Stability study of the alanine epr dosimetry system at IPEN/CNEN SP

Detalhes bibliográficos
Autor(a) principal: RODRIGUES JR ., ORLANDO
Data de Publicação: 2019
Outros Autores: SANTOS, SILAS C. dos, INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
Tipo de documento: Artigo de conferência
Título da fonte: Repositório Institucional do IPEN
Texto Completo: http://repositorio.ipen.br/handle/123456789/30350
Resumo: The dosimetry system using alanine EPR is a standard secondary calibration system for high doses. IPEN's High Doses Dosimetry Laboratory (LDA) has an Electron Paramagnetic Resonance equipment operating in the X band which performs high dose dosimetry service for electron and X ray fields for the dose range of 10Gy to 200kGy. IPEN has established a dosimetry system based on an alanine dosimeter developed at the institute. The system is composed of pure DL alanine encapsulated in a polyethylene microtube, which allows estimated doses of 10Gy to 200kGy. The advantages of this system are the low signal fading over a long period of time (over 90 days) and dose measurements above 150kGy. The main disadvantage is the difficulty in measures below 10Gy. In order to improve quality assurance services delivered in radiotherapy, commercial alanine dosimeters are being tested and compared with the already established system for measurements with time intervals longer than 90 days. The present work is an evaluation of the commercial L alanine dosimeter signal compared to the alanine dosimeter produced in IPEN. The dosimeters were irradiated with X ray and E beam and doses of 1, 10 and 20Gy. All measurements were performed following the same laboratory protocol based in ISO/ASTM guidelines. Environmental effects such as temperature, sensitivity to light and air humidity, and other storage conditions were evaluated. The results show that the decay of commercial dosimeters over 36 months was 11% higher than that observed in the dosimeter developed in IPEN, indicating the need for a correction factor for the comparison of dosimetry systems.
id IPEN_59b34787ccc699966efea6284e24e323
oai_identifier_str oai:repositorio.ipen.br:123456789/30350
network_acronym_str IPEN
network_name_str Repositório Institucional do IPEN
repository_id_str 4510
spelling 2019-11-27T17:18:24Z2019-11-27T17:18:24ZOctober 21-25, 2019http://repositorio.ipen.br/handle/123456789/30350The dosimetry system using alanine EPR is a standard secondary calibration system for high doses. IPEN's High Doses Dosimetry Laboratory (LDA) has an Electron Paramagnetic Resonance equipment operating in the X band which performs high dose dosimetry service for electron and X ray fields for the dose range of 10Gy to 200kGy. IPEN has established a dosimetry system based on an alanine dosimeter developed at the institute. The system is composed of pure DL alanine encapsulated in a polyethylene microtube, which allows estimated doses of 10Gy to 200kGy. The advantages of this system are the low signal fading over a long period of time (over 90 days) and dose measurements above 150kGy. The main disadvantage is the difficulty in measures below 10Gy. In order to improve quality assurance services delivered in radiotherapy, commercial alanine dosimeters are being tested and compared with the already established system for measurements with time intervals longer than 90 days. The present work is an evaluation of the commercial L alanine dosimeter signal compared to the alanine dosimeter produced in IPEN. The dosimeters were irradiated with X ray and E beam and doses of 1, 10 and 20Gy. All measurements were performed following the same laboratory protocol based in ISO/ASTM guidelines. Environmental effects such as temperature, sensitivity to light and air humidity, and other storage conditions were evaluated. The results show that the decay of commercial dosimeters over 36 months was 11% higher than that observed in the dosimeter developed in IPEN, indicating the need for a correction factor for the comparison of dosimetry systems.Submitted by Celia Satomi Uehara (celia.u-topservice@ipen.br) on 2019-11-27T17:18:24Z No. of bitstreams: 1 26138.pdf: 397530 bytes, checksum: 9a46fa717fd96ac2ae3443d9ff22036a (MD5)Made available in DSpace on 2019-11-27T17:18:24Z (GMT). No. of bitstreams: 1 26138.pdf: 397530 bytes, checksum: 9a46fa717fd96ac2ae3443d9ff22036a (MD5)259-264Associa????o Brasileira de Energia Nuclearalanine-lcalibrationdosemetersdosimetryirradiationradiation dose rangesradiation dosesStability study of the alanine epr dosimetry system at IPEN/CNEN SPinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/conferenceObjectINACIRio de JaneiroSantos, SP6357600RODRIGUES JR ., ORLANDOSANTOS, SILAS C. dosINTERNATIONAL NUCLEAR ATLANTIC CONFERENCEinfo:eu-repo/semantics/openAccessreponame:Repositório Institucional do IPENinstname:Instituto de Pesquisas Energéticas e Nucleares (IPEN)instacron:IPEN261382019RODRIGUES JR ., ORLANDOSANTOS, SILAS C. dos19-11Proceedings3226961RODRIGUES JR ., ORLANDO:322:-1:SSANTOS, SILAS C. dos:6961:330:NORIGINAL26138.pdf26138.pdfapplication/pdf397530http://repositorio.ipen.br/bitstream/123456789/30350/1/26138.pdf9a46fa717fd96ac2ae3443d9ff22036aMD51LICENSElicense.txtlicense.txttext/plain; charset=utf-81748http://repositorio.ipen.br/bitstream/123456789/30350/2/license.txt8a4605be74aa9ea9d79846c1fba20a33MD52123456789/303502021-04-12 13:33:09.94oai:repositorio.ipen.br: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Repositório InstitucionalPUBhttp://repositorio.ipen.br/oai/requestbibl@ipen.bropendoar:45102021-04-12T13:33:09Repositório Institucional do IPEN - Instituto de Pesquisas Energéticas e Nucleares (IPEN)false
