Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)

Detalhes bibliográficos
Autor(a) principal: Souza, Paulo Ricardo de Souza e
Data de Publicação: 2013
Outros Autores: Carvalho, Juliana Machado de, Albert, André Luis Mazzei, Moreira, Josino Costa, Leandro, Katia Christina
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Vigilância Sanitária em Debate
Texto Completo: https://visaemdebate.incqs.fiocruz.br/index.php/visaemdebate/article/view/16
Resumo: Valproic acid (VA) is used as an anticonvulsant and is used in the treatment of bipolar disorder and depression. The official analytical method for this drug in the compendia is gas chromatography, which is unavailable in many quality control laboratories. Here, we report a validated alternative method using high performance liquid chromatography with diode array detection (HPLC-DAD). The optimized parameters and conditions were as follows: C18 column (5 µm; 250 x 4 mm d.i.); flow 1.0 mL.min-1; wavelength: 210 nm; mobile phase: 55% acetonitrile (ACN) in water containing 0.05% v/v trifluoroacetic acid (TFA) (v/v). The analytical parameters that were validated included the selectivity and matrix effects, linearity, repeatability and intermediate, precision, accuracy, recovery and robustness. This method identifies VA unambiguously. In validation, the following results were obtained: good linearity in concentrations between 0.7 and 1.3 mg.mL-1 (r2 = 0.9998), relative standard deviations (RSDs) of 0.68% for repeatability and 1.23% for intermediate precision, a recovery of 99.42 to 101.55% (RSD 0.14 to 0.74%) and an accuracy of 100.68% (RSD = 0.79%). This method is robust to small variations in procedure.
id FIOCRUZ-9_7236345b900dfdb6b6a71f64bc682ee5
oai_identifier_str oai:ojs.visaemdebate.incqs.fiocruz.br:article/16
network_acronym_str FIOCRUZ-9
network_name_str Vigilância Sanitária em Debate
repository_id_str
spelling Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)cromatografia líquida de alta eficiência acoplada com detector de conjunto de fotodiodosácido valpróicovalidaçãohigh performance liquid chromatographyHPLC-DADvalproic acidvalidationValproic acid (VA) is used as an anticonvulsant and is used in the treatment of bipolar disorder and depression. The official analytical method for this drug in the compendia is gas chromatography, which is unavailable in many quality control laboratories. Here, we report a validated alternative method using high performance liquid chromatography with diode array detection (HPLC-DAD). The optimized parameters and conditions were as follows: C18 column (5 µm; 250 x 4 mm d.i.); flow 1.0 mL.min-1; wavelength: 210 nm; mobile phase: 55% acetonitrile (ACN) in water containing 0.05% v/v trifluoroacetic acid (TFA) (v/v). The analytical parameters that were validated included the selectivity and matrix effects, linearity, repeatability and intermediate, precision, accuracy, recovery and robustness. This method identifies VA unambiguously. In validation, the following results were obtained: good linearity in concentrations between 0.7 and 1.3 mg.mL-1 (r2 = 0.9998), relative standard deviations (RSDs) of 0.68% for repeatability and 1.23% for intermediate precision, a recovery of 99.42 to 101.55% (RSD 0.14 to 0.74%) and an accuracy of 100.68% (RSD = 0.79%). This method is robust to small variations in procedure.Título em português: Desenvolvimento e validação de um método para a determinação de ácido valpróico em formulações farmacêuticas por cromatografia líquida de alta eficiência acoplada com detector de conjunto de fotodiodosO ácido valpróico (VA) é utilizado como anticonvulsivante e também na terapêutica da desordem bipolar e depressão. O método de análise deste medicamento nos compêndios oficiais é por cromatografia gasosa e muitos laboratórios oficiais de controle de qualidade não possuem tal equipamento. A proposta deste estudo foi desenvolver e validar um método analítico por cromatografia líquida de alta eficiência acoplada com detector de conjunto de fotodiodos para a análise deste fármaco. Os parâmetros e condições otimizados foram: coluna C18 (5 µm; 250 x 4 mm d.i.), fluxo de 1,0 mL min-1, comprimento de onda 210 nm, fase móvel 55% ACN e 45% solução 0,05% v/v de ácido trifluoracético (v/v). Os parâmetros analíticos avaliados na validação foram: seletividade e efeito matriz, linearidade, repetitividade e precisão intermediária, exatidão e robustez. A metodologia desenvolvida mostrou identificar de forma inequívoca o analito de interesse. No que concerne aos parâmetros de validação foram obtidos: faixa linear nas concentrações de 0,7 a 1,3 mg mL-1 (r2 = 0,9998), repetitividade: desvio padrão relativo (DPR) = 0,68%, precisão intermediária: DPR = 1,23%, recuperação de 99,42 a 101,55% (DPR de 0,14 a 0,74%) e exatidão de 100,68% (DPR = 0,79%). Quando submetido a pequenas variações, a metodologia mostrou-se robusta.Instituto Nacional de Controle de Qualidade em Saúde2013-02-25info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersion"Peer-reviewed article""Artículo revisado por pares""Artigo avaliado pelos pares"application/pdfhttps://visaemdebate.incqs.fiocruz.br/index.php/visaemdebate/article/view/1610.3395/vd.v1n1.16Health Surveillance under Debate: Society, Science & Technology ; Vol. 1 No. 1 (2013): February; 52-58Vigilancia en Salud en Debate: Sociedad, Ciencia y Tecnología; Vol. 1 Núm. 1 (2013): Febrero; 52-58Vigil Sanit Debate, Rio de Janeiro; v. 1 n. 1 (2013): Fevereiro; 52-582317-269Xreponame:Vigilância Sanitária em Debateinstname:Fundação Oswaldo Cruz (FIOCRUZ)instacron:FIOCRUZenghttps://visaemdebate.incqs.fiocruz.br/index.php/visaemdebate/article/view/16/14Copyright (c) 2013 Vigilância Sanitária em Debate: Sociedade, Ciência & Tecnologia (Health Surveillance under Debate: Society, Science & Technology) – Visa em Debatehttps://creativecommons.org/licenses/by-nc-nd/4.0info:eu-repo/semantics/openAccessSouza, Paulo Ricardo de Souza eCarvalho, Juliana Machado deAlbert, André Luis MazzeiMoreira, Josino CostaLeandro, Katia Christina2023-06-27T14:54:35Zoai:ojs.visaemdebate.incqs.fiocruz.br:article/16Revistahttps://visaemdebate.incqs.fiocruz.br/index.php/visaemdebatePUBhttps://visaemdebate.incqs.fiocruz.br/index.php/visaemdebate/oaiincqs.visaemdebate@fiocruz.br || gisele.neves@fiocruz.br2317-269X2317-269Xopendoar:2023-06-27T14:54:35Vigilância Sanitária em Debate - Fundação Oswaldo Cruz (FIOCRUZ)false
dc.title.none.fl_str_mv Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
title Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
spellingShingle Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
Souza, Paulo Ricardo de Souza e
cromatografia líquida de alta eficiência acoplada com detector de conjunto de fotodiodos
ácido valpróico
validação
high performance liquid chromatography
HPLC-DAD
valproic acid
validation
title_short Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
title_full Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
title_fullStr Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
title_full_unstemmed Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
title_sort Development and validation of a method for the determination of valproic acid in pharmaceutical formulations by high performance liquid chromatography with diode array detection (HPLC-DAD)
author Souza, Paulo Ricardo de Souza e
author_facet Souza, Paulo Ricardo de Souza e
Carvalho, Juliana Machado de
Albert, André Luis Mazzei
Moreira, Josino Costa
Leandro, Katia Christina
author_role author
author2 Carvalho, Juliana Machado de
Albert, André Luis Mazzei
Moreira, Josino Costa
Leandro, Katia Christina
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Souza, Paulo Ricardo de Souza e
Carvalho, Juliana Machado de
Albert, André Luis Mazzei
Moreira, Josino Costa
Leandro, Katia Christina
dc.subject.por.fl_str_mv cromatografia líquida de alta eficiência acoplada com detector de conjunto de fotodiodos
ácido valpróico
validação
high performance liquid chromatography
HPLC-DAD
valproic acid
validation
topic cromatografia líquida de alta eficiência acoplada com detector de conjunto de fotodiodos
ácido valpróico
validação
high performance liquid chromatography
HPLC-DAD
valproic acid
validation
description Valproic acid (VA) is used as an anticonvulsant and is used in the treatment of bipolar disorder and depression. The official analytical method for this drug in the compendia is gas chromatography, which is unavailable in many quality control laboratories. Here, we report a validated alternative method using high performance liquid chromatography with diode array detection (HPLC-DAD). The optimized parameters and conditions were as follows: C18 column (5 µm; 250 x 4 mm d.i.); flow 1.0 mL.min-1; wavelength: 210 nm; mobile phase: 55% acetonitrile (ACN) in water containing 0.05% v/v trifluoroacetic acid (TFA) (v/v). The analytical parameters that were validated included the selectivity and matrix effects, linearity, repeatability and intermediate, precision, accuracy, recovery and robustness. This method identifies VA unambiguously. In validation, the following results were obtained: good linearity in concentrations between 0.7 and 1.3 mg.mL-1 (r2 = 0.9998), relative standard deviations (RSDs) of 0.68% for repeatability and 1.23% for intermediate precision, a recovery of 99.42 to 101.55% (RSD 0.14 to 0.74%) and an accuracy of 100.68% (RSD = 0.79%). This method is robust to small variations in procedure.
publishDate 2013
dc.date.none.fl_str_mv 2013-02-25
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
"Peer-reviewed article"
"Artículo revisado por pares"
"Artigo avaliado pelos pares"
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://visaemdebate.incqs.fiocruz.br/index.php/visaemdebate/article/view/16
10.3395/vd.v1n1.16
url https://visaemdebate.incqs.fiocruz.br/index.php/visaemdebate/article/view/16
identifier_str_mv 10.3395/vd.v1n1.16
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://visaemdebate.incqs.fiocruz.br/index.php/visaemdebate/article/view/16/14
dc.rights.driver.fl_str_mv https://creativecommons.org/licenses/by-nc-nd/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by-nc-nd/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Instituto Nacional de Controle de Qualidade em Saúde
publisher.none.fl_str_mv Instituto Nacional de Controle de Qualidade em Saúde
dc.source.none.fl_str_mv Health Surveillance under Debate: Society, Science & Technology ; Vol. 1 No. 1 (2013): February; 52-58
Vigilancia en Salud en Debate: Sociedad, Ciencia y Tecnología; Vol. 1 Núm. 1 (2013): Febrero; 52-58
Vigil Sanit Debate, Rio de Janeiro; v. 1 n. 1 (2013): Fevereiro; 52-58
2317-269X
reponame:Vigilância Sanitária em Debate
instname:Fundação Oswaldo Cruz (FIOCRUZ)
instacron:FIOCRUZ
instname_str Fundação Oswaldo Cruz (FIOCRUZ)
instacron_str FIOCRUZ
institution FIOCRUZ
reponame_str Vigilância Sanitária em Debate
collection Vigilância Sanitária em Debate
repository.name.fl_str_mv Vigilância Sanitária em Debate - Fundação Oswaldo Cruz (FIOCRUZ)
repository.mail.fl_str_mv incqs.visaemdebate@fiocruz.br || gisele.neves@fiocruz.br
_version_ 1797042043310047232