Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers

Detalhes bibliográficos
Autor(a) principal: Buoro,Sabrina
Data de Publicação: 2016
Outros Autores: Mecca,Tommaso, Seghezzi,Michela, Manenti,Barbara, Cerutti,Lorenzo, Dominoni,Paola, Napolitano,Gavino, Resmini,Stefano, Crippa,Alberto, Ottomano,Cosimo, Lippi,Giuseppe
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista brasileira de hematologia e hemoterapia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842016000300225
Resumo: ABSTRACT BACKGROUND: Different hematological analyzers have different analytical performances that are often reflected in the criteria for sample stability of the complete blood count. This study aimed to assess the stability of several hematological parameters using the XN-9000 Sysmex and BC-6800 Mindray analyzers. METHODS: The impact of storage at room temperature and 4 °C was evaluated after 2, 4, 6, 8, 24, 36 and 48 h using ten normal and 40 abnormal blood samples. The variation from the baseline measurement was evaluated by the Steel-Dwass-Critchlow-Fligner test and by Bland-Altman plots, using quality specifications and critical difference as the total allowable variation. RESULTS: Red blood cells and reticulocyte parameters (i.e. hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration, red blood cell distribution width, immature reticulocyte fractions, low-fluorescence reticulocytes, middle-fluorescence reticulocytes, high fluorescence mononuclear cells) showed less stability compared to leukocyte and platelet parameters (except for monocyte count and mean platelet volume). The bias for hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration and red blood cell distribution width coefficient of variation was higher than the critical difference after 8 h using both analyzers. CONCLUSION: Blood samples measured with both analyzers do not show analytically significant changes in up to 2 h of storage at room temperature and 4 °C. However, the maximum time for analysis can be extended for up to 8 h when the bias is compared to the critical difference.
id ABHHTC-1_d3cc61b57a6d05b7902575d5e5bead94
oai_identifier_str oai:scielo:S1516-84842016000300225
network_acronym_str ABHHTC-1
network_name_str Revista brasileira de hematologia e hemoterapia (Online)
repository_id_str
spelling Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzersXN-9000BC-6800Sample stabilityComplete blood countPre-analytical phaseABSTRACT BACKGROUND: Different hematological analyzers have different analytical performances that are often reflected in the criteria for sample stability of the complete blood count. This study aimed to assess the stability of several hematological parameters using the XN-9000 Sysmex and BC-6800 Mindray analyzers. METHODS: The impact of storage at room temperature and 4 °C was evaluated after 2, 4, 6, 8, 24, 36 and 48 h using ten normal and 40 abnormal blood samples. The variation from the baseline measurement was evaluated by the Steel-Dwass-Critchlow-Fligner test and by Bland-Altman plots, using quality specifications and critical difference as the total allowable variation. RESULTS: Red blood cells and reticulocyte parameters (i.e. hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration, red blood cell distribution width, immature reticulocyte fractions, low-fluorescence reticulocytes, middle-fluorescence reticulocytes, high fluorescence mononuclear cells) showed less stability compared to leukocyte and platelet parameters (except for monocyte count and mean platelet volume). The bias for hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration and red blood cell distribution width coefficient of variation was higher than the critical difference after 8 h using both analyzers. CONCLUSION: Blood samples measured with both analyzers do not show analytically significant changes in up to 2 h of storage at room temperature and 4 °C. However, the maximum time for analysis can be extended for up to 8 h when the bias is compared to the critical difference.Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular2016-09-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842016000300225Revista Brasileira de Hematologia e Hemoterapia v.38 n.3 2016reponame:Revista brasileira de hematologia e hemoterapia (Online)instname:Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)instacron:ABHHTC10.1016/j.bjhh.2016.05.010info:eu-repo/semantics/openAccessBuoro,SabrinaMecca,TommasoSeghezzi,MichelaManenti,BarbaraCerutti,LorenzoDominoni,PaolaNapolitano,GavinoResmini,StefanoCrippa,AlbertoOttomano,CosimoLippi,Giuseppeeng2016-12-12T00:00:00Zoai:scielo:S1516-84842016000300225Revistahttp://www.rbhh.org/pt/archivo/https://old.scielo.br/oai/scielo-oai.phpsbhh@terra.com.br||secretaria@rbhh.org1806-08701516-8484opendoar:2016-12-12T00:00Revista brasileira de hematologia e hemoterapia (Online) - Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)false
