High peak of carbamylated hemoglobina discordant with urea level: a case report

Bibliographic Details
Main Author: Ab-Rahim,Siti N.
Publication Date: 2021
Other Authors: Tuan-Ismail,Tuan S., Ibrahim,Hanim A., Hassan,Mohd N.
Format: Report
Language: eng
Source: Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
Download full: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442021000100706
Summary: ABSTRACT Hemoglobin A1c (HbA1c) measurement is commonly performed in diabetes mellitus patients to monitor glycemic control over the last three to four months. Carbamylated hemoglobin, which is the hemoglobin that binds to isocyanic acid derived from urea, is one of the possible analytical interference in the uremic patient. When measured by ion-exchange high-performance liquid chromatography (HPLC), carbamylated hemoglobin forms a peak that overlaps the peak of HbA1c, causing a falsely elevated HbA1c result. We report a case of a 60-years-old man who had a spurious increase in HbA1c, with a high carbamylated hemoglobin peak disproportionate to the urea value. Subsequent hemoglobin analysis using hemoglobin electrophoresis and HPLC hemoglobin testing system suggested hemoglobin J (Hb J) variant. Our case highlighted the possibility of misleading HbA1c interpretation in the presence of a high carbamylated hemoglobin peak, but not proportionate to urea value. In this study, Hb J was detected. A method free from hemoglobin variant interference should be used ideally, and monitoring glycemic control should be performed using alternative methods, such as serum fructosamine or continuous glucose monitoring.
id SBP-1_ef379a389ed3329313ce9bbb663fef50
oai_identifier_str oai:scielo:S1676-24442021000100706
network_acronym_str SBP-1
network_name_str Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
repository_id_str
spelling High peak of carbamylated hemoglobina discordant with urea level: a case reportglycated hemoglobin Achromatography high-pressure liquidcarbamylated hemoglobinureaABSTRACT Hemoglobin A1c (HbA1c) measurement is commonly performed in diabetes mellitus patients to monitor glycemic control over the last three to four months. Carbamylated hemoglobin, which is the hemoglobin that binds to isocyanic acid derived from urea, is one of the possible analytical interference in the uremic patient. When measured by ion-exchange high-performance liquid chromatography (HPLC), carbamylated hemoglobin forms a peak that overlaps the peak of HbA1c, causing a falsely elevated HbA1c result. We report a case of a 60-years-old man who had a spurious increase in HbA1c, with a high carbamylated hemoglobin peak disproportionate to the urea value. Subsequent hemoglobin analysis using hemoglobin electrophoresis and HPLC hemoglobin testing system suggested hemoglobin J (Hb J) variant. Our case highlighted the possibility of misleading HbA1c interpretation in the presence of a high carbamylated hemoglobin peak, but not proportionate to urea value. In this study, Hb J was detected. A method free from hemoglobin variant interference should be used ideally, and monitoring glycemic control should be performed using alternative methods, such as serum fructosamine or continuous glucose monitoring.Sociedade Brasileira de Patologia Clínica2021-01-01info:eu-repo/semantics/reportinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442021000100706Jornal Brasileiro de Patologia e Medicina Laboratorial v.57 2021reponame:Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)instname:Sociedade Brasileira de Patologia (SBP)instacron:SBP10.5935/1676-2444.20210029info:eu-repo/semantics/openAccessAb-Rahim,Siti N.Tuan-Ismail,Tuan S.Ibrahim,Hanim A.Hassan,Mohd N.eng2021-07-06T00:00:00Zoai:scielo:S1676-24442021000100706Revistahttp://www.scielo.br/jbpmlhttps://old.scielo.br/oai/scielo-oai.php||jbpml@sbpc.org.br1678-47741676-2444opendoar:2021-07-06T00:00Jornal Brasileiro de Patologia e Medicina Laboratorial (Online) - Sociedade Brasileira de Patologia (SBP)false
dc.title.none.fl_str_mv High peak of carbamylated hemoglobina discordant with urea level: a case report
title High peak of carbamylated hemoglobina discordant with urea level: a case report
spellingShingle High peak of carbamylated hemoglobina discordant with urea level: a case report
Ab-Rahim,Siti N.
glycated hemoglobin A
chromatography high-pressure liquid
carbamylated hemoglobin
urea
title_short High peak of carbamylated hemoglobina discordant with urea level: a case report
title_full High peak of carbamylated hemoglobina discordant with urea level: a case report
title_fullStr High peak of carbamylated hemoglobina discordant with urea level: a case report
title_full_unstemmed High peak of carbamylated hemoglobina discordant with urea level: a case report
title_sort High peak of carbamylated hemoglobina discordant with urea level: a case report
author Ab-Rahim,Siti N.
author_facet Ab-Rahim,Siti N.
Tuan-Ismail,Tuan S.
Ibrahim,Hanim A.
Hassan,Mohd N.
author_role author
author2 Tuan-Ismail,Tuan S.
Ibrahim,Hanim A.
Hassan,Mohd N.
author2_role author
author
author
dc.contributor.author.fl_str_mv Ab-Rahim,Siti N.
Tuan-Ismail,Tuan S.
Ibrahim,Hanim A.
Hassan,Mohd N.
dc.subject.por.fl_str_mv glycated hemoglobin A
chromatography high-pressure liquid
carbamylated hemoglobin
urea
topic glycated hemoglobin A
chromatography high-pressure liquid
carbamylated hemoglobin
urea
description ABSTRACT Hemoglobin A1c (HbA1c) measurement is commonly performed in diabetes mellitus patients to monitor glycemic control over the last three to four months. Carbamylated hemoglobin, which is the hemoglobin that binds to isocyanic acid derived from urea, is one of the possible analytical interference in the uremic patient. When measured by ion-exchange high-performance liquid chromatography (HPLC), carbamylated hemoglobin forms a peak that overlaps the peak of HbA1c, causing a falsely elevated HbA1c result. We report a case of a 60-years-old man who had a spurious increase in HbA1c, with a high carbamylated hemoglobin peak disproportionate to the urea value. Subsequent hemoglobin analysis using hemoglobin electrophoresis and HPLC hemoglobin testing system suggested hemoglobin J (Hb J) variant. Our case highlighted the possibility of misleading HbA1c interpretation in the presence of a high carbamylated hemoglobin peak, but not proportionate to urea value. In this study, Hb J was detected. A method free from hemoglobin variant interference should be used ideally, and monitoring glycemic control should be performed using alternative methods, such as serum fructosamine or continuous glucose monitoring.
publishDate 2021
dc.date.none.fl_str_mv 2021-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/report
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format report
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442021000100706
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442021000100706
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.5935/1676-2444.20210029
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv
Sociedade Brasileira de Patologia Clínica
publisher.none.fl_str_mv
Sociedade Brasileira de Patologia Clínica
dc.source.none.fl_str_mv Jornal Brasileiro de Patologia e Medicina Laboratorial v.57 2021
reponame:Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
instname:Sociedade Brasileira de Patologia (SBP)
instacron:SBP
instname_str Sociedade Brasileira de Patologia (SBP)
instacron_str SBP
institution SBP
reponame_str Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
collection Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
repository.name.fl_str_mv Jornal Brasileiro de Patologia e Medicina Laboratorial (Online) - Sociedade Brasileira de Patologia (SBP)
repository.mail.fl_str_mv ||jbpml@sbpc.org.br
_version_ 1752122297881198592