[pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA

Detalhes bibliográficos
Autor(a) principal: ALINE MARIA DE CARVALHO DRUMMOND
Data de Publicação: 2007
Tipo de documento: Outros
Idioma: por
Título da fonte: Repositório Institucional da PUC-RIO (Projeto Maxwell)
Texto Completo: https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@1
https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@2
http://doi.org/10.17771/PUCRio.acad.10706
Resumo: [pt] Para lançar equipamentos eletromédicos no mercado brasileiro, é necessária a obtenção de registro da Agência Nacional de Vigilância Sanitária, que tem por requisito a apresentação da certificação de conformidade. Na presente dissertação foram realizados estudos dos documentos normativos necessários para a avaliação da conformidade de equipamentos a laser e Luz Intensa Pulsada (IPL) para uso terapêutico, visando a análise da abrangência das normas para a garantia da confiabilidade metrológica de tais equipamentos biomédicos. Os equipamentos terapêuticos a laser precisam estar em conformidade com a Norma Geral NBR IEC 601-1, colaterais e com a Norma Particular NBR IEC 601- 2-22. O IPL, por sua vez, consiste em instrumentação óptica para uso terapêutico que não é referida na Norma Particular NBR IEC 601-2-22 e ainda não possui norma técnica específica. Verificou-se ao longo dos trabalhos para esta dissertação que, apesar da norma particular NBR IEC 601-2-22 prescrever a necessidade da medição de energia, no Brasil não existem padrões rastreáveis para calibração de medidor de energia laser. Também foi identificada a necessidade da inclusão na norma particular a avaliação das seguintes grandezas para os equipamentos a laser: divergência do feixe, duração do pulso, diâmetro do feixe de trabalho, taxa de repetição de pulso, e definição do limite máximo para o diâmetro do feixe de trabalho. Além disso, embora a norma particular não inclua os equipamentos IPL em seu escopo, esta poderia ser ampliada para adaptar-se também à avaliação deste equipamento. Para tanto, além das grandezas mencionadas acima, sugeridas para a avaliação dos equipamentos a laser, seria necessária também a avaliação de filtros que limitam os comprimentos de onda de emissão da radiação. Concluiuse também que, como a norma particular estudada recomenda o atendimento a várias cláusulas de norma não incorporada no Brasil pela ABNT - a IEC 60825-1 - cujo conteúdo é exigido para a avaliação da conformidade do equipamento, recomenda-se a complementação da norma nacional com a informação contida na IEC 60825-1. Além da inclusão de itens ao documento normativo, o presente trabalho recomenda a avaliação da incerteza da medição não somente para constar nos documentos acompanhantes, como também na avaliação de conformidade do produto. Também considerando a tolerância descrita na norma de 20% de erro da saída laser, e considerando desde já evidências apresentadas no trabalho indicando que é excessiva, recomendou-se um estudo detalhado objetivando, a partir da análise das diferentes aplicações biomédicas dos equipamentos, melhor definição de tal tolerância. Discute-se ainda sobre a importância da avaliação periódica durante o tempo de sua utilização clínica, para garantia da confiabilidade também ao longo da vida útil destes equipamentos.
id PUC_RIO-1_ebdd8531f9e5a7b1a8d0ed8f811ae085
oai_identifier_str oai:MAXWELL.puc-rio.br:10706
network_acronym_str PUC_RIO-1
network_name_str Repositório Institucional da PUC-RIO (Projeto Maxwell)
repository_id_str 534
spelling [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA [en] METROLOGICAL RELIABILITY OF LASER AND INTENSE PULSED LIGHT ELECTROMEDICAL EQUIPMENT [pt] METROLOGIA[pt] EQUIPAMENTO ELETROMEDICO[pt] LUZ[pt] LASER[en] METROLOGY[en] ELECTROMEDICAL EQUIPMENT[en] LIGHT[en] LASER[pt] Para lançar equipamentos eletromédicos no mercado brasileiro, é necessária a obtenção de registro da Agência Nacional de Vigilância Sanitária, que tem por requisito a apresentação da certificação de conformidade. Na presente dissertação foram realizados estudos dos documentos normativos necessários para a avaliação da conformidade de equipamentos a laser e Luz Intensa Pulsada (IPL) para uso terapêutico, visando a análise da abrangência das normas para a garantia da confiabilidade metrológica de tais equipamentos biomédicos. Os equipamentos terapêuticos a laser precisam estar em conformidade com a Norma Geral NBR IEC 601-1, colaterais e com a Norma Particular NBR IEC 601- 2-22. O IPL, por sua vez, consiste em instrumentação óptica para uso terapêutico que não é referida na Norma Particular NBR IEC 601-2-22 e ainda não possui norma técnica específica. Verificou-se ao longo dos trabalhos para esta dissertação que, apesar da norma particular NBR IEC 601-2-22 prescrever a necessidade da medição de energia, no Brasil não existem padrões rastreáveis para calibração de medidor de energia laser. Também foi identificada a necessidade da inclusão na norma particular a avaliação das seguintes grandezas para os equipamentos a laser: divergência do