Continuous positive airway pressure setups evaluated at simulated exercise conditions

Detalhes bibliográficos
Autor(a) principal: Nunes,Marcelo Vieira Leão
Data de Publicação: 2014
Outros Autores: Giannella-Neto,Antonio, Tavares,Frederico Caetano Jandre de Assis
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista Brasileira de Engenharia Biomédica (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009
Resumo: INTRODUCTION: Studies have shown increases in airway opening pressure (Pao) swings and work of breathing (WOB) by different continuous positive airway pressure (CPAP) devices at rest, but few address this issue during exercise. The aim of the present work was to analyze the imposed WOB (WOBi), the apparent resistance (Rapp) and swings of Pao (deltaP) of 3 CPAP assemblies at simulated exercise conditions. METHODS: The CPAP measures were obtained from: a commercial CPAP (Assembly 1), a high flow CPAP (Assembly 2) and the parallel association of these devices (Assembly 3). In each assembly the spring-loaded positive end-expiratory pressure (PEEP) valve was set to fully opened (mode A) or at the same CPAP pressure (mode B). The exercise protocol simulation, performed manually by a calibrated syringe and a metronome, employed a respiratory frequency of 30 bpm, tidal volume of 2.7 L and inspiratory-to-expiratory ratio of 1. The setups were evaluated at CPAP settings of 5, 10 and 15 cmH2O. RESULTS: The lowest deltaP as well as Rapp and WOBi were obtained with Assembly 3 in mode A with an adjusted CPAP of 10 cmH2O (deltaP=8.1 (0.5) cmH2O, WOBi=1.4 (0.14) cmH2O/L/s, Rapp= 1.3 (0.07) J/s) showed as median (interquartile range). CONCLUSION: For the conditions studied, the best CPAP setup was obtained with mode A.
id SBEB-2_78eb14d931278033703f23a959401984
oai_identifier_str oai:scielo:S1517-31512014000200009
network_acronym_str SBEB-2
network_name_str Revista Brasileira de Engenharia Biomédica (Online)
repository_id_str
spelling Continuous positive airway pressure setups evaluated at simulated exercise conditionsCPAPExerciseImposed work of breathingPEEP valveINTRODUCTION: Studies have shown increases in airway opening pressure (Pao) swings and work of breathing (WOB) by different continuous positive airway pressure (CPAP) devices at rest, but few address this issue during exercise. The aim of the present work was to analyze the imposed WOB (WOBi), the apparent resistance (Rapp) and swings of Pao (deltaP) of 3 CPAP assemblies at simulated exercise conditions. METHODS: The CPAP measures were obtained from: a commercial CPAP (Assembly 1), a high flow CPAP (Assembly 2) and the parallel association of these devices (Assembly 3). In each assembly the spring-loaded positive end-expiratory pressure (PEEP) valve was set to fully opened (mode A) or at the same CPAP pressure (mode B). The exercise protocol simulation, performed manually by a calibrated syringe and a metronome, employed a respiratory frequency of 30 bpm, tidal volume of 2.7 L and inspiratory-to-expiratory ratio of 1. The setups were evaluated at CPAP settings of 5, 10 and 15 cmH2O. RESULTS: The lowest deltaP as well as Rapp and WOBi were obtained with Assembly 3 in mode A with an adjusted CPAP of 10 cmH2O (deltaP=8.1 (0.5) cmH2O, WOBi=1.4 (0.14) cmH2O/L/s, Rapp= 1.3 (0.07) J/s) showed as median (interquartile range). CONCLUSION: For the conditions studied, the best CPAP setup was obtained with mode A.SBEB - Sociedade Brasileira de Engenharia Biomédica2014-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009Revista Brasileira de Engenharia Biomédica v.30 n.2 2014reponame:Revista Brasileira de Engenharia Biomédica (Online)instname:Sociedade Brasileira de Engenharia Biomédica (SBEB)instacron:SBEB10.1590/rbeb.2014.014info:eu-repo/semantics/openAccessNunes,Marcelo Vieira LeãoGiannella-Neto,AntonioTavares,Frederico Caetano Jandre de Assiseng2014-07-14T00:00:00Zoai:scielo:S1517-31512014000200009Revistahttp://www.scielo.br/rbebONGhttps://old.scielo.br/oai/scielo-oai.php||rbeb@rbeb.org.br1984-77421517-3151opendoar:2014-07-14T00:00Revista Brasileira de Engenharia Biomédica (Online) - Sociedade Brasileira de Engenharia Biomédica (SBEB)false
