Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope

Detalhes bibliográficos
Autor(a) principal: Barros Junior, Tadeu Vieira de
Data de Publicação: 2018
Tipo de documento: Dissertação
Idioma: por
Título da fonte: Repositório do Centro Universitário Braz Cubas
Texto Completo: https://repositorio.cruzeirodosul.edu.br/handle/123456789/2235
Resumo: Controlling the level of forces released by the appliances during orthodontic treatment is important to ensure treatment efficiency and minimize side effects. The objective of this work was to compare the forces of loading and unloading released by the Forsus and Powerscope devices. Two groups of devices - Group F (Forsus; n = 12) and Group P (Powerscope; n = 12) - underwent loading and unloading tests after being immersed in artificial saliva for 5 minutes. The devices were loaded to the maximum point of loading (Group F 12 mm, Group P 4 mm) and unloaded at the rate of 0.1 mm per second. The amount of forces released by each device was recorded at intervals of 0.5 mm. Data were analyzed using Student's t-test (p <0.05). Group F had on average 111.79 g of loading force and 90.58 g of unloading force (variation of 22.21 g). Group P presented on average 203.5 g of loading force and 140.87 g of unloading force (variation of 62.63 g). There was no significant difference between loading and stiffness for both Group F (p = 0.070) and Group P (p = 0.143). Group P presented higher loading stiffness (p <0.001) and unloading (p <0.001) compared to Group F. Powerscope released more force than Forsus. However, Forsus lost less force during unloading. When comparing the characteristics between them, we can conclude that the spring of the Powescope is stiffer and less resilient than that of the Forsus.
id CUB_ec60b8704d14a0fdb69bcb8402716ef2
oai_identifier_str oai:repositorio.cruzeirodosul.edu.br:123456789/2235
network_acronym_str CUB
network_name_str Repositório do Centro Universitário Braz Cubas
repository_id_str
spelling Estudo comparativo entre os propulsores mandibulares Forsus e PowerScopeOdontologiaOclusão dentáriaDescompressão cirúrgicaTeste de materiaisCNPQ::CIENCIAS DA SAUDE::ODONTOLOGIAControlling the level of forces released by the appliances during orthodontic treatment is important to ensure treatment efficiency and minimize side effects. The objective of this work was to compare the forces of loading and unloading released by the Forsus and Powerscope devices. Two groups of devices - Group F (Forsus; n = 12) and Group P (Powerscope; n = 12) - underwent loading and unloading tests after being immersed in artificial saliva for 5 minutes. The devices were loaded to the maximum point of loading (Group F 12 mm, Group P 4 mm) and unloaded at the rate of 0.1 mm per second. The amount of forces released by each device was recorded at intervals of 0.5 mm. Data were analyzed using Student's t-test (p <0.05). Group F had on average 111.79 g of loading force and 90.58 g of unloading force (variation of 22.21 g). Group P presented on average 203.5 g of loading force and 140.87 g of unloading force (variation of 62.63 g). There was no significant difference between loading and stiffness for both Group F (p = 0.070) and Group P (p = 0.143). Group P presented higher loading stiffness (p <0.001) and unloading (p <0.001) compared to Group F. Powerscope released more force than Forsus. However, Forsus lost less force during unloading. When comparing the characteristics between them, we can conclude that the spring of the Powescope is stiffer and less resilient than that of the Forsus.O controle do nível de forças liberadas pelos aparelhos durante o tratamento ortodôntico é importante para garantir a eficiência do tratamento e minimizar efeitos indesejados. O objetivo deste trabalho foi comparar as forças de compressão e descompressão geradas pelos aparelhos Forsus e Powerscope. Dois grupos de aparelhos – Grupo F (Forsus; n=12) e Grupo P (Powerscope; n=12) – foram submetidos a testes de compressão e descompressão após terem sido imersos em saliva artificial por 5 minutos. Os aparelhos foram comprimidos até o ponto máximo de compressão (Grupo F 12 mm; Grupo P 4 mm) e descomprimidos numa velocidade de 0,1 mm por segundo. Os valores das forças liberadas por cada aparelho foram registrados a cada intervalo de 0,5 mm. Os dados foram analisados por meio do teste t de Student (p<0,05). O Grupo F apresentou em média 111,79 g de força de compressão e 90,58 g de força de descompressão (variação de 22,21 g). O Grupo P apresentou em média 203,5 g de força de compressão e 140,87 g de força de descompressão (variação de 62,63 g). Não houve diferença significativa entre a rigidez de compressão e descompressão, tanto para o Grupo F (p=0,070) como para o grupo P (p=0,143). O Grupo P apresentou maior rigidez de compressão (p<0,001) e descompressão (p<0,001) em comparação ao Grupo F. O Powerscope liberou mais força que o Forsus. Porém, o Forsus perdeu menos força durante a descompressão. Ao se comparar as características entre eles pode-se concluir que a mola do Powescope é mais rígida e menos resiliente que a do Forsus.Universidade PositivoBrasilPós-GraduaçãoPrograma de Pós-Graduação em Odontologia ClínicaUPMoro, Alexandrehttp://lattes.cnpq.br/4812180363054225Barros Junior, Tadeu Vieira de2021-05-24T19:54:30Z20182021-05-24T19:54:30Z2018info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/masterThesisapplication/pdfhttps://repositorio.cruzeirodosul.edu.br/handle/123456789/2235porinfo:eu-repo/semantics/openAccessreponame:Repositório do Centro Universitário Braz Cubasinstname:Centro Universitário Braz Cubas (CUB)instacron:CUB2021-06-17T17:40:39Zoai:repositorio.cruzeirodosul.edu.br:123456789/2235Repositório InstitucionalPUBhttps://repositorio.brazcubas.edu.br/oai/requestbibli@brazcubas.edu.bropendoar:2021-06-17T17:40:39Repositório do Centro Universitário Braz Cubas - Centro Universitário Braz Cubas (CUB)false
dc.title.none.fl_str_mv Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
title Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
spellingShingle Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
Barros Junior, Tadeu Vieira de
Odontologia
Oclusão dentária
Descompressão cirúrgica
Teste de materiais
CNPQ::CIENCIAS DA SAUDE::ODONTOLOGIA
title_short Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
title_full Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
title_fullStr Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
title_full_unstemmed Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
title_sort Estudo comparativo entre os propulsores mandibulares Forsus e PowerScope
author Barros Junior, Tadeu Vieira de
author_facet Barros Junior, Tadeu Vieira de
author_role author
dc.contributor.none.fl_str_mv Moro, Alexandre
http://lattes.cnpq.br/4812180363054225
dc.contributor.author.fl_str_mv Barros Junior, Tadeu Vieira de
dc.subject.por.fl_str_mv Odontologia
Oclusão dentária
Descompressão cirúrgica
Teste de materiais
CNPQ::CIENCIAS DA SAUDE::ODONTOLOGIA
topic Odontologia
Oclusão dentária
Descompressão cirúrgica
Teste de materiais
CNPQ::CIENCIAS DA SAUDE::ODONTOLOGIA
description Controlling the level of forces released by the appliances during orthodontic treatment is important to ensure treatment efficiency and minimize side effects. The objective of this work was to compare the forces of loading and unloading released by the Forsus and Powerscope devices. Two groups of devices - Group F (Forsus; n = 12) and Group P (Powerscope; n = 12) - underwent loading and unloading tests after being immersed in artificial saliva for 5 minutes. The devices were loaded to the maximum point of loading (Group F 12 mm, Group P 4 mm) and unloaded at the rate of 0.1 mm per second. The amount of forces released by each device was recorded at intervals of 0.5 mm. Data were analyzed using Student's t-test (p <0.05). Group F had on average 111.79 g of loading force and 90.58 g of unloading force (variation of 22.21 g). Group P presented on average 203.5 g of loading force and 140.87 g of unloading force (variation of 62.63 g). There was no significant difference between loading and stiffness for both Group F (p = 0.070) and Group P (p = 0.143). Group P presented higher loading stiffness (p <0.001) and unloading (p <0.001) compared to Group F. Powerscope released more force than Forsus. However, Forsus lost less force during unloading. When comparing the characteristics between them, we can conclude that the spring of the Powescope is stiffer and less resilient than that of the Forsus.
publishDate 2018
dc.date.none.fl_str_mv 2018
2018
2021-05-24T19:54:30Z
2021-05-24T19:54:30Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/masterThesis
format masterThesis
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://repositorio.cruzeirodosul.edu.br/handle/123456789/2235
url https://repositorio.cruzeirodosul.edu.br/handle/123456789/2235
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade Positivo
Brasil
Pós-Graduação
Programa de Pós-Graduação em Odontologia Clínica
UP
publisher.none.fl_str_mv Universidade Positivo
Brasil
Pós-Graduação
Programa de Pós-Graduação em Odontologia Clínica
UP
dc.source.none.fl_str_mv reponame:Repositório do Centro Universitário Braz Cubas
instname:Centro Universitário Braz Cubas (CUB)
instacron:CUB
instname_str Centro Universitário Braz Cubas (CUB)
instacron_str CUB
institution CUB
reponame_str Repositório do Centro Universitário Braz Cubas
collection Repositório do Centro Universitário Braz Cubas
repository.name.fl_str_mv Repositório do Centro Universitário Braz Cubas - Centro Universitário Braz Cubas (CUB)
repository.mail.fl_str_mv bibli@brazcubas.edu.br
_version_ 1798311355537686528