Model of single-sized endotracheal tube for adults

Detalhes bibliográficos
Autor(a) principal: Calderon,Luiz Guilherme
Data de Publicação: 2020
Outros Autores: Moreira,Marcos Mello, Barreto,Gilson, Tincani,Alfio José
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Einstein (São Paulo)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082020000100204
Resumo: ABSTRACT Objective To simulate different diameters of endotracheal tubes and to verify the fluid dynamics aspects by means of flow and resistance measurements. Methods Fluid dynamics software was used to calculate mean flow and airway resistance in endotracheal tube with a diameter of 6.0, 7.0, 7.5, 8.0, 9.0 and 10.0mm at normal body temperature and under constant pressure. The same measurements were taken in the fusion of the first 22cm of a 9.0mm endotracheal tube with 10.0mm diameter, and with the end part in 12cm of a 6.0mm endotracheal tube with 7.0mm diameter. Results The fusion of the first 22cm of an endotracheal tube of 10.0mm diameter with the terminal part in 12cm of an endotracheal tube of 6.0mm diameter, preserving the total length of 34cm, generated average flow and airway resistance similar to that of a conventional 7.5mm endotracheal tube. Conclusion This simulation study demonstrates that a single-sized endotracheal tube may facilitate endotracheal intubation without causing increased airway resistance.
id IIEPAE-1_b90ee9839fd1490c4a2679993545e959
oai_identifier_str oai:scielo:S1679-45082020000100204
network_acronym_str IIEPAE-1
network_name_str Einstein (São Paulo)
repository_id_str
spelling Model of single-sized endotracheal tube for adultsAirway resistanceIntubation, intratrachealTracheal intubationFlow rateABSTRACT Objective To simulate different diameters of endotracheal tubes and to verify the fluid dynamics aspects by means of flow and resistance measurements. Methods Fluid dynamics software was used to calculate mean flow and airway resistance in endotracheal tube with a diameter of 6.0, 7.0, 7.5, 8.0, 9.0 and 10.0mm at normal body temperature and under constant pressure. The same measurements were taken in the fusion of the first 22cm of a 9.0mm endotracheal tube with 10.0mm diameter, and with the end part in 12cm of a 6.0mm endotracheal tube with 7.0mm diameter. Results The fusion of the first 22cm of an endotracheal tube of 10.0mm diameter with the terminal part in 12cm of an endotracheal tube of 6.0mm diameter, preserving the total length of 34cm, generated average flow and airway resistance similar to that of a conventional 7.5mm endotracheal tube. Conclusion This simulation study demonstrates that a single-sized endotracheal tube may facilitate endotracheal intubation without causing increased airway resistance.Instituto Israelita de Ensino e Pesquisa Albert Einstein2020-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082020000100204einstein (São Paulo) v.18 2020reponame:Einstein (São Paulo)instname:Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE)instacron:IIEPAE10.31744/einstein_journal/2020ao4805info:eu-repo/semantics/openAccessCalderon,Luiz GuilhermeMoreira,Marcos MelloBarreto,GilsonTincani,Alfio Joséeng2019-10-14T00:00:00Zoai:scielo:S1679-45082020000100204Revistahttps://journal.einstein.br/pt-br/ONGhttps://old.scielo.br/oai/scielo-oai.php||revista@einstein.br2317-63851679-4508opendoar:2019-10-14T00:00Einstein (São Paulo) - Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE)false
dc.title.none.fl_str_mv Model of single-sized endotracheal tube for adults
title Model of single-sized endotracheal tube for adults
spellingShingle Model of single-sized endotracheal tube for adults
Calderon,Luiz Guilherme
Airway resistance
Intubation, intratracheal
Tracheal intubation
Flow rate
title_short Model of single-sized endotracheal tube for adults
title_full Model of single-sized endotracheal tube for adults
title_fullStr Model of single-sized endotracheal tube for adults
title_full_unstemmed Model of single-sized endotracheal tube for adults
title_sort Model of single-sized endotracheal tube for adults
author Calderon,Luiz Guilherme
author_facet Calderon,Luiz Guilherme
Moreira,Marcos Mello
Barreto,Gilson
Tincani,Alfio José
author_role author
author2 Moreira,Marcos Mello
Barreto,Gilson
Tincani,Alfio José
author2_role author
author
author
dc.contributor.author.fl_str_mv Calderon,Luiz Guilherme
Moreira,Marcos Mello
Barreto,Gilson
Tincani,Alfio José
dc.subject.por.fl_str_mv Airway resistance
Intubation, intratracheal
Tracheal intubation
Flow rate
topic Airway resistance
Intubation, intratracheal
Tracheal intubation
Flow rate
description ABSTRACT Objective To simulate different diameters of endotracheal tubes and to verify the fluid dynamics aspects by means of flow and resistance measurements. Methods Fluid dynamics software was used to calculate mean flow and airway resistance in endotracheal tube with a diameter of 6.0, 7.0, 7.5, 8.0, 9.0 and 10.0mm at normal body temperature and under constant pressure. The same measurements were taken in the fusion of the first 22cm of a 9.0mm endotracheal tube with 10.0mm diameter, and with the end part in 12cm of a 6.0mm endotracheal tube with 7.0mm diameter. Results The fusion of the first 22cm of an endotracheal tube of 10.0mm diameter with the terminal part in 12cm of an endotracheal tube of 6.0mm diameter, preserving the total length of 34cm, generated average flow and airway resistance similar to that of a conventional 7.5mm endotracheal tube. Conclusion This simulation study demonstrates that a single-sized endotracheal tube may facilitate endotracheal intubation without causing increased airway resistance.
publishDate 2020
dc.date.none.fl_str_mv 2020-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082020000100204
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082020000100204
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.31744/einstein_journal/2020ao4805
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Instituto Israelita de Ensino e Pesquisa Albert Einstein
publisher.none.fl_str_mv Instituto Israelita de Ensino e Pesquisa Albert Einstein
dc.source.none.fl_str_mv einstein (São Paulo) v.18 2020
reponame:Einstein (São Paulo)
instname:Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE)
instacron:IIEPAE
instname_str Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE)
instacron_str IIEPAE
institution IIEPAE
reponame_str Einstein (São Paulo)
collection Einstein (São Paulo)
repository.name.fl_str_mv Einstein (São Paulo) - Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE)
repository.mail.fl_str_mv ||revista@einstein.br
_version_ 1752129909993504768