[pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES

Detalhes bibliográficos
Autor(a) principal: MARCELO LUCIO LESSA
Data de Publicação: 2018
Tipo de documento: Outros
Idioma: por
Título da fonte: Repositório Institucional da PUC-RIO (Projeto Maxwell)
Texto Completo: https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@1
https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@2
http://doi.org/10.17771/PUCRio.acad.35477
Resumo: [pt] O presente trabalho visa desenvolver um sistema de avaliação metrológica de autoclaves, que atenda às normas técnicas vigentes, de baixo custo e acessível a todos Estabelecimentos Assistenciais de Saúde (EAS), visando à validação do processo de esterilização por vapor saturado e a melhoria no controle de infecção hospitalar associada a problemas decorrentes do processo de esterilização. Por meio de uma abordagem matemática, o sistema deverá ser capaz de validar o processo de esterilização, garantindo assim o nível de esterilidade necessário e com isso reduzir as barreiras tecnológicas e financeiras que hoje impedem a garantia da qualidade da esterilização de materiais, sobretudo, em EAS, tanto público como privado. Apesar de instituições como a Organização Internacional de Metrologia Legal (OIML) externarem sua preocupação no que diz respeito à necessidade do acompanhamento da qualidade das medições realizadas por Equipamentos Eletromédicos (EEM) na fase pós-comercialização, a maioria dos países, entre eles o Brasil, apesar de já terem incorporado algumas recomendações publicadas pela OIML, têm concentrado suas exigências da fase de fabricação ao registro do produto. As iniciativas de calibração e ensaios de EEM dependem da cultura metrológica dos interessados ou na busca de algum tipo de acreditação, para a qual o controle metrológico dos equipamentos vem a ser exigência obrigatória. Aliado à falta de cultura metrológica depara-se com a existência de poucas instituições capacitadas para a realização desta tarefa e com a falta de recursos disponíveis. A exemplo dos EEM, no caso dos esterilizadores por vapor saturado ou autoclaves, existe apenas um único laboratório acreditado junto à Rede Brasileira de Laboratórios de Ensaio (RBLE), acreditado em janeiro de 2008. Utilizando-se dos estudos da esterilização, em particular das autoclaves e da Norma NBR ISO 11134/2001. Foi desenvolvido um sistema destinado à validação metrológica de autoclaves. Este sistema é composto de um Hardware (sensores, sistema de aquisição e computador portátil) e de um software desenvolvido em LabView. Neste sistema os sensores são posicionados no interior da câmara de esterilização, coletando a temperatura ao longo de todo o ciclo de esterilização. Com as temperaturas coletadas, tanto nos estudos de distribuição de temperatura como nos estudos de penetração de calor, foi possível calcular as taxas de letalidade do processo e apresentar os dados estatísticos, que confirmarão a efetividade do processo ou não. Após a realização de testes visando à comprovação do funcionamento do sistema, foi realizado o teste em um equipamento utilizado em um hospital público do Rio de Janeiro. Os testes compreenderam a realização de quatro estudos sendo, dois de distribuição de temperatura e dois de penetração de calor. Foram encontradas algumas limitações decorrentes da necessidade de desenvolvimento de alguns acessórios como, por exemplo, um flange para a utilização do sistema em autoclaves com acesso à câmara através de orifícios laterais, e a necessidade da construção de uma armação articulada para a perfeita disposição dos sensores no interior da câmara. O sistema desenvolvido no presente trabalho contribui para a disseminação da realização da validação do processo de esterilização por vapor saturado, avaliação fundamental para o controle da infecção hospitalar.
