Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers

Detalhes bibliográficos
Autor(a) principal: Marcelo Giannini
Data de Publicação: 2019
Outros Autores: Carolina Bosso André, Vanessa Cavalli Gobbo, Frederick Allen Rueggeberg
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UFMG
Texto Completo: http://hdl.handle.net/1843/45436
Resumo: This study measured and compared exitance irradiance and power of 4 commercial dental light-curing units (LCU) (Elipar S10, Elipar DeepCure-S, Corded VALO and Bluephase Style) using different types of radiometers. The devices used to analyze the LCU were classified as either handheld analog (Henry Schein, Spring, Demetron 100A, Demetron 100B and Demetron 200), handheld digital (Bluephase 1, Bluephase II, Coltolux, CureRite and Hilux), or laboratory instruments (Thermopile and Integrating Sphere). The laboratory instruments and the Bluephase II radiometer were also used to measure the LCU’s power (mW). The LCU’s were activated for 20 s (n=5). Data were analyzed using Kruskal-Wallis and Student-Newman-Keuls multiple comparison test (a=0.05). Among the LCU, the laboratory instruments presented different irradiance values, except for Corded VALO. The Coltolux and Hilux radiometers measured greater irradiance values compared to the laboratory instruments for the four LCUs tested. Within a given LCU, handheld analog units measured lower irradiance values, compared to handheld digital and laboratory instruments, except using the Spring radiometer for the Elipar S10 LCU. None of the handheld radiometers were able to measure similar irradiance values compared to laboratory instruments, except for Elipar S10 when comparing Bluephase 1 and Thermopile. Regarding power measurement, Bluephase II always presented the lowest values compared to the laboratory instruments. These findings suggest that the handheld radiometers utilized by practitioners (analog or digital) exhibit a wide range of irradiance values and may show lower outcomes compared to laboratory based instruments
id UFMG_bc67488da0eaebd8872920f6b714db96
oai_identifier_str oai:repositorio.ufmg.br:1843/45436
network_acronym_str UFMG
network_name_str Repositório Institucional da UFMG
repository_id_str
spelling 2022-09-24T16:39:25Z2022-09-24T16:39:25Z2019-07-2230439740310.1590/0103-644020190243018064760http://hdl.handle.net/1843/45436This study measured and compared exitance irradiance and power of 4 commercial dental light-curing units (LCU) (Elipar S10, Elipar DeepCure-S, Corded VALO and Bluephase Style) using different types of radiometers. The devices used to analyze the LCU were classified as either handheld analog (Henry Schein, Spring, Demetron 100A, Demetron 100B and Demetron 200), handheld digital (Bluephase 1, Bluephase II, Coltolux, CureRite and Hilux), or laboratory instruments (Thermopile and Integrating Sphere). The laboratory instruments and the Bluephase II radiometer were also used to measure the LCU’s power (mW). The LCU’s were activated for 20 s (n=5). Data were analyzed using Kruskal-Wallis and Student-Newman-Keuls multiple comparison test (a=0.05). Among the LCU, the laboratory instruments presented different irradiance values, except for Corded VALO. The Coltolux and Hilux radiometers measured greater irradiance values compared to the laboratory instruments for the four LCUs tested. Within a given LCU, handheld analog units measured lower irradiance values, compared to handheld digital and laboratory instruments, except using the Spring radiometer for the Elipar S10 LCU. None of the handheld radiometers were able to measure similar irradiance values compared to laboratory instruments, except for Elipar S10 when comparing Bluephase 1 and Thermopile. Regarding power measurement, Bluephase II always presented the lowest values compared to the laboratory instruments. These findings suggest that the handheld radiometers utilized by practitioners (analog or digital) exhibit a wide range of irradiance values and may show lower outcomes compared to laboratory