Patient-ventilator asynchrony

Detalhes bibliográficos
Autor(a) principal: Holanda,Marcelo Alcantara
Data de Publicação: 2018
Outros Autores: Vasconcelos,Renata dos Santos, Ferreira,Juliana Carvalho, Pinheiro,Bruno Valle
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Jornal Brasileiro de Pneumologia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-37132018000400321
Resumo: ABSTRACT Patient-v entilator asynchrony (PVA) is a mismatch between the patient, regarding time, flow, volume, or pressure demands of the patient respiratory system, and the ventilator, which supplies such demands, during mechanical ventilation (MV). It is a common phenomenon, with incidence rates ranging from 10% to 85%. PVA might be due to factors related to the patient, to the ventilator, or both. The most common PVA types are those related to triggering, such as ineffective effort, auto-triggering, and double triggering; those related to premature or delayed cycling; and those related to insufficient or excessive flow. Each of these types can be detected by visual inspection of volume, flow, and pressure waveforms on the mechanical ventilator display. Specific ventilatory strategies can be used in combination with clinical management, such as controlling patient pain, anxiety, fever, etc. Deep sedation should be avoided whenever possible. PVA has been associated with unwanted outcomes, such as discomfort, dyspnea, worsening of pulmonary gas exchange, increased work of breathing, diaphragmatic injury, sleep impairment, and increased use of sedation or neuromuscular blockade, as well as increases in the duration of MV, weaning time, and mortality. Proportional assist ventilation and neurally adjusted ventilatory assist are modalities of partial ventilatory support that reduce PVA and have shown promise. This article reviews the literature on the types and causes of PVA, as well as the methods used in its evaluation, its potential implications in the recovery process of critically ill patients, and strategies for its resolution.
id SBPT-1_dcf6339d84f5d9c98348742503eea476
oai_identifier_str oai:scielo:S1806-37132018000400321
network_acronym_str SBPT-1
network_name_str Jornal Brasileiro de Pneumologia (Online)
repository_id_str
spelling Patient-ventilator asynchronyRespiration, artificialRespiratory insufficiencyInteractive ventilatory supportABSTRACT Patient-v entilator asynchrony (PVA) is a mismatch between the patient, regarding time, flow, volume, or pressure demands of the patient respiratory system, and the ventilator, which supplies such demands, during mechanical ventilation (MV). It is a common phenomenon, with incidence rates ranging from 10% to 85%. PVA might be due to factors related to the patient, to the ventilator, or both. The most common PVA types are those related to triggering, such as ineffective effort, auto-triggering, and double triggering; those related to premature or delayed cycling; and those related to insufficient or excessive flow. Each of these types can be detected by visual inspection of volume, flow, and pressure waveforms on the mechanical ventilator display. Specific ventilatory strategies can be used in combination with clinical management, such as controlling patient pain, anxiety, fever, etc. Deep sedation should be avoided whenever possible. PVA has been associated with unwanted outcomes, such as discomfort, dyspnea, worsening of pulmonary gas exchange, increased work of breathing, diaphragmatic injury, sleep impairment, and increased use of sedation or neuromuscular blockade, as well as increases in the duration of MV, weaning time, and mortality. Proportional assist ventilation and neurally adjusted ventilatory assist are modalities of partial ventilatory support that reduce PVA and have shown promise. This article reviews the literature on the types and causes of PVA, as well as the methods used in its evaluation, its potential implications in the recovery process of critically ill patients, and strategies for its resolution.Sociedade Brasileira de Pneumologia e Tisiologia2018-08-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-37132018000400321Jornal Brasileiro de Pneumologia v.44 n.4 2018reponame:Jornal Brasileiro de Pneumologia (Online)instname:Sociedade Brasileira de Pneumologia e Tisiologia (SBPT)instacron:SBPT10.1590/s1806-37562017000000185info:eu-repo/semantics/openAccessHolanda,Marcelo AlcantaraVasconcelos,Renata dos SantosFerreira,Juliana CarvalhoPinheiro,Bruno Valleeng2018-10-08T00:00:00Zoai:scielo:S1806-37132018000400321Revistahttp://www.jornaldepneumologia.com.br/default.aspONGhttps://old.scielo.br/oai/scielo-oai.php||jbp@jbp.org.br|| jpneumo@jornaldepneumologia.com.br1806-37561806-3713opendoar:2018-10-08T00:00Jornal Brasileiro de Pneumologia (Online) - Sociedade Brasileira de Pneumologia e Tisiologia (SBPT)false
