Testing electrode suitability for field stimulation of high-threshold biological preparations
Autor(a) principal: | |
---|---|
Data de Publicação: | 2015 |
Outros Autores: | , |
Tipo de documento: | Relatório |
Idioma: | eng |
Título da fonte: | Research on Biomedical Engineering (Online) |
Texto Completo: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S2446-47402015000300273 |
Resumo: | IntroductionA problem posed by electrical field (E) stimulation of biological preparations with high excitation threshold is that the E intensity required for excitation is likely to induce water electrolysis at the electrode surface, which can alter the extracellular medium and cause deleterious effects on the cells. In this study, different electrode materials and geometries were tested aiming at identifying electrode configurations that could transduce the E intensity required for exciting ventricular cardiomyocytes isolated from neonatal rats (threshold E ~30 V/cm) without causing water electrolysis.MethodsWire and plate electrodes made of platinum, stainless steel and nickel/chrome alloy were used. The effect of blasting the electrode surface with sand and NaHCO3 solution was also tested. Electrodes were inserted into a cell perfusion chamber containing the saline solution routinely used for physiological experiments. During E application for 5 min, the electrode surface and its surroundings were examined at high magnification for the presence of microbubbles, which indicates the occurrence of water electrolysis. The greatest E intensity applied that failed to generate microbubbles (En) was estimated.ResultsWhile nickel/chrome and stainless steel electrodes resulted in low En values, the best performance was observed for sandblasted platinum wire (2 mm diameter) and plate (25 mm x 5 mm; 0.1 mm thickness) electrodes, for which Enwas ≥40 V/cm.ConclusionThese electrode configurations are suitable for effective and safe stimulation of isolated neonatal cardiomyocytes. |
id |
SBEB-1_c7417d3000260d0fef3f33c41664faa3 |
---|---|
oai_identifier_str |
oai:scielo:S2446-47402015000300273 |
network_acronym_str |
SBEB-1 |
network_name_str |
Research on Biomedical Engineering (Online) |
repository_id_str |
|
spelling |
Testing electrode suitability for field stimulation of high-threshold biological preparationsElectrodesField stimulationIsolated neonatal cardiomyocyteWater electrolysisIntroductionA problem posed by electrical field (E) stimulation of biological preparations with high excitation threshold is that the E intensity required for excitation is likely to induce water electrolysis at the electrode surface, which can alter the extracellular medium and cause deleterious effects on the cells. In this study, different electrode materials and geometries were tested aiming at identifying electrode configurations that could transduce the E intensity required for exciting ventricular cardiomyocytes isolated from neonatal rats (threshold E ~30 V/cm) without causing water electrolysis.MethodsWire and plate electrodes made of platinum, stainless steel and nickel/chrome alloy were used. The effect of blasting the electrode surface with sand and NaHCO3 solution was also tested. Electrodes were inserted into a cell perfusion chamber containing the saline solution routinely used for physiological experiments. During E application for 5 min, the electrode surface and its surroundings were examined at high magnification for the presence of microbubbles, which indicates the occurrence of water electrolysis. The greatest E intensity applied that failed to generate microbubbles (En) was estimated.ResultsWhile nickel/chrome and stainless steel electrodes resulted in low En values, the best performance was observed for sandblasted platinum wire (2 mm diameter) and plate (25 mm x 5 mm; 0.1 mm thickness) electrodes, for which Enwas ≥40 V/cm.ConclusionThese electrode configurations are suitable for effective and safe stimulation of isolated neonatal cardiomyocytes.Sociedade Brasileira de Engenharia Biomédica2015-09-01info:eu-repo/semantics/reportinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S2446-47402015000300273Research on Biomedical Engineering v.31 n.3 2015reponame:Research on Biomedical Engineering (Online)instname:Sociedade Brasileira de Engenharia Biomédica (SBEB)instacron:SBEB10.1590/2446-4740.0718info:eu-repo/semantics/openAccessMilan,Hugo Fernando MaiaBassani,Rosana AlmadaBassani,José Wilson Magalhãeseng2015-10-20T00:00:00Zoai:scielo:S2446-47402015000300273Revistahttp://www.rbejournal.org/https://old.scielo.br/oai/scielo-oai.php||rbe@rbejournal.org2446-47402446-4732opendoar:2015-10-20T00:00Research on Biomedical Engineering (Online) - Sociedade Brasileira de Engenharia Biomédica (SBEB)false |
