Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks

Detalhes bibliográficos
Autor(a) principal: Triches, Diego Fernandes
Data de Publicação: 2009
Outros Autores: Alonso, Fernando Rizzo, Seitenfus, Daniela Disconzi, Beck, João Carlos Pinheiro, Hirakata, Luciana Mayumi, Teixeira, Eduardo Rolim
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista odonto ciência (Online)
Texto Completo: https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428
Resumo: Purpose: This study aimed to determine the effect of using calcinable cylinders on the passivity of a framework that simulates a three-unit fixed partial prosthesis on two implants. Methods: Two 3.75×10 mm external hex implants were used, with their abutments set on a steel base forming the master model. Ten cobalt-chromium alloy frameworks were manufactured by the induction casting technique. Two groups were established: Group 1 used premachined cylinders to cast five frameworks; Group 2 used plastic calcinable cylinders to cast five frameworks. Passivity evaluation was accomplished by using strain gauges placed in the cervical and occlusal regions of the framework pontics. The measurement was performed during screwing of the second prosthetic screw with a torque of 10 Ncm. Data were analyzed by Student’s t test (α=0.05). Results: The mean (standard deviation) values were 39.16 (24.74) mV/V for Group 1 and 43.76 (21.13) mV/V for Group 2. There was no statistically significant difference between groups (P=0.815). Conclusion: The results suggest that the use of calcinable cylinders had a similar degree of passivity compared to the use of premachined prosthetic cylinders.
id PUC_RS-22_f93a0125dc33f7a10990904e8a15ce73
oai_identifier_str oai:ojs.revistaseletronicas.pucrs.br:article/4428
network_acronym_str PUC_RS-22
network_name_str Revista odonto ciência (Online)
repository_id_str
spelling Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworksInfluência de cilindros calcináveis fundidos em cobalto-cromo na passividade de próteses implanto-suportadasAdaptationcalcinable cylinderimplantsstrain gaugesImplant-supported prosthesisPurpose: This study aimed to determine the effect of using calcinable cylinders on the passivity of a framework that simulates a three-unit fixed partial prosthesis on two implants. Methods: Two 3.75×10 mm external hex implants were used, with their abutments set on a steel base forming the master model. Ten cobalt-chromium alloy frameworks were manufactured by the induction casting technique. Two groups were established: Group 1 used premachined cylinders to cast five frameworks; Group 2 used plastic calcinable cylinders to cast five frameworks. Passivity evaluation was accomplished by using strain gauges placed in the cervical and occlusal regions of the framework pontics. The measurement was performed during screwing of the second prosthetic screw with a torque of 10 Ncm. Data were analyzed by Student’s t test (α=0.05). Results: The mean (standard deviation) values were 39.16 (24.74) mV/V for Group 1 and 43.76 (21.13) mV/V for Group 2. There was no statistically significant difference between groups (P=0.815). Conclusion: The results suggest that the use of calcinable cylinders had a similar degree of passivity compared to the use of premachined prosthetic cylinders.Objetivo: Este trabalho avaliou a influência da utilização de cilindros calcináveis no grau de passividade de infra-estruturas que simulam uma prótese parcial fixa de três elementos sobre dois implantes. Metodologia: Foram utilizados dois implantes de hexágono externo de 3,75 mm×10 mm, com seus respectivos pilares intermediários, fixados numa base de aço, formando o modelomestre. A partir disto, foram fabricadas dez infra-estruturas em liga de cobalto-cromo através da técnica de fundição por indução. Foram criados dois grupos: Grupo 1 utilizando cilindros pré-usinados para a fabricação de cinco infra-estruturas; Grupo 2 cinco infra-estruturas fundidas com cilindros de plástico calcináveis. Para avaliação da passividade foram utilizados extensômetros colados na região oclusal e cervical do pôntico das infra-estruturas. A medição foi realizada no momento do aperto do segundo parafuso protético com torque de 10Ncm. Os dados foram analisados pelo teste t de Student (α=0,05). Resultados: As médias (desvio-padrão) dos grupos foram de 39,16 (24,74) mV/V no Grupo 1 e 43,76 (21,13) mV/V no Grupo 2. Não houve diferença estatisticamente significante entre os grupos (P=0,815). Conclusão: Considerando-se as limitações deste trabalho, sugere-se que o uso de cilindros calcináveis obtém o mesmo grau de passividade que o uso de cilindros pré-usinados.EDIPUCRS - Editora Universitária da PUCRS2009-04-10info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPeer-reviewed Articlein