based instrumentsEsse estudo mensurou e comparou a irradiância e a energia de quatro fotopolimerizadores comerciais (Elipar S10, Elipar DeepCure-S, Corded VALO e Bluephase Style) utilizando diferentes tipos de radiômetros. Os dispositivos utilizados para analisar os fotopolimerizadores foram classificados em portáteis analógicos (Henry Schein, Spring, Demetron 100A, Demetron 100B e Demetron 200), portáteis digitais (Bluephase 1, Bluephase II, Coltolux, CureRite e Hilux), ou instrumentos laboratoriais (Thermopile e Integrating Sphere). Os instrumentos laboratoriais e o radiômetro Bluephase II também foram utilizados para medir a energia dos fotopolimerizadores (mW). Os fotopolimerizadores foram ativados por 20 s (n=5). Os dados foram analisados utilizando Kruskal-Wallis e teste de Student-Newman-Keuls (a=0,05). Dentre os fotopolimerizadores, os instrumentos laboratoriais apresentaram diferentes valores de irradiância, exceto para o Corded VALO. O Coltolux e o Hilux mensuraram uma irradiância maior comparado aos instrumentos de laboratório para os quatro fotopolimerizadores testados. Para o mesmo fotopolimerizador, os radiômetros analógicos portáteis mensuraram menores valores de irradiância quando comparados aos digitais portáteis ou aos instrumentos laboratoriais, exceto quando utilizado o Spring para o Elipar S10. Nenhuma diferença foi observada entre os instrumentos laboratoriais na irradiância do Corded VALO. Para a medição da energia, nenhuma diferença foi observada usando Thermopile e a Integrating Sphere para o Corded VALO. Para os outros fotopolimerizadores, cada instrumento indicou uma energia diferente. Esses achados sugerem que radiômetros portáteis utilizados pelos cirurgiões-dentistas (analógico ou digital) exibem uma ampla gama de valores de irradiância e podem mostrar medidas inferiores comparados aos instrumentos de laboratórioCNPq - Conselho Nacional de Desenvolvimento Científico e TecnológicoengUniversidade Federal de Minas GeraisUFMGBrasilFAO - DEPARTAMENTO DE ODONTOLOGIA RESTAURADORABrazilian Dental JournalCuring lights dentalPolymerizationLED dental curing lightsPowerIrradianceAccuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometersinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articlehttps://www.scielo.br/j/bdj/a/rnfw5BWvTvJxZRTgLBjsYxx/?lang=enMarcelo GianniniCarolina Bosso AndréVanessa Cavalli GobboFrederick Allen Rueggebergapplication/pdfinfo:eu-repo/semantics/openAccessreponame:Repositório Institucional da UFMGinstname:Universidade Federal de Minas Gerais (UFMG)instacron:UFMGORIGINALAccuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers.pdfAccuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers.pdfapplication/pdf673496https://repositorio.ufmg.br/bitstream/1843/45436/2/Accuracy%20of%20irradiance%20and%20power%20of%20light-curing%20units%20measured%20with%20handheld%20or%20laboratory%20grade%20radiometers.pdf0c0e05a2163a36297264ebfe5507825cMD52LICENSELicense.txtLicense.txttext/plain; charset=utf-82042https://repositorio.ufmg.br/bitstream/1843/45436/1/License.txtfa505098d172de0bc8864fc1287ffe22MD511843/454362022-09-24 13:39:25.491oai:repositorio.ufmg.br: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Repositório de PublicaçõesPUBhttps://repositorio.ufmg.br/oaiopendoar:2022-09-24T16:39:25Repositório Institucional da UFMG - Universidade Federal de Minas Gerais (UFMG)false
dc.title.pt_BR.fl_str_mv Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
title Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
spellingShingle Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
Marcelo Giannini
Polymerization
LED dental curing lights
Power
Irradiance
Curing lights dental
title_short Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
title_full Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
title_fullStr Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
title_full_unstemmed Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
title_sort Accuracy of irradiance and power of light-curing units measured with handheld or laboratory grade radiometers