feixe, duração do pulso, diâmetro do feixe de trabalho, taxa de repetição de pulso, e definição do limite máximo para o diâmetro do feixe de trabalho. Além disso, embora a norma particular não inclua os equipamentos IPL em seu escopo, esta poderia ser ampliada para adaptar-se também à avaliação deste equipamento. Para tanto, além das grandezas mencionadas acima, sugeridas para a avaliação dos equipamentos a laser, seria necessária também a avaliação de filtros que limitam os comprimentos de onda de emissão da radiação. Concluiuse também que, como a norma particular estudada recomenda o atendimento a várias cláusulas de norma não incorporada no Brasil pela ABNT - a IEC 60825-1 - cujo conteúdo é exigido para a avaliação da conformidade do equipamento, recomenda-se a complementação da norma nacional com a informação contida na IEC 60825-1. Além da inclusão de itens ao documento normativo, o presente trabalho recomenda a avaliação da incerteza da medição não somente para constar nos documentos acompanhantes, como também na avaliação de conformidade do produto. Também considerando a tolerância descrita na norma de 20% de erro da saída laser, e considerando desde já evidências apresentadas no trabalho indicando que é excessiva, recomendou-se um estudo detalhado objetivando, a partir da análise das diferentes aplicações biomédicas dos equipamentos, melhor definição de tal tolerância. Discute-se ainda sobre a importância da avaliação periódica durante o tempo de sua utilização clínica, para garantia da confiabilidade também ao longo da vida útil destes equipamentos.[en] Manufacturers must present a Conformity Certificate and register electrical medical products with the National Health Vigilance Agency (ANVISA) before releasing them in the Brazilian market. For medical laser equipment, the mandatory standards are the general safety standard for medical electrical equipment (NBR IEC 601-1) and the particular safety standard for diagnostic and therapeutic laser equipment (NBR IEC 601-2-22). Despite not having a particular standard with safety requirements, medical IPL equipment used for therapeutic applications are also required by ANVISA to comply with NBR IEC 601-1 and NBR IEC 601-2-22. In this thesis, the standards used for the conformity evaluation of laser and intense pulsed light (IPL) electromedical equipment were examined with the purpose of assessing their effectiveness regarding metrological reliability. It was found that, despite the energy measurement requirements in the particular standard NBR IEC 601-2-22, there are no traceable standards for optical energy measurements in Brazil. It was also found that the measurement of other quantities (such as, beam divergence and diameter, pulse duration, pulse repetition rate, among others) should be included in the particular standard. Additionally, it was found that, in order to evaluate safety requirements for medical IPL equipment, the particular standard should include requirements concerning the filters that limit the bandwidth of the radiation. In addition to including the parameters mentioned above, this dissertation also recommends the inclusion of the uncertainty in measurement not only in accompanying documents, but also in the conformity evaluation of the product. Moreover, based on evidences, discussed in this thesis, that the current acceptable error of 20% for the laser power output is extremely high, an in-depth evaluation of the actual acceptable error is recommended. Furthermore, since the particular safety standard (NBR IEC 601-2-22) refers to several clauses in the IEC 60825-1 standard, which has not yet been adopted in Brazil, it is recommended that either the IEC 60825-1 standard be adopted in Brazil, or that the clauses mentioned in NBR IEC 601-2-22 be incorporated to it. Finally, the importance of periodic postsale verifications of electromedical equipment in order to ensure metrological reliability is also discussed.MAXWELLELISABETH COSTA MONTEIROALINE MARIA DE CARVALHO DRUMMOND2007-10-11info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/otherhttps://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@1https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@2http://doi.org/10.17771/PUCRio.acad.10706porreponame:Repositório Institucional da PUC-RIO (Projeto Maxwell)instname:Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)instacron:PUC_RIOinfo:eu-repo/semantics/openAccess2018-10-31T00:00:00Zoai:MAXWELL.puc-rio.br:10706Repositório InstitucionalPRIhttps://www.maxwell.vrac.puc-rio.br/ibict.phpopendoar:5342018-10-31T00:00Repositório Institucional da PUC-RIO (Projeto Maxwell) - Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)false
dc.title.none.fl_str_mv [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
[en] METROLOGICAL RELIABILITY OF LASER AND INTENSE PULSED LIGHT ELECTROMEDICAL EQUIPMENT
title [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
spellingShingle [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
ALINE MARIA DE CARVALHO DRUMMOND
[pt] METROLOGIA
[pt] EQUIPAMENTO ELETROMEDICO
