Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
Autor(a) principal: | |
---|---|
Data de Publicação: | 2021 |
Outros Autores: | , , , , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Repositório Institucional da UFMG |
Texto Completo: | http://hdl.handle.net/1843/57386 |
Resumo: | The small animal positron emission tomography (PET) scanner from Molecular Imaging Laboratory (LIM/CDTN) is dedicated to pre-clinical studies on new 18F and 11C-based radiopharmaceuticals and to the development of novel applications for traditional radiopharmaceuticals. Quality control tests recommended by the NEMA publication NU 4-2008 are routinely carried out to ensure the proper performance of the PET scanner. The aim of this work was to evaluate the influence of image reconstruction protocols on the image quality, accuracy of attenuation and scatter correction parameters for 18F and 11C PET images. 18F-FDG and 11C-PK1122 PET images of the image quality phantom were acquired and then reconstructed using different reconstruction protocols varying algorithms (FBP, MLEM-3D, OSEM-3D), resolution mode (high/standard) and number of iterations (10 to 150). Uniformity, spill-over ratio (SOR) and recovery coefficient (RC) tests were performed for each reconstructed image according to NEMA NU 4-2008. FBP based protocol generated noisier images compared to iterative algorithms (MLEM-3D or OSEM-3D) based protocols. The increase in the number of iterations resulted in higher standard deviation of the analyzed parameters for all images. MLEM-3D and OSEM-3D based protocols generates similar results when the number of iterations and resolution mode were identical. SOR and RC mean values remained stable when the number of iterations ranged from 40 to 150. This study allowed the evaluation of different image reconstruction protocols on important parameters of 18F and 11C PET image quality. Additionally, standard protocols to be adopted in LIM/CDTN for 18F and 11C images reconstruction in preclinical studies were defined. |
id |
UFMG_1c6fb85b62c5c9bae745391781b6f271 |
---|---|
oai_identifier_str |
oai:repositorio.ufmg.br:1843/57386 |
network_acronym_str |
UFMG |
network_name_str |
Repositório Institucional da UFMG |
repository_id_str |
|
spelling |
2023-08-02T20:53:13Z2023-08-02T20:53:13Z2021-09-2093011910.15392/bjrs.v9i3.173423190612http://hdl.handle.net/1843/573860000-0001-6479-3087The small animal positron emission tomography (PET) scanner from Molecular Imaging Laboratory (LIM/CDTN) is dedicated to pre-clinical studies on new 18F and 11C-based radiopharmaceuticals and to the development of novel applications for traditional radiopharmaceuticals. Quality control tests recommended by the NEMA publication NU 4-2008 are routinely carried out to ensure the proper performance of the PET scanner. The aim of this work was to evaluate the influence of image reconstruction protocols on the image quality, accuracy of attenuation and scatter correction parameters for 18F and 11C PET images. 18F-FDG and 11C-PK1122 PET images of the image quality phantom were acquired and then reconstructed using different reconstruction protocols varying algorithms (FBP, MLEM-3D, OSEM-3D), resolution mode (high/standard) and number of iterations (10 to 150). Uniformity, spill-over ratio (SOR) and recovery coefficient (RC) tests were performed for each reconstructed image according to NEMA NU 4-2008. FBP based protocol generated noisier images compared to iterative algorithms (MLEM-3D or OSEM-3D) based protocols. The increase in the number of iterations resulted in higher standard deviation of the analyzed parameters for all images. MLEM-3D and OSEM-3D based protocols generates similar results when the number of iterations and resolution mode were identical. SOR and RC mean values remained stable when the number of iterations ranged from 40 to 150. This study allowed the evaluation of different image reconstruction protocols on important parameters of 18F and 11C PET image quality. Additionally, standard protocols to be adopted in LIM/CDTN for 18F and 11C images reconstruction in preclinical studies were defined.engUniversidade Federal de Minas GeraisUFMGBrasilMED - DEPARTAMENTO DE ANATOMIA E IMAGEMBrazilian Journal of Radiation SciencesTomografia por Emissão de PósitronsDiagnóstico por ImagemSmall Animal PETImage QualityNEMA NU 4-2008Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11Cinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articlehttps://bjrs.org.br/revista/index.php/REVISTA/article/view/1734João Vitor do Carmo BarbosaRodrigo Modesto Gadelha GontijoGuilherme Cavalcante de Albuquerque SouzaBruno Melo MendesJuliana Batista da SilvaMarcelo MamedeAndréa Vidal Ferreiraapplication/pdfinfo:eu-repo/semantics/openAccessreponame:Repositório Institucional da UFMGinstname:Universidade Federal de Minas Gerais (UFMG)instacron:UFMGLICENSELicense.txtLicense.txttext/plain; charset=utf-82042https://repositorio.ufmg.br/bitstream/1843/57386/1/License.txtfa505098d172de0bc8864fc1287ffe22MD51ORIGINALInfluence of image reconstruction protocols on the pdfa.pdfInfluence of image reconstruction protocols on the pdfa.pdfapplication/pdf1210493https://repositorio.ufmg.br/bitstream/1843/57386/2/Influence%20of%20image%20reconstruction%20protocols%20on%20the%20pdfa.pdfa8a4aa987b53a2265c65fdd2bc2e07abMD521843/573862023-08-02 18:10:08.595oai:repositorio.ufmg.br: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Repositório de PublicaçõesPUBhttps://repositorio.ufmg.br/oaiopendoar:2023-08-02T21:10:08Repositório Institucional da UFMG - Universidade Federal de Minas Gerais (UFMG)false |
