Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment
Autor(a) principal: | |
---|---|
Data de Publicação: | 2018 |
Outros Autores: | , , , , , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Einstein (São Paulo) |
Texto Completo: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082018000400204 |
Resumo: | ABSTRACT Objective To evaluate the difference in transepidermal water loss in patients diagnosed with hyperhidrosis and healthy subjects, in an air-conditioned environment. Methods Twenty patients diagnosed with hyperhidrosis and 20 healthy subjects were subjected to quantitative assessment using a closed-chamber device, in six previously established sites. Results The measurements showed different transepidermal water loss values for healthy subjects and patients with hyperhidrosis, especially in the hands and feet. In the Control Group, the median for the hands was 46.4g/m2/hour (p25: 36.0; p75: 57.6), while in the Hyperhidrosis Group, the median was 123.5g/m2/hour (p25: 54.3; p75: 161.2) – p<0.001. For the feet, the Control Group had a median of 41.5g/m2/hour (p25: 31.3; p75: 63.5) and the Hyperhidrosis Group, 61.2g/m2/hour (p25: 32.3; p75: 117) – p<0.02. Measurements of the axillas also showed differences. In the Control Group, the median was 14.8g/m2/hour (p25: 11.8; p75: 19.0) and, in the Hyperhidrosis Group, 83.5g/m2/hour (p25: 29.5; p75: 161.7) – p<0.001. Conclusion Measuring transepidermal water loss is sufficient for diagnosis and follow-up of patients with hyperhidrosis. |
id |
IIEPAE-1_da542a2711a9539b46f291148d1f3d35 |
---|---|
oai_identifier_str |
oai:scielo:S1679-45082018000400204 |
network_acronym_str |
IIEPAE-1 |
network_name_str |
Einstein (São Paulo) |
repository_id_str |
|
spelling |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environmentThoracic surgeryHyperhidrosisSympathectomyABSTRACT Objective To evaluate the difference in transepidermal water loss in patients diagnosed with hyperhidrosis and healthy subjects, in an air-conditioned environment. Methods Twenty patients diagnosed with hyperhidrosis and 20 healthy subjects were subjected to quantitative assessment using a closed-chamber device, in six previously established sites. Results The measurements showed different transepidermal water loss values for healthy subjects and patients with hyperhidrosis, especially in the hands and feet. In the Control Group, the median for the hands was 46.4g/m2/hour (p25: 36.0; p75: 57.6), while in the Hyperhidrosis Group, the median was 123.5g/m2/hour (p25: 54.3; p75: 161.2) – p<0.001. For the feet, the Control Group had a median of 41.5g/m2/hour (p25: 31.3; p75: 63.5) and the Hyperhidrosis Group, 61.2g/m2/hour (p25: 32.3; p75: 117) – p<0.02. Measurements of the axillas also showed differences. In the Control Group, the median was 14.8g/m2/hour (p25: 11.8; p75: 19.0) and, in the Hyperhidrosis Group, 83.5g/m2/hour (p25: 29.5; p75: 161.7) – p<0.001. Conclusion Measuring transepidermal water loss is sufficient for diagnosis and follow-up of patients with hyperhidrosis.Instituto Israelita de Ensino e Pesquisa Albert Einstein2018-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082018000400204einstein (São Paulo) v.16 n.4 2018reponame:Einstein (São Paulo)instname:Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE)instacron:IIEPAE10.31744/einstein_journal/2018ao4312info:eu-repo/semantics/openAccessMiotto,AndreHonda,Pedro Augusto AntunesBachichi,Thiago GangiHolanda,Caio SantosEvangelista Neto,ErnestoPerfeito,João Alessio JulianoLeão,Luiz Eduardo VillaçaCosta Jr,Altair da Silvaeng2018-10-31T00:00:00Zoai:scielo:S1679-45082018000400204Revistahttps://journal.einstein.br/pt-br/ONGhttps://old.scielo.br/oai/scielo-oai.php||revista@einstein.br2317-63851679-4508opendoar:2018-10-31T00:00Einstein (São Paulo) - Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE)false |
