Mechanical testing of thin-walled zirconia abutments

Detalhes bibliográficos
Autor(a) principal: CANULLO, Luigi
Data de Publicação: 2013
Outros Autores: COELHO, Paulo G., BONFANTE, Estevam A.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Journal of applied oral science (Online)
Texto Completo: https://www.revistas.usp.br/jaos/article/view/53402
Resumo: Although the use of zirconia abutments for implant-supported restorations has gained momentum with the increasing demand for esthetics, little informed design rationale has been developed to characterize their fatigue behavior under different clinical scenarios. However, to prevent the zirconia from fracturing, the use of a titanium connection in bi-component aesthetic abutments has been suggested. Objective Mechanical testing of customized thin-walled titanium-zirconia abutments at the connection with the implant was performed in order to characterize the fatigue behavior and the failure modes for straight and angled abutments. Material and Methods Twenty custom-made bi-component abutments were tested according to ISO 14801:2007 either at a straight or a 25° angle inclination (n=10 each group). Fatigue was conducted at 15 Hz for 5 million cycles in dry conditions at 20°C±5°C. Mean values and standard deviations were calculated for each group. All comparisons were performed by t-tests assuming unequal variances. The level of statistical significance was set at p≤0.05. Failed samples were inspected in a polarized-light and then in a scanning electron microscope. Results Straight and angled abutments mean maximum load was 296.7 N and 1,145 N, the dynamic loading mean Fmax was 237.4 N and 240.7 N, respectively. No significant differences resulted between the straight and angled bi-component abutments in both static (p=0.253) and dynamic testing (p=0.135). A significant difference in the bending moment required for fracture was detected between the groups (p=0.01). Fractures in the angled group occurred mainly at the point of load application, whereas in the straight abutments, fractures were located coronally and close to the thinly designed areas at the cervical region. Conclusion Angled or straight thin-walled zirconia abutments presented similar Fmax under fatigue testing despite the different bending moments required for fracture. The main implication is that although zirconia angled or straight abutments presented similar mechanical behavior, the failure mode tended to be more catastrophic in straight (fracture at the cervical region) compared to angled abutments. ^len
id USP-17_c81c20cae94a5987c01b23ba48c3280e
oai_identifier_str oai:revistas.usp.br:article/53402
network_acronym_str USP-17
network_name_str Journal of applied oral science (Online)
repository_id_str
spelling Mechanical testing of thin-walled zirconia abutmentsDental implants^Zirconium^Dental implant-abutment design^Fatigue^Although the use of zirconia abutments for implant-supported restorations has gained momentum with the increasing demand for esthetics, little informed design rationale has been developed to characterize their fatigue behavior under different clinical scenarios. However, to prevent the zirconia from fracturing, the use of a titanium connection in bi-component aesthetic abutments has been suggested. Objective Mechanical testing of customized thin-walled titanium-zirconia abutments at the connection with the implant was performed in order to characterize the fatigue behavior and the failure modes for straight and angled abutments. Material and Methods Twenty custom-made bi-component abutments were tested according to ISO 14801:2007 either at a straight or a 25° angle inclination (n=10 each group). Fatigue was conducted at 15 Hz for 5 million cycles in dry conditions at 20°C±5°C. Mean values and standard deviations were calculated for each group. All comparisons were performed by t-tests assuming unequal variances. The level of statistical significance was set at p≤0.05. Failed samples were inspected in a polarized-light and then in a scanning electron microscope. Results Straight and angled abutments mean maximum load was 296.7 N and 1,145 N, the dynamic loading mean Fmax was 237.4 N and 240.7 N, respectively. No significant differences resulted between the straight and angled bi-component abutments in both static (p=0.253) and dynamic testing (p=0.135). A significant difference in the bending moment required for fracture was detected between the groups (p=0.01). Fractures in the angled group occurred mainly at the point of load application, whereas in the straight abutments, fractures were located coronally and close to the thinly designed areas at the cervical region. Conclusion Angled or straight thin-walled zirconia abutments presented similar Fmax under fatigue testing despite the different bending moments required for fracture. The main implication is that although zirconia angled or straight abutments presented similar mechanical behavior, the failure mode tended to be more catastrophic in straight (fracture at the cervical region) compared to angled abutments. ^lenUniversidade de São Paulo. Faculdade de Odontologia de Bauru2013-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://www.revistas.usp.br/jaos/article/view/5340210.1590/1678-7757201302124Journal of Applied Oral Science; Vol. 21 No. 1 (2013); 20-24Journal of Applied Oral Science; Vol. 21 Núm. 1 (2013); 20-24Journal of Applied Oral Science; v. 21 n. 1 (2013); 20-241678-77651678-7757reponame:Journal of applied oral science (Online)instname:Universidade de São Paulo (USP)instacron:USPenghttps://www.revistas.usp.br/jaos/article/view/53402/57381Copyright (c) 2013 Journal of Applied Oral Scienceinfo:eu-repo/semantics/openAccessCANULLO, LuigiCOELHO, Paulo G.BONFANTE, Estevam A.2014-05-08T13:21:16Zoai:revistas.usp.br:article/53402Revistahttp://www.scielo.br/jaosPUBhttps://www.revistas.usp.br/jaos/oai||jaos@usp.br1678-77651678-7757opendoar:2014-05-08T13:21:16Journal of applied oral science (Online) - Universidade de São Paulo (USP)false
dc.title.none.fl_str_mv Mechanical testing of thin-walled zirconia abutments
title Mechanical testing of thin-walled zirconia abutments
spellingShingle Mechanical testing of thin-walled zirconia abutments
CANULLO, Luigi
Dental implants^
Zirconium^
Dental implant-abutment design^
Fatigue^
title_short Mechanical testing of thin-walled zirconia abutments
title_full Mechanical testing of thin-walled zirconia abutments
title_fullStr Mechanical testing of thin-walled zirconia abutments
title_full_unstemmed Mechanical testing of thin-walled zirconia abutments
title_sort Mechanical testing of thin-walled zirconia abutments
author CANULLO, Luigi
author_facet CANULLO, Luigi
COELHO, Paulo G.
