Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone

Detalhes bibliográficos
Autor(a) principal: NEUMANN,Eduardo Aloisio Fleck
Data de Publicação: 2014
Outros Autores: VILLAR,Cristina Cunha, FRANÇA,Fabiana Mantovani Gomes
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Oral Research
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242014000100239
Resumo: Fractured abutment screws may be replaced; however, sometimes, the screw cannot be removed and the entire implant must be surgically removed and replaced. The aim of this study was to compare the fracture resistance of abutment retention screws made of titanium, polyetheretherketone (PEEK) and 30% carbon fiber-reinforced PEEK, using an external hexagonal implant/UCLA-type abutment interface assembly. UCLA-type abutments were fixed to implants using titanium screws (Group 1), polyetheretherketone (PEEK) screws (Group 2), and 30% carbon fiber-reinforced PEEK screws (Group 3). The assemblies were placed on a stainless steel holding apparatus to allow for loading at 45o off-axis, in a universal testing machine. A 200 N load (static load) was applied at the central point of the abutment extremity, at a crosshead speed of 5 mm/minute, until failure. Data was analyzed by ANOVA and Tukey’s range test. The titanium screws had higher fracture resistance, compared with PEEK and 30% carbon fiber-reinforced PEEK screws (p < 0.05). In contrast, no statistically significant difference was observed between the fracture resistance of the PEEK and the 30% carbon fiber-reinforced PEEK screws (p> 0.05). Finally, visual analysis of the fractions revealed that 100% of them occurred at the neck of the abutment screw, suggesting that this is the weakest point of this unit. PEEK abutment screws have lower fracture resistance, in comparison with titanium abutment screws.
id SBPQO-1_0f1b03d8ee308154c2e927cfc04a74cb
oai_identifier_str oai:scielo:S1806-83242014000100239
network_acronym_str SBPQO-1
network_name_str Brazilian Oral Research
repository_id_str
spelling Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketoneDental ImplantsTitaniumCompressive StrengthFractured abutment screws may be replaced; however, sometimes, the screw cannot be removed and the entire implant must be surgically removed and replaced. The aim of this study was to compare the fracture resistance of abutment retention screws made of titanium, polyetheretherketone (PEEK) and 30% carbon fiber-reinforced PEEK, using an external hexagonal implant/UCLA-type abutment interface assembly. UCLA-type abutments were fixed to implants using titanium screws (Group 1), polyetheretherketone (PEEK) screws (Group 2), and 30% carbon fiber-reinforced PEEK screws (Group 3). The assemblies were placed on a stainless steel holding apparatus to allow for loading at 45o off-axis, in a universal testing machine. A 200 N load (static load) was applied at the central point of the abutment extremity, at a crosshead speed of 5 mm/minute, until failure. Data was analyzed by ANOVA and Tukey’s range test. The titanium screws had higher fracture resistance, compared with PEEK and 30% carbon fiber-reinforced PEEK screws (p < 0.05). In contrast, no statistically significant difference was observed between the fracture resistance of the PEEK and the 30% carbon fiber-reinforced PEEK screws (p> 0.05). Finally, visual analysis of the fractions revealed that 100% of them occurred at the neck of the abutment screw, suggesting that this is the weakest point of this unit. PEEK abutment screws have lower fracture resistance, in comparison with titanium abutment screws.Sociedade Brasileira de Pesquisa Odontológica - SBPqO2014-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242014000100239Brazilian Oral Research v.28 n.1 2014reponame:Brazilian Oral Researchinstname:Sociedade Brasileira de Pesquisa Odontológica (SBPqO)instacron:SBPQO10.1590/1807-3107BOR-2014.vol28.0028info:eu-repo/semantics/openAccessNEUMANN,Eduardo Aloisio FleckVILLAR,Cristina CunhaFRANÇA,Fabiana Mantovani Gomeseng2018-09-17T00:00:00Zoai:scielo:S1806-83242014000100239Revistahttps://www.scielo.br/j/bor/https://old.scielo.br/oai/scielo-oai.phppob@edu.usp.br||bor@sbpqo.org.br1807-31071806-8324opendoar:2018-09-17T00:00Brazilian Oral Research - Sociedade Brasileira de Pesquisa Odontológica (SBPqO)false
dc.title.none.fl_str_mv Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
title Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
spellingShingle Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
NEUMANN,Eduardo Aloisio Fleck
Dental Implants
Titanium
Compressive Strength
title_short Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
title_full Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
title_fullStr Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
title_full_unstemmed Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
title_sort Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone
author NEUMANN,Eduardo Aloisio Fleck
author_facet NEUMANN,Eduardo Aloisio Fleck
VILLAR,Cristina Cunha
FRANÇA,Fabiana Mantovani Gomes
author_role author
author2 VILLAR,Cristina Cunha
FRANÇA,Fabiana Mantovani Gomes
author2_role author
author
dc.contributor.author.fl_str_mv NEUMANN,Eduardo Aloisio Fleck
VILLAR,Cristina Cunha
FRANÇA,Fabiana Mantovani Gomes
dc.subject.por.fl_str_mv Dental Implants
Titanium
Compressive Strength
topic Dental Implants
Titanium
Compressive Strength
description Fractured abutment screws may be replaced; however, sometimes, the screw cannot be removed and the entire implant must be surgically removed and replaced. The aim of this study was to compare the fracture resistance of abutment retention screws made of titanium, polyetheretherketone (PEEK) and 30% carbon fiber-reinforced PEEK, using an external hexagonal implant/UCLA-type abutment interface assembly. UCLA-type abutments were fixed to implants using titanium screws (Group 1), polyetheretherketone (PEEK) screws (Group 2), and 30% carbon fiber-reinforced PEEK screws (Group 3). The assemblies were placed on a stainless steel holding apparatus to allow for loading at 45o off-axis, in a universal testing machine. A 200 N load (static load) was applied at the central point of the abutment extremity, at a crosshead speed of 5 mm/minute, until failure. Data was analyzed by ANOVA and Tukey’s range test. The titanium screws had higher fracture resistance, compared with PEEK and 30% carbon fiber-reinforced PEEK screws (p < 0.05). In contrast, no statistically significant difference was observed between the fracture resistance of the PEEK and the 30% carbon fiber-reinforced PEEK screws (p> 0.05). Finally, visual analysis of the fractions revealed that 100% of them occurred at the neck of the abutment screw, suggesting that this is the weakest point of this unit. PEEK abutment screws have lower fracture resistance, in comparison with titanium abutment screws.
publishDate 2014
dc.date.none.fl_str_mv 2014-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242014000100239
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242014000100239
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/1807-3107BOR-2014.vol28.0028
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Pesquisa Odontológica - SBPqO
publisher.none.fl_str_mv Sociedade Brasileira de Pesquisa Odontológica - SBPqO
dc.source.none.fl_str_mv Brazilian Oral Research v.28 n.1 2014
reponame:Brazilian Oral Research
instname:Sociedade Brasileira de Pesquisa Odontológica (SBPqO)
instacron:SBPQO
instname_str Sociedade Brasileira de Pesquisa Odontológica (SBPqO)
instacron_str SBPQO
institution SBPQO
reponame_str Brazilian Oral Research
collection Brazilian Oral Research
repository.name.fl_str_mv Brazilian Oral Research - Sociedade Brasileira de Pesquisa Odontológica (SBPqO)
repository.mail.fl_str_mv pob@edu.usp.br||bor@sbpqo.org.br
_version_ 1750318323811221504