Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos

Bibliographic Details
Main Author: Ricci, Cláudia Lizandra
Publication Date: 2015
Format: Master thesis
Language: por
Source: Biblioteca Digital de Teses e Dissertações da UNOESTE
Download full: http://bdtd.unoeste.br:8080/tede/handle/tede/314
Summary: The objective of this study was to measurement of the intraocular pressure (IOP) with Kowa HA-2 tonometer in cats analyzing the calibration, the accuracy and the validation of ambulatory clinical use. For calibration the post-mortem study was accomplished in 10 healthy eyes of 5 cats comparing the ocular manometry with the values of the IOP checked with the tonometer. For evaluation of the accuracy an in vivo study was accomplished in 20 healthy eyes of 10 anesthetized cats being compared the ocular manometry with the IOP obtained with the tonometer. For validation of the ambulatory clinical study of the IOP measurement was accomplished in 78 eyes of 39 healthy cats, in 7 eyes with clinical signs of glaucoma and in 20 eyes with clinical signs of uveitis. The correlation coefficient (r²) between the manometer and the tonometer was 0.993 and the equation of lineal regression was y=0.0915x+0.0878 in postmortem study. In the in vivo study the medium values of IOP in the manometry were 15.6±1.1 mmHg and in the tonometry were 15.5±1.2 mmHg, there was no statistics significant difference between the manometry and the tonometry. In the ambulatory clinical study with healthy cats the medium values of IOP with the tonometer were 15.0±1.5 mmHg, in the eyes with clinical signs of glaucoma were 38,4±8,1 mmHg and in the eyes with clinical signs of uveitis were of 10,4±2,0 mmHg. Therefore, there was a satisfactory correlation and accuracy between the IOP values obtained by direct ocular manometry and the tonometer in question. In the ambulatory clinical study the IOP values obtained with the tonometer were compatible for animals with healthy eyes and with clinical signs of glaucoma and uveitis. So, we can conclude that the Kowa HA-2 tonometer can be used in the routine ophthalmic examination, as it is a practical method for IOP measurement in cats.
id UOES_7eab4a90335848cb6175cc2fbfd530a6
oai_identifier_str oai:bdtd.unoeste.br:tede/314
network_acronym_str UOES
network_name_str Biblioteca Digital de Teses e Dissertações da UNOESTE
repository_id_str
spelling Andrade, Silvia Maria Caldeira FrancoCPF:10513528865http://lattes.cnpq.br/1455954245374682Andrade, Alexandre Lima deCPF:10837297866http://lattes.cnpq.br/6404176495306171Laposy, Cecília BragaCPF:11651891869CPF:36934863879http://lattes.cnpq.br/0010548031240899Ricci, Cláudia Lizandra2016-01-26T18:55:43Z2015-05-272015-03-17RICCI, Cláudia Lizandra. Use of portable applanation tonometer Kowa HA-2 in the measurement of intraocular pressure in cats. 2015. 32 f. Dissertação (Mestrado em Ciências Agrárias) - Universidade do Oeste Paulista, Presidente Prudente, 2015.http://bdtd.unoeste.br:8080/tede/handle/tede/314The objective of this study was to measurement of the intraocular pressure (IOP) with Kowa HA-2 tonometer in cats analyzing the calibration, the accuracy and the validation of ambulatory clinical use. For calibration the post-mortem study was accomplished in 10 healthy eyes of 5 cats comparing the ocular manometry with the values of the IOP checked with the tonometer. For evaluation of the accuracy an in vivo study was accomplished in 20 healthy eyes of 10 anesthetized cats being compared the ocular manometry with the IOP obtained with the tonometer. For validation of the ambulatory clinical study of the IOP measurement was accomplished in 78 eyes of 39 healthy cats, in 7 eyes with