dc.title.pt_BR.fl_str_mv Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
title Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
spellingShingle Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
RODRIGUES JR ., ORLANDO
alanine-l
calibration
dosemeters
dosimetry
irradiation
radiation dose ranges
radiation doses
title_short Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
title_full Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
title_fullStr Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
title_full_unstemmed Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
title_sort Stability study of the alanine epr dosimetry system at IPEN/CNEN SP
author RODRIGUES JR ., ORLANDO
author_facet RODRIGUES JR ., ORLANDO
SANTOS, SILAS C. dos
INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
author_role author
author2 SANTOS, SILAS C. dos
INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
author2_role author
author
dc.contributor.author.fl_str_mv RODRIGUES JR ., ORLANDO
SANTOS, SILAS C. dos
INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE
dc.subject.por.fl_str_mv alanine-l
calibration
dosemeters
dosimetry
irradiation
radiation dose ranges
radiation doses
topic alanine-l
calibration
dosemeters
dosimetry
irradiation
radiation dose ranges
radiation doses
description The dosimetry system using alanine EPR is a standard secondary calibration system for high doses. IPEN's High Doses Dosimetry Laboratory (LDA) has an Electron Paramagnetic Resonance equipment operating in the X band which performs high dose dosimetry service for electron and X ray fields for the dose range of 10Gy to 200kGy. IPEN has established a dosimetry system based on an alanine dosimeter developed at the institute. The system is composed of pure DL alanine encapsulated in a polyethylene microtube, which allows estimated doses of 10Gy to 200kGy. The advantages of this system are the low signal fading over a long period of time (over 90 days) and dose measurements above 150kGy. The main disadvantage is the difficulty in measures below 10Gy. In order to improve quality assurance services delivered in radiotherapy, commercial alanine dosimeters are being tested and compared with the already established system for measurements with time intervals longer than 90 days. The present work is an evaluation of the commercial L alanine dosimeter signal compared to the alanine dosimeter produced in IPEN. The dosimeters were irradiated with X ray and E beam and doses of 1, 10 and 20Gy. All measurements were performed following the same laboratory protocol based in ISO/ASTM guidelines. Environmental effects such as temperature, sensitivity to light and air humidity, and other storage conditions were evaluated. The results show that the decay of commercial dosimeters over 36 months was 11% higher than that observed in the dosimeter developed in IPEN, indicating the need for a correction factor for the comparison of dosimetry systems.
publishDate 2019
dc.date.evento.pt_BR.fl_str_mv October 21-25, 2019
dc.date.accessioned.fl_str_mv 2019-11-27T17:18:24Z
dc.date.available.fl_str_mv 2019-11-27T17:18:24Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/conferenceObject
format conferenceObject
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://repositorio.ipen.br/handle/123456789/30350
url http://repositorio.ipen.br/handle/123456789/30350
dc.relation.authority.fl_str_mv 6357
dc.relation.confidence.fl_str_mv 600
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 259-264
dc.coverage.pt_BR.fl_str_mv I
dc.publisher.none.fl_str_mv Associa????o Brasileira de Energia Nuclear
publisher.none.fl_str_mv Associa????o Brasileira de Energia Nuclear
dc.source.none.fl_str_mv reponame:Repositório Institucional do IPEN
instname:Instituto de Pesquisas Energéticas e Nucleares (IPEN)
instacron:IPEN
instname_str Instituto de Pesquisas Energéticas e Nucleares (IPEN)
instacron_str IPEN
institution IPEN
reponame_str Repositório Institucional do IPEN
collection Repositório Institucional do IPEN
bitstream.url.fl_str_mv http://repositorio.ipen.br/bitstream/123456789/30350/1/26138.pdf
http://repositorio.ipen.br/bitstream/123456789/30350/2/license.txt
bitstream.checksum.fl_str_mv 9a46fa717fd96ac2ae3443d9ff22036a
8a4605be74aa9ea9d79846c1fba20a33
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Repositório Institucional do IPEN - Instituto de Pesquisas Energéticas e Nucleares (IPEN)
repository.mail.fl_str_mv bibl@ipen.br
_version_ 1767254248656994304