Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis

Detalhes bibliográficos
Autor(a) principal: Completo, António
Data de Publicação: 2016
Outros Autores: Nascimento, Abel, Neto, Fernando
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/10773/23264
Resumo: The reported outcomes of the Elektra thumb carpo-metacarpal joint implant have been very variable. This study evaluates the influence of daily cyclic loads and the type of the screw fit cup insertion technique in the trapezium, with and without prior threading, on the structural bone behaviour. The study was performed experimentally to predict initial implant stability and cortical bone strains. Computational models were developed to assess the structural cancellous bone behaviour. The use of Elektra implant changed considerably the bone strain behaviour comparing to the intact joint. This may be associated with risks of cancellous bone fatigue failure due to overload, particularly in the trapezium. The joint load magnitude has a more important structural role than that of the screw-fit cup insertion technique. Limiting the magnitude of thumb loads after arthroplasty may contribute positively to the longevity of this procedure.
id RCAP_19ee9501c7dfe727696e681e53d7c574
oai_identifier_str oai:ria.ua.pt:10773/23264
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysisCarpometacarpal joint,Uncemented trapeziometacarpal prosthesis,Finite element modelsStress-shieldingPrincipal cortex strainsThe reported outcomes of the Elektra thumb carpo-metacarpal joint implant have been very variable. This study evaluates the influence of daily cyclic loads and the type of the screw fit cup insertion technique in the trapezium, with and without prior threading, on the structural bone behaviour. The study was performed experimentally to predict initial implant stability and cortical bone strains. Computational models were developed to assess the structural cancellous bone behaviour. The use of Elektra implant changed considerably the bone strain behaviour comparing to the intact joint. This may be associated with risks of cancellous bone fatigue failure due to overload, particularly in the trapezium. The joint load magnitude has a more important structural role than that of the screw-fit cup insertion technique. Limiting the magnitude of thumb loads after arthroplasty may contribute positively to the longevity of this procedure.Sage2018-05-28T09:11:51Z2016-01-01T00:00:00Z2016info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10773/23264eng1753-193410.1177/1753193416659230Completo, AntónioNascimento, AbelNeto, Fernandoinfo:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2024-02-22T11:45:10Zoai:ria.ua.pt:10773/23264Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-20T02:57:02.481763Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
title Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
spellingShingle Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
Completo, António
Carpometacarpal joint,
Uncemented trapeziometacarpal prosthesis,
Finite element models
Stress-shielding
Principal cortex strains
title_short Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
title_full Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
title_fullStr Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
title_full_unstemmed Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
title_sort Total arthroplasty of basal thumb joint with Elektra prosthesis: An in vitro analysis
author Completo, António
author_facet Completo, António
Nascimento, Abel
Neto, Fernando
author_role author
author2 Nascimento, Abel
Neto, Fernando
author2_role author
author
dc.contributor.author.fl_str_mv Completo, António
Nascimento, Abel
Neto, Fernando
dc.subject.por.fl_str_mv Carpometacarpal joint,
Uncemented trapeziometacarpal prosthesis,
Finite element models
Stress-shielding
Principal cortex strains
topic Carpometacarpal joint,
Uncemented trapeziometacarpal prosthesis,
Finite element models
Stress-shielding
Principal cortex strains
description The reported outcomes of the Elektra thumb carpo-metacarpal joint implant have been very variable. This study evaluates the influence of daily cyclic loads and the type of the screw fit cup insertion technique in the trapezium, with and without prior threading, on the structural bone behaviour. The study was performed experimentally to predict initial implant stability and cortical bone strains. Computational models were developed to assess the structural cancellous bone behaviour. The use of Elektra implant changed considerably the bone strain behaviour comparing to the intact joint. This may be associated with risks of cancellous bone fatigue failure due to overload, particularly in the trapezium. The joint load magnitude has a more important structural role than that of the screw-fit cup insertion technique. Limiting the magnitude of thumb loads after arthroplasty may contribute positively to the longevity of this procedure.
publishDate 2016
dc.date.none.fl_str_mv 2016-01-01T00:00:00Z
2016
2018-05-28T09:11:51Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10773/23264
url http://hdl.handle.net/10773/23264
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 1753-1934
10.1177/1753193416659230
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Sage
publisher.none.fl_str_mv Sage
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799137624744525824