Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius

Detalhes bibliográficos
Autor(a) principal: Favre,Tereza Cristina
Data de Publicação: 2022
Outros Autores: Beck,Lilian Christina Nóbrega Holsback, Bezerra,Fernando Schemelzer Moraes, Graeff-Teixeira,Carlos, Coelho,Paulo Marcos Zech, Enk,Martin Johannes, Katz,Naftale, Oliveira,Ricardo Riccio, Reis,Mitermayer Galvão dos, Pieri,Otávio Sarmento
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista da Sociedade Brasileira de Medicina Tropical
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0037-86822022000100303
Resumo: ABSTRACT Background The World Health Organization recommends reliable point-of-care (POC) diagnostic testing to eliminate schistosomiasis. Lateral flow immunoassay that detects schistosome circulating cathodic antigen (CCA) in urine to establish prevalence thresholds for intervention in endemic areas is recommended. Stored urine may be useful if surveying at-risk populations is delayed or interrupted by unforeseen circumstances, such as the current COVID-19 pandemic. This study evaluated the manufacturer’s claim that Schistosoma mansoni infection can be reliably diagnosed in urine samples stored at -20°C for one year. Methods Two-hundred-forty-two subjects from an endemic site in Brazil provided one urine sample each for testing with URINE CCA (SCHISTO) ECO TESTE® (POC-ECO) and one stool sample each for testing with Kato-Katz (KK) and Helmintex® (HTX) as a robust reference standard for infection status. At least 2 ml of urine from each participant was stored at -20°C; after one year, 76 samples were randomly selected for POC-ECO retesting. Results: The POC-ECO agreement between freshly collected and stored urine was inadequate considering trace results as positive (Cohen’s kappa coefficient κ = 0.08) and negative (κ = 0.36). POC-ECO accuracy was not significantly greater than that of routine KK (54%; 95% confidence interval: 42.1%-65.5%). Conclusions The precision and accuracy of POC-ECO have to be optimized in both freshly collected and stored urine before it can be recommended for use in control programs in Brazil.
id SBMT-1_825c4d0fa28665c5face7db1a255d784
oai_identifier_str oai:scielo:S0037-86822022000100303
network_acronym_str SBMT-1
network_name_str Revista da Sociedade Brasileira de Medicina Tropical
repository_id_str
spelling Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees CelsiusSchistosoma mansoniPoint-of-Care (POC) testingCirculating Cathodic Antigen (CCA), frozen urineKato-KatzHelmintex®BrazilABSTRACT Background The World Health Organization recommends reliable point-of-care (POC) diagnostic testing to eliminate schistosomiasis. Lateral flow immunoassay that detects schistosome circulating cathodic antigen (CCA) in urine to establish prevalence thresholds for intervention in endemic areas is recommended. Stored urine may be useful if surveying at-risk populations is delayed or interrupted by unforeseen circumstances, such as the current COVID-19 pandemic. This study evaluated the manufacturer’s claim that Schistosoma mansoni infection can be reliably diagnosed in urine samples stored at -20°C for one year. Methods Two-hundred-forty-two subjects from an endemic site in Brazil provided one urine sample each for testing with URINE CCA (SCHISTO) ECO TESTE® (POC-ECO) and one stool sample each for testing with Kato-Katz (KK) and Helmintex® (HTX) as a robust reference standard for infection status. At least 2 ml of urine from each participant was stored at -20°C; after one year, 76 samples were randomly selected for POC-ECO retesting. Results: The POC-ECO agreement between freshly collected and stored urine was inadequate considering trace results as positive (Cohen’s kappa coefficient κ = 0.08) and negative (κ = 0.36). POC-ECO accuracy was not significantly greater than that of routine KK (54%; 95% confidence interval: 42.1%-65.5%). Conclusions The precision and accuracy of POC-ECO have to be optimized in both freshly collected and stored urine before it can be recommended for use in control programs in Brazil.Sociedade Brasileira de Medicina Tropical - SBMT2022-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0037-86822022000100303Revista da Sociedade Brasileira de Medicina Tropical v.55 2022reponame:Revista da Sociedade Brasileira de Medicina Tropicalinstname:Sociedade Brasileira de Medicina Tropical (SBMT)instacron:SBMT10.1590/0037-8682-0389-2021info:eu-repo/semantics/openAccessFavre,Tereza CristinaBeck,Lilian Christina Nóbrega HolsbackBezerra,Fernando Schemelzer MoraesGraeff-Teixeira,CarlosCoelho,Paulo Marcos ZechEnk,Martin JohannesKatz,NaftaleOliveira,Ricardo RiccioReis,Mitermayer Galvão dosPieri,Otávio Sarmentoeng2022-02-23T00:00:00Zoai:scielo:S0037-86822022000100303Revistahttps://www.sbmt.org.br/portal/revista/ONGhttps://old.scielo.br/oai/scielo-oai.php||dalmo@rsbmt.uftm.edu.br|| rsbmt@rsbmt.uftm.edu.br1678-98490037-8682opendoar:2022-02-23T00:00Revista da Sociedade Brasileira de Medicina Tropical - Sociedade Brasileira de Medicina Tropical (SBMT)false
