Stress distribution of different implant connections associated with multiple implant-supported prostheses

Detalhes bibliográficos
Autor(a) principal: Goiato, Marcelo Coelho [UNESP]
Data de Publicação: 2018
Outros Autores: Sarauza Arsufi, Guilherme [UNESP], de Medeiros, Rodrigo Antonio, Pesqueira, Aldiéris Alves [UNESP], da Silva, Emily Vivianne Freitas [UNESP], Sonego, Mariana Vilela [UNESP], dos Santos, Daniela Micheline [UNESP]
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://dx.doi.org/10.1080/03091902.2018.1513575
http://hdl.handle.net/11449/188110
Resumo: In some clinical situations, the clinician may encounter previously installed implants that should be associated with other implants for a proper rehabilitation. The aim of this study was to evaluate the biomechanical behaviour of a multiple prosthesis joined by different implant connections using photoelasticity. Photoelastic models with a screwed fixed prosthesis supported by implants with different connection systems (Morse taper, external hexagon, internal hexagon, and Flexcone), and different combinations among them, were fabricated. Each assembly was placed in a circular polariscope, and axial and oblique (45°) loads of 100 N were applied on the occlusal surface of the crowns. The fringe patterns were photographed and the analysis was performed by counting the number of high-intensity fringes and also according to the stress distribution region where they appeared. Among implants of the same connection, the external connections obtained a greater number of high intensity fringes when compared to the internal connections. From the biomechanical point of view, the association between different types of connections obtained positive results.
id UNSP_b6d5cd6bf1a213683e478cc649435a30
oai_identifier_str oai:repositorio.unesp.br:11449/188110
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Stress distribution of different implant connections associated with multiple implant-supported prosthesesDental implantsdental prosthesisimplant-supportedphotoelastic analysisIn some clinical situations, the clinician may encounter previously installed implants that should be associated with other implants for a proper rehabilitation. The aim of this study was to evaluate the biomechanical behaviour of a multiple prosthesis joined by different implant connections using photoelasticity. Photoelastic models with a screwed fixed prosthesis supported by implants with different connection systems (Morse taper, external hexagon, internal hexagon, and Flexcone), and different combinations among them, were fabricated. Each assembly was placed in a circular polariscope, and axial and oblique (45°) loads of 100 N were applied on the occlusal surface of the crowns. The fringe patterns were photographed and the analysis was performed by counting the number of high-intensity fringes and also according to the stress distribution region where they appeared. Among implants of the same connection, the external connections obtained a greater number of high intensity fringes when compared to the internal connections. From the biomechanical point of view, the association between different types of connections obtained positive results.Department of Dental Materials and Prosthodontics Aracatuba Dental School UNESP Sao Paulo State UniversityDepartment of Dental Materials and Prosthodontics UnB University of BrasiliaDepartment of Dental Materials and Prosthodontics Aracatuba Dental School UNESP Sao Paulo State UniversityUniversidade Estadual Paulista (Unesp)University of BrasiliaGoiato, Marcelo Coelho [UNESP]Sarauza Arsufi, Guilherme [UNESP]de Medeiros, Rodrigo AntonioPesqueira, Aldiéris Alves [UNESP]da Silva, Emily Vivianne Freitas [UNESP]Sonego, Mariana Vilela [UNESP]dos Santos, Daniela Micheline [UNESP]2019-10-06T15:57:37Z2019-10-06T15:57:37Z2018-07-04info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/article359-367http://dx.doi.org/10.1080/03091902.2018.1513575Journal of Medical Engineering and Technology, v. 42, n. 5, p. 359-367, 2018.1464-522X0309-1902http://hdl.handle.net/11449/18811010.1080/03091902.2018.15135752-s2.0-850539106459719883814872582Scopusreponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengJournal of Medical Engineering and Technologyinfo:eu-repo/semantics/openAccess2021-10-22T23:59:55Zoai:repositorio.unesp.br:11449/188110Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462021-10-22T23:59:55Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Stress distribution of different implant connections associated with multiple implant-supported prostheses
