Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C

Detalhes bibliográficos
Autor(a) principal: João Vitor do Carmo Barbosa
Data de Publicação: 2021
Outros Autores: Rodrigo Modesto Gadelha Gontijo, Guilherme Cavalcante de Albuquerque Souza, Bruno Melo Mendes, Juliana Batista da Silva, Marcelo Mamede, Andréa Vidal Ferreira
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UFMG
Texto Completo: http://hdl.handle.net/1843/57386
Resumo: The small animal positron emission tomography (PET) scanner from Molecular Imaging Laboratory (LIM/CDTN) is dedicated to pre-clinical studies on new 18F and 11C-based radiopharmaceuticals and to the development of novel applications for traditional radiopharmaceuticals. Quality control tests recommended by the NEMA publication NU 4-2008 are routinely carried out to ensure the proper performance of the PET scanner. The aim of this work was to evaluate the influence of image reconstruction protocols on the image quality, accuracy of attenuation and scatter correction parameters for 18F and 11C PET images. 18F-FDG and 11C-PK1122 PET images of the image quality phantom were acquired and then reconstructed using different reconstruction protocols varying algorithms (FBP, MLEM-3D, OSEM-3D), resolution mode (high/standard) and number of iterations (10 to 150). Uniformity, spill-over ratio (SOR) and recovery coefficient (RC) tests were performed for each reconstructed image according to NEMA NU 4-2008. FBP based protocol generated noisier images compared to iterative algorithms (MLEM-3D or OSEM-3D) based protocols. The increase in the number of iterations resulted in higher standard deviation of the analyzed parameters for all images. MLEM-3D and OSEM-3D based protocols generates similar results when the number of iterations and resolution mode were identical. SOR and RC mean values remained stable when the number of iterations ranged from 40 to 150. This study allowed the evaluation of different image reconstruction protocols on important parameters of 18F and 11C PET image quality. Additionally, standard protocols to be adopted in LIM/CDTN for 18F and 11C images reconstruction in preclinical studies were defined.
id UFMG_1c6fb85b62c5c9bae745391781b6f271
oai_identifier_str oai:repositorio.ufmg.br:1843/57386
network_acronym_str UFMG
network_name_str Repositório Institucional da UFMG
repository_id_str
spelling 2023-08-02T20:53:13Z2023-08-02T20:53:13Z2021-09-2093011910.15392/bjrs.v9i3.173423190612http://hdl.handle.net/1843/573860000-0001-6479-3087The small animal positron emission tomography (PET) scanner from Molecular Imaging Laboratory (LIM/CDTN) is dedicated to pre-clinical studies on new 18F and 11C-based radiopharmaceuticals and to the development of novel applications for traditional radiopharmaceuticals. Quality control tests recommended by the NEMA publication NU 4-2008 are routinely carried out to ensure the proper performance of the PET scanner. The aim of this work was to evaluate the influence of image reconstruction protocols on the image quality, accuracy of attenuation and scatter correction parameters for 18F and 11C PET images. 18F-FDG and 11C-PK1122 PET images of the image quality phantom were acquired and then reconstructed using different reconstruction protocols varying algorithms (FBP, MLEM-3D, OSEM-3D), resolution mode (high/standard) and number of iterations (10 to 150). Uniformity, spill-over ratio (SOR) and recovery coefficient (RC) tests were performed for each reconstructed image according to NEMA NU 4-2008. FBP based protocol generated noisier images compared to iterative algorithms (MLEM-3D or OSEM-3D) based protocols. The increase in the number of iterations resulted in higher standard deviation of the analyzed parameters for all images. MLEM-3D and OSEM-3D based protocols generates similar results when the number of iterations and resolution mode were identical. SOR and RC mean values remained stable when the number of iterations ranged from 40 to 150. This study allowed the evaluation of different image reconstruction protocols on important parameters of 18F and 11C PET image quality. Additionally, standard protocols to be adopted in LIM/CDTN for 18F and 11C images reconstruction in preclinical studies were defined.engUniversidade Federal de Minas GeraisUFMGBrasilMED - DEPARTAMENTO DE ANATOMIA E IMAGEMBrazilian Journal of Radiation SciencesTomografia por Emissão de PósitronsDiagnóstico por ImagemSmall Animal PETImage QualityNEMA NU 4-2008Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11Cinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articlehttps://bjrs.org.br/revista/index.php/REVISTA/article/view/1734João Vitor do Carmo BarbosaRodrigo Modesto Gadelha GontijoGuilherme Cavalcante de Albuquerque SouzaBruno Melo MendesJuliana Batista da SilvaMarcelo MamedeAndréa Vidal Ferreiraapplication/pdfinfo:eu-repo/semantics/openAccessreponame:Repositório Institucional da UFMGinstname:Universidade Federal de Minas Gerais (UFMG)instacron:UFMGLICENSELicense.txtLicense.txttext/plain; charset=utf-82042https://repositorio.ufmg.br/bitstream/1843/57386/1/License.txtfa505098d172de0bc8864fc1287ffe22MD51ORIGINALInfluence of image reconstruction protocols on the pdfa.pdfInfluence of image reconstruction protocols on the pdfa.pdfapplication/pdf1210493https://repositorio.ufmg.br/bitstream/1843/57386/2/Influence%20of%20image%20reconstruction%20protocols%20on%20the%20pdfa.pdfa8a4aa987b53a2265c65fdd2bc2e07abMD521843/573862023-08-02 18:10:08.595oai:repositorio.ufmg.br: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Repositório de PublicaçõesPUBhttps://repositorio.ufmg.br/oaiopendoar:2023-08-02T21:10:08Repositório Institucional da UFMG - Universidade Federal de Minas Gerais (UFMG)false