clinical signs of glaucoma and in 20 eyes with clinical signs of uveitis. The correlation coefficient (r²) between the manometer and the tonometer was 0.993 and the equation of lineal regression was y=0.0915x+0.0878 in postmortem study. In the in vivo study the medium values of IOP in the manometry were 15.6±1.1 mmHg and in the tonometry were 15.5±1.2 mmHg, there was no statistics significant difference between the manometry and the tonometry. In the ambulatory clinical study with healthy cats the medium values of IOP with the tonometer were 15.0±1.5 mmHg, in the eyes with clinical signs of glaucoma were 38,4±8,1 mmHg and in the eyes with clinical signs of uveitis were of 10,4±2,0 mmHg. Therefore, there was a satisfactory correlation and accuracy between the IOP values obtained by direct ocular manometry and the tonometer in question. In the ambulatory clinical study the IOP values obtained with the tonometer were compatible for animals with healthy eyes and with clinical signs of glaucoma and uveitis. So, we can conclude that the Kowa HA-2 tonometer can be used in the routine ophthalmic examination, as it is a practical method for IOP measurement in cats.O objetivo deste estudo foi mensurar a pressão intraocular (PIO) com o uso do tonômetro Kowa HA-2 em gatos, analisando a calibração, a acurácia e a validação do seu uso clínico ambulatorial. Para calibração foi realizado o estudo post-mortem em 10 olhos sadios de 5 gatos comparando a manometria ocular com os valores da PIO aferida com o tonômetro. Para avaliação de sua acurácia foi realizado um estudo in vivo em 20 olhos sadios de 10 gatos anestesiados comparando-se a manometria ocular com a PIO obtida com o tonômetro. Para validação do seu uso clínico ambulatorial foi realizado um estudo da mensuração da PIO em 78 olhos sadios de 39 gatos, em 7 olhos com sinais clínicos de glaucoma e em 20 olhos com sinais clínicos de uveíte. O coeficiente de correlação (r²) entre o manômetro e o tonômetro Kowa HA-2 foi de 0,993 e a equação de regressão linear foi y=0,0915x+0,0878 no estudo postmortem. No estudo in vivo, os valores médios de PIO na manometria foram de 15,6±1,1 mmHg e na tonometria foram de 15,5±1,2 mmHg não havendo diferença estatística significativa entre a manometria e a tonometria. No estudo ambulatorial com os gatos sadios, os valores médios de PIO com o tonômetro Kowa HA-2 foram 15,0±1,5 mmHg, nos olhos com sinais clínicos de glaucoma foram 38,4±8,1 mmHg e nos olhos com sinais clínicos de uveíte foram 10,4±2,0 mmHg. Houve, portanto, correlação e acurácia satisfatória entre os valores de PIO com a manometria e o tonômetro em questão. No estudo ambulatorial os valores de PIO obtidos com o tonômetro foram compatíveis para animais com olhos sadios e com sinais clínicos de glaucoma e uveíte. Desta maneira, podemos concluir que o tonômetro Kowa HA-2 pode ser empregado no exame oftálmico de rotina, pois trata-se de um método prático na aferição da PIO em gatos.Made available in DSpace on 2016-01-26T18:55:43Z (GMT). No. of bitstreams: 1 Claudia Lizandra Ricci.pdf: 291013 bytes, checksum: b9dccb042e091da74b7d28825ae3639d (MD5) Previous issue date: 2015-03-17application/pdfhttp://bdtd.unoeste.br:8080/tede/retrieve/800/Claudia%20Lizandra%20Ricci.pdf.jpgporUniversidade do Oeste PaulistaMestrado em Ciência AnimalUNOESTEBRCiências Agráriastonômetro de aplanaçãoKowa HA-2manometria ocularpressão intraoculargatosapplanation tonometerKowa HA-2ocular manometryintraocular pressurecatsCNPQ::CIENCIAS AGRARIAS::MEDICINA VETERINARIAUso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatosUse of portable applanation tonometer Kowa HA-2 in the measurement of intraocular pressure in catsinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/masterThesis1886725639743614135500info:eu-repo/semantics/openAccessreponame:Biblioteca Digital de Teses e Dissertações da UNOESTEinstname:Universidade do Oeste Paulista (UNOESTE)instacron:UNOESTEORIGINALClaudia Lizandra Ricci.pdfapplication/pdf291013http://bdtd.unoeste.br:8080/tede/bitstream/tede/314/1/Claudia+Lizandra+Ricci.pdfb9dccb042e091da74b7d28825ae3639dMD51THUMBNAILClaudia Lizandra Ricci.pdf.jpgClaudia Lizandra Ricci.pdf.jpgimage/jpeg2179http://bdtd.unoeste.br:8080/tede/bitstream/tede/314/2/Claudia+Lizandra+Ricci.pdf.jpg01a3bf5869d25756f643066aef98b98bMD52tede/3142016-01-27 01:03:46.052oai:bdtd.unoeste.br:tede/314Biblioteca Digital de Teses e Dissertaçõeshttp://bdtd.unoeste.br:8080/jspui/PUBhttp://bdtd.unoeste.br:8080/oai/requestbdtd@unoeste.bropendoar:2016-01-27T03:03:46Biblioteca Digital de Teses e Dissertações da UNOESTE - Universidade do Oeste Paulista (UNOESTE)false
dc.title.por.fl_str_mv Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
dc.title.alternative.eng.fl_str_mv Use of portable applanation tonometer Kowa HA-2 in the measurement of intraocular pressure in cats
title Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
spellingShingle Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
Ricci, Cláudia Lizandra
tonômetro de aplanação
Kowa HA-2
manometria ocular
pressão intraocular
gatos
applanation tonometer
Kowa HA-2
ocular manometry
intraocular pressure
cats
CNPQ::CIENCIAS AGRARIAS::MEDICINA VETERINARIA
title_short Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
title_full Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
title_fullStr Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
title_full_unstemmed Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
title_sort Uso do tonômetro de aplanação portátil Kowa HA-2 na mensuração da pressão intraocular em gatos
author Ricci, Cláudia Lizandra
author_facet Ricci, Cláudia Lizandra
author_role author
dc.contributor.advisor1.fl_str_mv Andrade, Silvia Maria Caldeira Franco
dc.contributor.advisor1ID.fl_str_mv CPF:10513528865
dc.contributor.advisor1Lattes.fl_str_mv http://lattes.cnpq.br/1455954245374682
dc.contributor.referee1.fl_str_mv Andrade, Alexandre Lima de
dc.contributor.referee1ID.fl_str_mv CPF:10837297866
dc.contributor.referee1Lattes.fl_str_mv http://lattes.cnpq.br/6404176495306171
dc.contributor.referee2.fl_str_mv Laposy, Cecília Braga
dc.contributor.referee2ID.fl_str_mv CPF:11651891869
dc.contributor.authorID.fl_str_mv CPF:36934863879
dc.contributor.authorLattes.fl_str_mv http://lattes.cnpq.br/0010548031240899
dc.contributor.author.fl_str_mv Ricci, Cláudia Lizandra
contributor_str_mv Andrade, Silvia Maria Caldeira Franco
Andrade, Alexandre Lima de
Laposy, Cecília Braga
dc.subject.por.fl_str_mv tonômetro de aplanação
Kowa HA-2
manometria ocular
pressão intraocular
gatos
topic tonômetro de aplanação
Kowa HA-2
manometria ocular
pressão intraocular
gatos
applanation tonometer
Kowa HA-2
ocular manometry
intraocular pressure
cats
CNPQ::CIENCIAS AGRARIAS::MEDICINA VETERINARIA
dc.subject.eng.fl_str_mv applanation tonometer
Kowa HA-2
ocular manometry
intraocular pressure
cats
dc.subject.cnpq.fl_str_mv CNPQ::CIENCIAS AGRARIAS::MEDICINA VETERINARIA
