Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging

Detalhes bibliográficos
Autor(a) principal: Correia,Fernando Faria
Data de Publicação: 2013
Outros Autores: Ramos,Isaac, Roberts,Cynthia J., Steinmueller,Andreas, Krug,Matthias, Ambrósio Jr.,Renato
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Arquivos brasileiros de oftalmologia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492013000500005
Resumo: PURPOSE: To study the deformation response of three distinct contact lenses with known structures, which served as corneal models, under different chamber pressures using ultra-high-speed (UHS) Scheimpflug imaging. METHODS: Three hydrophilic contact lenses were mounted on a sealed water chamber with precisely adjustable pressure: TAN-G5X (41% hydroxyethylmethacrylate/glycolmethacrylate, 550 µm thick), TAN-40 (62% hydroxyethylmethacrylate, 525 µm thick) and TAN-58 (42% methylmethacrylate, 258 µm thick). Each model was tested five times under different pressures (5, 15, 25, 35 and 45 mmHg), using ultra-high-speed Scheimpflug imaging during non-contact tonometry. 140 Scheimpflug images were taken with the UHS camera in each measurement. The deformation amplitude during non-contact tonometry was determined as the highest displacement of the apex at the highest concavity (HC) moment. RESULTS: At each pressure level, the deformation amplitude was statistically different for each lens tested (p<0.001, ANOVA). Each lens had different deformation amplitudes under different pressure levels (p<0.001; Bonferroni post-hoc test). The thicker lens with less polymer (TAN-G5X) had a higher deformation (less stiff behavior) than the one that was thinner but with more polymer (TAN-40), when measured at the same internal pressure. The thinnest lens with less polymers (TAN-58) had a lower deformation amplitude (stiffer behavior) at higher pressures than the thicker ones with more polymer (TAN-40 and TAN-G5X) at lower pressures. CONCLUSIONS: UHS Scheimpflug imaging allowed for biomechanical assessment through deformation characterization of corneal models. Biomechanical behavior was more influenced by material composition than by thickness. Chamber pressure had a significant impact on deformation response of each lens.
id CBO-2_4bb7f1a25c1b7c8fb49af65ffc4c37e3
oai_identifier_str oai:scielo:S0004-27492013000500005
network_acronym_str CBO-2
network_name_str Arquivos brasileiros de oftalmologia (Online)
repository_id_str
spelling Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imagingCornea/physiologyCorneal topography/methodsBiomechanicsTonometry, ocular/methodsPressurePURPOSE: To study the deformation response of three distinct contact lenses with known structures, which served as corneal models, under different chamber pressures using ultra-high-speed (UHS) Scheimpflug imaging. METHODS: Three hydrophilic contact lenses were mounted on a sealed water chamber with precisely adjustable pressure: TAN-G5X (41% hydroxyethylmethacrylate/glycolmethacrylate, 550 µm thick), TAN-40 (62% hydroxyethylmethacrylate, 525 µm thick) and TAN-58 (42% methylmethacrylate, 258 µm thick). Each model was tested five times under different pressures (5, 15, 25, 35 and 45 mmHg), using ultra-high-speed Scheimpflug imaging during non-contact tonometry. 140 Scheimpflug images were taken with the UHS camera in each measurement. The deformation amplitude during non-contact tonometry was determined as the highest displacement of the apex at the highest concavity (HC) moment. RESULTS: At each pressure level, the deformation amplitude was statistically different for each lens tested (p<0.001, ANOVA). Each lens had different deformation amplitudes under different pressure levels (p<0.001; Bonferroni post-hoc test). The thicker lens with less polymer (TAN-G5X) had a higher deformation (less stiff behavior) than the one that was thinner but with more polymer (TAN-40), when measured at the same internal pressure. The thinnest lens with less polymers (TAN-58) had a lower deformation amplitude (stiffer behavior) at higher pressures than the thicker ones with more polymer (TAN-40 and TAN-G5X) at lower pressures. CONCLUSIONS: UHS Scheimpflug imaging allowed for biomechanical assessment through deformation characterization of corneal models. Biomechanical behavior was more influenced by material composition than by thickness. Chamber pressure had a significant impact on deformation response of each lens.Conselho Brasileiro de Oftalmologia2013-10-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492013000500005Arquivos Brasileiros de Oftalmologia v.76 n.5 2013reponame:Arquivos brasileiros de oftalmologia (Online)instname:Conselho Brasileiro de Oftalmologia (CBO)instacron:CBO10.1590/S0004-27492013000500005info:eu-repo/semantics/openAccessCorreia,Fernando FariaRamos,IsaacRoberts,Cynthia J.Steinmueller,AndreasKrug,MatthiasAmbrósio Jr.,Renatoeng2013-11-12T00:00:00Zoai:scielo:S0004-27492013000500005Revistahttp://aboonline.org.br/https://old.scielo.br/oai/scielo-oai.phpaboonline@cbo.com.br||abo@cbo.com.br1678-29250004-2749opendoar:2013-11-12T00:00Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)false
dc.title.none.fl_str_mv Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
title Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
spellingShingle Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
Correia,Fernando Faria
Cornea/physiology
Corneal topography/methods
Biomechanics
Tonometry, ocular/methods
Pressure
title_short Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
title_full Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
title_fullStr Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
title_full_unstemmed Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
title_sort Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging
author Correia,Fernando Faria
author_facet Correia,Fernando Faria
Ramos,Isaac
Roberts,Cynthia J.
