In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes

Detalhes bibliográficos
Autor(a) principal: Coelho,Rafael de Macedo
Data de Publicação: 2016
Outros Autores: Paiva,Thiago Trigueiro Morais de, Mathias,Ligia Andrade da Silva Telles
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista Brasileira de Anestesiologia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942016000200120
Resumo: ABSTRACT BACKGROUND AND OBJECTIVE: Cuffs of tracheal tubes protect the lower airway from aspiration of gastric contents and facilitate ventilation, but may cause many complications, especially when the cuff pressure exceeds 30 cm H2O. This occurs in over 30% of conventional insufflations, so it is recommended to limit this pressure. In this study we evaluated the in vitro effectiveness of a method of limiting the cuff pressure to a range between 20 and 30 cm H2O. METHOD: Using an adapter to connect the tested tube to the anesthesia machine, the relief valve was regulated to 30 cm H2O, inflating the cuff by operating the rapid flow of oxygen button. There were 33 trials for each tube of three manufacturers, of five sizes (6.5-8.5), using three times inflation (10, 15 and 20 s), totaling 1485 tests. After inflation, the pressure obtained was measured with a manometer. Pressure >30 cm H2O or <20 cm H2O were considered failures. RESULTS: There were eight failures (0.5%, 95% CI: 0.1-0.9%), with all by pressures <20 cm H2O and after 10 s inflation (1.6%, 95% CI: 0 5-2.7%). One failure occurred with a 6.5 tube (0.3%, 95% CI: -0.3 to 0.9%), six with 7.0 tubes (2%, 95% CI: 0.4-3.6%), and one with a 7.5 tube (0.3%, 95% CI: -0.3 to 0.9%). CONCLUSION: This method was effective for inflating tracheal tube cuffs of different sizes and manufacturers, limiting its pressure to a range between 20 and 30 cm H2O, with a success rate of 99.5% (95% CI: 99.1-99.9%).
id SBA-1_887185ec81d514631dc163a38c2de8b0
oai_identifier_str oai:scielo:S0034-70942016000200120
network_acronym_str SBA-1
network_name_str Revista Brasileira de Anestesiologia (Online)
repository_id_str
spelling In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubesMethodInflationCuffsTubePressureABSTRACT BACKGROUND AND OBJECTIVE: Cuffs of tracheal tubes protect the lower airway from aspiration of gastric contents and facilitate ventilation, but may cause many complications, especially when the cuff pressure exceeds 30 cm H2O. This occurs in over 30% of conventional insufflations, so it is recommended to limit this pressure. In this study we evaluated the in vitro effectiveness of a method of limiting the cuff pressure to a range between 20 and 30 cm H2O. METHOD: Using an adapter to connect the tested tube to the anesthesia machine, the relief valve was regulated to 30 cm H2O, inflating the cuff by operating the rapid flow of oxygen button. There were 33 trials for each tube of three manufacturers, of five sizes (6.5-8.5), using three times inflation (10, 15 and 20 s), totaling 1485 tests. After inflation, the pressure obtained was measured with a manometer. Pressure >30 cm H2O or <20 cm H2O were considered failures. RESULTS: There were eight failures (0.5%, 95% CI: 0.1-0.9%), with all by pressures <20 cm H2O and after 10 s inflation (1.6%, 95% CI: 0 5-2.7%). One failure occurred with a 6.5 tube (0.3%, 95% CI: -0.3 to 0.9%), six with 7.0 tubes (2%, 95% CI: 0.4-3.6%), and one with a 7.5 tube (0.3%, 95% CI: -0.3 to 0.9%). CONCLUSION: This method was effective for inflating tracheal tube cuffs of different sizes and manufacturers, limiting its pressure to a range between 20 and 30 cm H2O, with a success rate of 99.5% (95% CI: 99.1-99.9%).Sociedade Brasileira de Anestesiologia2016-04-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942016000200120Revista Brasileira de Anestesiologia v.66 n.2 2016reponame:Revista Brasileira de Anestesiologia (Online)instname:Sociedade Brasileira de Anestesiologia (SBA)instacron:SBA10.1016/j.bjane.2014.06.012info:eu-repo/semantics/openAccessCoelho,Rafael de MacedoPaiva,Thiago Trigueiro Morais deMathias,Ligia Andrade da Silva Telleseng2016-03-29T00:00:00Zoai:scielo:S0034-70942016000200120Revistahttps://www.sbahq.org/revista/https://old.scielo.br/oai/scielo-oai.php||sba2000@openlink.com.br1806-907X0034-7094opendoar:2016-03-29T00:00Revista Brasileira de Anestesiologia (Online) - Sociedade Brasileira de Anestesiologia (SBA)false