author Marcelo Giannini
author_facet Marcelo Giannini
Carolina Bosso André
Vanessa Cavalli Gobbo
Frederick Allen Rueggeberg
author_role author
author2 Carolina Bosso André
Vanessa Cavalli Gobbo
Frederick Allen Rueggeberg
author2_role author
author
author
dc.contributor.author.fl_str_mv Marcelo Giannini
Carolina Bosso André
Vanessa Cavalli Gobbo
Frederick Allen Rueggeberg
dc.subject.por.fl_str_mv Polymerization
LED dental curing lights
Power
Irradiance
topic Polymerization
LED dental curing lights
Power
Irradiance
Curing lights dental
dc.subject.other.pt_BR.fl_str_mv Curing lights dental
description This study measured and compared exitance irradiance and power of 4 commercial dental light-curing units (LCU) (Elipar S10, Elipar DeepCure-S, Corded VALO and Bluephase Style) using different types of radiometers. The devices used to analyze the LCU were classified as either handheld analog (Henry Schein, Spring, Demetron 100A, Demetron 100B and Demetron 200), handheld digital (Bluephase 1, Bluephase II, Coltolux, CureRite and Hilux), or laboratory instruments (Thermopile and Integrating Sphere). The laboratory instruments and the Bluephase II radiometer were also used to measure the LCU’s power (mW). The LCU’s were activated for 20 s (n=5). Data were analyzed using Kruskal-Wallis and Student-Newman-Keuls multiple comparison test (a=0.05). Among the LCU, the laboratory instruments presented different irradiance values, except for Corded VALO. The Coltolux and Hilux radiometers measured greater irradiance values compared to the laboratory instruments for the four LCUs tested. Within a given LCU, handheld analog units measured lower irradiance values, compared to handheld digital and laboratory instruments, except using the Spring radiometer for the Elipar S10 LCU. None of the handheld radiometers were able to measure similar irradiance values compared to laboratory instruments, except for Elipar S10 when comparing Bluephase 1 and Thermopile. Regarding power measurement, Bluephase II always presented the lowest values compared to the laboratory instruments. These findings suggest that the handheld radiometers utilized by practitioners (analog or digital) exhibit a wide range of irradiance values and may show lower outcomes compared to laboratory based instruments
publishDate 2019
dc.date.issued.fl_str_mv 2019-07-22
dc.date.accessioned.fl_str_mv 2022-09-24T16:39:25Z
dc.date.available.fl_str_mv 2022-09-24T16:39:25Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/1843/45436
dc.identifier.doi.pt_BR.fl_str_mv 10.1590/0103-6440201902430
dc.identifier.issn.pt_BR.fl_str_mv 18064760
identifier_str_mv 10.1590/0103-6440201902430
18064760
url http://hdl.handle.net/1843/45436
dc.language.iso.fl_str_mv eng
language eng
dc.relation.ispartof.pt_BR.fl_str_mv Brazilian Dental Journal
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade Federal de Minas Gerais
dc.publisher.initials.fl_str_mv UFMG
dc.publisher.country.fl_str_mv Brasil
dc.publisher.department.fl_str_mv FAO - DEPARTAMENTO DE ODONTOLOGIA RESTAURADORA
publisher.none.fl_str_mv Universidade Federal de Minas Gerais
dc.source.none.fl_str_mv reponame:Repositório Institucional da UFMG
instname:Universidade Federal de Minas Gerais (UFMG)
instacron:UFMG
instname_str Universidade Federal de Minas Gerais (UFMG)
instacron_str UFMG
institution UFMG
reponame_str Repositório Institucional da UFMG
collection Repositório Institucional da UFMG
bitstream.url.fl_str_mv https://repositorio.ufmg.br/bitstream/1843/45436/2/Accuracy%20of%20irradiance%20and%20power%20of%20light-curing%20units%20measured%20with%20handheld%20or%20laboratory%20grade%20radiometers.pdf
https://repositorio.ufmg.br/bitstream/1843/45436/1/License.txt
bitstream.checksum.fl_str_mv 0c0e05a2163a36297264ebfe5507825c
fa505098d172de0bc8864fc1287ffe22
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Repositório Institucional da UFMG - Universidade Federal de Minas Gerais (UFMG)
repository.mail.fl_str_mv
_version_ 1801676887941971968