Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
Autor(a) principal: | |
---|---|
Data de Publicação: | 2009 |
Outros Autores: | , , , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Revista odonto ciência (Online) |
Texto Completo: | https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428 |
Resumo: | Purpose: This study aimed to determine the effect of using calcinable cylinders on the passivity of a framework that simulates a three-unit fixed partial prosthesis on two implants. Methods: Two 3.75×10 mm external hex implants were used, with their abutments set on a steel base forming the master model. Ten cobalt-chromium alloy frameworks were manufactured by the induction casting technique. Two groups were established: Group 1 used premachined cylinders to cast five frameworks; Group 2 used plastic calcinable cylinders to cast five frameworks. Passivity evaluation was accomplished by using strain gauges placed in the cervical and occlusal regions of the framework pontics. The measurement was performed during screwing of the second prosthetic screw with a torque of 10 Ncm. Data were analyzed by Student’s t test (α=0.05). Results: The mean (standard deviation) values were 39.16 (24.74) mV/V for Group 1 and 43.76 (21.13) mV/V for Group 2. There was no statistically significant difference between groups (P=0.815). Conclusion: The results suggest that the use of calcinable cylinders had a similar degree of passivity compared to the use of premachined prosthetic cylinders. |
id |
PUC_RS-22_f93a0125dc33f7a10990904e8a15ce73 |
---|---|
oai_identifier_str |
oai:ojs.revistaseletronicas.pucrs.br:article/4428 |
network_acronym_str |
PUC_RS-22 |
network_name_str |
Revista odonto ciência (Online) |
repository_id_str |
|
spelling |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworksInfluência de cilindros calcináveis fundidos em cobalto-cromo na passividade de próteses implanto-suportadasAdaptationcalcinable cylinderimplantsstrain gaugesImplant-supported prosthesisPurpose: This study aimed to determine the effect of using calcinable cylinders on the passivity of a framework that simulates a three-unit fixed partial prosthesis on two implants. Methods: Two 3.75×10 mm external hex implants were used, with their abutments set on a steel base forming the master model. Ten cobalt-chromium alloy frameworks were manufactured by the induction casting technique. Two groups were established: Group 1 used premachined cylinders to cast five frameworks; Group 2 used plastic calcinable cylinders to cast five frameworks. Passivity evaluation was accomplished by using strain gauges placed in the cervical and occlusal regions of the framework pontics. The measurement was performed during screwing of the second prosthetic screw with a torque of 10 Ncm. Data were analyzed by Student’s t test (α=0.05). Results: The mean (standard deviation) values were 39.16 (24.74) mV/V for Group 1 and 43.76 (21.13) mV/V for Group 2. There was no statistically significant difference between groups (P=0.815). Conclusion: The results suggest that the use of calcinable cylinders had a similar degree of passivity compared to the use of premachined prosthetic cylinders.Objetivo: Este trabalho avaliou a influência da utilização de cilindros calcináveis no grau de passividade de infra-estruturas que simulam uma prótese parcial fixa de três elementos sobre dois implantes. Metodologia: Foram utilizados dois implantes de hexágono externo de 3,75 mm×10 mm, com seus respectivos pilares intermediários, fixados numa base de aço, formando o modelomestre. A partir disto, foram fabricadas dez infra-estruturas em liga de cobalto-cromo através da técnica de fundição por indução. Foram criados dois grupos: Grupo 1 utilizando cilindros pré-usinados para a fabricação de cinco infra-estruturas; Grupo 2 cinco infra-estruturas fundidas com cilindros de plástico calcináveis. Para avaliação da passividade foram utilizados extensômetros colados na região oclusal e cervical do pôntico das infra-estruturas. A medição foi realizada no momento do aperto do segundo parafuso protético com torque de 10Ncm. Os dados foram analisados pelo teste t de Student (α=0,05). Resultados: As médias (desvio-padrão) dos grupos foram de 39,16 (24,74) mV/V no Grupo 1 e 43,76 (21,13) mV/V no Grupo 2. Não houve diferença estatisticamente significante entre os grupos (P=0,815). Conclusão: Considerando-se as limitações deste trabalho, sugere-se que o uso de cilindros calcináveis obtém o mesmo grau de passividade que o uso de cilindros pré-usinados.EDIPUCRS - Editora Universitária da PUCRS2009-04-10info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPeer-reviewed Articlein vitro, experimentalapplication/pdfhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428Revista