Continuous positive airway pressure setups evaluated at simulated exercise conditions
Autor(a) principal: | |
---|---|
Data de Publicação: | 2014 |
Outros Autores: | , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Revista Brasileira de Engenharia Biomédica (Online) |
Texto Completo: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009 |
Resumo: | INTRODUCTION: Studies have shown increases in airway opening pressure (Pao) swings and work of breathing (WOB) by different continuous positive airway pressure (CPAP) devices at rest, but few address this issue during exercise. The aim of the present work was to analyze the imposed WOB (WOBi), the apparent resistance (Rapp) and swings of Pao (deltaP) of 3 CPAP assemblies at simulated exercise conditions. METHODS: The CPAP measures were obtained from: a commercial CPAP (Assembly 1), a high flow CPAP (Assembly 2) and the parallel association of these devices (Assembly 3). In each assembly the spring-loaded positive end-expiratory pressure (PEEP) valve was set to fully opened (mode A) or at the same CPAP pressure (mode B). The exercise protocol simulation, performed manually by a calibrated syringe and a metronome, employed a respiratory frequency of 30 bpm, tidal volume of 2.7 L and inspiratory-to-expiratory ratio of 1. The setups were evaluated at CPAP settings of 5, 10 and 15 cmH2O. RESULTS: The lowest deltaP as well as Rapp and WOBi were obtained with Assembly 3 in mode A with an adjusted CPAP of 10 cmH2O (deltaP=8.1 (0.5) cmH2O, WOBi=1.4 (0.14) cmH2O/L/s, Rapp= 1.3 (0.07) J/s) showed as median (interquartile range). CONCLUSION: For the conditions studied, the best CPAP setup was obtained with mode A. |
id |
SBEB-2_78eb14d931278033703f23a959401984 |
---|---|
oai_identifier_str |
oai:scielo:S1517-31512014000200009 |
network_acronym_str |
SBEB-2 |
network_name_str |
Revista Brasileira de Engenharia Biomédica (Online) |
repository_id_str |
|
spelling |
Continuous positive airway pressure setups evaluated at simulated exercise conditionsCPAPExerciseImposed work of breathingPEEP valveINTRODUCTION: Studies have shown increases in airway opening pressure (Pao) swings and work of breathing (WOB) by different continuous positive airway pressure (CPAP) devices at rest, but few address this issue during exercise. The aim of the present work was to analyze the imposed WOB (WOBi), the apparent resistance (Rapp) and swings of Pao (deltaP) of 3 CPAP assemblies at simulated exercise conditions. METHODS: The CPAP measures were obtained from: a commercial CPAP (Assembly 1), a high flow CPAP (Assembly 2) and the parallel association of these devices (Assembly 3). In each assembly the spring-loaded positive end-expiratory pressure (PEEP) valve was set to fully opened (mode A) or at the same CPAP pressure (mode B). The exercise protocol simulation, performed manually by a calibrated syringe and a metronome, employed a respiratory frequency of 30 bpm, tidal volume of 2.7 L and inspiratory-to-expiratory ratio of 1. The setups were evaluated at CPAP settings of 5, 10 and 15 cmH2O. RESULTS: The lowest deltaP as well as Rapp and WOBi were obtained with Assembly 3 in mode A with an adjusted CPAP of 10 cmH2O (deltaP=8.1 (0.5) cmH2O, WOBi=1.4 (0.14) cmH2O/L/s, Rapp= 1.3 (0.07) J/s) showed as median (interquartile range). CONCLUSION: For the conditions studied, the best CPAP setup was obtained with mode A.SBEB - Sociedade Brasileira de Engenharia Biomédica2014-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009Revista Brasileira de Engenharia Biomédica v.30 n.2 2014reponame:Revista Brasileira de Engenharia Biomédica (Online)instname:Sociedade Brasileira de Engenharia Biomédica (SBEB)instacron:SBEB10.1590/rbeb.2014.014info:eu-repo/semantics/openAccessNunes,Marcelo Vieira LeãoGiannella-Neto,AntonioTavares,Frederico Caetano Jandre de Assiseng2014-07-14T00:00:00Zoai:scielo:S1517-31512014000200009Revistahttp://www.scielo.br/rbebONGhttps://old.scielo.br/oai/scielo-oai.php||rbeb@rbeb.org.br1984-77421517-3151opendoar:2014-07-14T00:00Revista Brasileira de Engenharia Biomédica (Online) - Sociedade Brasileira de Engenharia Biomédica (SBEB)false |