dc.title.none.fl_str_mv Patient-ventilator asynchrony
title Patient-ventilator asynchrony
spellingShingle Patient-ventilator asynchrony
Holanda,Marcelo Alcantara
Respiration, artificial
Respiratory insufficiency
Interactive ventilatory support
title_short Patient-ventilator asynchrony
title_full Patient-ventilator asynchrony
title_fullStr Patient-ventilator asynchrony
title_full_unstemmed Patient-ventilator asynchrony
title_sort Patient-ventilator asynchrony
author Holanda,Marcelo Alcantara
author_facet Holanda,Marcelo Alcantara
Vasconcelos,Renata dos Santos
Ferreira,Juliana Carvalho
Pinheiro,Bruno Valle
author_role author
author2 Vasconcelos,Renata dos Santos
Ferreira,Juliana Carvalho
Pinheiro,Bruno Valle
author2_role author
author
author
dc.contributor.author.fl_str_mv Holanda,Marcelo Alcantara
Vasconcelos,Renata dos Santos
Ferreira,Juliana Carvalho
Pinheiro,Bruno Valle
dc.subject.por.fl_str_mv Respiration, artificial
Respiratory insufficiency
Interactive ventilatory support
topic Respiration, artificial
Respiratory insufficiency
Interactive ventilatory support
description ABSTRACT Patient-v entilator asynchrony (PVA) is a mismatch between the patient, regarding time, flow, volume, or pressure demands of the patient respiratory system, and the ventilator, which supplies such demands, during mechanical ventilation (MV). It is a common phenomenon, with incidence rates ranging from 10% to 85%. PVA might be due to factors related to the patient, to the ventilator, or both. The most common PVA types are those related to triggering, such as ineffective effort, auto-triggering, and double triggering; those related to premature or delayed cycling; and those related to insufficient or excessive flow. Each of these types can be detected by visual inspection of volume, flow, and pressure waveforms on the mechanical ventilator display. Specific ventilatory strategies can be used in combination with clinical management, such as controlling patient pain, anxiety, fever, etc. Deep sedation should be avoided whenever possible. PVA has been associated with unwanted outcomes, such as discomfort, dyspnea, worsening of pulmonary gas exchange, increased work of breathing, diaphragmatic injury, sleep impairment, and increased use of sedation or neuromuscular blockade, as well as increases in the duration of MV, weaning time, and mortality. Proportional assist ventilation and neurally adjusted ventilatory assist are modalities of partial ventilatory support that reduce PVA and have shown promise. This article reviews the literature on the types and causes of PVA, as well as the methods used in its evaluation, its potential implications in the recovery process of critically ill patients, and strategies for its resolution.
publishDate 2018
dc.date.none.fl_str_mv 2018-08-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-37132018000400321
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1806-37132018000400321
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/s1806-37562017000000185
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Pneumologia e Tisiologia
publisher.none.fl_str_mv Sociedade Brasileira de Pneumologia e Tisiologia
dc.source.none.fl_str_mv Jornal Brasileiro de Pneumologia v.44 n.4 2018
reponame:Jornal Brasileiro de Pneumologia (Online)
instname:Sociedade Brasileira de Pneumologia e Tisiologia (SBPT)
instacron:SBPT
instname_str Sociedade Brasileira de Pneumologia e Tisiologia (SBPT)
instacron_str SBPT
institution SBPT
reponame_str Jornal Brasileiro de Pneumologia (Online)
collection Jornal Brasileiro de Pneumologia (Online)
repository.name.fl_str_mv Jornal Brasileiro de Pneumologia (Online) - Sociedade Brasileira de Pneumologia e Tisiologia (SBPT)
repository.mail.fl_str_mv ||jbp@jbp.org.br|| jpneumo@jornaldepneumologia.com.br
_version_ 1750318347297226752