Resumo: | The aim of this study was to evaluate the effect of pitch distance on the primary stability (PS) of orthodontic mini-implants (MIs) in artificial bone. Twenty experimental MIs were allocated to two groups, according to their geometric design: G1(30oX 0.6 mm) and G2(45o X 0.8 mm), and inserted into artificial bone of different densities: D1 (0.32 g/cm3) and D2 (0.64 g/cm3). The maximum insertion torque (IT) and removal torque (RT) values were recorded in N.cm. Loss of torque (LT) values were obtained by calculating the difference between the IT and RT. MI mobility was measured by means of a Periotest assessment. Normality and homogeneity were determined by means of the Kolmogorov-Smirnov and Shapiro-Wilk tests, respectively. A two-way ANOVA was used to detect differences between the mini-implant design and density factors. The ANOVA/Tukey tests were used to determine the intergroup difference. Higher IT values were observed for G2 (p ≤ 0.05) in D2. No statistical difference for RT was observed between the groups, whereas G2 presented higher values only for LT (p ≤ 0.05). The Periotest values (PTV) were higher for G1, in comparison with G2, in D1. G1 presented better PS in D2 (IT, RT and LT), whereas G2 was found to be more stable in D1, after evaluation with Periotest. |
id |
SBPQO-1_d44a079195e42a5829fbaf89ec2ee0df |
---|---|
oai_identifier_str |
oai:scielo:S1806-83242015000100231 |
network_acronym_str |
SBPQO-1 |
network_name_str |
Brazilian Oral Research |
repository_id_str |
|
spelling |
Mechanical influence of thread pitch on orthodontic mini-implant stabilityMechanicsOrthodontic Anchorage ProceduresMaterials TestingThe aim of this study was to evaluate the effect of pitch distance on the primary stability (PS) of orthodontic mini-implants (MIs) in artificial bone. Twenty experimental MIs were allocated to two groups, according to their geometric design: G1(30oX 0.6 mm) and G2(45o X 0.8 mm), and inserted into artificial bone of different densities: D1 (0.32 g/cm3) and D2 (0.64 g/cm3). The maximum insertion torque (IT) and removal torque (RT) values were recorded in N.cm. Loss of torque (LT) values were obtained by calculating the difference between the IT and RT. MI mobility was measured by means of a Periotest assessment. Normality and homogeneity were determined by means of the Kolmogorov-Smirnov and Shapiro-Wilk tests, respectively. A two-way ANOVA was used to detect differences between the mini-implant design and density factors. The ANOVA/Tukey tests were used to determine the intergroup difference. Higher IT values were observed for G2 (p ≤ 0.05) in D2. No statistical difference for RT was observed between the groups, whereas G2 presented higher values only for LT (p ≤ 0.05). The Periotest values (PTV) were higher for G1, in comparison with G2, in D1. G1 presented better PS in D2 (IT, RT and LT), whereas G2 was found to be more stable in D1, after evaluation with Periotest.Sociedade Brasileira de Pesquisa Odontológica - SBPqO2015-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242015000100231Brazilian Oral Research v.29 n.1 2015reponame:Brazilian Oral Researchinstname:Sociedade Brasileira de Pesquisa Odontológica (SBPqO)instacron:SBPQO10.1590/1807-3107BOR-2015.vol29.0042info:eu-repo/semantics/openAccessCUNHA,Amanda Carneiro daFREITAS,Amanda Osório Ayres deMARQUEZAN,MarianaNOJIMA,Lincoln Issamueng2018-09-17T00:00:00Zoai:scielo:S1806-83242015000100231Revistahttps://www.scielo.br/j/bor/https://old.scielo.br/oai/scielo-oai.phppob@edu.usp.br||bor@sbpqo.org.br1807-31071806-8324opendoar:2018-09-17T00:00Brazilian Oral Research - Sociedade Brasileira de Pesquisa Odontológica (SBPqO)false |
dc.title.none.fl_str_mv |
