Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light

Detalhes bibliográficos
Autor(a) principal: Torres, Carlos Rocha Gomes
Data de Publicação: 2008
Outros Autores: Torres, Adriana Cristina de Mello, Lima, Valdeci Ferreira, Ribeiro, Carolina Ferraz, Santos, Joyce Roma Correia dos, Gama, Luciana Maria Fortes
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista odonto ciência (Online)
Texto Completo: https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/3750
Resumo: Purpose: To evaluate temperature changes of pulpal chamber (PC) and infrared blocking bleaching gel on dental surface (SUR) during the activation with photocuring units with halogen lights that emit radiation in the infrared (IR) range. Methods: The bleaching gels Whiteness HP Maxx (IR-blocker – MX) and Whiteness HP (control – CON) were used in ten maxillary central incisors. Light-activation was performed by using the photocuring units (FC) Optilight (600mW/cm2, OP) or Jet Lite (1200mW/cm2, JL) with four 40s-activation procedures with no interval (total time: 3min 20s). The temperature was measured each 40s using a digital thermometer with two type K thermocouples placed inside the PC and on dental surface (SUR). Data were analyzed with 3-way ANOVA (gel type – GT, FC and activation time) and Tukey’s test. Results: A significant effect was found for the main factors FC and GT, and for their interaction. The means (SD) (in Celsius degree) were: PC:[GT={MX-6,38(3,25)a, CON-7,70(2,91)b}, FC={OP-5,51(2,54)a, JL-8,56(2,93)b}]; SUR: [GT={CON-9,85(3,89)a, MX-14,21(7,76)b}, FC={OP-7,32(2,50)a, JL-16,74(5,82)b}]. Conclusions: MX resulted in lower pulp chamber heating and higher gel heating than the control group. JL promoted greater heating than OP. Key words: Dental bleaching; halogen light; infrared light
id PUC_RS-22_274631b037bc021926f1b54b4a7cb2da
oai_identifier_str oai:ojs.revistaseletronicas.pucrs.br:article/3750
network_acronym_str PUC_RS-22
network_name_str Revista odonto ciência (Online)
repository_id_str
spelling Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen lightVariação térmica da câmara pulpar e do gel clareador bloqueador de infravermelho ativado com lâmpada halógenaDental bleachinghalogen lightinfrared lightDetnal bleachingPurpose: To evaluate temperature changes of pulpal chamber (PC) and infrared blocking bleaching gel on dental surface (SUR) during the activation with photocuring units with halogen lights that emit radiation in the infrared (IR) range. Methods: The bleaching gels Whiteness HP Maxx (IR-blocker – MX) and Whiteness HP (control – CON) were used in ten maxillary central incisors. Light-activation was performed by using the photocuring units (FC) Optilight (600mW/cm2, OP) or Jet Lite (1200mW/cm2, JL) with four 40s-activation procedures with no interval (total time: 3min 20s). The temperature was measured each 40s using a digital thermometer with two type K thermocouples placed inside the PC and on dental surface (SUR). Data were analyzed with 3-way ANOVA (gel type – GT, FC and activation time) and Tukey’s test. Results: A significant effect was found for the main factors FC and GT, and for their interaction. The means (SD) (in Celsius degree) were: PC:[GT={MX-6,38(3,25)a, CON-7,70(2,91)b}, FC={OP-5,51(2,54)a, JL-8,56(2,93)b}]; SUR: [GT={CON-9,85(3,89)a, MX-14,21(7,76)b}, FC={OP-7,32(2,50)a, JL-16,74(5,82)b}]. Conclusions: MX resulted in lower pulp chamber heating and higher gel heating than the control group. JL promoted greater heating than OP. Key words: Dental bleaching; halogen light; infrared lightObjetivo: Avaliar a temperatura da câmara pulpar (CP) e do gel clareador bloqueador de infravermelho na superfície dental (SUP), durante a ativação com fotopolimerizadores com lâmpadas halógenas que emitem radiação na faixa do infravermelho (IV). Metodologia: Os géis clareadores Whiteness HP Maxx (com bloqueador de IV – MX) e Whiteness HP (controle – CON) foram aplicados em 10 incisivos centrais superiores. A ativação com luz foi realizada com os fotopolimerizadores (FP) Optilight (600mW/cm2, OP) e Jet Lite (1200mW/cm2, JL), sendo quatro ativações de 40s, sem intervalo (total de 3min e 20s). As temperaturas foram aferidas a cada 40s com um termômetro digital com dois termopares tipo K, um no interior da CP e outro na superfície dental (SUP). Os dados foram analisados com ANOVA a 3 fatores (tipo de gel – TG, FP e tempo de ativação) e teste de Tukey. Resultados: Houve efeito significativo de TG e FP e para a interação FP/TG. As médias (DP) (em graus Celsius) foram: PC:[GT={MX-6,38(3,25)a, CON-7,70(2,91)b}, FC={OP- 5,51(2,54)a, JL-8,56(2,93)b}]; SUR: [GT={CON-9,85(3,89)a, MX-14,21(7,76)b}, FC={OP-7,32(2,50)a, JL-16,74(5,82)b}]. Conclusões: O gel bloqueador de IV resultou em menor aquecimento da CP e maior aquecimento do gel que o controle. O FP Jet Lite promoveu maior aquecimento que o Optilight. Palavras-chave: Clareamento dental; lâmpada halógena; radiação infravermelhoEDIPUCRS - Editora Universitária da