dc.title.none.fl_str_mv |
Testing electrode suitability for field stimulation of high-threshold biological preparations |
title |
Testing electrode suitability for field stimulation of high-threshold biological preparations |
spellingShingle |
Testing electrode suitability for field stimulation of high-threshold biological preparations Milan,Hugo Fernando Maia Electrodes Field stimulation Isolated neonatal cardiomyocyte Water electrolysis |
title_short |
Testing electrode suitability for field stimulation of high-threshold biological preparations |
title_full |
Testing electrode suitability for field stimulation of high-threshold biological preparations |
title_fullStr |
Testing electrode suitability for field stimulation of high-threshold biological preparations |
title_full_unstemmed |
Testing electrode suitability for field stimulation of high-threshold biological preparations |
title_sort |
Testing electrode suitability for field stimulation of high-threshold biological preparations |
author |
Milan,Hugo Fernando Maia |
author_facet |
Milan,Hugo Fernando Maia Bassani,Rosana Almada Bassani,José Wilson Magalhães |
author_role |
author |
author2 |
Bassani,Rosana Almada Bassani,José Wilson Magalhães |
author2_role |
author author |
dc.contributor.author.fl_str_mv |
Milan,Hugo Fernando Maia Bassani,Rosana Almada Bassani,José Wilson Magalhães |
dc.subject.por.fl_str_mv |
Electrodes Field stimulation Isolated neonatal cardiomyocyte Water electrolysis |
topic |
Electrodes Field stimulation Isolated neonatal cardiomyocyte Water electrolysis |
description |
IntroductionA problem posed by electrical field (E) stimulation of biological preparations with high excitation threshold is that the E intensity required for excitation is likely to induce water electrolysis at the electrode surface, which can alter the extracellular medium and cause deleterious effects on the cells. In this study, different electrode materials and geometries were tested aiming at identifying electrode configurations that could transduce the E intensity required for exciting ventricular cardiomyocytes isolated from neonatal rats (threshold E ~30 V/cm) without causing water electrolysis.MethodsWire and plate electrodes made of platinum, stainless steel and nickel/chrome alloy were used. The effect of blasting the electrode surface with sand and NaHCO3 solution was also tested. Electrodes were inserted into a cell perfusion chamber containing the saline solution routinely used for physiological experiments. During E application for 5 min, the electrode surface and its surroundings were examined at high magnification for the presence of microbubbles, which indicates the occurrence of water electrolysis. The greatest E intensity applied that failed to generate microbubbles (En) was estimated.ResultsWhile nickel/chrome and stainless steel electrodes resulted in low En values, the best performance was observed for sandblasted platinum wire (2 mm diameter) and plate (25 mm x 5 mm; 0.1 mm thickness) electrodes, for which Enwas ≥40 V/cm.ConclusionThese electrode configurations are suitable for effective and safe stimulation of isolated neonatal cardiomyocytes. |
publishDate |
2015 |
dc.date.none.fl_str_mv |
2015-09-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/report |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
report |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S2446-47402015000300273 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S2446-47402015000300273 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.1590/2446-4740.0718 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
Sociedade Brasileira de Engenharia Biomédica |
publisher.none.fl_str_mv |
Sociedade Brasileira de Engenharia Biomédica |
dc.source.none.fl_str_mv |
Research on Biomedical Engineering v.31 n.3 2015 reponame:Research on Biomedical Engineering (Online) instname:Sociedade Brasileira de Engenharia Biomédica (SBEB) instacron:SBEB |
instname_str |
Sociedade Brasileira de Engenharia Biomédica (SBEB) |
instacron_str |
SBEB |
institution |
SBEB |
reponame_str |
Research on Biomedical Engineering (Online) |
collection |
Research on Biomedical Engineering (Online) |
repository.name.fl_str_mv |
Research on Biomedical Engineering (Online) - Sociedade Brasileira de Engenharia Biomédica (SBEB) |
repository.mail.fl_str_mv |
||rbe@rbejournal.org |
_version_ |
1752126288207806464 |