id PUC_RIO-1_cee368cb362bf5031f06bca826334be0
oai_identifier_str oai:MAXWELL.puc-rio.br:35477
network_acronym_str PUC_RIO-1
network_name_str Repositório Institucional da PUC-RIO (Projeto Maxwell)
repository_id_str 534
spelling [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES [en] SYSTEM FOR METROLOGICAL EVALUATION OF AUTOCLAVES [pt] METROLOGIA[pt] CALOR UMIDO[pt] ESTERILIZACAO[pt] BIOMETROLOGIA[en] METROLOGY[en] STEAM HEAT[en] STERILIZATION[en] BIOMETROLOGY[pt] O presente trabalho visa desenvolver um sistema de avaliação metrológica de autoclaves, que atenda às normas técnicas vigentes, de baixo custo e acessível a todos Estabelecimentos Assistenciais de Saúde (EAS), visando à validação do processo de esterilização por vapor saturado e a melhoria no controle de infecção hospitalar associada a problemas decorrentes do processo de esterilização. Por meio de uma abordagem matemática, o sistema deverá ser capaz de validar o processo de esterilização, garantindo assim o nível de esterilidade necessário e com isso reduzir as barreiras tecnológicas e financeiras que hoje impedem a garantia da qualidade da esterilização de materiais, sobretudo, em EAS, tanto público como privado. Apesar de instituições como a Organização Internacional de Metrologia Legal (OIML) externarem sua preocupação no que diz respeito à necessidade do acompanhamento da qualidade das medições realizadas por Equipamentos Eletromédicos (EEM) na fase pós-comercialização, a maioria dos países, entre eles o Brasil, apesar de já terem incorporado algumas recomendações publicadas pela OIML, têm concentrado suas exigências da fase de fabricação ao registro do produto. As iniciativas de calibração e ensaios de EEM dependem da cultura metrológica dos interessados ou na busca de algum tipo de acreditação, para a qual o controle metrológico dos equipamentos vem a ser exigência obrigatória. Aliado à falta de cultura metrológica depara-se com a existência de poucas instituições capacitadas para a realização desta tarefa e com a falta de recursos disponíveis. A exemplo dos EEM, no caso dos esterilizadores por vapor saturado ou autoclaves, existe apenas um único laboratório acreditado junto à Rede Brasileira de Laboratórios de Ensaio (RBLE), acreditado em janeiro de 2008. Utilizando-se dos estudos da esterilização, em particular das autoclaves e da Norma NBR ISO 11134/2001. Foi desenvolvido um sistema destinado à validação metrológica de autoclaves. Este sistema é composto de um Hardware (sensores, sistema de aquisição e computador portátil) e de um software desenvolvido em LabView. Neste sistema os sensores são posicionados no interior da câmara de esterilização, coletando a temperatura ao longo de todo o ciclo de esterilização. Com as temperaturas coletadas, tanto nos estudos de distribuição de temperatura como nos estudos de penetração de calor, foi possível calcular as taxas de letalidade do processo e apresentar os dados estatísticos, que confirmarão a efetividade do processo ou não. Após a realização de testes visando à comprovação do funcionamento do sistema, foi realizado o teste em um equipamento utilizado em um hospital público do Rio de Janeiro. Os testes compreenderam a realização de quatro estudos sendo, dois de distribuição de temperatura e dois de penetração de calor. Foram encontradas algumas limitações decorrentes da necessidade de desenvolvimento de alguns acessórios como, por exemplo, um flange para a utilização do sistema em autoclaves com acesso à câmara através de orifícios laterais, e a necessidade da construção de uma armação articulada para a perfeita disposição dos sensores no interior da câmara. O sistema desenvolvido no presente trabalho contribui para a disseminação da realização da validação do processo de esterilização por vapor saturado, avaliação fundamental para o controle da infecção hospitalar.