Assessment of the light intensity of otoscopes utilized in teaching hospitals

Detalhes bibliográficos
Autor(a) principal: Fonseca,Vinicius Ribas
Data de Publicação: 2015
Outros Autores: Basso,Giovana Bittencourt, Cavalheiro,Mariana Nagata
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Journal of Otorhinolaryngology
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-86942015000500479
Resumo: ABSTRACT INTRODUCTION: To attain an accurate otoscopic diagnosis, a functioning device with adequate capacity must be used. OBJECTIVE: Evaluate the light intensity of otoscopes, comparing it utilizing the batteries present at the moment of calibration and after new batteries were supplied. METHODS: Cross-sectional study of a historical cohort, which assessed 38 otoscopes, measuring the light intensity with the batteries present at the moment of assessment compared to the intensity with new batteries, as well as charge of the test batteries and the new batteries. RESULTS: The mean of the sum of new batteries' charge was 3.19 V, and of the test batteries was 2.70 V, representing a decrease of 18.02% in charge. The mean luminosity with the new batteries was 366.89 lumens, whereas in the test batteries it was 188.32 lumens, representing a decrease of 83.75% in the light intensity. Student's t-test was applied for data comparison, showing a statistical difference between the light intensity with the original batteries and the new batteries (p = 0.0001; CI = 0.95). CONCLUSION: There was a statistically significant difference between the proportions of light intensity in the otoscopes. A small variation in battery charge results in a great variation in light.
id ABORL-F-1_16cc7dbe6054482dfd1e273053bf4995
oai_identifier_str oai:scielo:S1808-86942015000500479
network_acronym_str ABORL-F-1
network_name_str Brazilian Journal of Otorhinolaryngology
repository_id_str
spelling Assessment of the light intensity of otoscopes utilized in teaching hospitalsOtoscopyOtoscopesLightABSTRACT INTRODUCTION: To attain an accurate otoscopic diagnosis, a functioning device with adequate capacity must be used. OBJECTIVE: Evaluate the light intensity of otoscopes, comparing it utilizing the batteries present at the moment of calibration and after new batteries were supplied. METHODS: Cross-sectional study of a historical cohort, which assessed 38 otoscopes, measuring the light intensity with the batteries present at the moment of assessment compared to the intensity with new batteries, as well as charge of the test batteries and the new batteries. RESULTS: The mean of the sum of new batteries' charge was 3.19 V, and of the test batteries was 2.70 V, representing a decrease of 18.02% in charge. The mean luminosity with the new batteries was 366.89 lumens, whereas in the test batteries it was 188.32 lumens, representing a decrease of 83.75% in the light intensity. Student's t-test was applied for data comparison, showing a statistical difference between the light intensity with the original batteries and the new batteries (p = 0.0001; CI = 0.95). CONCLUSION: There was a statistically significant difference between the proportions of light intensity in the otoscopes. A small variation in battery charge results in a great variation in light.Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial.2015-10-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-86942015000500479Brazilian Journal of Otorhinolaryngology v.81 n.5 2015reponame:Brazilian Journal of Otorhinolaryngologyinstname:Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial (ABORL-CCF)instacron:ABORL-CCF10.1016/j.bjorl.2015.07.006info:eu-repo/semantics/openAccessFonseca,Vinicius RibasBasso,Giovana BittencourtCavalheiro,Mariana Nagataeng2015-11-24T00:00:00Zoai:scielo:S1808-86942015000500479Revistahttp://www.bjorl.org.br/https://old.scielo.br/oai/scielo-oai.phprevista@aborlccf.org.br||revista@aborlccf.org.br1808-86861808-8686opendoar:2015-11-24T00:00Brazilian Journal of Otorhinolaryngology - Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial (ABORL-CCF)false
dc.title.none.fl_str_mv Assessment of the light intensity of otoscopes utilized in teaching hospitals
title Assessment of the light intensity of otoscopes utilized in teaching hospitals
spellingShingle Assessment of the light intensity of otoscopes utilized in teaching hospitals
Fonseca,Vinicius Ribas
Otoscopy
Otoscopes
Light
title_short Assessment of the light intensity of otoscopes utilized in teaching hospitals
title_full Assessment of the light intensity of otoscopes utilized in teaching hospitals
title_fullStr Assessment of the light intensity of otoscopes utilized in teaching hospitals
title_full_unstemmed Assessment of the light intensity of otoscopes utilized in teaching hospitals
title_sort Assessment of the light intensity of otoscopes utilized in teaching hospitals
author Fonseca,Vinicius Ribas
author_facet Fonseca,Vinicius Ribas
Basso,Giovana Bittencourt
Cavalheiro,Mariana Nagata
author_role author
author2 Basso,Giovana Bittencourt
Cavalheiro,Mariana Nagata
author2_role author
author
dc.contributor.author.fl_str_mv Fonseca,Vinicius Ribas
Basso,Giovana Bittencourt
Cavalheiro,Mariana Nagata
dc.subject.por.fl_str_mv Otoscopy
Otoscopes
Light
topic Otoscopy
Otoscopes
Light
description ABSTRACT INTRODUCTION: To attain an accurate otoscopic diagnosis, a functioning device with adequate capacity must be used. OBJECTIVE: Evaluate the light intensity of otoscopes, comparing it utilizing the batteries present at the moment of calibration and after new batteries were supplied. METHODS: Cross-sectional study of a historical cohort, which assessed 38 otoscopes, measuring the light intensity with the batteries present at the moment of assessment compared to the intensity with new batteries, as well as charge of the test batteries and the new batteries. RESULTS: The mean of the sum of new batteries' charge was 3.19 V, and of the test batteries was 2.70 V, representing a decrease of 18.02% in charge. The mean luminosity with the new batteries was 366.89 lumens, whereas in the test batteries it was 188.32 lumens, representing a decrease of 83.75% in the light intensity. Student's t-test was applied for data comparison, showing a statistical difference between the light intensity with the original batteries and the new batteries (p = 0.0001; CI = 0.95). CONCLUSION: There was a statistically significant difference between the proportions of light intensity in the otoscopes. A small variation in battery charge results in a great variation in light.
publishDate 2015
dc.date.none.fl_str_mv 2015-10-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-86942015000500479
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-86942015000500479
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1016/j.bjorl.2015.07.006
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial.
publisher.none.fl_str_mv Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial.
dc.source.none.fl_str_mv Brazilian Journal of Otorhinolaryngology v.81 n.5 2015
reponame:Brazilian Journal of Otorhinolaryngology
instname:Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial (ABORL-CCF)
instacron:ABORL-CCF
instname_str Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial (ABORL-CCF)
instacron_str ABORL-CCF
institution ABORL-CCF
reponame_str Brazilian Journal of Otorhinolaryngology
collection Brazilian Journal of Otorhinolaryngology
repository.name.fl_str_mv Brazilian Journal of Otorhinolaryngology - Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial (ABORL-CCF)
repository.mail.fl_str_mv revista@aborlccf.org.br||revista@aborlccf.org.br
_version_ 1754575991676600320