Mammography calibration qualities establishment in a Mo-Mo clinical system

Detalhes bibliográficos
Autor(a) principal: CORREA, E.L.
Data de Publicação: 2017
Outros Autores: SANTOS, L.R. dos, VIVOLO, V., POTIENS, M.P.A., BRAZILIAN CONGRESS ON METROLOGY, 8th
Tipo de documento: Artigo de conferência
Título da fonte: Repositório Institucional do IPEN
Texto Completo: http://repositorio.ipen.br/handle/123456789/27222
Resumo: In this study the mammography calibration qualities were established in a clinical mammography system. The objective is to provide the IPEN instruments calibration laboratory with both mammography calibration methods (using a clinical and an industrial system). The results showed a good behavior of mammography equipment, in terms of kVp, PPV and exposure time. The additional filtration of molybdenum is adequate, air-kerma rates were determined and spectra were obtained.
id IPEN_ecd046f8954be091aa0af3a84b6194f6
oai_identifier_str oai:repositorio.ipen.br:123456789/27222
network_acronym_str IPEN
network_name_str Repositório Institucional do IPEN
repository_id_str 4510
spelling 2017-03-20T14:38:54Z2017-03-20T14:38:54Z29 de novembro - 04 de dezembro, 2015http://repositorio.ipen.br/handle/123456789/27222In this study the mammography calibration qualities were established in a clinical mammography system. The objective is to provide the IPEN instruments calibration laboratory with both mammography calibration methods (using a clinical and an industrial system). The results showed a good behavior of mammography equipment, in terms of kVp, PPV and exposure time. The additional filtration of molybdenum is adequate, air-kerma rates were determined and spectra were obtained.Submitted by Marco Antonio Oliveira da Silva (maosilva@ipen.br) on 2017-03-20T14:38:54Z No. of bitstreams: 1 23585.pdf: 968802 bytes, checksum: 260e5506138a15fd9a07e136e64bba4e (MD5)Made available in DSpace on 2017-03-20T14:38:54Z (GMT). No. of bitstreams: 1 23585.pdf: 968802 bytes, checksum: 260e5506138a15fd9a07e136e64bba4e (MD5)1-5Bristol, UK: IOP PublishingJournal of Physics: Conference Series, v. 733, 2016reponame:Repositório Institucional do IPENinstname:Instituto de Pesquisas Energéticas e Nucleares (IPEN)instacron:IPENMammography calibration qualities establishment in a Mo-Mo clinical systeminfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/conferenceObjectIBento Gon??alves, RSCORREA, E.L.SANTOS, L.R. dosVIVOLO, V.POTIENS, M.P.A.BRAZILIAN CONGRESS ON METROLOGY, 8thinfo:eu-repo/semantics/openAccess235852016CORREA, E.L.SANTOS, L.R. dosVIVOLO, V.POTIENS, M.P.A.17-03Proceedings74988725406346CORREA, E.L.:7498:330:SSANTOS, L.R. DOS:8725:-1:NVIVOLO, V.:406:-1:NPOTIENS, M.P.A.:346:-1:NORIGINAL23585.pdf23585.pdfapplication/pdf968802http://repositorio.ipen.br/bitstream/123456789/27222/1/23585.pdf260e5506138a15fd9a07e136e64bba4eMD51LICENSElicense.txtlicense.txttext/plain; charset=utf-81748http://repositorio.ipen.br/bitstream/123456789/27222/2/license.txt8a4605be74aa9ea9d79846c1fba20a33MD52123456789/272222020-08-17 21:46:46.491oai:repositorio.ipen.br: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Repositório InstitucionalPUBhttp://repositorio.ipen.br/oai/requestbibl@ipen.bropendoar:45102020-08-17T21:46:46Repositório Institucional do IPEN - Instituto de Pesquisas Energéticas e Nucleares (IPEN)false
dc.title.pt_BR.fl_str_mv Mammography calibration qualities establishment in a Mo-Mo clinical system
title Mammography calibration qualities establishment in a Mo-Mo clinical system
spellingShingle Mammography calibration qualities establishment in a Mo-Mo clinical system
CORREA, E.L.
title_short Mammography calibration qualities establishment in a Mo-Mo clinical system
title_full Mammography calibration qualities establishment in a Mo-Mo clinical system
title_fullStr Mammography calibration qualities establishment in a Mo-Mo clinical system
title_full_unstemmed Mammography calibration qualities establishment in a Mo-Mo clinical system
title_sort Mammography calibration qualities establishment in a Mo-Mo clinical system
author CORREA, E.L.
author_facet CORREA, E.L.
SANTOS, L.R. dos
VIVOLO, V.
POTIENS, M.P.A.
BRAZILIAN CONGRESS ON METROLOGY, 8th
author_role author
author2 SANTOS, L.R. dos
VIVOLO, V.
POTIENS, M.P.A.
BRAZILIAN CONGRESS ON METROLOGY, 8th
author2_role author
author
author
author
dc.contributor.author.fl_str_mv CORREA, E.L.
SANTOS, L.R. dos
VIVOLO, V.
POTIENS, M.P.A.
BRAZILIAN CONGRESS ON METROLOGY, 8th
description In this study the mammography calibration qualities were established in a clinical mammography system. The objective is to provide the IPEN instruments calibration laboratory with both mammography calibration methods (using a clinical and an industrial system). The results showed a good behavior of mammography equipment, in terms of kVp, PPV and exposure time. The additional filtration of molybdenum is adequate, air-kerma rates were determined and spectra were obtained.
publishDate 2017
dc.date.evento.pt_BR.fl_str_mv 29 de novembro - 04 de dezembro, 2015
dc.date.accessioned.fl_str_mv 2017-03-20T14:38:54Z
dc.date.available.fl_str_mv 2017-03-20T14:38:54Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/conferenceObject
format conferenceObject
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://repositorio.ipen.br/handle/123456789/27222
url http://repositorio.ipen.br/handle/123456789/27222
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 1-5
dc.coverage.pt_BR.fl_str_mv I
dc.publisher.none.fl_str_mv Bristol, UK: IOP Publishing
publisher.none.fl_str_mv Bristol, UK: IOP Publishing
dc.source.pt_BR.fl_str_mv Journal of Physics: Conference Series, v. 733, 2016
dc.source.none.fl_str_mv reponame:Repositório Institucional do IPEN
instname:Instituto de Pesquisas Energéticas e Nucleares (IPEN)
instacron:IPEN
instname_str Instituto de Pesquisas Energéticas e Nucleares (IPEN)
instacron_str IPEN
institution IPEN
reponame_str Repositório Institucional do IPEN
collection Repositório Institucional do IPEN
bitstream.url.fl_str_mv http://repositorio.ipen.br/bitstream/123456789/27222/1/23585.pdf
http://repositorio.ipen.br/bitstream/123456789/27222/2/license.txt
bitstream.checksum.fl_str_mv 260e5506138a15fd9a07e136e64bba4e
8a4605be74aa9ea9d79846c1fba20a33
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Repositório Institucional do IPEN - Instituto de Pesquisas Energéticas e Nucleares (IPEN)
repository.mail.fl_str_mv bibl@ipen.br
_version_ 1767254240830423040