Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head

Detalhes bibliográficos
Autor(a) principal: Guimarães,Daniel Peres
Data de Publicação: 2020
Outros Autores: Caneo,Luiz Fernando, Matte,Gregory, Carletto,Luciana P., Policarpo,Valéria Camargo, Castro,Ana Vitória C. X., Miranda,Matheus H. C., Costa,Priscila S., Jatene,Marcelo B., Cestari,Idagene, Jatene,Fabio B.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Journal of Cardiovascular Surgery (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382020000200134
Resumo: Abstract Objective: To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter. Methods: The simulation circuit consisted of a Quadrox-I Pediatric oxygenator, a Rotaflow CP (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set primed with Lactated Ringer's solution and packed red blood cells. Venous line pressure, reservoir pressure, and arterial flow were measured with VAVD turned off to record baseline values. Four other conditions were tested with progressively higher vacuum pressures (-20, -40, -60, and -80 mmHg) applied to the baseline cardiotomy pressure. An arterial filter was placed into the circuit and arterial flow was measured with the purge line in both open and closed positions. These trials were repeated at set arterial flow rates of 1500, 2000, and 2500 mL/min. Results: The use of progressively higher vacuum caused a reduction in effective arterial flow from 1490±0.00 to 590±0.00, from 2020±0.01 to 1220±0.00, and from 2490±0.0 to 1830±0.01 mL/min. Effective forward flow decreased with increased levels of VAVD. Conclusion: The use of VAVD reduces arterial flow when a CP is used as the main arterial pump. The reduction in the forward arterial flow increases as the vacuum level increases. The loss of forward flow is further reduced when the arterial filter purge line is kept in the recommended open position. An independent flow probe is essential to monitor pump flow during cardiopulmonary bypass.
id SBCCV-1_ef60d2420f1955a33252e14604347bf4
oai_identifier_str oai:scielo:S0102-76382020000200134
network_acronym_str SBCCV-1
network_name_str Brazilian Journal of Cardiovascular Surgery (Online)
repository_id_str
spelling Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump HeadCongenital Heart DiseaseCardiopulmonary BypassChildVacuumDrainagePressureErythrocytesAbstract Objective: To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter. Methods: The simulation circuit consisted of a Quadrox-I Pediatric oxygenator, a Rotaflow CP (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set primed with Lactated Ringer's solution and packed red blood cells. Venous line pressure, reservoir pressure, and arterial flow were measured with VAVD turned off to record baseline values. Four other conditions were tested with progressively higher vacuum pressures (-20, -40, -60, and -80 mmHg) applied to the baseline cardiotomy pressure. An arterial filter was placed into the circuit and arterial flow was measured with the purge line in both open and closed positions. These trials were repeated at set arterial flow rates of 1500, 2000, and 2500 mL/min. Results: The use of progressively higher vacuum caused a reduction in effective arterial flow from 1490±0.00 to 590±0.00, from 2020±0.01 to 1220±0.00, and from 2490±0.0 to 1830±0.01 mL/min. Effective forward flow decreased with increased levels of VAVD. Conclusion: The use of VAVD reduces arterial flow when a CP is used as the main arterial pump. The reduction in the forward arterial flow increases as the vacuum level increases. The loss of forward flow is further reduced when the arterial filter purge line is kept in the recommended open position. An independent flow probe is essential to monitor pump flow during cardiopulmonary bypass.Sociedade Brasileira de Cirurgia Cardiovascular2020-04-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382020000200134Brazilian Journal of Cardiovascular Surgery v.35 n.2 2020reponame:Brazilian Journal of Cardiovascular Surgery (Online)instname:Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)instacron:SBCCV10.21470/1678-9741-2019-0311info:eu-repo/semantics/openAccessGuimarães,Daniel PeresCaneo,Luiz FernandoMatte,GregoryCarletto,Luciana P.Policarpo,Valéria CamargoCastro,Ana Vitória C. X.Miranda,Matheus H. C.Costa,Priscila S.Jatene,Marcelo B.Cestari,IdageneJatene,Fabio B.eng2020-04-28T00:00:00Zoai:scielo:S0102-76382020000200134Revistahttp://www.rbccv.org.br/https://old.scielo.br/oai/scielo-oai.php||rosangela.monteiro@incor.usp.br|| domingo@braile.com.br|| brandau@braile.com.br1678-97410102-7638opendoar:2020-04-28T00:00Brazilian Journal of Cardiovascular Surgery (Online) - Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)false
dc.title.none.fl_str_mv Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
spellingShingle Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
Guimarães,Daniel Peres
Congenital Heart Disease
Cardiopulmonary Bypass
Child
Vacuum
Drainage
Pressure
Erythrocytes
title_short Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_full Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_fullStr Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_full_unstemmed Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
title_sort Impact of Vacuum-Assisted Venous Drainage on Forward Flow in Simulated Pediatric Cardiopulmonary Bypass Circuits Utilizing a Centrifugal Arterial Pump Head
author Guimarães,Daniel Peres
author_facet Guimarães,Daniel Peres
Caneo,Luiz Fernando
Matte,Gregory
Carletto,Luciana P.