PUCRS2008-07-04info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPeer-reviewed Articlelaboratory; in vitro; experimentalapplication/pdfhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/3750Revista Odonto Ciência; Vol. 23 No. 1 (2008); 72 - 76Revista Odonto Ciência; v. 23 n. 1 (2008); 72 - 761980-65230102-9460reponame:Revista odonto ciência (Online)instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)instacron:PUC_RSenghttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/3750/2877Torres, Carlos Rocha GomesTorres, Adriana Cristina de MelloLima, Valdeci FerreiraRibeiro, Carolina FerrazSantos, Joyce Roma Correia dosGama, Luciana Maria Fortesinfo:eu-repo/semantics/openAccess2013-11-18T10:55:40Zoai:ojs.revistaseletronicas.pucrs.br:article/3750Revistahttps://revistaseletronicas.pucrs.br/ojs/index.php/foPRIhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/oai||odontociencia@pucrs.br1980-65230102-9460opendoar:2013-11-18T10:55:40Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)false
dc.title.none.fl_str_mv Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
Variação térmica da câmara pulpar e do gel clareador bloqueador de infravermelho ativado com lâmpada halógena
title Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
spellingShingle Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
Torres, Carlos Rocha Gomes
Dental bleaching
halogen light
infrared light
Detnal bleaching
title_short Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
title_full Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
title_fullStr Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
title_full_unstemmed Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
title_sort Temperature changes of pulpal chamber and infrared-blocking bleaching gel activated with halogen light
author Torres, Carlos Rocha Gomes
author_facet Torres, Carlos Rocha Gomes
Torres, Adriana Cristina de Mello
Lima, Valdeci Ferreira
Ribeiro, Carolina Ferraz
Santos, Joyce Roma Correia dos
Gama, Luciana Maria Fortes
author_role author
author2 Torres, Adriana Cristina de Mello
Lima, Valdeci Ferreira
Ribeiro, Carolina Ferraz
Santos, Joyce Roma Correia dos
Gama, Luciana Maria Fortes
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Torres, Carlos Rocha Gomes
Torres, Adriana Cristina de Mello
Lima, Valdeci Ferreira
Ribeiro, Carolina Ferraz
Santos, Joyce Roma Correia dos
Gama, Luciana Maria Fortes
dc.subject.por.fl_str_mv Dental bleaching
halogen light
infrared light
Detnal bleaching
topic Dental bleaching
halogen light
infrared light
Detnal bleaching
description Purpose: To evaluate temperature changes of pulpal chamber (PC) and infrared blocking bleaching gel on dental surface (SUR) during the activation with photocuring units with halogen lights that emit radiation in the infrared (IR) range. Methods: The bleaching gels Whiteness HP Maxx (IR-blocker – MX) and Whiteness HP (control – CON) were used in ten maxillary central incisors. Light-activation was performed by using the photocuring units (FC) Optilight (600mW/cm2, OP) or Jet Lite (1200mW/cm2, JL) with four 40s-activation procedures with no interval (total time: 3min 20s). The temperature was measured each 40s using a digital thermometer with two type K thermocouples placed inside the PC and on dental surface (SUR). Data were analyzed with 3-way ANOVA (gel type – GT, FC and activation time) and Tukey’s test. Results: A significant effect was found for the main factors FC and GT, and for their interaction. The means (SD) (in Celsius degree) were: PC:[GT={MX-6,38(3,25)a, CON-7,70(2,91)b}, FC={OP-5,51(2,54)a, JL-8,56(2,93)b}]; SUR: [GT={CON-9,85(3,89)a, MX-14,21(7,76)b}, FC={OP-7,32(2,50)a, JL-16,74(5,82)b}]. Conclusions: MX resulted in lower pulp chamber heating and higher gel heating than the control group. JL promoted greater heating than OP. Key words: Dental bleaching; halogen light; infrared light
publishDate 2008
dc.date.none.fl_str_mv 2008-07-04
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Peer-reviewed Article
laboratory; in vitro; experimental
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/3750
url https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/3750
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/3750/2877
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv EDIPUCRS - Editora Universitária da PUCRS
publisher.none.fl_str_mv EDIPUCRS - Editora Universitária da PUCRS
dc.source.none.fl_str_mv Revista Odonto Ciência; Vol. 23 No. 1 (2008); 72 - 76
Revista Odonto Ciência; v. 23 n. 1 (2008); 72 - 76
1980-6523
0102-9460
reponame:Revista odonto ciência (Online)
instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
instacron:PUC_RS
instname_str Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
instacron_str PUC_RS
institution PUC_RS
reponame_str Revista odonto ciência (Online)
collection Revista odonto ciência (Online)
repository.name.fl_str_mv Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
repository.mail.fl_str_mv ||odontociencia@pucrs.br
_version_ 1754820874337255424