Implant/abutment vertical misfit of one-piece cast frameworks made with different materials

Detalhes bibliográficos
Autor(a) principal: Barbosa,Gustavo Augusto Seabra
Data de Publicação: 2010
Outros Autores: Neves,Flávio Domingues das, Mattos,Maria da Gloria Chiarello de, Rodrigues,Renata Cristina Silveira, Ribeiro,Ricardo Faria
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Dental Journal
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402010000600006
Resumo: This study compared vertical and passive fit of one-piece cast frameworks made with 3 different materials: commercially pure titanium (CP Ti - G1), cobalt-chromium alloy (Co-Cr - G2) and nickel-chromium-titanium alloy (Ni-Cr-Ti - G3). Fifteen frameworks were obtained simulating bars for fixed prosthesis in a model with 5 implants. The passive and vertical fit of the framework interface was measured using an optical microscope at ×30 magnification. Data were statistically analyzed by ANOVA and LSD tests (α=0.05). Mean and standard deviation values for passive fit and vertical fit were, respectively: G1 [472.49 (109.88) µm and 29.9 (13.24) µm], G2 [584.84 (120.20) µm and 27.05 (10.30) µm], and G3 [462.70 (179.18) µm and 24.95 (11.14) µm]. For vertical fit, there were no significant differences among G1, G2 and G3 (p=0.285). There were no significant differences for passive fit between G1 and G3 (p=0.844), but both differed significantly from G2 (p=0.028 and p=0.035, respectively), which showed the highest misfit values. It may be concluded that the vertical fit of frameworks was not affected by the tested materials, and that one-piece cast frameworks resulted in inadequate passive fit. The Co-Cr alloy presented the worst values for passive fit.
id FUNORP-1_4eb854dec1c110c38ed8c87675d62e44
oai_identifier_str oai:scielo:S0103-64402010000600006
network_acronym_str FUNORP-1
network_name_str Brazilian Dental Journal
repository_id_str
spelling Implant/abutment vertical misfit of one-piece cast frameworks made with different materialsdental implantsprosthesis fixed by implantsprosthesis fitThis study compared vertical and passive fit of one-piece cast frameworks made with 3 different materials: commercially pure titanium (CP Ti - G1), cobalt-chromium alloy (Co-Cr - G2) and nickel-chromium-titanium alloy (Ni-Cr-Ti - G3). Fifteen frameworks were obtained simulating bars for fixed prosthesis in a model with 5 implants. The passive and vertical fit of the framework interface was measured using an optical microscope at ×30 magnification. Data were statistically analyzed by ANOVA and LSD tests (α=0.05). Mean and standard deviation values for passive fit and vertical fit were, respectively: G1 [472.49 (109.88) µm and 29.9 (13.24) µm], G2 [584.84 (120.20) µm and 27.05 (10.30) µm], and G3 [462.70 (179.18) µm and 24.95 (11.14) µm]. For vertical fit, there were no significant differences among G1, G2 and G3 (p=0.285). There were no significant differences for passive fit between G1 and G3 (p=0.844), but both differed significantly from G2 (p=0.028 and p=0.035, respectively), which showed the highest misfit values. It may be concluded that the vertical fit of frameworks was not affected by the tested materials, and that one-piece cast frameworks resulted in inadequate passive fit. The Co-Cr alloy presented the worst values for passive fit.Fundação Odontológica de Ribeirão Preto2010-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402010000600006Brazilian Dental Journal v.21 n.6 2010reponame:Brazilian Dental Journalinstname:Fundação Odontológica de Ribeirão Preto (FUNORP)instacron:FUNORP10.1590/S0103-64402010000600006info:eu-repo/semantics/openAccessBarbosa,Gustavo Augusto SeabraNeves,Flávio Domingues dasMattos,Maria da Gloria Chiarello deRodrigues,Renata Cristina SilveiraRibeiro,Ricardo Fariaeng2011-01-17T00:00:00Zoai:scielo:S0103-64402010000600006Revistahttps://www.scielo.br/j/bdj/https://old.scielo.br/oai/scielo-oai.phpbdj@forp.usp.br||sergio@fosjc.unesp.br1806-47600103-6440opendoar:2011-01-17T00:00Brazilian Dental Journal - Fundação Odontológica de Ribeirão Preto (FUNORP)false