title Stress distribution of different implant connections associated with multiple implant-supported prostheses
spellingShingle Stress distribution of different implant connections associated with multiple implant-supported prostheses
Goiato, Marcelo Coelho [UNESP]
Dental implants
dental prosthesis
implant-supported
photoelastic analysis
title_short Stress distribution of different implant connections associated with multiple implant-supported prostheses
title_full Stress distribution of different implant connections associated with multiple implant-supported prostheses
title_fullStr Stress distribution of different implant connections associated with multiple implant-supported prostheses
title_full_unstemmed Stress distribution of different implant connections associated with multiple implant-supported prostheses
title_sort Stress distribution of different implant connections associated with multiple implant-supported prostheses
author Goiato, Marcelo Coelho [UNESP]
author_facet Goiato, Marcelo Coelho [UNESP]
Sarauza Arsufi, Guilherme [UNESP]
de Medeiros, Rodrigo Antonio
Pesqueira, Aldiéris Alves [UNESP]
da Silva, Emily Vivianne Freitas [UNESP]
Sonego, Mariana Vilela [UNESP]
dos Santos, Daniela Micheline [UNESP]
author_role author
author2 Sarauza Arsufi, Guilherme [UNESP]
de Medeiros, Rodrigo Antonio
Pesqueira, Aldiéris Alves [UNESP]
da Silva, Emily Vivianne Freitas [UNESP]
Sonego, Mariana Vilela [UNESP]
dos Santos, Daniela Micheline [UNESP]
author2_role author
author
author
author
author
author
dc.contributor.none.fl_str_mv Universidade Estadual Paulista (Unesp)
University of Brasilia
dc.contributor.author.fl_str_mv Goiato, Marcelo Coelho [UNESP]
Sarauza Arsufi, Guilherme [UNESP]
de Medeiros, Rodrigo Antonio
Pesqueira, Aldiéris Alves [UNESP]
da Silva, Emily Vivianne Freitas [UNESP]
Sonego, Mariana Vilela [UNESP]
dos Santos, Daniela Micheline [UNESP]
dc.subject.por.fl_str_mv Dental implants
dental prosthesis
implant-supported
photoelastic analysis
topic Dental implants
dental prosthesis
implant-supported
photoelastic analysis
description In some clinical situations, the clinician may encounter previously installed implants that should be associated with other implants for a proper rehabilitation. The aim of this study was to evaluate the biomechanical behaviour of a multiple prosthesis joined by different implant connections using photoelasticity. Photoelastic models with a screwed fixed prosthesis supported by implants with different connection systems (Morse taper, external hexagon, internal hexagon, and Flexcone), and different combinations among them, were fabricated. Each assembly was placed in a circular polariscope, and axial and oblique (45°) loads of 100 N were applied on the occlusal surface of the crowns. The fringe patterns were photographed and the analysis was performed by counting the number of high-intensity fringes and also according to the stress distribution region where they appeared. Among implants of the same connection, the external connections obtained a greater number of high intensity fringes when compared to the internal connections. From the biomechanical point of view, the association between different types of connections obtained positive results.
publishDate 2018
dc.date.none.fl_str_mv 2018-07-04
2019-10-06T15:57:37Z
2019-10-06T15:57:37Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://dx.doi.org/10.1080/03091902.2018.1513575
Journal of Medical Engineering and Technology, v. 42, n. 5, p. 359-367, 2018.
1464-522X
0309-1902
http://hdl.handle.net/11449/188110
10.1080/03091902.2018.1513575
2-s2.0-85053910645
9719883814872582
url http://dx.doi.org/10.1080/03091902.2018.1513575
http://hdl.handle.net/11449/188110
identifier_str_mv Journal of Medical Engineering and Technology, v. 42, n. 5, p. 359-367, 2018.
1464-522X
0309-1902
10.1080/03091902.2018.1513575
2-s2.0-85053910645
9719883814872582
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Journal of Medical Engineering and Technology
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 359-367
dc.source.none.fl_str_mv Scopus
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1792962049554251776