Policarpo,Valéria Camargo
Castro,Ana Vitória C. X.
Miranda,Matheus H. C.
Costa,Priscila S.
Jatene,Marcelo B.
Cestari,Idagene
Jatene,Fabio B.
author_role author
author2 Caneo,Luiz Fernando
Matte,Gregory
Carletto,Luciana P.
Policarpo,Valéria Camargo
Castro,Ana Vitória C. X.
Miranda,Matheus H. C.
Costa,Priscila S.
Jatene,Marcelo B.
Cestari,Idagene
Jatene,Fabio B.
author2_role author
author
author
author
author
author
author
author
author
author
dc.contributor.author.fl_str_mv Guimarães,Daniel Peres
Caneo,Luiz Fernando
Matte,Gregory
Carletto,Luciana P.
Policarpo,Valéria Camargo
Castro,Ana Vitória C. X.
Miranda,Matheus H. C.
Costa,Priscila S.
Jatene,Marcelo B.
Cestari,Idagene
Jatene,Fabio B.
dc.subject.por.fl_str_mv Congenital Heart Disease
Cardiopulmonary Bypass
Child
Vacuum
Drainage
Pressure
Erythrocytes
topic Congenital Heart Disease
Cardiopulmonary Bypass
Child
Vacuum
Drainage
Pressure
Erythrocytes
description Abstract Objective: To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter. Methods: The simulation circuit consisted of a Quadrox-I Pediatric oxygenator, a Rotaflow CP (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set primed with Lactated Ringer's solution and packed red blood cells. Venous line pressure, reservoir pressure, and arterial flow were measured with VAVD turned off to record baseline values. Four other conditions were tested with progressively higher vacuum pressures (-20, -40, -60, and -80 mmHg) applied to the baseline cardiotomy pressure. An arterial filter was placed into the circuit and arterial flow was measured with the purge line in both open and closed positions. These trials were repeated at set arterial flow rates of 1500, 2000, and 2500 mL/min. Results: The use of progressively higher vacuum caused a reduction in effective arterial flow from 1490±0.00 to 590±0.00, from 2020±0.01 to 1220±0.00, and from 2490±0.0 to 1830±0.01 mL/min. Effective forward flow decreased with increased levels of VAVD. Conclusion: The use of VAVD reduces arterial flow when a CP is used as the main arterial pump. The reduction in the forward arterial flow increases as the vacuum level increases. The loss of forward flow is further reduced when the arterial filter purge line is kept in the recommended open position. An independent flow probe is essential to monitor pump flow during cardiopulmonary bypass.
publishDate 2020
dc.date.none.fl_str_mv 2020-04-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382020000200134
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382020000200134
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.21470/1678-9741-2019-0311
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Cirurgia Cardiovascular
publisher.none.fl_str_mv Sociedade Brasileira de Cirurgia Cardiovascular
dc.source.none.fl_str_mv Brazilian Journal of Cardiovascular Surgery v.35 n.2 2020
reponame:Brazilian Journal of Cardiovascular Surgery (Online)
instname:Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)
instacron:SBCCV
instname_str Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)
instacron_str SBCCV
institution SBCCV
reponame_str Brazilian Journal of Cardiovascular Surgery (Online)
collection Brazilian Journal of Cardiovascular Surgery (Online)
repository.name.fl_str_mv Brazilian Journal of Cardiovascular Surgery (Online) - Sociedade Brasileira de Cirurgia Cardiovascular (SBCCV)
repository.mail.fl_str_mv ||rosangela.monteiro@incor.usp.br|| domingo@braile.com.br|| brandau@braile.com.br
_version_ 1752126601718398976