Odonto Ciência; Vol. 24 No. 2 (2009); 151 - 155Revista Odonto Ciência; v. 24 n. 2 (2009); 151 - 1551980-65230102-9460reponame:Revista odonto ciência (Online)instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)instacron:PUC_RSenghttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428/3994Triches, Diego FernandesAlonso, Fernando RizzoSeitenfus, Daniela DisconziBeck, João Carlos PinheiroHirakata, Luciana MayumiTeixeira, Eduardo Roliminfo:eu-repo/semantics/openAccess2013-11-11T13:15:30Zoai:ojs.revistaseletronicas.pucrs.br:article/4428Revistahttps://revistaseletronicas.pucrs.br/ojs/index.php/foPRIhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/oai||odontociencia@pucrs.br1980-65230102-9460opendoar:2013-11-11T13:15:30Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)false |
dc.title.none.fl_str_mv |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks Influência de cilindros calcináveis fundidos em cobalto-cromo na passividade de próteses implanto-suportadas |
title |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks |
spellingShingle |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks Triches, Diego Fernandes Adaptation calcinable cylinder implants strain gauges Implant-supported prosthesis |
title_short |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks |
title_full |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks |
title_fullStr |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks |
title_full_unstemmed |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks |
title_sort |
Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks |
author |
Triches, Diego Fernandes |
author_facet |
Triches, Diego Fernandes Alonso, Fernando Rizzo Seitenfus, Daniela Disconzi Beck, João Carlos Pinheiro Hirakata, Luciana Mayumi Teixeira, Eduardo Rolim |
author_role |
author |
author2 |
Alonso, Fernando Rizzo Seitenfus, Daniela Disconzi Beck, João Carlos Pinheiro Hirakata, Luciana Mayumi Teixeira, Eduardo Rolim |
author2_role |
author author author author author |
dc.contributor.author.fl_str_mv |
Triches, Diego Fernandes Alonso, Fernando Rizzo Seitenfus, Daniela Disconzi Beck, João Carlos Pinheiro Hirakata, Luciana Mayumi Teixeira, Eduardo Rolim |
dc.subject.por.fl_str_mv |
Adaptation calcinable cylinder implants strain gauges Implant-supported prosthesis |
topic |
Adaptation calcinable cylinder implants strain gauges Implant-supported prosthesis |
description |
Purpose: This study aimed to determine the effect of using calcinable cylinders on the passivity of a framework that simulates a three-unit fixed partial prosthesis on two implants. Methods: Two 3.75×10 mm external hex implants were used, with their abutments set on a steel base forming the master model. Ten cobalt-chromium alloy frameworks were manufactured by the induction casting technique. Two groups were established: Group 1 used premachined cylinders to cast five frameworks; Group 2 used plastic calcinable cylinders to cast five frameworks. Passivity evaluation was accomplished by using strain gauges placed in the cervical and occlusal regions of the framework pontics. The measurement was performed during screwing of the second prosthetic screw with a torque of 10 Ncm. Data were analyzed by Student’s t test (α=0.05). Results: The mean (standard deviation) values were 39.16 (24.74) mV/V for Group 1 and 43.76 (21.13) mV/V for Group 2. There was no statistically significant difference between groups (P=0.815). Conclusion: The results suggest that the use of calcinable cylinders had a similar degree of passivity compared to the use of premachined prosthetic cylinders. |
publishDate |
2009 |
dc.date.none.fl_str_mv |
2009-04-10 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion Peer-reviewed Article in vitro, experimental |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428 |
url |
https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428/3994 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf |
dc.publisher.none.fl_str_mv |
EDIPUCRS - Editora Universitária da PUCRS |
publisher.none.fl_str_mv |
EDIPUCRS - Editora Universitária da PUCRS |
dc.source.none.fl_str_mv |
Revista Odonto Ciência; Vol. 24 No. 2 (2009); 151 - 155 Revista Odonto Ciência; v. 24 n. 2 (2009); 151 - 155 1980-6523 0102-9460 reponame:Revista odonto ciência (Online) instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS) instacron:PUC_RS |
instname_str |
Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS) |
instacron_str |
PUC_RS |
institution |
PUC_RS |
reponame_str |
Revista odonto ciência (Online) |
collection |
Revista odonto ciência (Online) |
repository.name.fl_str_mv |
Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS) |
repository.mail.fl_str_mv |
||odontociencia@pucrs.br |
_version_ |
1754820874792337408 |