dc.title.pt_BR.fl_str_mv |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C |
title |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C |
spellingShingle |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C João Vitor do Carmo Barbosa Small Animal PET Image Quality NEMA NU 4-2008 Tomografia por Emissão de Pósitrons Diagnóstico por Imagem |
title_short |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C |
title_full |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C |
title_fullStr |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C |
title_full_unstemmed |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C |
title_sort |
Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C |
author |
João Vitor do Carmo Barbosa |
author_facet |
João Vitor do Carmo Barbosa Rodrigo Modesto Gadelha Gontijo Guilherme Cavalcante de Albuquerque Souza Bruno Melo Mendes Juliana Batista da Silva Marcelo Mamede Andréa Vidal Ferreira |
author_role |
author |
author2 |
Rodrigo Modesto Gadelha Gontijo Guilherme Cavalcante de Albuquerque Souza Bruno Melo Mendes Juliana Batista da Silva Marcelo Mamede Andréa Vidal Ferreira |
author2_role |
author author author author author author |
dc.contributor.author.fl_str_mv |
João Vitor do Carmo Barbosa Rodrigo Modesto Gadelha Gontijo Guilherme Cavalcante de Albuquerque Souza Bruno Melo Mendes Juliana Batista da Silva Marcelo Mamede Andréa Vidal Ferreira |
dc.subject.por.fl_str_mv |
Small Animal PET Image Quality NEMA NU 4-2008 |
topic |
Small Animal PET Image Quality NEMA NU 4-2008 Tomografia por Emissão de Pósitrons Diagnóstico por Imagem |
dc.subject.other.pt_BR.fl_str_mv |
Tomografia por Emissão de Pósitrons Diagnóstico por Imagem |
description |
The small animal positron emission tomography (PET) scanner from Molecular Imaging Laboratory (LIM/CDTN) is dedicated to pre-clinical studies on new 18F and 11C-based radiopharmaceuticals and to the development of novel applications for traditional radiopharmaceuticals. Quality control tests recommended by the NEMA publication NU 4-2008 are routinely carried out to ensure the proper performance of the PET scanner. The aim of this work was to evaluate the influence of image reconstruction protocols on the image quality, accuracy of attenuation and scatter correction parameters for 18F and 11C PET images. 18F-FDG and 11C-PK1122 PET images of the image quality phantom were acquired and then reconstructed using different reconstruction protocols varying algorithms (FBP, MLEM-3D, OSEM-3D), resolution mode (high/standard) and number of iterations (10 to 150). Uniformity, spill-over ratio (SOR) and recovery coefficient (RC) tests were performed for each reconstructed image according to NEMA NU 4-2008. FBP based protocol generated noisier images compared to iterative algorithms (MLEM-3D or OSEM-3D) based protocols. The increase in the number of iterations resulted in higher standard deviation of the analyzed parameters for all images. MLEM-3D and OSEM-3D based protocols generates similar results when the number of iterations and resolution mode were identical. SOR and RC mean values remained stable when the number of iterations ranged from 40 to 150. This study allowed the evaluation of different image reconstruction protocols on important parameters of 18F and 11C PET image quality. Additionally, standard protocols to be adopted in LIM/CDTN for 18F and 11C images reconstruction in preclinical studies were defined. |
publishDate |
2021 |
dc.date.issued.fl_str_mv |
2021-09-20 |
dc.date.accessioned.fl_str_mv |
2023-08-02T20:53:13Z |
dc.date.available.fl_str_mv |
2023-08-02T20:53:13Z |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://hdl.handle.net/1843/57386 |
dc.identifier.doi.pt_BR.fl_str_mv |
10.15392/bjrs.v9i3.1734 |
dc.identifier.issn.pt_BR.fl_str_mv |
23190612 |
dc.identifier.orcid.pt_BR.fl_str_mv |
0000-0001-6479-3087 |
identifier_str_mv |
10.15392/bjrs.v9i3.1734 23190612 0000-0001-6479-3087 |
url |
http://hdl.handle.net/1843/57386 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.ispartof.none.fl_str_mv |
Brazilian Journal of Radiation Sciences |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf |
dc.publisher.none.fl_str_mv |
Universidade Federal de Minas Gerais |
dc.publisher.initials.fl_str_mv |
UFMG |
dc.publisher.country.fl_str_mv |
Brasil |
dc.publisher.department.fl_str_mv |
MED - DEPARTAMENTO DE ANATOMIA E IMAGEM |
publisher.none.fl_str_mv |
Universidade Federal de Minas Gerais |
dc.source.none.fl_str_mv |
reponame:Repositório Institucional da UFMG instname:Universidade Federal de Minas Gerais (UFMG) instacron:UFMG |
instname_str |
Universidade Federal de Minas Gerais (UFMG) |
instacron_str |
UFMG |
institution |
UFMG |
reponame_str |
Repositório Institucional da UFMG |
collection |
Repositório Institucional da UFMG |
bitstream.url.fl_str_mv |
https://repositorio.ufmg.br/bitstream/1843/57386/1/License.txt https://repositorio.ufmg.br/bitstream/1843/57386/2/Influence%20of%20image%20reconstruction%20protocols%20on%20the%20pdfa.pdf |
bitstream.checksum.fl_str_mv |
fa505098d172de0bc8864fc1287ffe22 a8a4aa987b53a2265c65fdd2bc2e07ab |
bitstream.checksumAlgorithm.fl_str_mv |
MD5 MD5 |
repository.name.fl_str_mv |
Repositório Institucional da UFMG - Universidade Federal de Minas Gerais (UFMG) |
repository.mail.fl_str_mv |
|
_version_ |
1797971377232281600 |