dc.title.none.fl_str_mv Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
title Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
spellingShingle Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
Favre,Tereza Cristina
Schistosoma mansoni
Point-of-Care (POC) testing
Circulating Cathodic Antigen (CCA), frozen urine
Kato-Katz
Helmintex®
Brazil
title_short Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
title_full Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
title_fullStr Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
title_full_unstemmed Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
title_sort Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
author Favre,Tereza Cristina
author_facet Favre,Tereza Cristina
Beck,Lilian Christina Nóbrega Holsback
Bezerra,Fernando Schemelzer Moraes
Graeff-Teixeira,Carlos
Coelho,Paulo Marcos Zech
Enk,Martin Johannes
Katz,Naftale
Oliveira,Ricardo Riccio
Reis,Mitermayer Galvão dos
Pieri,Otávio Sarmento
author_role author
author2 Beck,Lilian Christina Nóbrega Holsback
Bezerra,Fernando Schemelzer Moraes
Graeff-Teixeira,Carlos
Coelho,Paulo Marcos Zech
Enk,Martin Johannes
Katz,Naftale
Oliveira,Ricardo Riccio
Reis,Mitermayer Galvão dos
Pieri,Otávio Sarmento
author2_role author
author
author
author
author
author
author
author
author
dc.contributor.author.fl_str_mv Favre,Tereza Cristina
Beck,Lilian Christina Nóbrega Holsback
Bezerra,Fernando Schemelzer Moraes
Graeff-Teixeira,Carlos
Coelho,Paulo Marcos Zech
Enk,Martin Johannes
Katz,Naftale
Oliveira,Ricardo Riccio
Reis,Mitermayer Galvão dos
Pieri,Otávio Sarmento
dc.subject.por.fl_str_mv Schistosoma mansoni
Point-of-Care (POC) testing
Circulating Cathodic Antigen (CCA), frozen urine
Kato-Katz
Helmintex®
Brazil
topic Schistosoma mansoni
Point-of-Care (POC) testing
Circulating Cathodic Antigen (CCA), frozen urine
Kato-Katz
Helmintex®
Brazil
description ABSTRACT Background The World Health Organization recommends reliable point-of-care (POC) diagnostic testing to eliminate schistosomiasis. Lateral flow immunoassay that detects schistosome circulating cathodic antigen (CCA) in urine to establish prevalence thresholds for intervention in endemic areas is recommended. Stored urine may be useful if surveying at-risk populations is delayed or interrupted by unforeseen circumstances, such as the current COVID-19 pandemic. This study evaluated the manufacturer’s claim that Schistosoma mansoni infection can be reliably diagnosed in urine samples stored at -20°C for one year. Methods Two-hundred-forty-two subjects from an endemic site in Brazil provided one urine sample each for testing with URINE CCA (SCHISTO) ECO TESTE® (POC-ECO) and one stool sample each for testing with Kato-Katz (KK) and Helmintex® (HTX) as a robust reference standard for infection status. At least 2 ml of urine from each participant was stored at -20°C; after one year, 76 samples were randomly selected for POC-ECO retesting. Results: The POC-ECO agreement between freshly collected and stored urine was inadequate considering trace results as positive (Cohen’s kappa coefficient κ = 0.08) and negative (κ = 0.36). POC-ECO accuracy was not significantly greater than that of routine KK (54%; 95% confidence interval: 42.1%-65.5%). Conclusions The precision and accuracy of POC-ECO have to be optimized in both freshly collected and stored urine before it can be recommended for use in control programs in Brazil.
publishDate 2022
dc.date.none.fl_str_mv 2022-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0037-86822022000100303
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0037-86822022000100303
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/0037-8682-0389-2021
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Medicina Tropical - SBMT
publisher.none.fl_str_mv Sociedade Brasileira de Medicina Tropical - SBMT
dc.source.none.fl_str_mv Revista da Sociedade Brasileira de Medicina Tropical v.55 2022
reponame:Revista da Sociedade Brasileira de Medicina Tropical
instname:Sociedade Brasileira de Medicina Tropical (SBMT)
instacron:SBMT
instname_str Sociedade Brasileira de Medicina Tropical (SBMT)
instacron_str SBMT
institution SBMT
reponame_str Revista da Sociedade Brasileira de Medicina Tropical
collection Revista da Sociedade Brasileira de Medicina Tropical
repository.name.fl_str_mv Revista da Sociedade Brasileira de Medicina Tropical - Sociedade Brasileira de Medicina Tropical (SBMT)
repository.mail.fl_str_mv ||dalmo@rsbmt.uftm.edu.br|| rsbmt@rsbmt.uftm.edu.br
_version_ 1752122163016499200