Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron

Detalhes bibliográficos
Autor(a) principal: Bortoleto,Gisele G.
Data de Publicação: 2004
Outros Autores: Macarovscha,Greice T., Cadore,Solange
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Journal of the Brazilian Chemical Society (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000200024
Resumo: A method for the determination of cadmium by flame atomic absorption spectrometry (FAAS) after its preconcentration onto a column containing silica gel modified with cupferron was developed. The pH, amount of adsorbent material, concentration of eluent, volume of sample and amount of Cd were optimized. The effect of several foreign ions was also investigated and showed that the retention of cadmium depended on the amounts of Zn(II) and Cu(II) present and that these interferences could be overcome by using a 0.05 mmol KI. An enrichment factor of up to 30 was obtained , the LOD was 0.5 µg L-1 (3sigma) and the LOQ was 2.0 µg L-1 (10sigma) with rsd of 1.1% (n = 10). The accuracy of the proposed method was ascertained by using certified reference material and the obtained result (3.93 ± 0.01 µg g-1) agrees with the certified value (4.15 ± 0.38 µg g-1). The determination of Cd in nail polish showed quantitative recoveries for the spiked samples. The proposed method is characterized by simplicity, efficiency and low cost.
id SBQ-2_f592bdeb2e6fcc701dece9e37097126d
oai_identifier_str oai:scielo:S0103-50532004000200024
network_acronym_str SBQ-2
network_name_str Journal of the Brazilian Chemical Society (Online)
repository_id_str
spelling Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferroncadmiumsolid-liquid extractioncupferronFAASA method for the determination of cadmium by flame atomic absorption spectrometry (FAAS) after its preconcentration onto a column containing silica gel modified with cupferron was developed. The pH, amount of adsorbent material, concentration of eluent, volume of sample and amount of Cd were optimized. The effect of several foreign ions was also investigated and showed that the retention of cadmium depended on the amounts of Zn(II) and Cu(II) present and that these interferences could be overcome by using a 0.05 mmol KI. An enrichment factor of up to 30 was obtained , the LOD was 0.5 µg L-1 (3sigma) and the LOQ was 2.0 µg L-1 (10sigma) with rsd of 1.1% (n = 10). The accuracy of the proposed method was ascertained by using certified reference material and the obtained result (3.93 ± 0.01 µg g-1) agrees with the certified value (4.15 ± 0.38 µg g-1). The determination of Cd in nail polish showed quantitative recoveries for the spiked samples. The proposed method is characterized by simplicity, efficiency and low cost.Sociedade Brasileira de Química2004-04-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000200024Journal of the Brazilian Chemical Society v.15 n.2 2004reponame:Journal of the Brazilian Chemical Society (Online)instname:Sociedade Brasileira de Química (SBQ)instacron:SBQ10.1590/S0103-50532004000200024info:eu-repo/semantics/openAccessBortoleto,Gisele G.Macarovscha,Greice T.Cadore,Solangeeng2004-06-30T00:00:00Zoai:scielo:S0103-50532004000200024Revistahttp://jbcs.sbq.org.brONGhttps://old.scielo.br/oai/scielo-oai.php||office@jbcs.sbq.org.br1678-47900103-5053opendoar:2004-06-30T00:00Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)false
dc.title.none.fl_str_mv Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
title Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
spellingShingle Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
Bortoleto,Gisele G.
cadmium
solid-liquid extraction
cupferron
FAAS
title_short Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
title_full Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
title_fullStr Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
title_full_unstemmed Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
title_sort Determination of cadmium by flame-atomic absorption spectrometry after preconcentration on silica gel modified with cupferron
author Bortoleto,Gisele G.
author_facet Bortoleto,Gisele G.
Macarovscha,Greice T.
Cadore,Solange
author_role author
author2 Macarovscha,Greice T.
Cadore,Solange
author2_role author
author
dc.contributor.author.fl_str_mv Bortoleto,Gisele G.
Macarovscha,Greice T.
Cadore,Solange
dc.subject.por.fl_str_mv cadmium
solid-liquid extraction
cupferron
FAAS
topic cadmium
solid-liquid extraction
cupferron
FAAS
description A method for the determination of cadmium by flame atomic absorption spectrometry (FAAS) after its preconcentration onto a column containing silica gel modified with cupferron was developed. The pH, amount of adsorbent material, concentration of eluent, volume of sample and amount of Cd were optimized. The effect of several foreign ions was also investigated and showed that the retention of cadmium depended on the amounts of Zn(II) and Cu(II) present and that these interferences could be overcome by using a 0.05 mmol KI. An enrichment factor of up to 30 was obtained , the LOD was 0.5 µg L-1 (3sigma) and the LOQ was 2.0 µg L-1 (10sigma) with rsd of 1.1% (n = 10). The accuracy of the proposed method was ascertained by using certified reference material and the obtained result (3.93 ± 0.01 µg g-1) agrees with the certified value (4.15 ± 0.38 µg g-1). The determination of Cd in nail polish showed quantitative recoveries for the spiked samples. The proposed method is characterized by simplicity, efficiency and low cost.
publishDate 2004
dc.date.none.fl_str_mv 2004-04-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000200024
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000200024
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0103-50532004000200024
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Química
publisher.none.fl_str_mv Sociedade Brasileira de Química
dc.source.none.fl_str_mv Journal of the Brazilian Chemical Society v.15 n.2 2004
reponame:Journal of the Brazilian Chemical Society (Online)
instname:Sociedade Brasileira de Química (SBQ)
instacron:SBQ
instname_str Sociedade Brasileira de Química (SBQ)
instacron_str SBQ
institution SBQ
reponame_str Journal of the Brazilian Chemical Society (Online)
collection Journal of the Brazilian Chemical Society (Online)
repository.name.fl_str_mv Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)
repository.mail.fl_str_mv ||office@jbcs.sbq.org.br
_version_ 1750318165724758016