Steinmueller,Andreas
Krug,Matthias
Ambrósio Jr.,Renato
author_role author
author2 Ramos,Isaac
Roberts,Cynthia J.
Steinmueller,Andreas
Krug,Matthias
Ambrósio Jr.,Renato
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Correia,Fernando Faria
Ramos,Isaac
Roberts,Cynthia J.
Steinmueller,Andreas
Krug,Matthias
Ambrósio Jr.,Renato
dc.subject.por.fl_str_mv Cornea/physiology
Corneal topography/methods
Biomechanics
Tonometry, ocular/methods
Pressure
topic Cornea/physiology
Corneal topography/methods
Biomechanics
Tonometry, ocular/methods
Pressure
description PURPOSE: To study the deformation response of three distinct contact lenses with known structures, which served as corneal models, under different chamber pressures using ultra-high-speed (UHS) Scheimpflug imaging. METHODS: Three hydrophilic contact lenses were mounted on a sealed water chamber with precisely adjustable pressure: TAN-G5X (41% hydroxyethylmethacrylate/glycolmethacrylate, 550 µm thick), TAN-40 (62% hydroxyethylmethacrylate, 525 µm thick) and TAN-58 (42% methylmethacrylate, 258 µm thick). Each model was tested five times under different pressures (5, 15, 25, 35 and 45 mmHg), using ultra-high-speed Scheimpflug imaging during non-contact tonometry. 140 Scheimpflug images were taken with the UHS camera in each measurement. The deformation amplitude during non-contact tonometry was determined as the highest displacement of the apex at the highest concavity (HC) moment. RESULTS: At each pressure level, the deformation amplitude was statistically different for each lens tested (p<0.001, ANOVA). Each lens had different deformation amplitudes under different pressure levels (p<0.001; Bonferroni post-hoc test). The thicker lens with less polymer (TAN-G5X) had a higher deformation (less stiff behavior) than the one that was thinner but with more polymer (TAN-40), when measured at the same internal pressure. The thinnest lens with less polymers (TAN-58) had a lower deformation amplitude (stiffer behavior) at higher pressures than the thicker ones with more polymer (TAN-40 and TAN-G5X) at lower pressures. CONCLUSIONS: UHS Scheimpflug imaging allowed for biomechanical assessment through deformation characterization of corneal models. Biomechanical behavior was more influenced by material composition than by thickness. Chamber pressure had a significant impact on deformation response of each lens.
publishDate 2013
dc.date.none.fl_str_mv 2013-10-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492013000500005
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492013000500005
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0004-27492013000500005
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
dc.source.none.fl_str_mv Arquivos Brasileiros de Oftalmologia v.76 n.5 2013
reponame:Arquivos brasileiros de oftalmologia (Online)
instname:Conselho Brasileiro de Oftalmologia (CBO)
instacron:CBO
instname_str Conselho Brasileiro de Oftalmologia (CBO)
instacron_str CBO
institution CBO
reponame_str Arquivos brasileiros de oftalmologia (Online)
collection Arquivos brasileiros de oftalmologia (Online)
repository.name.fl_str_mv Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)
repository.mail.fl_str_mv aboonline@cbo.com.br||abo@cbo.com.br
_version_ 1754209027760324608