Remote programming of cochlear implants

Detalhes bibliográficos
Autor(a) principal: Samuel,Paola Angelica
Data de Publicação: 2014
Outros Autores: Goffi-Gomez,Maria Valéria Schmidt, Bittencourt,Aline Gomes, Tsuji,Robinson Koji, Brito,Rubens de
Tipo de documento: Artigo
Idioma: eng
Título da fonte: CoDAS
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S2317-17822014000600481
Resumo: PURPOSE: To verify the effectiveness of remote programming of cochlear implants by stimulation levels and results in the perception of speech and free-field audiometry tests. METHODS: Twelve patients from both genders, aged between 18 and 59 years, users of internal cochlear implant and speech processor of the same model for at least 12 months, were selected. Both the remote programming (RP) and the live programming (LP) were performed on the same day, measuring the minimum (T) and maximum (C) stimulation levels of five electrodes with the interpolation of the remaining ones. Speech perception tests were applied using 65 dBSPL (recorded open context sentences and monosyllables). The patients were submitted to free-field audiometry at 250-8,000 Hz frequencies. The results for the RP and LP were compared. RESULTS: Differences in mean of the T levels for three electrodes and the C levels for one electrode were found. No difference between the results was obtained in the speech perception tests and audiometric thresholds in the RP and LP. CONCLUSION: The RP is a simple and effective procedure for programming cochlear implant devices and, although there were differences in the stimulation levels of some electrodes, it did not interfere in the speech perception outcomes.
id SBFA-1_712dcd8a96b4f3869c817a0c537c0374
oai_identifier_str oai:scielo:S2317-17822014000600481
network_acronym_str SBFA-1
network_name_str CoDAS
repository_id_str
spelling Remote programming of cochlear implants PURPOSE: To verify the effectiveness of remote programming of cochlear implants by stimulation levels and results in the perception of speech and free-field audiometry tests. METHODS: Twelve patients from both genders, aged between 18 and 59 years, users of internal cochlear implant and speech processor of the same model for at least 12 months, were selected. Both the remote programming (RP) and the live programming (LP) were performed on the same day, measuring the minimum (T) and maximum (C) stimulation levels of five electrodes with the interpolation of the remaining ones. Speech perception tests were applied using 65 dBSPL (recorded open context sentences and monosyllables). The patients were submitted to free-field audiometry at 250-8,000 Hz frequencies. The results for the RP and LP were compared. RESULTS: Differences in mean of the T levels for three electrodes and the C levels for one electrode were found. No difference between the results was obtained in the speech perception tests and audiometric thresholds in the RP and LP. CONCLUSION: The RP is a simple and effective procedure for programming cochlear implant devices and, although there were differences in the stimulation levels of some electrodes, it did not interfere in the speech perception outcomes. Sociedade Brasileira de Fonoaudiologia2014-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S2317-17822014000600481CoDAS v.26 n.6 2014reponame:CoDASinstname:Sociedade Brasileira de Fonoaudiologia (SBFA)instacron:SBFA10.1590/2317-1782/20142014007info:eu-repo/semantics/openAccessSamuel,Paola AngelicaGoffi-Gomez,Maria Valéria SchmidtBittencourt,Aline GomesTsuji,Robinson KojiBrito,Rubens deeng2015-08-18T00:00:00Zoai:scielo:S2317-17822014000600481Revistahttps://www.codas.org.br/ONGhttps://old.scielo.br/oai/scielo-oai.phpcodas@editoracubo.com.br||revista@codas.org.br2317-17822317-1782opendoar:2015-08-18T00:00CoDAS - Sociedade Brasileira de Fonoaudiologia (SBFA)false
dc.title.none.fl_str_mv Remote programming of cochlear implants
title Remote programming of cochlear implants
spellingShingle Remote programming of cochlear implants
Samuel,Paola Angelica
title_short Remote programming of cochlear implants
title_full Remote programming of cochlear implants
title_fullStr Remote programming of cochlear implants
title_full_unstemmed Remote programming of cochlear implants
title_sort Remote programming of cochlear implants
author Samuel,Paola Angelica
author_facet Samuel,Paola Angelica
Goffi-Gomez,Maria Valéria Schmidt
Bittencourt,Aline Gomes
Tsuji,Robinson Koji
Brito,Rubens de
author_role author
author2 Goffi-Gomez,Maria Valéria Schmidt
Bittencourt,Aline Gomes
Tsuji,Robinson Koji
Brito,Rubens de
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Samuel,Paola Angelica
Goffi-Gomez,Maria Valéria Schmidt
Bittencourt,Aline Gomes
Tsuji,Robinson Koji
Brito,Rubens de
description PURPOSE: To verify the effectiveness of remote programming of cochlear implants by stimulation levels and results in the perception of speech and free-field audiometry tests. METHODS: Twelve patients from both genders, aged between 18 and 59 years, users of internal cochlear implant and speech processor of the same model for at least 12 months, were selected. Both the remote programming (RP) and the live programming (LP) were performed on the same day, measuring the minimum (T) and maximum (C) stimulation levels of five electrodes with the interpolation of the remaining ones. Speech perception tests were applied using 65 dBSPL (recorded open context sentences and monosyllables). The patients were submitted to free-field audiometry at 250-8,000 Hz frequencies. The results for the RP and LP were compared. RESULTS: Differences in mean of the T levels for three electrodes and the C levels for one electrode were found. No difference between the results was obtained in the speech perception tests and audiometric thresholds in the RP and LP. CONCLUSION: The RP is a simple and effective procedure for programming cochlear implant devices and, although there were differences in the stimulation levels of some electrodes, it did not interfere in the speech perception outcomes.
publishDate 2014
dc.date.none.fl_str_mv 2014-12-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S2317-17822014000600481
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S2317-17822014000600481
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/2317-1782/20142014007
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Fonoaudiologia
publisher.none.fl_str_mv Sociedade Brasileira de Fonoaudiologia
dc.source.none.fl_str_mv CoDAS v.26 n.6 2014
reponame:CoDAS
instname:Sociedade Brasileira de Fonoaudiologia (SBFA)
instacron:SBFA
instname_str Sociedade Brasileira de Fonoaudiologia (SBFA)
instacron_str SBFA
institution SBFA
reponame_str CoDAS
collection CoDAS
repository.name.fl_str_mv CoDAS - Sociedade Brasileira de Fonoaudiologia (SBFA)
repository.mail.fl_str_mv codas@editoracubo.com.br||revista@codas.org.br
_version_ 1752122440416231424