dc.title.none.fl_str_mv Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
title Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
spellingShingle Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
Buoro,Sabrina
XN-9000
BC-6800
Sample stability
Complete blood count
Pre-analytical phase
title_short Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
title_full Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
title_fullStr Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
title_full_unstemmed Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
title_sort Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
author Buoro,Sabrina
author_facet Buoro,Sabrina
Mecca,Tommaso
Seghezzi,Michela
Manenti,Barbara
Cerutti,Lorenzo
Dominoni,Paola
Napolitano,Gavino
Resmini,Stefano
Crippa,Alberto
Ottomano,Cosimo
Lippi,Giuseppe
author_role author
author2 Mecca,Tommaso
Seghezzi,Michela
Manenti,Barbara
Cerutti,Lorenzo
Dominoni,Paola
Napolitano,Gavino
Resmini,Stefano
Crippa,Alberto
Ottomano,Cosimo
Lippi,Giuseppe
author2_role author
author
author
author
author
author
author
author
author
author
dc.contributor.author.fl_str_mv Buoro,Sabrina
Mecca,Tommaso
Seghezzi,Michela
Manenti,Barbara
Cerutti,Lorenzo
Dominoni,Paola
Napolitano,Gavino
Resmini,Stefano
Crippa,Alberto
Ottomano,Cosimo
Lippi,Giuseppe
dc.subject.por.fl_str_mv XN-9000
BC-6800
Sample stability
Complete blood count
Pre-analytical phase
topic XN-9000
BC-6800
Sample stability
Complete blood count
Pre-analytical phase
description ABSTRACT BACKGROUND: Different hematological analyzers have different analytical performances that are often reflected in the criteria for sample stability of the complete blood count. This study aimed to assess the stability of several hematological parameters using the XN-9000 Sysmex and BC-6800 Mindray analyzers. METHODS: The impact of storage at room temperature and 4 °C was evaluated after 2, 4, 6, 8, 24, 36 and 48 h using ten normal and 40 abnormal blood samples. The variation from the baseline measurement was evaluated by the Steel-Dwass-Critchlow-Fligner test and by Bland-Altman plots, using quality specifications and critical difference as the total allowable variation. RESULTS: Red blood cells and reticulocyte parameters (i.e. hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration, red blood cell distribution width, immature reticulocyte fractions, low-fluorescence reticulocytes, middle-fluorescence reticulocytes, high fluorescence mononuclear cells) showed less stability compared to leukocyte and platelet parameters (except for monocyte count and mean platelet volume). The bias for hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration and red blood cell distribution width coefficient of variation was higher than the critical difference after 8 h using both analyzers. CONCLUSION: Blood samples measured with both analyzers do not show analytically significant changes in up to 2 h of storage at room temperature and 4 °C. However, the maximum time for analysis can be extended for up to 8 h when the bias is compared to the critical difference.
publishDate 2016
dc.date.none.fl_str_mv 2016-09-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842016000300225
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-84842016000300225
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1016/j.bjhh.2016.05.010
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular
publisher.none.fl_str_mv Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular
dc.source.none.fl_str_mv Revista Brasileira de Hematologia e Hemoterapia v.38 n.3 2016
reponame:Revista brasileira de hematologia e hemoterapia (Online)
instname:Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)
instacron:ABHHTC
instname_str Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)
instacron_str ABHHTC
institution ABHHTC
reponame_str Revista brasileira de hematologia e hemoterapia (Online)
collection Revista brasileira de hematologia e hemoterapia (Online)
repository.name.fl_str_mv Revista brasileira de hematologia e hemoterapia (Online) - Associação Brasileira de Hematologia e Hemoterapia e Terapia Celular (ABHHTC)
repository.mail.fl_str_mv sbhh@terra.com.br||secretaria@rbhh.org
_version_ 1754213112829968384