vitro, experimentalapplication/pdfhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428Revista Odonto Ciência; Vol. 24 No. 2 (2009); 151 - 155Revista Odonto Ciência; v. 24 n. 2 (2009); 151 - 1551980-65230102-9460reponame:Revista odonto ciência (Online)instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)instacron:PUC_RSenghttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428/3994Triches, Diego FernandesAlonso, Fernando RizzoSeitenfus, Daniela DisconziBeck, João Carlos PinheiroHirakata, Luciana MayumiTeixeira, Eduardo Roliminfo:eu-repo/semantics/openAccess2013-11-11T13:15:30Zoai:ojs.revistaseletronicas.pucrs.br:article/4428Revistahttps://revistaseletronicas.pucrs.br/ojs/index.php/foPRIhttps://revistaseletronicas.pucrs.br/ojs/index.php/fo/oai||odontociencia@pucrs.br1980-65230102-9460opendoar:2013-11-11T13:15:30Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)false
dc.title.none.fl_str_mv Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
Influência de cilindros calcináveis fundidos em cobalto-cromo na passividade de próteses implanto-suportadas
title Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
spellingShingle Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
Triches, Diego Fernandes
Adaptation
calcinable cylinder
implants
strain gauges
Implant-supported prosthesis
title_short Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
title_full Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
title_fullStr Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
title_full_unstemmed Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
title_sort Influence of calcinable cylinders cast in cobalt-chromium alloy on the passivity of implant-supported frameworks
author Triches, Diego Fernandes
author_facet Triches, Diego Fernandes
Alonso, Fernando Rizzo
Seitenfus, Daniela Disconzi
Beck, João Carlos Pinheiro
Hirakata, Luciana Mayumi
Teixeira, Eduardo Rolim
author_role author
author2 Alonso, Fernando Rizzo
Seitenfus, Daniela Disconzi
Beck, João Carlos Pinheiro
Hirakata, Luciana Mayumi
Teixeira, Eduardo Rolim
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Triches, Diego Fernandes
Alonso, Fernando Rizzo
Seitenfus, Daniela Disconzi
Beck, João Carlos Pinheiro
Hirakata, Luciana Mayumi
Teixeira, Eduardo Rolim
dc.subject.por.fl_str_mv Adaptation
calcinable cylinder
implants
strain gauges
Implant-supported prosthesis
topic Adaptation
calcinable cylinder
implants
strain gauges
Implant-supported prosthesis
description Purpose: This study aimed to determine the effect of using calcinable cylinders on the passivity of a framework that simulates a three-unit fixed partial prosthesis on two implants. Methods: Two 3.75×10 mm external hex implants were used, with their abutments set on a steel base forming the master model. Ten cobalt-chromium alloy frameworks were manufactured by the induction casting technique. Two groups were established: Group 1 used premachined cylinders to cast five frameworks; Group 2 used plastic calcinable cylinders to cast five frameworks. Passivity evaluation was accomplished by using strain gauges placed in the cervical and occlusal regions of the framework pontics. The measurement was performed during screwing of the second prosthetic screw with a torque of 10 Ncm. Data were analyzed by Student’s t test (α=0.05). Results: The mean (standard deviation) values were 39.16 (24.74) mV/V for Group 1 and 43.76 (21.13) mV/V for Group 2. There was no statistically significant difference between groups (P=0.815). Conclusion: The results suggest that the use of calcinable cylinders had a similar degree of passivity compared to the use of premachined prosthetic cylinders.
publishDate 2009
dc.date.none.fl_str_mv 2009-04-10
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Peer-reviewed Article
in vitro, experimental
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428
url https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://revistaseletronicas.pucrs.br/ojs/index.php/fo/article/view/4428/3994
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv EDIPUCRS - Editora Universitária da PUCRS
publisher.none.fl_str_mv EDIPUCRS - Editora Universitária da PUCRS
dc.source.none.fl_str_mv Revista Odonto Ciência; Vol. 24 No. 2 (2009); 151 - 155
Revista Odonto Ciência; v. 24 n. 2 (2009); 151 - 155
1980-6523
0102-9460
reponame:Revista odonto ciência (Online)
instname:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
instacron:PUC_RS
instname_str Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
instacron_str PUC_RS
institution PUC_RS
reponame_str Revista odonto ciência (Online)
collection Revista odonto ciência (Online)
repository.name.fl_str_mv Revista odonto ciência (Online) - Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
repository.mail.fl_str_mv ||odontociencia@pucrs.br
_version_ 1754820874792337408