Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse

Detalhes bibliográficos
Autor(a) principal: Kim,Ji-Hyun
Data de Publicação: 2022
Outros Autores: Choi,Jin-Young, Yoon,Hun-Young
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Acta Cirúrgica Brasileira (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-86502022000500202
Resumo: ABSTRACT Purpose: To identify an optimal self-expandable metallic stent (SEMS) and verify whether a mechanically superior SEMS would result in better clinical outcomes in the treatment of tracheal collapse. Methods: We selected three SEMS (n = 8 each), including an S-type stent with a wire diameter of 0.006 inches (S6) and two D-type stents with wire diameters of 0.006 (D6) and 0.007 inches (D7). Twenty-four New Zealand White rabbits were divided into three equal groups. After the stents were deployed, the clinical signs were recorded daily, and radiographic examinations were performed monthly. All rabbits were euthanized after three months. Results: Two rabbits with S6 stents and one rabbit with a D7 stent died within three months because of stent migration or pneumonia. All rabbits with D6 stents survived for three months. On histological examination, the D6 group had the lowest inflammation score. Conclusions: Both clinically and histopathologically, the results with D-type stents with a wire diameter of 0.006 inches were superior to those of the other groups (p = 0.001). The use of an optimal intraluminal stent may improve the long-term clinical outcomes in the treatment of tracheal collapse in dogs.
id SBDPC-1_f6009013bcb230cc27cd6287ef1d2b62
oai_identifier_str oai:scielo:S0102-86502022000500202
network_acronym_str SBDPC-1
network_name_str Acta Cirúrgica Brasileira (Online)
repository_id_str
spelling Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapseSelf Expandable Metallic StentsTracheaRabbitsABSTRACT Purpose: To identify an optimal self-expandable metallic stent (SEMS) and verify whether a mechanically superior SEMS would result in better clinical outcomes in the treatment of tracheal collapse. Methods: We selected three SEMS (n = 8 each), including an S-type stent with a wire diameter of 0.006 inches (S6) and two D-type stents with wire diameters of 0.006 (D6) and 0.007 inches (D7). Twenty-four New Zealand White rabbits were divided into three equal groups. After the stents were deployed, the clinical signs were recorded daily, and radiographic examinations were performed monthly. All rabbits were euthanized after three months. Results: Two rabbits with S6 stents and one rabbit with a D7 stent died within three months because of stent migration or pneumonia. All rabbits with D6 stents survived for three months. On histological examination, the D6 group had the lowest inflammation score. Conclusions: Both clinically and histopathologically, the results with D-type stents with a wire diameter of 0.006 inches were superior to those of the other groups (p = 0.001). The use of an optimal intraluminal stent may improve the long-term clinical outcomes in the treatment of tracheal collapse in dogs.Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia2022-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-86502022000500202Acta Cirúrgica Brasileira v.37 n.5 2022reponame:Acta Cirúrgica Brasileira (Online)instname:Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia (SBDPC)instacron:SBDPC10.1590/acb370502info:eu-repo/semantics/openAccessKim,Ji-HyunChoi,Jin-YoungYoon,Hun-Youngeng2022-08-09T00:00:00Zoai:scielo:S0102-86502022000500202Revistahttps://www.bvs-vet.org.br/vetindex/periodicos/acta-cirurgica-brasileira/https://old.scielo.br/oai/scielo-oai.php||sgolden@terra.com.br0102-86501678-2674opendoar:2022-08-09T00:00Acta Cirúrgica Brasileira (Online) - Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia (SBDPC)false
dc.title.none.fl_str_mv Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
title Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
spellingShingle Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
Kim,Ji-Hyun
Self Expandable Metallic Stents
Trachea
Rabbits
title_short Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
title_full Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
title_fullStr Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
title_full_unstemmed Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
title_sort Comparison of three different self-expanding metal stents using rabbit models for the treatment of tracheal collapse
author Kim,Ji-Hyun
author_facet Kim,Ji-Hyun
Choi,Jin-Young
Yoon,Hun-Young
author_role author
author2 Choi,Jin-Young
Yoon,Hun-Young
author2_role author
author
dc.contributor.author.fl_str_mv Kim,Ji-Hyun
Choi,Jin-Young
Yoon,Hun-Young
dc.subject.por.fl_str_mv Self Expandable Metallic Stents
Trachea
Rabbits
topic Self Expandable Metallic Stents
Trachea
Rabbits
description ABSTRACT Purpose: To identify an optimal self-expandable metallic stent (SEMS) and verify whether a mechanically superior SEMS would result in better clinical outcomes in the treatment of tracheal collapse. Methods: We selected three SEMS (n = 8 each), including an S-type stent with a wire diameter of 0.006 inches (S6) and two D-type stents with wire diameters of 0.006 (D6) and 0.007 inches (D7). Twenty-four New Zealand White rabbits were divided into three equal groups. After the stents were deployed, the clinical signs were recorded daily, and radiographic examinations were performed monthly. All rabbits were euthanized after three months. Results: Two rabbits with S6 stents and one rabbit with a D7 stent died within three months because of stent migration or pneumonia. All rabbits with D6 stents survived for three months. On histological examination, the D6 group had the lowest inflammation score. Conclusions: Both clinically and histopathologically, the results with D-type stents with a wire diameter of 0.006 inches were superior to those of the other groups (p = 0.001). The use of an optimal intraluminal stent may improve the long-term clinical outcomes in the treatment of tracheal collapse in dogs.
publishDate 2022
dc.date.none.fl_str_mv 2022-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-86502022000500202
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0102-86502022000500202
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/acb370502
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia
publisher.none.fl_str_mv Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia
dc.source.none.fl_str_mv Acta Cirúrgica Brasileira v.37 n.5 2022
reponame:Acta Cirúrgica Brasileira (Online)
instname:Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia (SBDPC)
instacron:SBDPC
instname_str Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia (SBDPC)
instacron_str SBDPC
institution SBDPC
reponame_str Acta Cirúrgica Brasileira (Online)
collection Acta Cirúrgica Brasileira (Online)
repository.name.fl_str_mv Acta Cirúrgica Brasileira (Online) - Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia (SBDPC)
repository.mail.fl_str_mv ||sgolden@terra.com.br
_version_ 1752126446529150976