Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory

Detalhes bibliográficos
Autor(a) principal: Bitencourt,Eberson Damiao dos Santos de
Data de Publicação: 2013
Outros Autores: Voegeli,Carlos Franco, Onzi,Gabriela dos Santos, Boscato,Sara Cardoso, Ghem,Carine, Munhoz,Terezinha
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista brasileira de hematologia e hemoterapia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842013000600404
Resumo: Background: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex®SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process. Objective: To evaluate the performance and reliability of the Sysmex® SP-1000i slide preparer-stainer incorporated into the routine of a hospital laboratory in Porto Alegre. Methods: Peripheral blood samples of patients attending the laboratory for ambulatory exams with leukocyte counts between 7000/°L and 12,000/°L were evaluated, independent of gender and age. Two slides were prepared for each sample using the Sysmex® SP-1000i equipment; one of the slides was used to perform quality control tests using the CellaVision® DM96 device, and the other slide was used to compare pre-classification by the same device and the classification performed by a pharmacist-biochemist. Results: The results of all the slides used as controls were acceptable according to the quality control test as established by the manufacturer of the device. In the comparison between the automated pre-classification and the classification made by the professional, there was an acceptable variation in the differential counts of leukocytes for 90% of the analyzed slides. Pearson correlation coefficient showed a strong correlation for band neutrophils (r = 0.802; p-value < 0.001), segmented neutrophils (r = 0.963; p-value < 0.001), eosinophils (r = 0.958; p-value < 0.001), lymphocytes (r = 0.985; p-value < 0.001) and atypical lymphocytes (r = 0.866; p-value < 0.001) using both methods. The red blood cell analysis was adequate for all slides analyzed by the equipment and by the professional. Conclusion: The new Sysmex®SP1000-i methodology was found to be reliable, fast and safe for the routines of medium and large laboratories, improving the quality of microscopic analysis in complete blood counts.
id ABHHTC-1_b9a1a4bd09e61a8eb8def035498d0443
oai_identifier_str oai:scielo:S1516-84842013000600404
network_acronym_str ABHHTC-1
network_name_str Revista brasileira de hematologia e hemoterapia (Online)
repository_id_str
spelling Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratoryBlood cell count/methodsDiagnostic equipmentFlow cytometryValidation studiesAutomation, laboratory Background: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex®SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process. Objective: To evaluate the performance and reliability of the Sysmex® SP-1000i slide preparer-stainer incorporated into the routine of a hospital laboratory in Porto Alegre. Methods: Peripheral blood samples of patients attending the laboratory for ambulatory exams with leukocyte counts between 7000/°L and 12,000/°L were evaluated, independent of gender and age. Two slides were prepared for each sample using the Sysmex® SP-1000i equipment; one of the slides was used to perform quality control tests using the CellaVision® DM96 device, and the other slide was used to compare pre-classification by the same device and the classification performed by a pharmacist-biochemist. Results: The results of all the slides used as controls were acceptable according to the quality control test as established by the manufacturer of the device. In the comparison between the automated pre-classification and the classification made by the professional, there was an acceptable variation in the differential counts of leukocytes for 90% of the analyzed slides. Pearson correlation coefficient showed a strong correlation for band neutrophils (r = 0.802; p-value < 0.001), segmented neutrophils (r = 0.963; p-value < 0.001), eosinophils (r = 0.958; p-value < 0.001), lymphocytes (r = 0.985; p-value < 0.001) and atypical lymphocytes (r = 0.866; p-value < 0.001) using both methods. The red blood cell analysis was adequate for all slides analyzed by the equipment and by the professional. Conclusion: The new Sysmex®SP1000-i methodology was found to be reliable, fast and safe for the routines of medium and large laboratories, improving the quality of microscopic analysis in complete blood counts. Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular2013-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842013000600404Revista Brasileira de Hematologia e Hemoterapia v.35 n.6 2013reponame:Revista brasileira de hematologia e hemoterapia (Online)instname:Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)instacron:ABHHTC10.5581/1516-8484.20130121info:eu-repo/semantics/openAccessBitencourt,Eberson Damiao dos Santos deVoegeli,Carlos FrancoOnzi,Gabriela dos SantosBoscato,Sara CardosoGhem,CarineMunhoz,Terezinhaeng2014-01-29T00:00:00Zoai:scielo:S1516-84842013000600404Revistahttp://www.rbhh.org/pt/archivo/https://old.scielo.br/oai/scielo-oai.phpsbhh@terra.com.br||secretaria@rbhh.org1806-08701516-8484opendoar:2014-01-29T00:00Revista brasileira de hematologia e hemoterapia (Online) - Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)false
