A technique for the absolute measurement of the W-value for X-rays in counting gases

Detalhes bibliográficos
Autor(a) principal: Vinagre, F. L. R.
Data de Publicação: 2000
Outros Autores: Conde, C. A. N.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/10316/4527
Resumo: A technique was developed for the absolute measurement of the W-value (the mean energy for the production of an electron-ion pair) for low-energy X-rays in a wide range of gases at atmospheric pressures, with a standard uncertainty better than 1%. This technique is based on the absolute measurement of the primary ionization charge produced by X-ray photons from a constant intensity monoenergetic X-ray source, e.g. a long lifetime radioactive source. The ionization charge is calibrated by the number of X-ray photons absorbed in the gas, counted with a photon detector. For this purpose, a hybrid detector system was tested and its use in W-value measurements was investigated. The technique was applied to pure xenon at 825 Torr with 5.9 keV X-rays and a W-value of 21.61-0.10+0.14 eV was obtained for a 68% confidence level. The required corrections and the different factors contributing to the accuracy of the results are discussed. The advantages and limitations of this technique are explored and future developments are discussed.
id RCAP_302cfc4ec4de5d62e827a50c645fb4db
oai_identifier_str oai:estudogeral.uc.pt:10316/4527
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling A technique for the absolute measurement of the W-value for X-rays in counting gasesW-valueIonization yieldRadiation physicsDosimetryX-ray detectorsA technique was developed for the absolute measurement of the W-value (the mean energy for the production of an electron-ion pair) for low-energy X-rays in a wide range of gases at atmospheric pressures, with a standard uncertainty better than 1%. This technique is based on the absolute measurement of the primary ionization charge produced by X-ray photons from a constant intensity monoenergetic X-ray source, e.g. a long lifetime radioactive source. The ionization charge is calibrated by the number of X-ray photons absorbed in the gas, counted with a photon detector. For this purpose, a hybrid detector system was tested and its use in W-value measurements was investigated. The technique was applied to pure xenon at 825 Torr with 5.9 keV X-rays and a W-value of 21.61-0.10+0.14 eV was obtained for a 68% confidence level. The required corrections and the different factors contributing to the accuracy of the results are discussed. The advantages and limitations of this technique are explored and future developments are discussed.http://www.sciencedirect.com/science/article/B6TJM-40WDR8N-N/1/6c1c7baa3952be96f48aa99c589e7a182000info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleaplication/PDFhttp://hdl.handle.net/10316/4527http://hdl.handle.net/10316/4527engNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 450:2-3 (2000) 365-372Vinagre, F. L. R.Conde, C. A. N.info:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2020-05-29T10:04:20Zoai:estudogeral.uc.pt:10316/4527Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-19T20:59:38.824457Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv A technique for the absolute measurement of the W-value for X-rays in counting gases
title A technique for the absolute measurement of the W-value for X-rays in counting gases
spellingShingle A technique for the absolute measurement of the W-value for X-rays in counting gases
Vinagre, F. L. R.
W-value
Ionization yield
Radiation physics
Dosimetry
X-ray detectors
title_short A technique for the absolute measurement of the W-value for X-rays in counting gases
title_full A technique for the absolute measurement of the W-value for X-rays in counting gases
title_fullStr A technique for the absolute measurement of the W-value for X-rays in counting gases
title_full_unstemmed A technique for the absolute measurement of the W-value for X-rays in counting gases
title_sort A technique for the absolute measurement of the W-value for X-rays in counting gases
author Vinagre, F. L. R.
author_facet Vinagre, F. L. R.
Conde, C. A. N.
author_role author
author2 Conde, C. A. N.
author2_role author
dc.contributor.author.fl_str_mv Vinagre, F. L. R.
Conde, C. A. N.
dc.subject.por.fl_str_mv W-value
Ionization yield
Radiation physics
Dosimetry
X-ray detectors
topic W-value
Ionization yield
Radiation physics
Dosimetry
X-ray detectors
description A technique was developed for the absolute measurement of the W-value (the mean energy for the production of an electron-ion pair) for low-energy X-rays in a wide range of gases at atmospheric pressures, with a standard uncertainty better than 1%. This technique is based on the absolute measurement of the primary ionization charge produced by X-ray photons from a constant intensity monoenergetic X-ray source, e.g. a long lifetime radioactive source. The ionization charge is calibrated by the number of X-ray photons absorbed in the gas, counted with a photon detector. For this purpose, a hybrid detector system was tested and its use in W-value measurements was investigated. The technique was applied to pure xenon at 825 Torr with 5.9 keV X-rays and a W-value of 21.61-0.10+0.14 eV was obtained for a 68% confidence level. The required corrections and the different factors contributing to the accuracy of the results are discussed. The advantages and limitations of this technique are explored and future developments are discussed.
publishDate 2000
dc.date.none.fl_str_mv 2000
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10316/4527
http://hdl.handle.net/10316/4527
url http://hdl.handle.net/10316/4527
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 450:2-3 (2000) 365-372
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv aplication/PDF
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799133887845105664