[pt] LUZ
[pt] LASER
[en] METROLOGY
[en] ELECTROMEDICAL EQUIPMENT
[en] LIGHT
[en] LASER
title_short [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
title_full [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
title_fullStr [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
title_full_unstemmed [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
title_sort [pt] CONFIABILIDADE METROLÓGICA DE EQUIPAMENTOS ELETROMÉDICOS A LASER E A LUZ INTENSA PULSADA
author ALINE MARIA DE CARVALHO DRUMMOND
author_facet ALINE MARIA DE CARVALHO DRUMMOND
author_role author
dc.contributor.none.fl_str_mv ELISABETH COSTA MONTEIRO
dc.contributor.author.fl_str_mv ALINE MARIA DE CARVALHO DRUMMOND
dc.subject.por.fl_str_mv [pt] METROLOGIA
[pt] EQUIPAMENTO ELETROMEDICO
[pt] LUZ
[pt] LASER
[en] METROLOGY
[en] ELECTROMEDICAL EQUIPMENT
[en] LIGHT
[en] LASER
topic [pt] METROLOGIA
[pt] EQUIPAMENTO ELETROMEDICO
[pt] LUZ
[pt] LASER
[en] METROLOGY
[en] ELECTROMEDICAL EQUIPMENT
[en] LIGHT
[en] LASER
description [pt] Para lançar equipamentos eletromédicos no mercado brasileiro, é necessária a obtenção de registro da Agência Nacional de Vigilância Sanitária, que tem por requisito a apresentação da certificação de conformidade. Na presente dissertação foram realizados estudos dos documentos normativos necessários para a avaliação da conformidade de equipamentos a laser e Luz Intensa Pulsada (IPL) para uso terapêutico, visando a análise da abrangência das normas para a garantia da confiabilidade metrológica de tais equipamentos biomédicos. Os equipamentos terapêuticos a laser precisam estar em conformidade com a Norma Geral NBR IEC 601-1, colaterais e com a Norma Particular NBR IEC 601- 2-22. O IPL, por sua vez, consiste em instrumentação óptica para uso terapêutico que não é referida na Norma Particular NBR IEC 601-2-22 e ainda não possui norma técnica específica. Verificou-se ao longo dos trabalhos para esta dissertação que, apesar da norma particular NBR IEC 601-2-22 prescrever a necessidade da medição de energia, no Brasil não existem padrões rastreáveis para calibração de medidor de energia laser. Também foi identificada a necessidade da inclusão na norma particular a avaliação das seguintes grandezas para os equipamentos a laser: divergência do feixe, duração do pulso, diâmetro do feixe de trabalho, taxa de repetição de pulso, e definição do limite máximo para o diâmetro do feixe de trabalho. Além disso, embora a norma particular não inclua os equipamentos IPL em seu escopo, esta poderia ser ampliada para adaptar-se também à avaliação deste equipamento. Para tanto, além das grandezas mencionadas acima, sugeridas para a avaliação dos equipamentos a laser, seria necessária também a avaliação de filtros que limitam os comprimentos de onda de emissão da radiação. Concluiuse também que, como a norma particular estudada recomenda o atendimento a várias cláusulas de norma não incorporada no Brasil pela ABNT - a IEC 60825-1 - cujo conteúdo é exigido para a avaliação da conformidade do equipamento, recomenda-se a complementação da norma nacional com a informação contida na IEC 60825-1. Além da inclusão de itens ao documento normativo, o presente trabalho recomenda a avaliação da incerteza da medição não somente para constar nos documentos acompanhantes, como também na avaliação de conformidade do produto. Também considerando a tolerância descrita na norma de 20% de erro da saída laser, e considerando desde já evidências apresentadas no trabalho indicando que é excessiva, recomendou-se um estudo detalhado objetivando, a partir da análise das diferentes aplicações biomédicas dos equipamentos, melhor definição de tal tolerância. Discute-se ainda sobre a importância da avaliação periódica durante o tempo de sua utilização clínica, para garantia da confiabilidade também ao longo da vida útil destes equipamentos.
publishDate 2007
dc.date.none.fl_str_mv 2007-10-11
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/other
format other
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@1
https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@2
http://doi.org/10.17771/PUCRio.acad.10706
url https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@1
https://www.maxwell.vrac.puc-rio.br/colecao.php?strSecao=resultado&nrSeq=10706@2
http://doi.org/10.17771/PUCRio.acad.10706
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.publisher.none.fl_str_mv MAXWELL
publisher.none.fl_str_mv MAXWELL
dc.source.none.fl_str_mv reponame:Repositório Institucional da PUC-RIO (Projeto Maxwell)
instname:Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)
instacron:PUC_RIO
instname_str Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)
instacron_str PUC_RIO
institution PUC_RIO
reponame_str Repositório Institucional da PUC-RIO (Projeto Maxwell)
collection Repositório Institucional da PUC-RIO (Projeto Maxwell)
repository.name.fl_str_mv Repositório Institucional da PUC-RIO (Projeto Maxwell) - Pontifícia Universidade Católica do Rio de Janeiro (PUC-RIO)
repository.mail.fl_str_mv
_version_ 1814822565242208256