Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph

Detalhes bibliográficos
Autor(a) principal: Vilaso,J. E.
Data de Publicação: 2016
Outros Autores: Arada,M. A., Baez,J. A., Cespedes,A.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://scielo.pt/scielo.php?script=sci_arttext&pid=S0872-19042016000500002
Resumo: A minimal instrumentation micropolarograph was built, and its metrological characterization was performed. The oxidation of K4[Fe(CN)6] was studied by linear voltammetry. A microdisc and a bar of carbon were used as working and auxiliary electrodes, respectively. A bar divided into compartments of copper-internal solution- cotton was used as reference. The linear coefficient of determination between the limiting current and the concentration was 0.9926, in the range of 10-2 - 10-1 mol L-1 . The coefficient of variation of ten voltammograms at 0.08 mol L-1 was 1.8 %, being inferior to the maximum value reported for these methods. Mood and Kruskal-Wallis tests did not show significant differences between voltammograms. In the range of linear correlation coefficients of potential sweep 0.9488 to 0.9989, no sweep influence on voltammograms was observed. The expanded uncertainty associated with the determination of the limiting current was 1.9 μA. The linear voltammograms for oxidation of ascorbic acid and KI were obtained, showing correspondence with the reported results.
id RCAP_bb6602f631480c7f98af781bf1fd96cb
oai_identifier_str oai:scielo:S0872-19042016000500002
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling Construction and Metrological Characterization of a Minimal Instrumentation MicropolarographMinimal Instrumentation Micropolarographlinear voltammetrymetrologyA minimal instrumentation micropolarograph was built, and its metrological characterization was performed. The oxidation of K4[Fe(CN)6] was studied by linear voltammetry. A microdisc and a bar of carbon were used as working and auxiliary electrodes, respectively. A bar divided into compartments of copper-internal solution- cotton was used as reference. The linear coefficient of determination between the limiting current and the concentration was 0.9926, in the range of 10-2 - 10-1 mol L-1 . The coefficient of variation of ten voltammograms at 0.08 mol L-1 was 1.8 %, being inferior to the maximum value reported for these methods. Mood and Kruskal-Wallis tests did not show significant differences between voltammograms. In the range of linear correlation coefficients of potential sweep 0.9488 to 0.9989, no sweep influence on voltammograms was observed. The expanded uncertainty associated with the determination of the limiting current was 1.9 μA. The linear voltammograms for oxidation of ascorbic acid and KI were obtained, showing correspondence with the reported results.Sociedade Portuguesa de Electroquímica2016-09-01info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articletext/htmlhttp://scielo.pt/scielo.php?script=sci_arttext&pid=S0872-19042016000500002Portugaliae Electrochimica Acta v.34 n.5 2016reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAPenghttp://scielo.pt/scielo.php?script=sci_arttext&pid=S0872-19042016000500002Vilaso,J. E.Arada,M. A.Baez,J. A.Cespedes,A.info:eu-repo/semantics/openAccess2024-02-06T17:07:22Zoai:scielo:S0872-19042016000500002Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-20T02:20:18.792231Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
title Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
spellingShingle Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
Vilaso,J. E.
Minimal Instrumentation Micropolarograph
linear voltammetry
metrology
title_short Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
title_full Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
title_fullStr Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
title_full_unstemmed Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
title_sort Construction and Metrological Characterization of a Minimal Instrumentation Micropolarograph
author Vilaso,J. E.
author_facet Vilaso,J. E.
Arada,M. A.
Baez,J. A.
Cespedes,A.
author_role author
author2 Arada,M. A.
Baez,J. A.
Cespedes,A.
author2_role author
author
author
dc.contributor.author.fl_str_mv Vilaso,J. E.
Arada,M. A.
Baez,J. A.
Cespedes,A.
dc.subject.por.fl_str_mv Minimal Instrumentation Micropolarograph
linear voltammetry
metrology
topic Minimal Instrumentation Micropolarograph
linear voltammetry
metrology
description A minimal instrumentation micropolarograph was built, and its metrological characterization was performed. The oxidation of K4[Fe(CN)6] was studied by linear voltammetry. A microdisc and a bar of carbon were used as working and auxiliary electrodes, respectively. A bar divided into compartments of copper-internal solution- cotton was used as reference. The linear coefficient of determination between the limiting current and the concentration was 0.9926, in the range of 10-2 - 10-1 mol L-1 . The coefficient of variation of ten voltammograms at 0.08 mol L-1 was 1.8 %, being inferior to the maximum value reported for these methods. Mood and Kruskal-Wallis tests did not show significant differences between voltammograms. In the range of linear correlation coefficients of potential sweep 0.9488 to 0.9989, no sweep influence on voltammograms was observed. The expanded uncertainty associated with the determination of the limiting current was 1.9 μA. The linear voltammograms for oxidation of ascorbic acid and KI were obtained, showing correspondence with the reported results.
publishDate 2016
dc.date.none.fl_str_mv 2016-09-01
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://scielo.pt/scielo.php?script=sci_arttext&pid=S0872-19042016000500002
url http://scielo.pt/scielo.php?script=sci_arttext&pid=S0872-19042016000500002
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv http://scielo.pt/scielo.php?script=sci_arttext&pid=S0872-19042016000500002
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Portuguesa de Electroquímica
publisher.none.fl_str_mv Sociedade Portuguesa de Electroquímica
dc.source.none.fl_str_mv Portugaliae Electrochimica Acta v.34 n.5 2016
reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799137291489247232