Tonometer calibration in Brasília, Brazil

Detalhes bibliográficos
Autor(a) principal: Abrão,Fernanda Pires da Silva
Data de Publicação: 2009
Outros Autores: Araújo,Warley Alves, Vieira,Geraldo Magela
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Arquivos brasileiros de oftalmologia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492009000300012
Resumo: PURPOSE: To determine calibration errors of Goldmann applanation tonometers in ophthalmic clinics of Brasília, Brazil, and correlate the findings with variables related to tonometers model and utilization. METHODS: Tonometers from ophthalmic clinics in Brasília, Brazil, were checked for calibration errors. A standard Goldmann applanation tonometer checking tool was used to asses the calibration error. Only one trained individual made all verifications, with a masked reading of the results. Data on the model, age, daily use, frequency of calibration checking and the nature of the ophthalmic department - private or public - were collected and correlated with the observed errors. RESULTS: One hundred tonometers were checked for calibration. Forty seven percent (47/100) were out of 1 mmHg range at least at one point checking. Tonometers mounted to slit lamp, with less than 5 years, used in less than 20 patients daily, that had a calibration check on a yearly basis, and those from private office exhibit a lower rate of inaccuracy, but only the first variable was statistically significant. Sixty one percent of tonometers on public hospitals were out of calibration. CONCLUSION: Calibration of tonometers in the capital of Brazil is poor; those from general hospitals are worst, and this fact can lead to inaccurate detection and assessment of glaucoma patients, overall in the population under government assistance.
id CBO-2_c46f13c27857a6b95887a3b21bdecf01
oai_identifier_str oai:scielo:S0004-27492009000300012
network_acronym_str CBO-2
network_name_str Arquivos brasileiros de oftalmologia (Online)
repository_id_str
spelling Tonometer calibration in Brasília, BrazilIntraocular pressureTonometry, ocular/instrumentationCalibrationEquipment failure analysisGlaucomaQuality controlPURPOSE: To determine calibration errors of Goldmann applanation tonometers in ophthalmic clinics of Brasília, Brazil, and correlate the findings with variables related to tonometers model and utilization. METHODS: Tonometers from ophthalmic clinics in Brasília, Brazil, were checked for calibration errors. A standard Goldmann applanation tonometer checking tool was used to asses the calibration error. Only one trained individual made all verifications, with a masked reading of the results. Data on the model, age, daily use, frequency of calibration checking and the nature of the ophthalmic department - private or public - were collected and correlated with the observed errors. RESULTS: One hundred tonometers were checked for calibration. Forty seven percent (47/100) were out of 1 mmHg range at least at one point checking. Tonometers mounted to slit lamp, with less than 5 years, used in less than 20 patients daily, that had a calibration check on a yearly basis, and those from private office exhibit a lower rate of inaccuracy, but only the first variable was statistically significant. Sixty one percent of tonometers on public hospitals were out of calibration. CONCLUSION: Calibration of tonometers in the capital of Brazil is poor; those from general hospitals are worst, and this fact can lead to inaccurate detection and assessment of glaucoma patients, overall in the population under government assistance.Conselho Brasileiro de Oftalmologia2009-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492009000300012Arquivos Brasileiros de Oftalmologia v.72 n.3 2009reponame:Arquivos brasileiros de oftalmologia (Online)instname:Conselho Brasileiro de Oftalmologia (CBO)instacron:CBO10.1590/S0004-27492009000300012info:eu-repo/semantics/openAccessAbrão,Fernanda Pires da SilvaAraújo,Warley AlvesVieira,Geraldo Magelaeng2009-08-04T00:00:00Zoai:scielo:S0004-27492009000300012Revistahttp://aboonline.org.br/https://old.scielo.br/oai/scielo-oai.phpaboonline@cbo.com.br||abo@cbo.com.br1678-29250004-2749opendoar:2009-08-04T00:00Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)false
dc.title.none.fl_str_mv Tonometer calibration in Brasília, Brazil
title Tonometer calibration in Brasília, Brazil
spellingShingle Tonometer calibration in Brasília, Brazil
Abrão,Fernanda Pires da Silva
Intraocular pressure
Tonometry, ocular/instrumentation
Calibration
Equipment failure analysis
Glaucoma
Quality control
title_short Tonometer calibration in Brasília, Brazil
title_full Tonometer calibration in Brasília, Brazil
title_fullStr Tonometer calibration in Brasília, Brazil
title_full_unstemmed Tonometer calibration in Brasília, Brazil
title_sort Tonometer calibration in Brasília, Brazil
author Abrão,Fernanda Pires da Silva
author_facet Abrão,Fernanda Pires da Silva
Araújo,Warley Alves
Vieira,Geraldo Magela
author_role author
author2 Araújo,Warley Alves
Vieira,Geraldo Magela
author2_role author
author
dc.contributor.author.fl_str_mv Abrão,Fernanda Pires da Silva
Araújo,Warley Alves
Vieira,Geraldo Magela
dc.subject.por.fl_str_mv Intraocular pressure
Tonometry, ocular/instrumentation
Calibration
Equipment failure analysis
Glaucoma
Quality control
topic Intraocular pressure
Tonometry, ocular/instrumentation
Calibration
Equipment failure analysis
Glaucoma
Quality control
description PURPOSE: To determine calibration errors of Goldmann applanation tonometers in ophthalmic clinics of Brasília, Brazil, and correlate the findings with variables related to tonometers model and utilization. METHODS: Tonometers from ophthalmic clinics in Brasília, Brazil, were checked for calibration errors. A standard Goldmann applanation tonometer checking tool was used to asses the calibration error. Only one trained individual made all verifications, with a masked reading of the results. Data on the model, age, daily use, frequency of calibration checking and the nature of the ophthalmic department - private or public - were collected and correlated with the observed errors. RESULTS: One hundred tonometers were checked for calibration. Forty seven percent (47/100) were out of 1 mmHg range at least at one point checking. Tonometers mounted to slit lamp, with less than 5 years, used in less than 20 patients daily, that had a calibration check on a yearly basis, and those from private office exhibit a lower rate of inaccuracy, but only the first variable was statistically significant. Sixty one percent of tonometers on public hospitals were out of calibration. CONCLUSION: Calibration of tonometers in the capital of Brazil is poor; those from general hospitals are worst, and this fact can lead to inaccurate detection and assessment of glaucoma patients, overall in the population under government assistance.
publishDate 2009
dc.date.none.fl_str_mv 2009-06-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492009000300012
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492009000300012
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0004-27492009000300012
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
dc.source.none.fl_str_mv Arquivos Brasileiros de Oftalmologia v.72 n.3 2009
reponame:Arquivos brasileiros de oftalmologia (Online)
instname:Conselho Brasileiro de Oftalmologia (CBO)
instacron:CBO
instname_str Conselho Brasileiro de Oftalmologia (CBO)
instacron_str CBO
institution CBO
reponame_str Arquivos brasileiros de oftalmologia (Online)
collection Arquivos brasileiros de oftalmologia (Online)
repository.name.fl_str_mv Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)
repository.mail.fl_str_mv aboonline@cbo.com.br||abo@cbo.com.br
_version_ 1754209025909587968