Mechanical influence of thread pitch on orthodontic mini-implant stability |
title |
Mechanical influence of thread pitch on orthodontic mini-implant stability |
spellingShingle |
Mechanical influence of thread pitch on orthodontic mini-implant stability CUNHA,Amanda Carneiro da Mechanics Orthodontic Anchorage Procedures Materials Testing |
title_short |
Mechanical influence of thread pitch on orthodontic mini-implant stability |
title_full |
Mechanical influence of thread pitch on orthodontic mini-implant stability |
title_fullStr |
Mechanical influence of thread pitch on orthodontic mini-implant stability |
title_full_unstemmed |
Mechanical influence of thread pitch on orthodontic mini-implant stability |
title_sort |
Mechanical influence of thread pitch on orthodontic mini-implant stability |
author |
CUNHA,Amanda Carneiro da |
author_facet |
CUNHA,Amanda Carneiro da FREITAS,Amanda Osório Ayres de MARQUEZAN,Mariana NOJIMA,Lincoln Issamu |
author_role |
author |
author2 |
FREITAS,Amanda Osório Ayres de MARQUEZAN,Mariana NOJIMA,Lincoln Issamu |
author2_role |
author author author |
dc.contributor.author.fl_str_mv |
CUNHA,Amanda Carneiro da FREITAS,Amanda Osório Ayres de MARQUEZAN,Mariana NOJIMA,Lincoln Issamu |
dc.subject.por.fl_str_mv |
Mechanics Orthodontic Anchorage Procedures Materials Testing |
topic |
Mechanics Orthodontic Anchorage Procedures Materials Testing |
description |
The aim of this study was to evaluate the effect of pitch distance on the primary stability (PS) of orthodontic mini-implants (MIs) in artificial bone. Twenty experimental MIs were allocated to two groups, according to their geometric design: G1(30oX 0.6 mm) and G2(45o X 0.8 mm), and inserted into artificial bone of different densities: D1 (0.32 g/cm3) and D2 (0.64 g/cm3). The maximum insertion torque (IT) and removal torque (RT) values were recorded in N.cm. Loss of torque (LT) values were obtained by calculating the difference between the IT and RT. MI mobility was measured by means of a Periotest assessment. Normality and homogeneity were determined by means of the Kolmogorov-Smirnov and Shapiro-Wilk tests, respectively. A two-way ANOVA was used to detect differences between the mini-implant design and density factors. The ANOVA/Tukey tests were used to determine the intergroup difference. Higher IT values were observed for G2 (p ≤ 0.05) in D2. No statistical difference for RT was observed between the groups, whereas G2 presented higher values only for LT (p ≤ 0.05). The Periotest values (PTV) were higher for G1, in comparison with G2, in D1. G1 presented better PS in D2 (IT, RT and LT), whereas G2 was found to be more stable in D1, after evaluation with Periotest. |
publishDate |
2015 |
dc.date.none.fl_str_mv |
2015-01-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242015000100231 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242015000100231 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.1590/1807-3107BOR-2015.vol29.0042 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
Sociedade Brasileira de Pesquisa Odontológica - SBPqO |
publisher.none.fl_str_mv |
Sociedade Brasileira de Pesquisa Odontológica - SBPqO |
dc.source.none.fl_str_mv |
Brazilian Oral Research v.29 n.1 2015 reponame:Brazilian Oral Research instname:Sociedade Brasileira de Pesquisa Odontológica (SBPqO) instacron:SBPQO |
instname_str |
Sociedade Brasileira de Pesquisa Odontológica (SBPqO) |
instacron_str |
SBPQO |
institution |
SBPQO |
reponame_str |
Brazilian Oral Research |
collection |
Brazilian Oral Research |
repository.name.fl_str_mv |
Brazilian Oral Research - Sociedade Brasileira de Pesquisa Odontológica (SBPqO) |
repository.mail.fl_str_mv |
pob@edu.usp.br||bor@sbpqo.org.br |
_version_ |
1750318324212826112 |