dc.title.none.fl_str_mv Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
title Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
spellingShingle Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
Bitencourt,Eberson Damiao dos Santos de
Blood cell count/methods
Diagnostic equipment
Flow cytometry
Validation studies
Automation, laboratory
title_short Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
title_full Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
title_fullStr Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
title_full_unstemmed Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
title_sort Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
author Bitencourt,Eberson Damiao dos Santos de
author_facet Bitencourt,Eberson Damiao dos Santos de
Voegeli,Carlos Franco
Onzi,Gabriela dos Santos
Boscato,Sara Cardoso
Ghem,Carine
Munhoz,Terezinha
author_role author
author2 Voegeli,Carlos Franco
Onzi,Gabriela dos Santos
Boscato,Sara Cardoso
Ghem,Carine
Munhoz,Terezinha
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Bitencourt,Eberson Damiao dos Santos de
Voegeli,Carlos Franco
Onzi,Gabriela dos Santos
Boscato,Sara Cardoso
Ghem,Carine
Munhoz,Terezinha
dc.subject.por.fl_str_mv Blood cell count/methods
Diagnostic equipment
Flow cytometry
Validation studies
Automation, laboratory
topic Blood cell count/methods
Diagnostic equipment
Flow cytometry
Validation studies
Automation, laboratory
description Background: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex®SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process. Objective: To evaluate the performance and reliability of the Sysmex® SP-1000i slide preparer-stainer incorporated into the routine of a hospital laboratory in Porto Alegre. Methods: Peripheral blood samples of patients attending the laboratory for ambulatory exams with leukocyte counts between 7000/°L and 12,000/°L were evaluated, independent of gender and age. Two slides were prepared for each sample using the Sysmex® SP-1000i equipment; one of the slides was used to perform quality control tests using the CellaVision® DM96 device, and the other slide was used to compare pre-classification by the same device and the classification performed by a pharmacist-biochemist. Results: The results of all the slides used as controls were acceptable according to the quality control test as established by the manufacturer of the device. In the comparison between the automated pre-classification and the classification made by the professional, there was an acceptable variation in the differential counts of leukocytes for 90% of the analyzed slides. Pearson correlation coefficient showed a strong correlation for band neutrophils (r = 0.802; p-value < 0.001), segmented neutrophils (r = 0.963; p-value < 0.001), eosinophils (r = 0.958; p-value < 0.001), lymphocytes (r = 0.985; p-value < 0.001) and atypical lymphocytes (r = 0.866; p-value < 0.001) using both methods. The red blood cell analysis was adequate for all slides analyzed by the equipment and by the professional. Conclusion: The new Sysmex®SP1000-i methodology was found to be reliable, fast and safe for the routines of medium and large laboratories, improving the quality of microscopic analysis in complete blood counts.
publishDate 2013
dc.date.none.fl_str_mv 2013-12-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842013000600404
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842013000600404
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.5581/1516-8484.20130121
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular
publisher.none.fl_str_mv Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular
dc.source.none.fl_str_mv Revista Brasileira de Hematologia e Hemoterapia v.35 n.6 2013
reponame:Revista brasileira de hematologia e hemoterapia (Online)
instname:Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)
instacron:ABHHTC
instname_str Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)
instacron_str ABHHTC
institution ABHHTC
reponame_str Revista brasileira de hematologia e hemoterapia (Online)
collection Revista brasileira de hematologia e hemoterapia (Online)
repository.name.fl_str_mv Revista brasileira de hematologia e hemoterapia (Online) - Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)
repository.mail.fl_str_mv sbhh@terra.com.br||secretaria@rbhh.org
_version_ 1754213112021516288