Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters
Autor(a) principal: | |
---|---|
Data de Publicação: | 2013 |
Outros Autores: | , , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Radiologia Brasileira (Online) |
Texto Completo: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-39842013000600358 |
Resumo: | Objective The authors have sought to study the calibration of a clinical PKA meter (Diamentor E2) and a calibrator for clinical meters (PDC) in the Laboratory of Ionizing Radiation Metrology at Instituto de Energia e Ambiente - Universidade de São Paulo. Materials and Methods Different qualities of both incident and transmitted beams were utilized in conditions similar to a clinical setting, analyzing the influence from the reference dosimeter, from the distance between meters, from the filtration and from the average beam energy. Calibrations were performed directly against a standard 30 cm3 cylindrical chamber or a parallel-plate monitor chamber, and indirectly against the PDC meter. Results The lowest energy dependence was observed for transmitted beams. The cross calibration between the Diamentor E2 and the PDC meters, and the PDC presented the greatest propagation of uncertainties. Conclusion The calibration coefficient of the PDC meter showed to be more stable with voltage, while the Diamentor E2 calibration coefficient was more variable. On the other hand, the PDC meter presented greater uncertainty in readings (5.0%) than with the use of the monitor chamber (3.5%) as a reference. |
id |
CBR-1_be152a55606161f1112e34169bd6e1b0 |
---|---|
oai_identifier_str |
oai:scielo:S0100-39842013000600358 |
network_acronym_str |
CBR-1 |
network_name_str |
Radiologia Brasileira (Online) |
repository_id_str |
|
spelling |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) metersKermaMetersCalibrationDosimetryRadiologyX-rays Objective The authors have sought to study the calibration of a clinical PKA meter (Diamentor E2) and a calibrator for clinical meters (PDC) in the Laboratory of Ionizing Radiation Metrology at Instituto de Energia e Ambiente - Universidade de São Paulo. Materials and Methods Different qualities of both incident and transmitted beams were utilized in conditions similar to a clinical setting, analyzing the influence from the reference dosimeter, from the distance between meters, from the filtration and from the average beam energy. Calibrations were performed directly against a standard 30 cm3 cylindrical chamber or a parallel-plate monitor chamber, and indirectly against the PDC meter. Results The lowest energy dependence was observed for transmitted beams. The cross calibration between the Diamentor E2 and the PDC meters, and the PDC presented the greatest propagation of uncertainties. Conclusion The calibration coefficient of the PDC meter showed to be more stable with voltage, while the Diamentor E2 calibration coefficient was more variable. On the other hand, the PDC meter presented greater uncertainty in readings (5.0%) than with the use of the monitor chamber (3.5%) as a reference. Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem2013-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-39842013000600358Radiologia Brasileira v.46 n.6 2013reponame:Radiologia Brasileira (Online)instname:Colégio Brasileiro de Radiologia e Diagnóstico por Imagem (CBR)instacron:CBR10.1590/S0100-39842013000600008info:eu-repo/semantics/openAccessTerini,Ricardo AndradeCampelo,Maria Carolina de SantanaAlmeida Jr.,José Neres deHerdade,Silvio BruniPereira,Marco Aurélio Guedeseng2015-09-03T00:00:00Zoai:scielo:S0100-39842013000600358Revistahttps://www.scielo.br/j/rb/https://old.scielo.br/oai/scielo-oai.phpradiologiabrasileira@cbr.org.br1678-70990100-3984opendoar:2015-09-03T00:00Radiologia Brasileira (Online) - Colégio Brasileiro de Radiologia e Diagnóstico por Imagem (CBR)false |
dc.title.none.fl_str_mv |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters |
title |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters |
spellingShingle |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters Terini,Ricardo Andrade Kerma Meters Calibration Dosimetry Radiology X-rays |
title_short |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters |
title_full |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters |
title_fullStr |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters |
title_full_unstemmed |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters |
title_sort |
Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters |
author |
Terini,Ricardo Andrade |
author_facet |
Terini,Ricardo Andrade Campelo,Maria Carolina de Santana Almeida Jr.,José Neres de Herdade,Silvio Bruni Pereira,Marco Aurélio Guedes |
author_role |
author |
author2 |
Campelo,Maria Carolina de Santana Almeida Jr.,José Neres de Herdade,Silvio Bruni Pereira,Marco Aurélio Guedes |
author2_role |
author author author author |
dc.contributor.author.fl_str_mv |
Terini,Ricardo Andrade Campelo,Maria Carolina de Santana Almeida Jr.,José Neres de Herdade,Silvio Bruni Pereira,Marco Aurélio Guedes |
dc.subject.por.fl_str_mv |
Kerma Meters Calibration Dosimetry Radiology X-rays |
topic |
Kerma Meters Calibration Dosimetry Radiology X-rays |
description |
Objective The authors have sought to study the calibration of a clinical PKA meter (Diamentor E2) and a calibrator for clinical meters (PDC) in the Laboratory of Ionizing Radiation Metrology at Instituto de Energia e Ambiente - Universidade de São Paulo. Materials and Methods Different qualities of both incident and transmitted beams were utilized in conditions similar to a clinical setting, analyzing the influence from the reference dosimeter, from the distance between meters, from the filtration and from the average beam energy. Calibrations were performed directly against a standard 30 cm3 cylindrical chamber or a parallel-plate monitor chamber, and indirectly against the PDC meter. Results The lowest energy dependence was observed for transmitted beams. The cross calibration between the Diamentor E2 and the PDC meters, and the PDC presented the greatest propagation of uncertainties. Conclusion The calibration coefficient of the PDC meter showed to be more stable with voltage, while the Diamentor E2 calibration coefficient was more variable. On the other hand, the PDC meter presented greater uncertainty in readings (5.0%) than with the use of the monitor chamber (3.5%) as a reference. |
publishDate |
2013 |
dc.date.none.fl_str_mv |
2013-12-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-39842013000600358 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-39842013000600358 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.1590/S0100-39842013000600008 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem |
publisher.none.fl_str_mv |
Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem |
dc.source.none.fl_str_mv |
Radiologia Brasileira v.46 n.6 2013 reponame:Radiologia Brasileira (Online) instname:Colégio Brasileiro de Radiologia e Diagnóstico por Imagem (CBR) instacron:CBR |
instname_str |
Colégio Brasileiro de Radiologia e Diagnóstico por Imagem (CBR) |
instacron_str |
CBR |
institution |
CBR |
reponame_str |
Radiologia Brasileira (Online) |
collection |
Radiologia Brasileira (Online) |
repository.name.fl_str_mv |
Radiologia Brasileira (Online) - Colégio Brasileiro de Radiologia e Diagnóstico por Imagem (CBR) |
repository.mail.fl_str_mv |
radiologiabrasileira@cbr.org.br |
_version_ |
1754208938490855424 |