description The objective of this study was to measurement of the intraocular pressure (IOP) with Kowa HA-2 tonometer in cats analyzing the calibration, the accuracy and the validation of ambulatory clinical use. For calibration the post-mortem study was accomplished in 10 healthy eyes of 5 cats comparing the ocular manometry with the values of the IOP checked with the tonometer. For evaluation of the accuracy an in vivo study was accomplished in 20 healthy eyes of 10 anesthetized cats being compared the ocular manometry with the IOP obtained with the tonometer. For validation of the ambulatory clinical study of the IOP measurement was accomplished in 78 eyes of 39 healthy cats, in 7 eyes with clinical signs of glaucoma and in 20 eyes with clinical signs of uveitis. The correlation coefficient (r²) between the manometer and the tonometer was 0.993 and the equation of lineal regression was y=0.0915x+0.0878 in postmortem study. In the in vivo study the medium values of IOP in the manometry were 15.6±1.1 mmHg and in the tonometry were 15.5±1.2 mmHg, there was no statistics significant difference between the manometry and the tonometry. In the ambulatory clinical study with healthy cats the medium values of IOP with the tonometer were 15.0±1.5 mmHg, in the eyes with clinical signs of glaucoma were 38,4±8,1 mmHg and in the eyes with clinical signs of uveitis were of 10,4±2,0 mmHg. Therefore, there was a satisfactory correlation and accuracy between the IOP values obtained by direct ocular manometry and the tonometer in question. In the ambulatory clinical study the IOP values obtained with the tonometer were compatible for animals with healthy eyes and with clinical signs of glaucoma and uveitis. So, we can conclude that the Kowa HA-2 tonometer can be used in the routine ophthalmic examination, as it is a practical method for IOP measurement in cats.
publishDate 2015
dc.date.available.fl_str_mv 2015-05-27
dc.date.issued.fl_str_mv 2015-03-17
dc.date.accessioned.fl_str_mv 2016-01-26T18:55:43Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/masterThesis
format masterThesis
status_str publishedVersion
dc.identifier.citation.fl_str_mv RICCI, Cláudia Lizandra. Use of portable applanation tonometer Kowa HA-2 in the measurement of intraocular pressure in cats. 2015. 32 f. Dissertação (Mestrado em Ciências Agrárias) - Universidade do Oeste Paulista, Presidente Prudente, 2015.
dc.identifier.uri.fl_str_mv http://bdtd.unoeste.br:8080/tede/handle/tede/314
identifier_str_mv RICCI, Cláudia Lizandra. Use of portable applanation tonometer Kowa HA-2 in the measurement of intraocular pressure in cats. 2015. 32 f. Dissertação (Mestrado em Ciências Agrárias) - Universidade do Oeste Paulista, Presidente Prudente, 2015.
url http://bdtd.unoeste.br:8080/tede/handle/tede/314
dc.language.iso.fl_str_mv por
language por
dc.relation.program.fl_str_mv 1886725639743614135
dc.relation.confidence.fl_str_mv 500
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade do Oeste Paulista
dc.publisher.program.fl_str_mv Mestrado em Ciência Animal
dc.publisher.initials.fl_str_mv UNOESTE
dc.publisher.country.fl_str_mv BR
dc.publisher.department.fl_str_mv Ciências Agrárias
publisher.none.fl_str_mv Universidade do Oeste Paulista
dc.source.none.fl_str_mv reponame:Biblioteca Digital de Teses e Dissertações da UNOESTE
instname:Universidade do Oeste Paulista (UNOESTE)
instacron:UNOESTE
instname_str Universidade do Oeste Paulista (UNOESTE)
instacron_str UNOESTE
institution UNOESTE
reponame_str Biblioteca Digital de Teses e Dissertações da UNOESTE
collection Biblioteca Digital de Teses e Dissertações da UNOESTE
bitstream.url.fl_str_mv http://bdtd.unoeste.br:8080/tede/bitstream/tede/314/1/Claudia+Lizandra+Ricci.pdf
http://bdtd.unoeste.br:8080/tede/bitstream/tede/314/2/Claudia+Lizandra+Ricci.pdf.jpg
bitstream.checksum.fl_str_mv b9dccb042e091da74b7d28825ae3639d
01a3bf5869d25756f643066aef98b98b
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Biblioteca Digital de Teses e Dissertações da UNOESTE - Universidade do Oeste Paulista (UNOESTE)
repository.mail.fl_str_mv bdtd@unoeste.br
_version_ 1766605142543564800