Phantom positioning variation in the gamma knife?? perfexion dosimetry

Detalhes bibliográficos
Autor(a) principal: COSTA, N.A.
Data de Publicação: 2017
Outros Autores: POTIENS, M.P.A., SARAIVA, C.W.C., BRAZILIAN CONGRESS ON METROLOGY, 8th
Tipo de documento: Artigo de conferência
Título da fonte: Repositório Institucional do IPEN
Texto Completo: http://repositorio.ipen.br/handle/123456789/27223
Resumo: The use of small volume ionization chamber has become required for the dosimetry of equipments that use small radiation fields such as the Gamma Knife?? Perfexion (GKP) unit. In this work, a pinpoint ionization chamber was inserted into the dosimetry phantom and measurements were performed with the phantom in different positions, in order to verify if the change in the phantom positioning affects the dosimetry of the GKP. Four different phantom positions were performed. The variation in the result is within the range allowed for the dosimetry of a GKP equipment.
id IPEN_21359d995a0693d4c133f39f41074362
oai_identifier_str oai:repositorio.ipen.br:123456789/27223
network_acronym_str IPEN
network_name_str Repositório Institucional do IPEN
repository_id_str 4510
spelling 2017-03-20T14:47:16Z2017-03-20T14:47:16Z29 de novembro - 04 de dezembro, 2015http://repositorio.ipen.br/handle/123456789/27223The use of small volume ionization chamber has become required for the dosimetry of equipments that use small radiation fields such as the Gamma Knife?? Perfexion (GKP) unit. In this work, a pinpoint ionization chamber was inserted into the dosimetry phantom and measurements were performed with the phantom in different positions, in order to verify if the change in the phantom positioning affects the dosimetry of the GKP. Four different phantom positions were performed. The variation in the result is within the range allowed for the dosimetry of a GKP equipment.Submitted by Marco Antonio Oliveira da Silva (maosilva@ipen.br) on 2017-03-20T14:47:16Z No. of bitstreams: 1 23586.pdf: 1037700 bytes, checksum: fa53552521b8e54a9ca2a1ada0246582 (MD5)Made available in DSpace on 2017-03-20T14:47:16Z (GMT). No. of bitstreams: 1 23586.pdf: 1037700 bytes, checksum: fa53552521b8e54a9ca2a1ada0246582 (MD5)1-5Bristol, UK: IOP PublishingJournal of Physics: Conference Series, v. 733, 2016reponame:Repositório Institucional do IPENinstname:Instituto de Pesquisas Energéticas e Nucleares (IPEN)instacron:IPENPhantom positioning variation in the gamma knife?? perfexion dosimetryinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/conferenceObjectIBento Gon??alves, RSCOSTA, N.A.POTIENS, M.P.A.SARAIVA, C.W.C.BRAZILIAN CONGRESS ON METROLOGY, 8thinfo:eu-repo/semantics/openAccess235862016COSTA, N.A.POTIENS, M.P.A.17-03Proceedings10182346COSTA, N.A.:10182:-1:SPOTIENS, M.P.A.:346:-1:NORIGINAL23586.pdf23586.pdfapplication/pdf1037700http://repositorio.ipen.br/bitstream/123456789/27223/1/23586.pdffa53552521b8e54a9ca2a1ada0246582MD51LICENSElicense.txtlicense.txttext/plain; charset=utf-81748http://repositorio.ipen.br/bitstream/123456789/27223/2/license.txt8a4605be74aa9ea9d79846c1fba20a33MD52123456789/272232017-03-20 14:47:16.574oai:repositorio.ipen.br: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Repositório InstitucionalPUBhttp://repositorio.ipen.br/oai/requestbibl@ipen.bropendoar:45102017-03-20T14:47:16Repositório Institucional do IPEN - Instituto de Pesquisas Energéticas e Nucleares (IPEN)false
dc.title.pt_BR.fl_str_mv Phantom positioning variation in the gamma knife?? perfexion dosimetry
title Phantom positioning variation in the gamma knife?? perfexion dosimetry
spellingShingle Phantom positioning variation in the gamma knife?? perfexion dosimetry
COSTA, N.A.
title_short Phantom positioning variation in the gamma knife?? perfexion dosimetry
title_full Phantom positioning variation in the gamma knife?? perfexion dosimetry
title_fullStr Phantom positioning variation in the gamma knife?? perfexion dosimetry
title_full_unstemmed Phantom positioning variation in the gamma knife?? perfexion dosimetry
title_sort Phantom positioning variation in the gamma knife?? perfexion dosimetry
author COSTA, N.A.
author_facet COSTA, N.A.
POTIENS, M.P.A.
SARAIVA, C.W.C.
BRAZILIAN CONGRESS ON METROLOGY, 8th
author_role author
author2 POTIENS, M.P.A.
SARAIVA, C.W.C.
BRAZILIAN CONGRESS ON METROLOGY, 8th
author2_role author
author
author
dc.contributor.author.fl_str_mv COSTA, N.A.
POTIENS, M.P.A.
SARAIVA, C.W.C.
BRAZILIAN CONGRESS ON METROLOGY, 8th
description The use of small volume ionization chamber has become required for the dosimetry of equipments that use small radiation fields such as the Gamma Knife?? Perfexion (GKP) unit. In this work, a pinpoint ionization chamber was inserted into the dosimetry phantom and measurements were performed with the phantom in different positions, in order to verify if the change in the phantom positioning affects the dosimetry of the GKP. Four different phantom positions were performed. The variation in the result is within the range allowed for the dosimetry of a GKP equipment.
publishDate 2017
dc.date.evento.pt_BR.fl_str_mv 29 de novembro - 04 de dezembro, 2015
dc.date.accessioned.fl_str_mv 2017-03-20T14:47:16Z
dc.date.available.fl_str_mv 2017-03-20T14:47:16Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/conferenceObject
format conferenceObject
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://repositorio.ipen.br/handle/123456789/27223
url http://repositorio.ipen.br/handle/123456789/27223
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 1-5
dc.coverage.pt_BR.fl_str_mv I
dc.publisher.none.fl_str_mv Bristol, UK: IOP Publishing
publisher.none.fl_str_mv Bristol, UK: IOP Publishing
dc.source.pt_BR.fl_str_mv Journal of Physics: Conference Series, v. 733, 2016
dc.source.none.fl_str_mv reponame:Repositório Institucional do IPEN
instname:Instituto de Pesquisas Energéticas e Nucleares (IPEN)
instacron:IPEN
instname_str Instituto de Pesquisas Energéticas e Nucleares (IPEN)
instacron_str IPEN
institution IPEN
reponame_str Repositório Institucional do IPEN
collection Repositório Institucional do IPEN
bitstream.url.fl_str_mv http://repositorio.ipen.br/bitstream/123456789/27223/1/23586.pdf
http://repositorio.ipen.br/bitstream/123456789/27223/2/license.txt
bitstream.checksum.fl_str_mv fa53552521b8e54a9ca2a1ada0246582
8a4605be74aa9ea9d79846c1fba20a33
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Repositório Institucional do IPEN - Instituto de Pesquisas Energéticas e Nucleares (IPEN)
repository.mail.fl_str_mv bibl@ipen.br
_version_ 1767254240834617344