[en] The present work aims at developing a system for metrological evaluation of autoclaves in accordance with the standards, having a low cost and being accessible to all the healthcare system, aiming at the validation of the process of sterilization and the improvement of the control of hospital infection. By means of a mathematical approach, the system has to be capable to validate the sterilization process, thus guaranteeing the level of sterility necessary and reducing the technological and financial barriers that today hinder the guarantee of the quality of the sterilization of materials, in public and private health system. Although organizations as the International Organization of Legal Metrology (OIML) recommend the follow-up of the quality of the measurements carried through Medical Electrical Equipment (MEE) in the after-sales phase, the majority of the countries, including Brazil, even though already having incorporated some recommendations published by OIML, have concentrated its requirements on the phase of manufacture and register of the product. The initiatives of calibration and essays of MEE depend on the metrological culture of the interested parties or in the search of some type of accreditation process, for which the metrological control of the equipment is a requirement. Ally to the lack of metrological culture comes across itself with the existence of few institutions enabled to the accomplishment of this task and with the lack of available resources. In the case of the autoclaves, in Brazil, there was only one laboratory accredited in last January. A system destined to the metrological Evaluation of autoclaves was developed. This system is composed by hardware (sensors, system of acquisition and computer) and software, developed in LabView. In this system the sensors are arranged inside of the sterilization chamber, collecting the temperature throughout the sterilization cycle. With the collected temperatures, as much in the studies of distribution of temperature as in the studies of penetration of heat, it is possible to calculate the taxes of lethality of the process and to present the statistical data, that will confirm or not the effectiveness of the process. After the tests aiming at evidencing the correct operation of the system, it has been carried out the test in an equipment used in a public hospital of Rio de Janeiro. The tests consisted of 4 (four) studies, 2 (two) of distribution of temperature and 2 (two) of penetration of heat. Some limitations due to the necessity of development of some accessories have been found, for example, a flange for the use of the system in sterilizers with access to the chamber through lateral orifices, and the necessity of the construction of a articulated scaffolding for the perfect placement of the sensors in the interior of the chamber. The system developed in the present work contributes for the dissemination of the accomplishment of the validation of the process of sterilization for saturated vapor, basic evaluation for the control of the hospital infection.MAXWELLELISABETH COSTA MONTEIROMARCELO LUCIO LESSA2018-10-31info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/otherhttps://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@1https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@2http://doi.org/10.17771/PUCRio.acad.35477porreponame:Repositório Institucional da PUC-RIO (Projeto Maxwell)instname:Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)instacron:PUC_RIOinfo:eu-repo/semantics/openAccess2018-11-01T00:00:00Zoai:MAXWELL.puc-rio.br:35477Repositório InstitucionalPRIhttps://www.maxwell.vrac.puc-rio.br/ibict.phpopendoar:5342018-11-01T00:00Repositório Institucional da PUC-RIO (Projeto Maxwell) - Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)false
dc.title.none.fl_str_mv [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
[en] SYSTEM FOR METROLOGICAL EVALUATION OF AUTOCLAVES
title [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
spellingShingle [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
MARCELO LUCIO LESSA
[pt] METROLOGIA
[pt] CALOR UMIDO
[pt] ESTERILIZACAO
[pt] BIOMETROLOGIA
[en] METROLOGY
[en] STEAM HEAT
[en] STERILIZATION
[en] BIOMETROLOGY
title_short [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
title_full [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
title_fullStr [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
title_full_unstemmed [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
title_sort [pt] SISTEMA PARA AVALIAÇÃO METROLÓGICA DE AUTOCLAVES
author MARCELO LUCIO LESSA
author_facet MARCELO LUCIO LESSA
author_role author
dc.contributor.none.fl_str_mv ELISABETH COSTA MONTEIRO
dc.contributor.author.fl_str_mv MARCELO LUCIO LESSA
dc.subject.por.fl_str_mv [pt] METROLOGIA
[pt] CALOR UMIDO
[pt] ESTERILIZACAO
[pt] BIOMETROLOGIA
[en] METROLOGY
[en] STEAM HEAT
[en] STERILIZATION
[en] BIOMETROLOGY
topic [pt] METROLOGIA
[pt] CALOR UMIDO
[pt] ESTERILIZACAO
[pt] BIOMETROLOGIA
[en] METROLOGY
[en] STEAM HEAT
[en] STERILIZATION
[en] BIOMETROLOGY
