Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry

Detalhes bibliográficos
Autor(a) principal: Toledo,Carla
Data de Publicação: 2007
Outros Autores: Valle,Luis, Narváez,Jessica, Richter,Pablo
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Journal of the Brazilian Chemical Society (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532007000500010
Resumo: A rapid and simple method was developed for screening polychlorinated biphenyls in transformer oil. The method is based on liquid-liquid extraction in methanol and subsequent direct gas chromatography-mass spectrometric determination. The extent of methanol polarity allows to extract the PCBs, co-extracting a minimum amount of the apolar oil components, thus avoiding the time-consuming steps involving in sample preparation. Since the regulation established internationally for wastes containing PCBs is 50 mg kg-1, the method rapidly determines whether the transformer oil is PCBs-contaminated or not. In order to reach the best conditions for extraction a multivariate optimization was carried out considering two significant variables: volume of methanol and extraction time. Under optimum conditions, recoveries of Aroclor 1256 and Aroclor 1260 were between 80 and 100%, and precision, expressed as relative standard deviation, was from 2 to 4%. The method was applied to real samples polluted with PCBs.
id SBQ-2_16e6647005a7b09e9d7e22a90ffbe9bb
oai_identifier_str oai:scielo:S0103-50532007000500010
network_acronym_str SBQ-2
network_name_str Journal of the Brazilian Chemical Society (Online)
repository_id_str
spelling Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometryPCBsoil transformerliquid-liquid extractionGC-MSA rapid and simple method was developed for screening polychlorinated biphenyls in transformer oil. The method is based on liquid-liquid extraction in methanol and subsequent direct gas chromatography-mass spectrometric determination. The extent of methanol polarity allows to extract the PCBs, co-extracting a minimum amount of the apolar oil components, thus avoiding the time-consuming steps involving in sample preparation. Since the regulation established internationally for wastes containing PCBs is 50 mg kg-1, the method rapidly determines whether the transformer oil is PCBs-contaminated or not. In order to reach the best conditions for extraction a multivariate optimization was carried out considering two significant variables: volume of methanol and extraction time. Under optimum conditions, recoveries of Aroclor 1256 and Aroclor 1260 were between 80 and 100%, and precision, expressed as relative standard deviation, was from 2 to 4%. The method was applied to real samples polluted with PCBs.Sociedade Brasileira de Química2007-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532007000500010Journal of the Brazilian Chemical Society v.18 n.5 2007reponame:Journal of the Brazilian Chemical Society (Online)instname:Sociedade Brasileira de Química (SBQ)instacron:SBQ10.1590/S0103-50532007000500010info:eu-repo/semantics/openAccessToledo,CarlaValle,LuisNarváez,JessicaRichter,Pabloeng2007-10-25T00:00:00Zoai:scielo:S0103-50532007000500010Revistahttp://jbcs.sbq.org.brONGhttps://old.scielo.br/oai/scielo-oai.php||office@jbcs.sbq.org.br1678-47900103-5053opendoar:2007-10-25T00:00Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)false
dc.title.none.fl_str_mv Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
title Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
spellingShingle Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
Toledo,Carla
PCBs
oil transformer
liquid-liquid extraction
GC-MS
title_short Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
title_full Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
title_fullStr Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
title_full_unstemmed Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
title_sort Screening method for rapid determination of polychlorinated biphenyls in transformer oil by liquid-liquid extraction and gas chromatography-mass spectrometry
author Toledo,Carla
author_facet Toledo,Carla
Valle,Luis
Narváez,Jessica
Richter,Pablo
author_role author
author2 Valle,Luis
Narváez,Jessica
Richter,Pablo
author2_role author
author
author
dc.contributor.author.fl_str_mv Toledo,Carla
Valle,Luis
Narváez,Jessica
Richter,Pablo
dc.subject.por.fl_str_mv PCBs
oil transformer
liquid-liquid extraction
GC-MS
topic PCBs
oil transformer
liquid-liquid extraction
GC-MS
description A rapid and simple method was developed for screening polychlorinated biphenyls in transformer oil. The method is based on liquid-liquid extraction in methanol and subsequent direct gas chromatography-mass spectrometric determination. The extent of methanol polarity allows to extract the PCBs, co-extracting a minimum amount of the apolar oil components, thus avoiding the time-consuming steps involving in sample preparation. Since the regulation established internationally for wastes containing PCBs is 50 mg kg-1, the method rapidly determines whether the transformer oil is PCBs-contaminated or not. In order to reach the best conditions for extraction a multivariate optimization was carried out considering two significant variables: volume of methanol and extraction time. Under optimum conditions, recoveries of Aroclor 1256 and Aroclor 1260 were between 80 and 100%, and precision, expressed as relative standard deviation, was from 2 to 4%. The method was applied to real samples polluted with PCBs.
publishDate 2007
dc.date.none.fl_str_mv 2007-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532007000500010
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532007000500010
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0103-50532007000500010
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Química
publisher.none.fl_str_mv Sociedade Brasileira de Química
dc.source.none.fl_str_mv Journal of the Brazilian Chemical Society v.18 n.5 2007
reponame:Journal of the Brazilian Chemical Society (Online)
instname:Sociedade Brasileira de Química (SBQ)
instacron:SBQ
instname_str Sociedade Brasileira de Química (SBQ)
instacron_str SBQ
institution SBQ
reponame_str Journal of the Brazilian Chemical Society (Online)
collection Journal of the Brazilian Chemical Society (Online)
repository.name.fl_str_mv Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)
repository.mail.fl_str_mv ||office@jbcs.sbq.org.br
_version_ 1750318168180523008