dc.title.none.fl_str_mv |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment |
title |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment |
spellingShingle |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment Miotto,Andre Thoracic surgery Hyperhidrosis Sympathectomy |
title_short |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment |
title_full |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment |
title_fullStr |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment |
title_full_unstemmed |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment |
title_sort |
Comparative study of transepidermal water loss in patients with and without hyperhidrosis by closed-chamber measurer in an air-conditioned environment |
author |
Miotto,Andre |
author_facet |
Miotto,Andre Honda,Pedro Augusto Antunes Bachichi,Thiago Gangi Holanda,Caio Santos Evangelista Neto,Ernesto Perfeito,João Alessio Juliano Leão,Luiz Eduardo Villaça Costa Jr,Altair da Silva |
author_role |
author |
author2 |
Honda,Pedro Augusto Antunes Bachichi,Thiago Gangi Holanda,Caio Santos Evangelista Neto,Ernesto Perfeito,João Alessio Juliano Leão,Luiz Eduardo Villaça Costa Jr,Altair da Silva |
author2_role |
author author author author author author author |
dc.contributor.author.fl_str_mv |
Miotto,Andre Honda,Pedro Augusto Antunes Bachichi,Thiago Gangi Holanda,Caio Santos Evangelista Neto,Ernesto Perfeito,João Alessio Juliano Leão,Luiz Eduardo Villaça Costa Jr,Altair da Silva |
dc.subject.por.fl_str_mv |
Thoracic surgery Hyperhidrosis Sympathectomy |
topic |
Thoracic surgery Hyperhidrosis Sympathectomy |
description |
ABSTRACT Objective To evaluate the difference in transepidermal water loss in patients diagnosed with hyperhidrosis and healthy subjects, in an air-conditioned environment. Methods Twenty patients diagnosed with hyperhidrosis and 20 healthy subjects were subjected to quantitative assessment using a closed-chamber device, in six previously established sites. Results The measurements showed different transepidermal water loss values for healthy subjects and patients with hyperhidrosis, especially in the hands and feet. In the Control Group, the median for the hands was 46.4g/m2/hour (p25: 36.0; p75: 57.6), while in the Hyperhidrosis Group, the median was 123.5g/m2/hour (p25: 54.3; p75: 161.2) – p<0.001. For the feet, the Control Group had a median of 41.5g/m2/hour (p25: 31.3; p75: 63.5) and the Hyperhidrosis Group, 61.2g/m2/hour (p25: 32.3; p75: 117) – p<0.02. Measurements of the axillas also showed differences. In the Control Group, the median was 14.8g/m2/hour (p25: 11.8; p75: 19.0) and, in the Hyperhidrosis Group, 83.5g/m2/hour (p25: 29.5; p75: 161.7) – p<0.001. Conclusion Measuring transepidermal water loss is sufficient for diagnosis and follow-up of patients with hyperhidrosis. |
publishDate |
2018 |
dc.date.none.fl_str_mv |
2018-01-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082018000400204 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1679-45082018000400204 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.31744/einstein_journal/2018ao4312 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
Instituto Israelita de Ensino e Pesquisa Albert Einstein |
publisher.none.fl_str_mv |
Instituto Israelita de Ensino e Pesquisa Albert Einstein |
dc.source.none.fl_str_mv |
einstein (São Paulo) v.16 n.4 2018 reponame:Einstein (São Paulo) instname:Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE) instacron:IIEPAE |
instname_str |
Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE) |
instacron_str |
IIEPAE |
institution |
IIEPAE |
reponame_str |
Einstein (São Paulo) |
collection |
Einstein (São Paulo) |
repository.name.fl_str_mv |
Einstein (São Paulo) - Instituto Israelita de Ensino e Pesquisa Albert Einstein (IIEPAE) |
repository.mail.fl_str_mv |
||revista@einstein.br |
_version_ |
1752129909541568512 |