BONFANTE, Estevam A.
author_role author
author2 COELHO, Paulo G.
BONFANTE, Estevam A.
author2_role author
author
dc.contributor.author.fl_str_mv CANULLO, Luigi
COELHO, Paulo G.
BONFANTE, Estevam A.
dc.subject.por.fl_str_mv Dental implants^
Zirconium^
Dental implant-abutment design^
Fatigue^
topic Dental implants^
Zirconium^
Dental implant-abutment design^
Fatigue^
description Although the use of zirconia abutments for implant-supported restorations has gained momentum with the increasing demand for esthetics, little informed design rationale has been developed to characterize their fatigue behavior under different clinical scenarios. However, to prevent the zirconia from fracturing, the use of a titanium connection in bi-component aesthetic abutments has been suggested. Objective Mechanical testing of customized thin-walled titanium-zirconia abutments at the connection with the implant was performed in order to characterize the fatigue behavior and the failure modes for straight and angled abutments. Material and Methods Twenty custom-made bi-component abutments were tested according to ISO 14801:2007 either at a straight or a 25° angle inclination (n=10 each group). Fatigue was conducted at 15 Hz for 5 million cycles in dry conditions at 20°C±5°C. Mean values and standard deviations were calculated for each group. All comparisons were performed by t-tests assuming unequal variances. The level of statistical significance was set at p≤0.05. Failed samples were inspected in a polarized-light and then in a scanning electron microscope. Results Straight and angled abutments mean maximum load was 296.7 N and 1,145 N, the dynamic loading mean Fmax was 237.4 N and 240.7 N, respectively. No significant differences resulted between the straight and angled bi-component abutments in both static (p=0.253) and dynamic testing (p=0.135). A significant difference in the bending moment required for fracture was detected between the groups (p=0.01). Fractures in the angled group occurred mainly at the point of load application, whereas in the straight abutments, fractures were located coronally and close to the thinly designed areas at the cervical region. Conclusion Angled or straight thin-walled zirconia abutments presented similar Fmax under fatigue testing despite the different bending moments required for fracture. The main implication is that although zirconia angled or straight abutments presented similar mechanical behavior, the failure mode tended to be more catastrophic in straight (fracture at the cervical region) compared to angled abutments. ^len
publishDate 2013
dc.date.none.fl_str_mv 2013-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://www.revistas.usp.br/jaos/article/view/53402
10.1590/1678-7757201302124
url https://www.revistas.usp.br/jaos/article/view/53402
identifier_str_mv 10.1590/1678-7757201302124
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://www.revistas.usp.br/jaos/article/view/53402/57381
dc.rights.driver.fl_str_mv Copyright (c) 2013 Journal of Applied Oral Science
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Copyright (c) 2013 Journal of Applied Oral Science
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade de São Paulo. Faculdade de Odontologia de Bauru
publisher.none.fl_str_mv Universidade de São Paulo. Faculdade de Odontologia de Bauru
dc.source.none.fl_str_mv Journal of Applied Oral Science; Vol. 21 No. 1 (2013); 20-24
Journal of Applied Oral Science; Vol. 21 Núm. 1 (2013); 20-24
Journal of Applied Oral Science; v. 21 n. 1 (2013); 20-24
1678-7765
1678-7757
reponame:Journal of applied oral science (Online)
instname:Universidade de São Paulo (USP)
instacron:USP
instname_str Universidade de São Paulo (USP)
instacron_str USP
institution USP
reponame_str Journal of applied oral science (Online)
collection Journal of applied oral science (Online)
repository.name.fl_str_mv Journal of applied oral science (Online) - Universidade de São Paulo (USP)
repository.mail.fl_str_mv ||jaos@usp.br
_version_ 1787713193283420160