman,Allen R.
Undar,Akif
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Marupudi,Neelima
Wang,Shigang
Canêo,Luiz Fernando
Jatene,Fabio Biscegli
Kunselman,Allen R.
Undar,Akif
dc.subject.por.fl_str_mv Cardiopulmonary Bypass
Pediatrics
Oxygenators, Membrane
topic Cardiopulmonary Bypass
Pediatrics
Oxygenators, Membrane
description Abstract Objective: Usually only FDA-approved oxygenators are subject of studies by the international scientific community. The objective of this study is to evaluate two types of neonatal membrane oxygenators in terms of transmembrane pressure gradient, hemodynamic energy transmission and gaseous microemboli capture in simulated cardiopulmonary bypass systems. Methods: We investigated the Braile Infant 1500 (Braile Biomédica, São José do Rio Preto, Brazil), an oxygenator commonly used in Brazilian operating rooms, and compared it to the Dideco Kids D100 (Sorin Group, Arvada, CO, USA), that is an FDA-approved and widely used model in the USA. Cardiopulmonary bypass circuits were primed with lactated Ringer's solution and packed red blood cells (Hematocrit 40%). Trials were conducted at flow rates of 500 ml/min and 700 ml/min at 35ºC and 25ºC. Real-time pressure and flow data were recorded using a custom-based data acquisition system. For gaseous microemboli testing, 5cc of air were manually injected into the venous line. Gaseous microemboli were recorded using the Emboli Detection and Classification Quantifier. Results: Braile Infant 1500 had a lower pressure drop (P<0.01) and a higher total hemodynamic energy delivered to the pseudopatient (P<0.01). However, there was a higher raw number of gaseous microemboli seen prior to oxygenator at lower temperatures with the Braile oxygenator compared to the Kids D100 (P<0.01). Conclusion: Braile Infant 1500 oxygenator had a better hemodynamic performance compared to the Dideco Kids D100 oxygenator. Braile had more gaseous microemboli detected at the pre-oxygenator site under hypothermia, but delivered a smaller percentage of air emboli to the pseudopatient than the Dideco oxygenator.
publishDate 2016
dc.date.none.fl_str_mv 2016-10-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382016000500343
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382016000500343
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.5935/1678-9741.20160075
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Cirurgia Cardiovascular
publisher.none.fl_str_mv Sociedade Brasileira de Cirurgia Cardiovascular
dc.source.none.fl_str_mv Brazilian Journal of Cardiovascular Surgery v.31 n.5 2016
reponame:Brazilian Journal of Cardiovascular Surgery (Online)
instname:Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)
instacron:SBCCV
instname_str Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)
instacron_str SBCCV
institution SBCCV
reponame_str Brazilian Journal of Cardiovascular Surgery (Online)
collection Brazilian Journal of Cardiovascular Surgery (Online)
repository.name.fl_str_mv Brazilian Journal of Cardiovascular Surgery (Online) - Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)
repository.mail.fl_str_mv ||rosangela.monteiro@incor.usp.br|| domingo@braile.com.br|| brandau@braile.com.br
_version_ 1752126599420968960