dc.title.none.fl_str_mv In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
title In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
spellingShingle In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
Coelho,Rafael de Macedo
Method
Inflation
Cuffs
Tube
Pressure
title_short In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
title_full In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
title_fullStr In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
title_full_unstemmed In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
title_sort In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes
author Coelho,Rafael de Macedo
author_facet Coelho,Rafael de Macedo
Paiva,Thiago Trigueiro Morais de
Mathias,Ligia Andrade da Silva Telles
author_role author
author2 Paiva,Thiago Trigueiro Morais de
Mathias,Ligia Andrade da Silva Telles
author2_role author
author
dc.contributor.author.fl_str_mv Coelho,Rafael de Macedo
Paiva,Thiago Trigueiro Morais de
Mathias,Ligia Andrade da Silva Telles
dc.subject.por.fl_str_mv Method
Inflation
Cuffs
Tube
Pressure
topic Method
Inflation
Cuffs
Tube
Pressure
description ABSTRACT BACKGROUND AND OBJECTIVE: Cuffs of tracheal tubes protect the lower airway from aspiration of gastric contents and facilitate ventilation, but may cause many complications, especially when the cuff pressure exceeds 30 cm H2O. This occurs in over 30% of conventional insufflations, so it is recommended to limit this pressure. In this study we evaluated the in vitro effectiveness of a method of limiting the cuff pressure to a range between 20 and 30 cm H2O. METHOD: Using an adapter to connect the tested tube to the anesthesia machine, the relief valve was regulated to 30 cm H2O, inflating the cuff by operating the rapid flow of oxygen button. There were 33 trials for each tube of three manufacturers, of five sizes (6.5-8.5), using three times inflation (10, 15 and 20 s), totaling 1485 tests. After inflation, the pressure obtained was measured with a manometer. Pressure >30 cm H2O or <20 cm H2O were considered failures. RESULTS: There were eight failures (0.5%, 95% CI: 0.1-0.9%), with all by pressures <20 cm H2O and after 10 s inflation (1.6%, 95% CI: 0 5-2.7%). One failure occurred with a 6.5 tube (0.3%, 95% CI: -0.3 to 0.9%), six with 7.0 tubes (2%, 95% CI: 0.4-3.6%), and one with a 7.5 tube (0.3%, 95% CI: -0.3 to 0.9%). CONCLUSION: This method was effective for inflating tracheal tube cuffs of different sizes and manufacturers, limiting its pressure to a range between 20 and 30 cm H2O, with a success rate of 99.5% (95% CI: 99.1-99.9%).
publishDate 2016
dc.date.none.fl_str_mv 2016-04-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942016000200120
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942016000200120
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1016/j.bjane.2014.06.012
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Anestesiologia
publisher.none.fl_str_mv Sociedade Brasileira de Anestesiologia
dc.source.none.fl_str_mv Revista Brasileira de Anestesiologia v.66 n.2 2016
reponame:Revista Brasileira de Anestesiologia (Online)
instname:Sociedade Brasileira de Anestesiologia (SBA)
instacron:SBA
instname_str Sociedade Brasileira de Anestesiologia (SBA)
instacron_str SBA
institution SBA
reponame_str Revista Brasileira de Anestesiologia (Online)
collection Revista Brasileira de Anestesiologia (Online)
repository.name.fl_str_mv Revista Brasileira de Anestesiologia (Online) - Sociedade Brasileira de Anestesiologia (SBA)
repository.mail.fl_str_mv ||sba2000@openlink.com.br
_version_ 1752126628639539200