dc.title.none.fl_str_mv Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
title Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
spellingShingle Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
Barbosa,Gustavo Augusto Seabra
dental implants
prosthesis fixed by implants
prosthesis fit
title_short Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
title_full Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
title_fullStr Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
title_full_unstemmed Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
title_sort Implant/abutment vertical misfit of one-piece cast frameworks made with different materials
author Barbosa,Gustavo Augusto Seabra
author_facet Barbosa,Gustavo Augusto Seabra
Neves,Flávio Domingues das
Mattos,Maria da Gloria Chiarello de
Rodrigues,Renata Cristina Silveira
Ribeiro,Ricardo Faria
author_role author
author2 Neves,Flávio Domingues das
Mattos,Maria da Gloria Chiarello de
Rodrigues,Renata Cristina Silveira
Ribeiro,Ricardo Faria
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Barbosa,Gustavo Augusto Seabra
Neves,Flávio Domingues das
Mattos,Maria da Gloria Chiarello de
Rodrigues,Renata Cristina Silveira
Ribeiro,Ricardo Faria
dc.subject.por.fl_str_mv dental implants
prosthesis fixed by implants
prosthesis fit
topic dental implants
prosthesis fixed by implants
prosthesis fit
description This study compared vertical and passive fit of one-piece cast frameworks made with 3 different materials: commercially pure titanium (CP Ti - G1), cobalt-chromium alloy (Co-Cr - G2) and nickel-chromium-titanium alloy (Ni-Cr-Ti - G3). Fifteen frameworks were obtained simulating bars for fixed prosthesis in a model with 5 implants. The passive and vertical fit of the framework interface was measured using an optical microscope at ×30 magnification. Data were statistically analyzed by ANOVA and LSD tests (α=0.05). Mean and standard deviation values for passive fit and vertical fit were, respectively: G1 [472.49 (109.88) µm and 29.9 (13.24) µm], G2 [584.84 (120.20) µm and 27.05 (10.30) µm], and G3 [462.70 (179.18) µm and 24.95 (11.14) µm]. For vertical fit, there were no significant differences among G1, G2 and G3 (p=0.285). There were no significant differences for passive fit between G1 and G3 (p=0.844), but both differed significantly from G2 (p=0.028 and p=0.035, respectively), which showed the highest misfit values. It may be concluded that the vertical fit of frameworks was not affected by the tested materials, and that one-piece cast frameworks resulted in inadequate passive fit. The Co-Cr alloy presented the worst values for passive fit.
publishDate 2010
dc.date.none.fl_str_mv 2010-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402010000600006
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-64402010000600006
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0103-64402010000600006
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Fundação Odontológica de Ribeirão Preto
publisher.none.fl_str_mv Fundação Odontológica de Ribeirão Preto
dc.source.none.fl_str_mv Brazilian Dental Journal v.21 n.6 2010
reponame:Brazilian Dental Journal
instname:Fundação Odontológica de Ribeirão Preto (FUNORP)
instacron:FUNORP
instname_str Fundação Odontológica de Ribeirão Preto (FUNORP)
instacron_str FUNORP
institution FUNORP
reponame_str Brazilian Dental Journal
collection Brazilian Dental Journal
repository.name.fl_str_mv Brazilian Dental Journal - Fundação Odontológica de Ribeirão Preto (FUNORP)
repository.mail.fl_str_mv bdj@forp.usp.br||sergio@fosjc.unesp.br
_version_ 1754204091676884992