dc.title.none.fl_str_mv |
Continuous positive airway pressure setups evaluated at simulated exercise conditions |
title |
Continuous positive airway pressure setups evaluated at simulated exercise conditions |
spellingShingle |
Continuous positive airway pressure setups evaluated at simulated exercise conditions Nunes,Marcelo Vieira Leão CPAP Exercise Imposed work of breathing PEEP valve |
title_short |
Continuous positive airway pressure setups evaluated at simulated exercise conditions |
title_full |
Continuous positive airway pressure setups evaluated at simulated exercise conditions |
title_fullStr |
Continuous positive airway pressure setups evaluated at simulated exercise conditions |
title_full_unstemmed |
Continuous positive airway pressure setups evaluated at simulated exercise conditions |
title_sort |
Continuous positive airway pressure setups evaluated at simulated exercise conditions |
author |
Nunes,Marcelo Vieira Leão |
author_facet |
Nunes,Marcelo Vieira Leão Giannella-Neto,Antonio Tavares,Frederico Caetano Jandre de Assis |
author_role |
author |
author2 |
Giannella-Neto,Antonio Tavares,Frederico Caetano Jandre de Assis |
author2_role |
author author |
dc.contributor.author.fl_str_mv |
Nunes,Marcelo Vieira Leão Giannella-Neto,Antonio Tavares,Frederico Caetano Jandre de Assis |
dc.subject.por.fl_str_mv |
CPAP Exercise Imposed work of breathing PEEP valve |
topic |
CPAP Exercise Imposed work of breathing PEEP valve |
description |
INTRODUCTION: Studies have shown increases in airway opening pressure (Pao) swings and work of breathing (WOB) by different continuous positive airway pressure (CPAP) devices at rest, but few address this issue during exercise. The aim of the present work was to analyze the imposed WOB (WOBi), the apparent resistance (Rapp) and swings of Pao (deltaP) of 3 CPAP assemblies at simulated exercise conditions. METHODS: The CPAP measures were obtained from: a commercial CPAP (Assembly 1), a high flow CPAP (Assembly 2) and the parallel association of these devices (Assembly 3). In each assembly the spring-loaded positive end-expiratory pressure (PEEP) valve was set to fully opened (mode A) or at the same CPAP pressure (mode B). The exercise protocol simulation, performed manually by a calibrated syringe and a metronome, employed a respiratory frequency of 30 bpm, tidal volume of 2.7 L and inspiratory-to-expiratory ratio of 1. The setups were evaluated at CPAP settings of 5, 10 and 15 cmH2O. RESULTS: The lowest deltaP as well as Rapp and WOBi were obtained with Assembly 3 in mode A with an adjusted CPAP of 10 cmH2O (deltaP=8.1 (0.5) cmH2O, WOBi=1.4 (0.14) cmH2O/L/s, Rapp= 1.3 (0.07) J/s) showed as median (interquartile range). CONCLUSION: For the conditions studied, the best CPAP setup was obtained with mode A. |
publishDate |
2014 |
dc.date.none.fl_str_mv |
2014-06-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1517-31512014000200009 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.1590/rbeb.2014.014 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
SBEB - Sociedade Brasileira de Engenharia Biomédica |
publisher.none.fl_str_mv |
SBEB - Sociedade Brasileira de Engenharia Biomédica |
dc.source.none.fl_str_mv |
Revista Brasileira de Engenharia Biomédica v.30 n.2 2014 reponame:Revista Brasileira de Engenharia Biomédica (Online) instname:Sociedade Brasileira de Engenharia Biomédica (SBEB) instacron:SBEB |
instname_str |
Sociedade Brasileira de Engenharia Biomédica (SBEB) |
instacron_str |
SBEB |
institution |
SBEB |
reponame_str |
Revista Brasileira de Engenharia Biomédica (Online) |
collection |
Revista Brasileira de Engenharia Biomédica (Online) |
repository.name.fl_str_mv |
Revista Brasileira de Engenharia Biomédica (Online) - Sociedade Brasileira de Engenharia Biomédica (SBEB) |
repository.mail.fl_str_mv |
||rbeb@rbeb.org.br |
_version_ |
1754820915119521792 |