Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population

Detalhes bibliográficos
Autor(a) principal: Fernandes, Paulo Rodrigues
Data de Publicação: 2005
Outros Autores: Diaz-Rey, José Alberto, González-Méijome, José Manuel, Jorge, Jorge, Queirós, A.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/1822/5792
Resumo: The aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare) in comparison with the Goldmann applanation tonometer (AT). The left eyes of 46 university students were assessed with the two tonometers, with induction tonometry being performed first. The ICare was handled by an optometrist and the Goldmann tonometer by an ophthalmologist. In this study, statistically significant differences were found when comparing the ICare rebound tonometer with applanation tonometry (AT) (p < 0.05). The mean difference between the two tonometers was 1.34 +/- 2.03 mmHg (mean +/- S.D.) and the 95% limits of agreement were +/-3.98 mmHg. A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <3 mmHg between the ICare and the AT. In the present population the ICare overestimates the IOP value by 1.34 mmHg on average when compared with Goldmann tonometer. Nevertheless, the ICare tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable or not recommended, as it is able to estimate IOP within a range of +/-3.00 mmHg in more than 80% of the population
id RCAP_cc3f759772c41c1d1f078f60e92edb2d
oai_identifier_str oai:repositorium.sdum.uminho.pt:1822/5792
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal populationInductionApplanation tonometryImpact tonometryIOPNon-contact tonometryRebound tonometerScience & TechnologyThe aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare) in comparison with the Goldmann applanation tonometer (AT). The left eyes of 46 university students were assessed with the two tonometers, with induction tonometry being performed first. The ICare was handled by an optometrist and the Goldmann tonometer by an ophthalmologist. In this study, statistically significant differences were found when comparing the ICare rebound tonometer with applanation tonometry (AT) (p < 0.05). The mean difference between the two tonometers was 1.34 +/- 2.03 mmHg (mean +/- S.D.) and the 95% limits of agreement were +/-3.98 mmHg. A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <3 mmHg between the ICare and the AT. In the present population the ICare overestimates the IOP value by 1.34 mmHg on average when compared with Goldmann tonometer. Nevertheless, the ICare tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable or not recommended, as it is able to estimate IOP within a range of +/-3.00 mmHg in more than 80% of the populationBlackwell PublishingUniversidade do MinhoFernandes, Paulo RodriguesDiaz-Rey, José AlbertoGonzález-Méijome, José ManuelJorge, JorgeQueirós, A.2005-092005-09-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/1822/5792eng"Ophthalmic and physiological optics". ISSN 1040-5488. 25:5 (Sept.2005) 436-440.1040-548810.1111/j.1475-1313.2005.00327.x16101950The definitive version is available at www.blackwell-synergy.cominfo:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2023-07-21T11:56:54Zoai:repositorium.sdum.uminho.pt:1822/5792Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-19T18:46:34.791880Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
title Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
spellingShingle Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
Fernandes, Paulo Rodrigues
Induction
Applanation tonometry
Impact tonometry
IOP
Non-contact tonometry
Rebound tonometer
Science & Technology
title_short Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
title_full Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
title_fullStr Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
title_full_unstemmed Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
title_sort Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
author Fernandes, Paulo Rodrigues
author_facet Fernandes, Paulo Rodrigues
Diaz-Rey, José Alberto
González-Méijome, José Manuel
Jorge, Jorge
Queirós, A.
author_role author
author2 Diaz-Rey, José Alberto
González-Méijome, José Manuel
Jorge, Jorge
Queirós, A.
author2_role author
author
author
author
dc.contributor.none.fl_str_mv Universidade do Minho
dc.contributor.author.fl_str_mv Fernandes, Paulo Rodrigues
Diaz-Rey, José Alberto
González-Méijome, José Manuel
Jorge, Jorge
Queirós, A.
dc.subject.por.fl_str_mv Induction
Applanation tonometry
Impact tonometry
IOP
Non-contact tonometry
Rebound tonometer
Science & Technology
topic Induction
Applanation tonometry
Impact tonometry
IOP
Non-contact tonometry
Rebound tonometer
Science & Technology
description The aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare) in comparison with the Goldmann applanation tonometer (AT). The left eyes of 46 university students were assessed with the two tonometers, with induction tonometry being performed first. The ICare was handled by an optometrist and the Goldmann tonometer by an ophthalmologist. In this study, statistically significant differences were found when comparing the ICare rebound tonometer with applanation tonometry (AT) (p < 0.05). The mean difference between the two tonometers was 1.34 +/- 2.03 mmHg (mean +/- S.D.) and the 95% limits of agreement were +/-3.98 mmHg. A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <3 mmHg between the ICare and the AT. In the present population the ICare overestimates the IOP value by 1.34 mmHg on average when compared with Goldmann tonometer. Nevertheless, the ICare tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable or not recommended, as it is able to estimate IOP within a range of +/-3.00 mmHg in more than 80% of the population
publishDate 2005
dc.date.none.fl_str_mv 2005-09
2005-09-01T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/1822/5792
url http://hdl.handle.net/1822/5792
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv "Ophthalmic and physiological optics". ISSN 1040-5488. 25:5 (Sept.2005) 436-440.
1040-5488
10.1111/j.1475-1313.2005.00327.x
16101950
The definitive version is available at www.blackwell-synergy.com
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Blackwell Publishing
publisher.none.fl_str_mv Blackwell Publishing
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799132222054203392