description [pt] O presente trabalho visa desenvolver um sistema de avaliação metrológica de autoclaves, que atenda às normas técnicas vigentes, de baixo custo e acessível a todos Estabelecimentos Assistenciais de Saúde (EAS), visando à validação do processo de esterilização por vapor saturado e a melhoria no controle de infecção hospitalar associada a problemas decorrentes do processo de esterilização. Por meio de uma abordagem matemática, o sistema deverá ser capaz de validar o processo de esterilização, garantindo assim o nível de esterilidade necessário e com isso reduzir as barreiras tecnológicas e financeiras que hoje impedem a garantia da qualidade da esterilização de materiais, sobretudo, em EAS, tanto público como privado. Apesar de instituições como a Organização Internacional de Metrologia Legal (OIML) externarem sua preocupação no que diz respeito à necessidade do acompanhamento da qualidade das medições realizadas por Equipamentos Eletromédicos (EEM) na fase pós-comercialização, a maioria dos países, entre eles o Brasil, apesar de já terem incorporado algumas recomendações publicadas pela OIML, têm concentrado suas exigências da fase de fabricação ao registro do produto. As iniciativas de calibração e ensaios de EEM dependem da cultura metrológica dos interessados ou na busca de algum tipo de acreditação, para a qual o controle metrológico dos equipamentos vem a ser exigência obrigatória. Aliado à falta de cultura metrológica depara-se com a existência de poucas instituições capacitadas para a realização desta tarefa e com a falta de recursos disponíveis. A exemplo dos EEM, no caso dos esterilizadores por vapor saturado ou autoclaves, existe apenas um único laboratório acreditado junto à Rede Brasileira de Laboratórios de Ensaio (RBLE), acreditado em janeiro de 2008. Utilizando-se dos estudos da esterilização, em particular das autoclaves e da Norma NBR ISO 11134/2001. Foi desenvolvido um sistema destinado à validação metrológica de autoclaves. Este sistema é composto de um Hardware (sensores, sistema de aquisição e computador portátil) e de um software desenvolvido em LabView. Neste sistema os sensores são posicionados no interior da câmara de esterilização, coletando a temperatura ao longo de todo o ciclo de esterilização. Com as temperaturas coletadas, tanto nos estudos de distribuição de temperatura como nos estudos de penetração de calor, foi possível calcular as taxas de letalidade do processo e apresentar os dados estatísticos, que confirmarão a efetividade do processo ou não. Após a realização de testes visando à comprovação do funcionamento do sistema, foi realizado o teste em um equipamento utilizado em um hospital público do Rio de Janeiro. Os testes compreenderam a realização de quatro estudos sendo, dois de distribuição de temperatura e dois de penetração de calor. Foram encontradas algumas limitações decorrentes da necessidade de desenvolvimento de alguns acessórios como, por exemplo, um flange para a utilização do sistema em autoclaves com acesso à câmara através de orifícios laterais, e a necessidade da construção de uma armação articulada para a perfeita disposição dos sensores no interior da câmara. O sistema desenvolvido no presente trabalho contribui para a disseminação da realização da validação do processo de esterilização por vapor saturado, avaliação fundamental para o controle da infecção hospitalar.
publishDate 2018
dc.date.none.fl_str_mv 2018-10-31
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/other
format other
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@1
https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@2
http://doi.org/10.17771/PUCRio.acad.35477
url https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@1
https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=35477@2
http://doi.org/10.17771/PUCRio.acad.35477
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.publisher.none.fl_str_mv MAXWELL
publisher.none.fl_str_mv MAXWELL
dc.source.none.fl_str_mv reponame:Repositório Institucional da PUC-RIO (Projeto Maxwell)
instname:Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)
instacron:PUC_RIO
instname_str Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)
instacron_str PUC_RIO
institution PUC_RIO
reponame_str Repositório Institucional da PUC-RIO (Projeto Maxwell)
collection Repositório Institucional da PUC-RIO (Projeto Maxwell)
repository.name.fl_str_mv Repositório Institucional da PUC-RIO (Projeto Maxwell) - Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)
repository.mail.fl_str_mv
_version_ 1814822611461341184