Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study

Detalhes bibliográficos
Autor(a) principal: Mattos,Cláudia Trindade
Data de Publicação: 2010
Outros Autores: Ruellas,Antônio Carlos de Oliveira, Elias,Carlos Nelson
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Materials research (São Carlos. Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-14392010000400015
Resumo: The aim of this study was to compare the surface morphology and fracture torque resistance of as-received, sterilized and retrieved mini-implants in order to evaluate the fracture risks of re-using orthodontic mini-implants after sterilization. Thirty mini-implants retrieved from 19 patients, twenty as-received and twenty autoclaved mini-implants from the same manufacturer were used in this study. Ten samples from each group were analyzed by scanning electron microscopy and the remaining specimens were tested for their fracture torque resistance. Results were statistically analyzed using one-way ANOVA. No defects or corrosion could be identified in autoclaved and retrieved mini-implants, but a worn surface and scratch marks could be observed on the retrieved ones. A statistical significant difference in the fracture torque was observed between the as-received and the retrieved groups. Re-use of retrieved mini-implants should not be recommended because their torque resistance is reduced. As-received mini-implants submitted to autoclave sterilization can be recommended to be used clinically.
id ABMABCABPOL-1_c4434817e28b826fe24ae9c232db7bd1
oai_identifier_str oai:scielo:S1516-14392010000400015
network_acronym_str ABMABCABPOL-1
network_name_str Materials research (São Carlos. Online)
repository_id_str
spelling Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro studyimplant fracturere-useScanning Electron Microscopy (SEM)sterilizationThe aim of this study was to compare the surface morphology and fracture torque resistance of as-received, sterilized and retrieved mini-implants in order to evaluate the fracture risks of re-using orthodontic mini-implants after sterilization. Thirty mini-implants retrieved from 19 patients, twenty as-received and twenty autoclaved mini-implants from the same manufacturer were used in this study. Ten samples from each group were analyzed by scanning electron microscopy and the remaining specimens were tested for their fracture torque resistance. Results were statistically analyzed using one-way ANOVA. No defects or corrosion could be identified in autoclaved and retrieved mini-implants, but a worn surface and scratch marks could be observed on the retrieved ones. A statistical significant difference in the fracture torque was observed between the as-received and the retrieved groups. Re-use of retrieved mini-implants should not be recommended because their torque resistance is reduced. As-received mini-implants submitted to autoclave sterilization can be recommended to be used clinically.ABM, ABC, ABPol2010-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-14392010000400015Materials Research v.13 n.4 2010reponame:Materials research (São Carlos. Online)instname:Universidade Federal de São Carlos (UFSCAR)instacron:ABM ABC ABPOL10.1590/S1516-14392010000400015info:eu-repo/semantics/openAccessMattos,Cláudia TrindadeRuellas,Antônio Carlos de OliveiraElias,Carlos Nelsoneng2011-01-24T00:00:00Zoai:scielo:S1516-14392010000400015Revistahttp://www.scielo.br/mrPUBhttps://old.scielo.br/oai/scielo-oai.phpdedz@power.ufscar.br1980-53731516-1439opendoar:2011-01-24T00:00Materials research (São Carlos. Online) - Universidade Federal de São Carlos (UFSCAR)false
dc.title.none.fl_str_mv Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
title Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
spellingShingle Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
Mattos,Cláudia Trindade
implant fracture
re-use
Scanning Electron Microscopy (SEM)
sterilization
title_short Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
title_full Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
title_fullStr Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
title_full_unstemmed Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
title_sort Is it possible to re-use mini-implants for orthodontic anchorage? Results of an in vitro study
author Mattos,Cláudia Trindade
author_facet Mattos,Cláudia Trindade
Ruellas,Antônio Carlos de Oliveira
Elias,Carlos Nelson
author_role author
author2 Ruellas,Antônio Carlos de Oliveira
Elias,Carlos Nelson
author2_role author
author
dc.contributor.author.fl_str_mv Mattos,Cláudia Trindade
Ruellas,Antônio Carlos de Oliveira
Elias,Carlos Nelson
dc.subject.por.fl_str_mv implant fracture
re-use
Scanning Electron Microscopy (SEM)
sterilization
topic implant fracture
re-use
Scanning Electron Microscopy (SEM)
sterilization
description The aim of this study was to compare the surface morphology and fracture torque resistance of as-received, sterilized and retrieved mini-implants in order to evaluate the fracture risks of re-using orthodontic mini-implants after sterilization. Thirty mini-implants retrieved from 19 patients, twenty as-received and twenty autoclaved mini-implants from the same manufacturer were used in this study. Ten samples from each group were analyzed by scanning electron microscopy and the remaining specimens were tested for their fracture torque resistance. Results were statistically analyzed using one-way ANOVA. No defects or corrosion could be identified in autoclaved and retrieved mini-implants, but a worn surface and scratch marks could be observed on the retrieved ones. A statistical significant difference in the fracture torque was observed between the as-received and the retrieved groups. Re-use of retrieved mini-implants should not be recommended because their torque resistance is reduced. As-received mini-implants submitted to autoclave sterilization can be recommended to be used clinically.
publishDate 2010
dc.date.none.fl_str_mv 2010-12-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-14392010000400015
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-14392010000400015
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S1516-14392010000400015
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv ABM, ABC, ABPol
publisher.none.fl_str_mv ABM, ABC, ABPol
dc.source.none.fl_str_mv Materials Research v.13 n.4 2010
reponame:Materials research (São Carlos. Online)
instname:Universidade Federal de São Carlos (UFSCAR)
instacron:ABM ABC ABPOL
instname_str Universidade Federal de São Carlos (UFSCAR)
instacron_str ABM ABC ABPOL
institution ABM ABC ABPOL
reponame_str Materials research (São Carlos. Online)
collection Materials research (São Carlos. Online)
repository.name.fl_str_mv Materials research (São Carlos. Online) - Universidade Federal de São Carlos (UFSCAR)
repository.mail.fl_str_mv dedz@power.ufscar.br
_version_ 1754212659714064384