dc.title.none.fl_str_mv Continuous positive airway pressure setups evaluated at simulated exercise conditions
title Continuous positive airway pressure setups evaluated at simulated exercise conditions
spellingShingle Continuous positive airway pressure setups evaluated at simulated exercise conditions
Nunes,Marcelo Vieira Leão
CPAP
Exercise
Imposed work of breathing
PEEP valve
title_short Continuous positive airway pressure setups evaluated at simulated exercise conditions
title_full Continuous positive airway pressure setups evaluated at simulated exercise conditions
title_fullStr Continuous positive airway pressure setups evaluated at simulated exercise conditions
title_full_unstemmed Continuous positive airway pressure setups evaluated at simulated exercise conditions
title_sort Continuous positive airway pressure setups evaluated at simulated exercise conditions
author Nunes,Marcelo Vieira Leão
author_facet Nunes,Marcelo Vieira Leão
Giannella-Neto,Antonio
Tavares,Frederico Caetano Jandre de Assis
author_role author
author2 Giannella-Neto,Antonio
Tavares,Frederico Caetano Jandre de Assis
author2_role author
author
dc.contributor.author.fl_str_mv Nunes,Marcelo Vieira Leão
Giannella-Neto,Antonio
Tavares,Frederico Caetano Jandre de Assis
dc.subject.por.fl_str_mv CPAP
Exercise
Imposed work of breathing
PEEP valve
topic CPAP
Exercise
Imposed work of breathing
PEEP valve
description INTRODUCTION: Studies have shown increases in airway opening pressure (Pao) swings and work of breathing (WOB) by different continuous positive airway pressure (CPAP) devices at rest, but few address this issue during exercise. The aim of the present work was to analyze the imposed WOB (WOBi), the apparent resistance (Rapp) and swings of Pao (deltaP) of 3 CPAP assemblies at simulated exercise conditions. METHODS: The CPAP measures were obtained from: a commercial CPAP (Assembly 1), a high flow CPAP (Assembly 2) and the parallel association of these devices (Assembly 3). In each assembly the spring-loaded positive end-expiratory pressure (PEEP) valve was set to fully opened (mode A) or at the same CPAP pressure (mode B). The exercise protocol simulation, performed manually by a calibrated syringe and a metronome, employed a respiratory frequency of 30 bpm, tidal volume of 2.7 L and inspiratory-to-expiratory ratio of 1. The setups were evaluated at CPAP settings of 5, 10 and 15 cmH2O. RESULTS: The lowest deltaP as well as Rapp and WOBi were obtained with Assembly 3 in mode A with an adjusted CPAP of 10 cmH2O (deltaP=8.1 (0.5) cmH2O, WOBi=1.4 (0.14) cmH2O/L/s, Rapp= 1.3 (0.07) J/s) showed as median (interquartile range). CONCLUSION: For the conditions studied, the best CPAP setup was obtained with mode A.
publishDate 2014
dc.date.none.fl_str_mv 2014-06-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/rbeb.2014.014
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv SBEB - Sociedade Brasileira de Engenharia Biomédica
publisher.none.fl_str_mv SBEB - Sociedade Brasileira de Engenharia Biomédica
dc.source.none.fl_str_mv Revista Brasileira de Engenharia Biomédica v.30 n.2 2014
reponame:Revista Brasileira de Engenharia Biomédica (Online)
instname:Sociedade Brasileira de Engenharia Biomédica (SBEB)
instacron:SBEB
instname_str Sociedade Brasileira de Engenharia Biomédica (SBEB)
instacron_str SBEB
institution SBEB
reponame_str Revista Brasileira de Engenharia Biomédica (Online)
collection Revista Brasileira de Engenharia Biomédica (Online)
repository.name.fl_str_mv Revista Brasileira de Engenharia Biomédica (Online) - Sociedade Brasileira de Engenharia Biomédica (SBEB)
repository.mail.fl_str_mv ||rbeb@rbeb.org.br
_version_ 1754820915119521792