dc.title.pt_BR.fl_str_mv Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
title Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
spellingShingle Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
João Vitor do Carmo Barbosa
Small Animal PET
Image Quality
NEMA NU 4-2008
Tomografia por Emissão de Pósitrons
Diagnóstico por Imagem
title_short Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
title_full Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
title_fullStr Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
title_full_unstemmed Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
title_sort Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
author João Vitor do Carmo Barbosa
author_facet João Vitor do Carmo Barbosa
Rodrigo Modesto Gadelha Gontijo
Guilherme Cavalcante de Albuquerque Souza
Bruno Melo Mendes
Juliana Batista da Silva
Marcelo Mamede
Andréa Vidal Ferreira
author_role author
author2 Rodrigo Modesto Gadelha Gontijo
Guilherme Cavalcante de Albuquerque Souza
Bruno Melo Mendes
Juliana Batista da Silva
Marcelo Mamede
Andréa Vidal Ferreira
author2_role author
author
author
author
author
author
dc.contributor.author.fl_str_mv João Vitor do Carmo Barbosa
Rodrigo Modesto Gadelha Gontijo
Guilherme Cavalcante de Albuquerque Souza
Bruno Melo Mendes
Juliana Batista da Silva
Marcelo Mamede
Andréa Vidal Ferreira
dc.subject.por.fl_str_mv Small Animal PET
Image Quality
NEMA NU 4-2008
topic Small Animal PET
Image Quality
NEMA NU 4-2008
Tomografia por Emissão de Pósitrons
Diagnóstico por Imagem
dc.subject.other.pt_BR.fl_str_mv Tomografia por Emissão de Pósitrons
Diagnóstico por Imagem
description The small animal positron emission tomography (PET) scanner from Molecular Imaging Laboratory (LIM/CDTN) is dedicated to pre-clinical studies on new 18F and 11C-based radiopharmaceuticals and to the development of novel applications for traditional radiopharmaceuticals. Quality control tests recommended by the NEMA publication NU 4-2008 are routinely carried out to ensure the proper performance of the PET scanner. The aim of this work was to evaluate the influence of image reconstruction protocols on the image quality, accuracy of attenuation and scatter correction parameters for 18F and 11C PET images. 18F-FDG and 11C-PK1122 PET images of the image quality phantom were acquired and then reconstructed using different reconstruction protocols varying algorithms (FBP, MLEM-3D, OSEM-3D), resolution mode (high/standard) and number of iterations (10 to 150). Uniformity, spill-over ratio (SOR) and recovery coefficient (RC) tests were performed for each reconstructed image according to NEMA NU 4-2008. FBP based protocol generated noisier images compared to iterative algorithms (MLEM-3D or OSEM-3D) based protocols. The increase in the number of iterations resulted in higher standard deviation of the analyzed parameters for all images. MLEM-3D and OSEM-3D based protocols generates similar results when the number of iterations and resolution mode were identical. SOR and RC mean values remained stable when the number of iterations ranged from 40 to 150. This study allowed the evaluation of different image reconstruction protocols on important parameters of 18F and 11C PET image quality. Additionally, standard protocols to be adopted in LIM/CDTN for 18F and 11C images reconstruction in preclinical studies were defined.
publishDate 2021
dc.date.issued.fl_str_mv 2021-09-20
dc.date.accessioned.fl_str_mv 2023-08-02T20:53:13Z
dc.date.available.fl_str_mv 2023-08-02T20:53:13Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/1843/57386
dc.identifier.doi.pt_BR.fl_str_mv 10.15392/bjrs.v9i3.1734
dc.identifier.issn.pt_BR.fl_str_mv 23190612
dc.identifier.orcid.pt_BR.fl_str_mv 0000-0001-6479-3087
identifier_str_mv 10.15392/bjrs.v9i3.1734
23190612
0000-0001-6479-3087
url http://hdl.handle.net/1843/57386
dc.language.iso.fl_str_mv eng
language eng
dc.relation.ispartof.none.fl_str_mv Brazilian Journal of Radiation Sciences
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade Federal de Minas Gerais
dc.publisher.initials.fl_str_mv UFMG
dc.publisher.country.fl_str_mv Brasil
dc.publisher.department.fl_str_mv MED - DEPARTAMENTO DE ANATOMIA E IMAGEM
publisher.none.fl_str_mv Universidade Federal de Minas Gerais
dc.source.none.fl_str_mv reponame:Repositório Institucional da UFMG
instname:Universidade Federal de Minas Gerais (UFMG)
instacron:UFMG
instname_str Universidade Federal de Minas Gerais (UFMG)
instacron_str UFMG
institution UFMG
reponame_str Repositório Institucional da UFMG
collection Repositório Institucional da UFMG
bitstream.url.fl_str_mv https://repositorio.ufmg.br/bitstream/1843/57386/1/License.txt
https://repositorio.ufmg.br/bitstream/1843/57386/2/Influence%20of%20image%20reconstruction%20protocols%20on%20the%20pdfa.pdf
bitstream.checksum.fl_str_mv fa505098d172de0bc8864fc1287ffe22
a8a4aa987b53a2265c65fdd2bc2e07ab
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Repositório Institucional da UFMG - Universidade Federal de Minas Gerais (UFMG)
repository.mail.fl_str_mv
_version_ 1797971377232281600